K160128 · Ethicon Endo-Surgery, LLC · GEI · Apr 28, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160128
Device Name
Endopath Electrosurgery Probe Plus II
Applicant
Ethicon Endo-Surgery, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Apr 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ENDOPATH Electrosurgery Probe Plus II system has application in minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation, and fluid evacuation through a common trocar sleeve.
Device Story
Monopolar electrosurgical probe used in minimally invasive surgery; facilitates tissue dissection, coagulation, irrigation, and fluid evacuation via trocar sleeve. Device consists of handle and interchangeable shafts; connects to standard monopolar electrosurgery unit (ESU). Surgeon operates device manually; activation via hand buttons or ESU foot pedal. Provides suction/irrigation and hemostasis; improves surgical ergonomics and activation force compared to predicate. Benefits patient by enabling precise tissue management and fluid control during laparoscopic procedures.
Clinical Evidence
No human clinical trial data. Evidence based on bench testing (flow rates, impedance, mechanical forces) and an acute animal study comparing tissue effects (dissection, hemostasis) against the predicate device. Results confirmed subject device performs as well as the predicate.
Technological Characteristics
Monopolar electrosurgical device; suction/irrigation modalities. Materials evaluated per ISO 10993:2009. Sterilization via gamma radiation. Complies with IEC 60601-1:2005 (electrical safety) and IEC 60601-1-2:2007 (EMC). Features include improved button force, ergonomic design, and dual hand/foot activation capability.
Indications for Use
Indicated for minimally invasive surgical procedures requiring tissue dissection, coagulation, irrigation, and fluid evacuation. Contraindicated for contraceptive coagulation of fallopian tissue (though permitted for hemostasis following transection) and when minimally invasive techniques are contraindicated.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Ethicon Endo-Surgery, LLC David Locke Senior Regulatory Affairs Specialist 4545 Creek Road Cincinnati, Ohio 45242
Re: K160128
Trade/Device Name: Endopath Electrosurgery Probe Plus II Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: January 19, 2016 Received: January 20, 2016
Dear David Locke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K160128
Device Name ENDOPATH Electrosurgery Probe Plus II
#### Indications for Use (Describe)
Indications
The ENDOPATH Electrosurgery Probe Plus II system has application in minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation, and fluid evacuation through a common trocar sleeve.
Contraindications
· These instruments are not intended for contraceptive coagulation of fallopian tissue but may be used to achieve
hemostasis following transection of the fallopian tube.
· These instruments are not intended for use when minimally invasive techniques are contraindicated.
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
# Company
Ethicon Endo-Surgery 475 Calle C Guaynabo, PR 00969
# Contact
David Locke, DRSc, MS, RAC Senior Regulatory Affairs Program Lead Ethicon Endo-Surgery, Inc. 4545 Creek Road Cincinnati, OH 45242 Telephone: (513) 337-1281 Fax: (513) 337- 4366 Email: dlocke1 @its.jnj.com
# Date Prepared 19 January 2016
#### Device Name
Trade Name: Evacuation/Irrigation/Electrosurgical Device Classification Name: Laparoscope, General & Plastic Surgery Common Name: Endopath® Electrosurgery Probe Plus II Catalog Number: EPH02, EPS01, EPS02, EPS03, EPS04, EPS05, EPS06, EPS07
Device Class Class II
Device Panel General & Plastic Surgery
Product Code GEI
# Predicate Device
Ethicon Evacuation/Irrigation/Electrosurgical Device cleared under K912492
#### Device Description
The Endopath® Electrosurgery Probe Plus II device (product codes beginning with EPH and EPS) work with a standard monopolar unit and has applications for minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation, and fluid evacuation through a common trocar sleeve. The subject device system offers one handle (EPH02) and seven shafts (EPS01-EPS07) which allow for a variety of configurations to best meet the surgeon's needs. Shafts EPS01 through EPS07 are for use through a 5 mm diameter surgical trocar or a larger trocar with a 5 mm reducer.
{4}------------------------------------------------
#### Indications for Use
The ENDOPATH Electrosurgery Probe Plus II system has application in minimally invasive procedures to facilitate tissue dissection, coagulation, irrigation and fluid evacuation through a common trocar sleeve.
#### Contraindications
- These instruments are not intended for contraceptive coagulation of fallopian tissue . but may be used to achieve hemostasis following transection of the fallopian tube.
- These instruments are not intended for use when minimally invasive techniques are . contraindicated.
# Technological Characteristics
The subject ENDOPATH Electrosurgery Probe Plus II device and predicate Ethicon Evacuation/Irrigation/Electrosurgical device use the same monopolar technology to perform their intended use. The subject and predicate devices also use the same standard suction and irrigation modalities to perform their intended use. The subject ENDOPATH Electrosurgery Probe Plus II device is similar to the predicate Ethicon Evacuation/Electrosurgical device with respect to the primary modes of action which include suction, irrigation and hemostasis. A few technological differences were identified between the subject device and the currently marketed predicate devices, attributable to the respective design of each. As compared with the predicate devices, the subject device does however incorporates several design enhancements which includes suction and irrigation button force improvement, monopolar activation force improvement, elimination of the button pin, improved ergonomics, and combined hand and foot activation capability (hand activation through the ENDOPATH Electrosurgery Probe Plus II device itself and foot activation through a monopolar electrosurgery unit (ESU) generator foot pedal).
# Performance Data
Bench testing and laboratory evaluations were conducted to demonstrate that the ENDOPATH Electrosurgery Probe Plus II performed as intended.
#### Sterilization
The subject device and the predicate device are sterilized via gamma radiation; both devices are sterilized to the same sterility assurance level.
#### Biocompatibility
The biocompatibility evaluation for the ENDOPATH Electrosurgery Probe Plus II device was conducted in accordance with the following standards: ISO 10993:2009. Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and FDA Blue Book Memorandum #G95-1: Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". The device passed all biocompatibility ISO testing which included cytotoxicity, sensitization, intracutaneous reactivity and acute system toxicity testing.
{5}------------------------------------------------
# EMC
Electrical safety and EMC testing were conducted on the ENDOPATH Electrosurgery Probe Plus II; the system complies with IEC 60601-1-2:2007 for electromagnetic compatibility and IEC 60601-1:2005 for electrical safety.
# Bench
Flow rate of suction and irrigation, impedance, minimum distal retraction force and sheath extension force were evaluated for the ENDOPATH Electrosurgery Probe Plus II to support substantial equivalent to the predicate device. Data generated from these tests met the predetermined acceptance criteria.
# Acute
Testing was performed in an acute study with the ENDOPATH Electrosurgery Probe Plus II vs. the currently marketed predicate device to demonstrate that the tissue effects were not different than the predicate device. The results of the study demonstrated the ability of the subject device to create tissue planes, dissect tissue, and provide hemostasis in different tissue types.
# Clinical
This premarket notification does not rely on human clinical trial data to demonstrate substantial equivalence.
# Conclusion
The results of the bench testing and laboratory evaluations in an animal model demonstrate that the Endopath® Electrosurgery Proble Plus II device is as safe and effective and performs as well as the identified legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.