K160114 · X-Spine Systems, Inc. · NQW · Mar 15, 2016 · Orthopedic
Device Facts
Record ID
K160114
Device Name
Xspan Laminoplasty Fixation System
Applicant
X-Spine Systems, Inc.
Product Code
NQW · Orthopedic
Decision Date
Mar 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3050
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Xspan™ Laminoplasty Fixation System is intended for use in the lower cervical and upper thoracic spine (C3-T3) after a laminoplasty has been performed. The Xspan Laminoplasty Fixation System holds or buttresses the allograft in place in order to prevent expulsion of the allograft or impingement of the spinal cord.
Device Story
Posterior spinal fixation system; consists of titanium alloy laminoplasty plates, hinge plates, and mounting screws. Used in lower cervical and upper thoracic spine (C3-T3) following laminoplasty. Surgeon implants device to hold/buttress allograft, preventing expulsion or spinal cord impingement. Provided sterile via gamma radiation; single-use only. Includes instrumentation for implantation. Mechanical performance validated via static and dynamic 4-point bend testing.
Clinical Evidence
Bench testing only. No clinical data provided. Mechanical performance validated via static and dynamic 4-point bend testing per ASTM F2193 and ASTM F382. Biocompatibility supported by history of use of ASTM F136 titanium alloy.
Technological Characteristics
Materials: Titanium-6Aluminum-4Vanadium ELI alloy (ASTM F136). Components: Laminoplasty plates, hinge plates, mounting screws. Energy: None (mechanical fixation). Connectivity: None. Sterilization: Gamma radiation. Form factor: Various sizes/forms for anatomical adaptation.
Indications for Use
Indicated for patients undergoing laminoplasty in the lower cervical and upper thoracic spine (C3-T3) to hold or buttress allograft and prevent expulsion or spinal cord impingement.
Regulatory Classification
Identification
A spinal interlaminal fixation orthosis is a device intended to be implanted made of an alloy, such as stainless steel, that consists of various hooks and a posteriorly placed compression or distraction rod. The device is implanted, usually across three adjacent vertebrae, to straighten and immobilize the spine to allow bone grafts to unite and fuse the vertebrae together. The device is used primarily in the treatment of scoliosis (a lateral curvature of the spine), but it also may be used in the treatment of fracture or dislocation of the spine, grades 3 and 4 of spondylolisthesis (a dislocation of the spinal column), and lower back syndrome.
Predicate Devices
Aesculap Implant Systems (AIS) Laminoplasty Plating System (K103284; K090354)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 15, 2016
X-spine Systems, Incorporated Kriss Anderson Director, Regulatory Affairs 452 Alexandersville Road Miamisburg, Ohio 45342
Re: K160114
Trade/Device Name: Xspan™ Laminoplasty Fixation System Regulation Number: 21 CFR 888.3050 Regulation Name: Spinal interlaminal fixation orthosis Regulatory Class: Class II Product Code: NQW Dated: January 15, 2016 Received: January 19, 2016
Dear Kriss Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160114
R10011+
Device Name Xspan™ Laminoplasty Fixation System
### Indications for Use (Describe)
The Xspan™ Laminoplasty Fixation System is intended for use in the lower cervical and upper thoracic spine (C3-T3) after a laminoplasty has been performed. The Xspan Laminoplasty Fixation System holds or buttresses the allograft in place in order to prevent expulsion of the allograft or impingement of the spinal cord.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY (21 CFR 807.92)
# Xspan™ Laminoplasty Fixation System January 15, 2016
#### l. SUBMITTER/MANUFACTURER:
X-spine Systems, Inc. 452 Alexandersville Rd. Miamisburg, OH 45342
Telephone (937) 847-8400 FAX (937) 847-8410
Establishment Registration Number: 3005031160
Mr. Kriss Anderson Official Contact: Director, Requlatory Affairs Email: kanderson@X-spine.com Telephone (937) 847-8400, ext. 137
#### ll. OWNER/OPERATOR:
Xtant Medical Inc. 604 Cruiser Lane Belgrade, MT 59714
Owner/Operator Number: 10028385
Official Correspondent:
Stephen Smith, Vice President Regulatory Assurance/ Quality Assurance Xtant Medical, Inc. Telephone (406) 388-0480
#### lll. DEVICE
Trade/Proprietary Name: Xspan™ Laminoplasty Fixation System Device Common Name: Orthosis, Spine, Plate, Laminoplasty, Metal Device Class: Class II Requlation Number: 21 CFR §888.3050 NQW - Orthosis, Spine, Plate, Laminoplasty, Product Code: Metal Orthopedic
Review Panel:
### X-Spine Systems, Inc.
{4}------------------------------------------------
### IV. PREDICATE DEVICES
- Aesculap Implant Systems (AIS): AIS Laminoplasty Plating System (K103284; . K090354) (Primary Predicate)
- o This predicate has not been subject to a design related recall.
- NuVasive Laminoplasty Fixation System (K091623) o
- o This predicate has not been subject to a design related recall.
### V. INTENDED USE
The Xspan™ Laminoplasty Fixation System is intended for use in the lower cervical and upper thoracic spine (C3-T3) after a laminoplasty has been performed. The Xspan™ Laminoplasty Fixation System holds or buttresses the allograft in place in order to prevent expulsion of the allograft or impingement of the spinal cord.
### VI. DEVICE DESCRIPTION
The Xspan™ Laminoplasty Fixation System is a posterior implant system comprised of laminoplasty plates, hinge plates, and mounting screws. Various forms and sizes of these implants are available so that adaptations can be made to take into account the pathology and anatomy of an individual patient. The system components are manufactured from Titanium based alloy in accordance with ASTM F136. The single use only implants are provided sterile [gamma radiation] and should not be reused under any circumstances.
The system also includes instrumentation to aid with implantation. The instruments are supplied in a tray which is used for instrument sterilization and storage.
### VII. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The technological principle for both the subject and predicate devices is posterior fixation in the lower cervical and upper thoracic spine (C3 to T3), used in laminoplasty procedures. At a high level, the subject device Xspan™ Laminoplasty Fixation System and the predicate devices AIS Laminoplasty Plating System (K103284; K090354) and NuVasive Laminoplasty Fixation System (K091623) are based on the following same or equivalent technological elements:
- FDA Product Code: NQW Orthosis, Spine, Plate, Laminoplasty, Metal. .
- Implants use the same material: Titanium alloy. ●
- Equivalent intended uses. ●
- Variety of implant sizes and forms providing a range of options to accommodate variations in patient anatomy.
- Same anatomical region. ●
- Same surgical approach. ●
- Equivalent mechanical performance. ●
{5}------------------------------------------------
The following difference exists between the subject device and the predicate devices:
- . Sterility - the Xspan™ implants are provided in sterile, single use packages. The predicate devices are provided non-sterile, and are sterilized by the user.
#### VIII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility Evaluation
The implants of the Xspan™ Laminoplasty Fixation System are made of titanium allov. The titanium alloy conforms to ASTM F136 - Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications. The titanium material has a long history of use as surgical implants and has been proven to be biocompatible, corrosion-resistant, and not toxic to the biologic environment. The comprehensive history of biocompatibility of the implant materials provides a rationale basis for not specifically performing additional testing for each additional biological effect.
# Mechanical Testing
Mechanical testing was performed with reference to ASTM F2193 Standard Specifications and Test Methods for Components Used in the Surgical Fixation of the Spinal Skeletal System and ASTM F382 Standard Specification and Test Method for Metallic Bone Plates. Performance data is provided in support of the substantial equivalence determination. The Xspan™ Laminoplasty Fixation System demonstrated equivalent performance to the predicate systems through static and dynamic 4-point bend testing.
#### IX. CONCLUSION
Based on a review of the information provided, X-spine finds that the Xspan™ Laminoplasty Fixation System is substantially equivalent to the referenced predicate device systems. The Xspan™ Laminoplasty Fixation System was found to have an equivalent safety and effectiveness profile compared to the referenced predicate systems.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.