← Product Code [MBI](/productcode/MBI) · K160118

# STROPP (Single Tunnel Repair of Plantar Plate) (K160118)

_Crossroads Extremity Systems, LLC · MBI · May 3, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K160118

## Device Facts

- **Applicant:** Crossroads Extremity Systems, LLC
- **Product Code:** [MBI](/productcode/MBI.md)
- **Decision Date:** May 3, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The STROPP™ System is intended for use with a suture in metatarsal ligament and tendon repairs.

## Device Story

The STROPP™ System is a sterile kit containing implants and manual surgical instruments (needle, needle passer, inserter, K-wires, distractor) for plantar plate repair. The device is a smooth, press-fit implant designed for use with a suture (not included) in a single bone tunnel technique. The surgeon places the implant into the bone tunnel and ties the suture knot over the shoulder of the device. It is used in a clinical/surgical setting by a physician. The device provides a secure anchor for ligament and tendon repair, facilitating tissue stabilization. The smooth design aims to prevent suture damage. It is available in 3.0 mm and 3.5 mm diameters to accommodate varying patient anatomy and bone quality.

## Clinical Evidence

Bench testing only. Non-clinical performance testing included construct pull-through testing in a simulated use model to evaluate force resistance and cadaveric simulated use testing to evaluate the interaction between the implant, bone, and suture. Results demonstrated statistical equivalence to current plantar plate repair options with no damage to bone, implant, or suture, and no new failure modes.

## Technological Characteristics

The device is composed of polyetheretherketone (PEEK) per ASTM F2026. It is a smooth, press-fit bone fixation fastener available in 3.0 mm and 3.5 mm diameters (6-12 mm lengths). It is a manual, non-powered, sterile-packaged surgical implant system.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Titanium Suture Anchor ([K133229](/device/K133229.md))
- Threaded PEEK K-Wire ([K133235](/device/K133235.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other, representing the department's focus on people and health.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 3, 2016

CrossRoads Extremity Systems, LLC Mr. Vernon Hartdegen Sr. Vice President of Operations 6055 Primacy Parkway, Suite 140 Memphis, Tennessee 38119

Re: K160118

Trade/Device Name: STROPP™ (Single Tunnel Repair of Plantar Plate) Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: March 14, 2016 Received: March 17, 2016

Dear Mr. Hartdegen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

K160118

Device Name STROPP(TM) System

Indications for Use (Describe)

The STROPP(TM) System is intended for use with a suture in metatarsal ligament and tendon repairs.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Date:             | April 27, 2016                                                                                                                                        |
|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name:      | Single Tunnel Repair of Plantar Plate (STROPPTM) System                                                                                               |
| Company:          | CrossRoads Extremity Systems<br>6055 Primacy Parkway, Suite 140<br>Memphis, TN 38119 USA<br>Phone: 901.221.8406                                       |
| Primary Contact:  | Vernon Hartdegen, Sr. VP of Operations<br>901.221.8406<br>vhartdegen@crextremity.com                                                                  |
| Trade Name:       | Single Tunnel Repair of Plantar Plate (STROPPTM) System                                                                                               |
| Common Name:      | Suture Retention Device                                                                                                                               |
| Classification:   | Class II                                                                                                                                              |
| Regulation:       | 21 CFR 888.3040 Smooth or threaded metallic bone<br>fixation fastener                                                                                 |
| Panel:            | 87 - Orthopedic                                                                                                                                       |
| Product Code:     | MBI                                                                                                                                                   |
| Predicate Device: | Primary Predicate<br>K133229 Titanium Suture Anchor<br>MTP Solutions (Smith Nephew Hat Trick Lesser Toe<br>Repair System)<br><br>Additional Predicate |
|                   | K133235 Threaded PEEK K-Wire<br>MTP Solutions (Smith Nephew Hat Trick Lesser Toe<br>Repair System)                                                    |

# Device Description:

The STROPP™ System is sterile packaged kit to include implants and instruments for plantar plate repair. Implants are 3.0 mm or 3.5 mm diameters and range in length from 6 mm to 12 mm. Implants are packaged with the following manual surgical

{4}------------------------------------------------

instruments for general use: a needle, needle passer, an inserter, Kwires and distractor.

The implants are compatible with Size 0 suture; suture is not included with the system.

### Indications for Use:

The STROPP™ System is intended for use with a suture in metatarsal ligament and tendon repairs.

### Materials:

The STRoPP™ device is composed of polyetheretherketone (PEEK) per ASTM F2026.

## Substantial Equivalence:

The subject STROPP™ System is substantially equivalent to the primary predicate device, K133229 (MTP Solutions Titanium Suture Anchor; Smith Nephew Hat Trick Lesser Toe Repair System) in intended use, surgical technique and principle of operation. The subject STROPP™ System is substantially equivalent to the additional predicate device, K133235 (MTP Solutions Threaded PEEK K-wire; Smith Nephew Hat Trick Lesser Toe Repair System) in its material of construction. The differences in technological characteristics between the predicates and the subject device do not raise any new or different questions of safety and effectiveness and is thereby substantially equivalent to the predicate devices.

The primary predicate device (K133229) contains threads for bone engagement that also interact with the suture. The subject device is a smooth press-fit device with no sharp edges or features to potentially impart damage to the suture. Both the subject and primary predicate (K133229) devices utilize a single bone tunnel technique and require the suture knot to be tied over the shoulder of the device. The primary predicate device (K133229) is manufactured from titanium alloy and the subject device is manufactured from PEEK (ASTM F2026), which is the same material as the additional predicate device (K133235). The primary predicate device (K133229) is available in one diameter (2.5mm) and the subject device is provided in two diameters (3.0mm and 3.5mm). The additional sizes of the subject device allow the surgeon to determine the optimal device size based on patient anatomy and bone quality. The primary predicate (K133229) is indicated for use with suture in the foot along with multiple expanded indications. While the indications for use of the subject device are encompassed in the indications for use of the primary predicate (K133229), they match exactly the indication for use of the additional predicate (K133235).

## Performance Testing:

Non-clinical performance testing of the STROPP™ System has been performed to demonstrate substantial equivalence. Testing has been performed on test samples that are representative of the finished device. Non-clinical testing included

{5}------------------------------------------------

construct pull through testing in a simulated use model utilizing various suture materials to determine if the STRoPP™ construct could withstand the pull through forces that may be present in vivo. The construct pull through testing demonstrated statistical equivalence to current plantar plate repair options and that the STRoPP™ System introduces no additional failure modes. In addition, cadaveric simulated use testing was conducted to evaluate the interaction between the implant, bone and suture. The cadaveric testing resulted in no damage to the bone, implant or suture. All testing met the specified acceptance criteria and raised no new issues of safety or effectiveness. Based on the non-clinical test results, it can be concluded that the subject STRoPP™ System is substantially equivalent and is safe and effective for the intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K160118](https://fda-staging.innolitics.com/device/K160118)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).