SinuBand is indicated for use in patients undergoing nasal/sinus surgery as a space-occupying dressing intended to separate mucosal surfaces and prevent formation of adhesions.
Device Story
SinuBand is a sterile, single-use, bioresorbable two-layer film; applied by a physician under endoscopic visualization in a clinical setting following nasal/sinus surgery. The device acts as a mechanical/physical barrier to separate mucosal surfaces and prevent adhesions. It is hygroscopic, forming a gelatinous mass upon contact with fluid; it remains in the nasal cavity for approximately 2 weeks before being resorbed or removed via gentle irrigation. The device benefits patients by supporting the wound healing environment and reducing post-surgical complications.
Clinical Evidence
No human clinical data provided. Evidence includes biocompatibility testing per ISO 10993-1, comparative in vitro bench testing against MeroGel, and two pre-clinical in vivo studies in a New Zealand white rabbit model (n=16 implanted, n=10 control). Rabbit studies evaluated histology (hemorrhage, inflammation, fibrosis, bone remodeling) over 28 days, demonstrating device resorption within 9-15 days and no significant adverse histologic changes compared to controls.
Technological Characteristics
Bioresorbable two-layer film; hygroscopic; conforms to nasal anatomy. Resorption time approximately 14 days. Biocompatibility tested per ISO 10993-1. Sterile, single-use. No electronic components or software.
Indications for Use
Indicated for patients undergoing nasal/sinus surgery requiring a space-occupying dressing to separate mucosal surfaces and prevent adhesion formation.
Regulatory Classification
Identification
An intranasal splint is intended to minimize bleeding and edema and to prevent adhesions between the septum and the nasal cavity. It is placed in the nasal cavity after surgery or trauma. The intranasal splint is constructed from plastic, silicone, or absorbent material.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services seal, featuring an eagle. To the right of the seal is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 12, 2020
Bioinspire Technologies, Incorporated Margaret Dillon 2468 Embarcadero Way Palo Alto, California 94303-3313
Re: K160101
Trade/Device Name: SinuBand Regulation Number: 21 CFR 874.4780 Regulation Name: Intranasal splint Regulatory Class: Class I Product Code: LYA
Dear Contact:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination. Specifically, FDA is assigning a new product code to replace the one listed on the original SE letter. FDA has created a new product code to better categorize your device technology. The new product code is as follows:
| Product Code | Device Name |
|--------------|----------------------------------------|
| QJL | Splint, Intranasal Septal,<br>CBER led |
In addition, the procode QJL has been assigned to CBER. Like CDRH, CBER also has regulatory authority of certain medical device led combination products. Assignment to CBER will better align this product with other products that include blood or cellular products. For questions regarding this letter please contact Sheryl Lard-Whiteford, Office of The Director, Center for Biologics Evaluation and Research, Food and Drug Administration, 240-402-7912, sherry.lard@fda.hhs.gov.
> Sincerely, Marjorie G. Shulman -S Marjorie Shulman Assistant Director, 510(k), De Novo, 513(g), Device Determinations and Custom Devices Team, Office of Regulatory Programs Office of Product Evaluation and Ouality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 27, 2016
BioInspire Technologies, Inc. Ms. Margaret Dillon Vice President, Regulatory Affairs and Quality 2468 Embarcadero Wav Palo Alto, CA 94303-3313
Re: K160101
Trade/Device Name: SinuBand Regulation Number: 21 CFR 874.4780 Regulation Name: Intranasal Splint Regulatory Class: Class I Product Code: LYA Dated: June 23, 2016 Received: June 27, 2016
Dear Ms. Dillon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160101
Device Name SinuBand
Indications for Use (Describe)
SinuBand is indicated for use in patients undergoing nasal/sinus surgery as a space-occupying dressing intended to separate mucosal surfaces and prevent formation of adhesions.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------|---------------------------------------------|
| × Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for BioInspire Technologies. The word "BioInspire" is written in a large, green font. Above and to the right of "BioInspire" is the word "Technologies" in a smaller, gray font. The logo is simple and clean, and the use of green suggests a connection to nature or biology.
## K160101 - SinuBand 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title 21 CFR Part 807.92(c).
#### I.SUBMITTER
BioInspire Technologies, Incorporated 2468 Embarcadero Way Palo Alto, CA 94303 Phone: 707-239-5707 Fax: 707-239-5259 Contact Person: Margaret Dillon mdillon@bioinspiretechnologies.com Date Prepared: July 26, 2016
### II. DEVICE
| • Name of Device: | SinuBand |
|---------------------------|-------------------------------------|
| • Common or Usual Name: | Temporary Nasal Dressing |
| • Regulation Description: | Intranasal Splint (21 CFR 874.4780) |
| • Regulatory Class: | I |
- LYA ● Product Code:
### III. PREDICATE DEVICES
- MeroGel Nasal Dressing and Sinus Stent (K982731) ●
- AdvaCoat Gel and Stent (K070496)
- Synthemed Device (K082276)
- NasoPore Nasal Dressing (K052099)
### IV. DEVICE DESCRIPTION
SinuBand is a sterile, single-use, bioresorbable two-layer film designed to be a temporary nasal dressing applied under endoscopic visualization after sinus surgery. SinuBand is composed of two layers, with one side designed to be placed against the nasal tissue and the other facing the nasal cavity.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for BioInspire Technologies. The word "BioInspire" is written in a large, green font. Above and to the right of "BioInspire" is the word "Technologies" in a smaller, black font. The logo is simple and clean, with a focus on the company name.
#### V. INDICATIONS FOR USE
SinuBand is indicated for use in patients undergoing nasal/sinus surgery as a spaceoccupying dressing intended to separate mucosal surfaces and prevent formation of adhesions.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
The SinuBand placement and method of action is the same as the predicate devices, using manual application to place the device in the nasal cavity after sinus/nasal surgery. The SinuBand and predicate devices are hygroscopic, forming a gelatinous mass when in contact with fluid and serving as mechanical/physical barrier which prevents adhesions from forming. The SinuBand and predicate devices separate the mucosal surfaces and perform as an adjunct to the healing process.
The SinuBand and predicate devices are resorbable within 14 days of placement. The SinuBand and predicate devices can be removed from the nasal cavity after endoscopic placement by gentle irrigation. Comparative testing of SinuBand and MeroGel Nasal Dressing and Sinus Stent was performed to demonstrate substantial equivalence.
Any differences in properties between the SinuBand and the identified predicate devices do not alter the intended use of the SinuBand nor affect the safety and effectiveness of the SinuBand relative to the predicate devices.
#### VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
#### Biocompatibility Testing
A biocompatibility evaluation for the patient-contacting materials of SinuBand was conducted in accordance with the International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing within a Risk Management Process," as recognized by FDA. All test results were acceptable.
#### In Vitro Bench Testing
In vitro bench testing on SinuBand and the predicate device. MeroGel Nasal Dressing, evaluated mechanical and physical properties that are important to an effective nasal/sinus dressing, including: maximum sustained load, flexibility, ability to absorb water, pH, ease of placement and removal from tissue, and conformability. All test results were comparable for the two devices.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for BioInspire Technologies. The text "BioInspire" is in a large, green, sans-serif font. Above and to the right of "BioInspire" is the word "Technologies" in a smaller, dark gray, serif font. The logo is simple and clean, with a focus on the company name.
#### Animal Studies
Two (2) pre-clinical in vivo studies in a New Zealand white rabbit model of maxillary sinus surgery were completed as part of the evaluation of the SinuBand device. SinuBand was implanted in 16 intact or injured rabbit sinuses followed by histologic evaluation (hemorrhage, inflammation, fibrosis, bone remodeling, presence of device) of the sinus tissue for up to 28 days; 10 intact or injured rabbit sinuses that were not implanted served as controls. The results showed that the device rapidly adhered to the sinus mucosa and remained visible after 48 hours, but was no longer visible and appeared to be resorbed after 9-15 days. No significant histologic changes were identified when SinuBand was placed on intact mucosa (no injury), and the device did not alter inflammation, fibrosis or bone remodeling at injury sites. No complications were encountered during the study. The rabbit studies demonstrated the safety, biocompatibility and performance of SinuBand.
### VIII. CONCLUSION
Testing on SinuBand has demonstrated the desired properties of a nasal/sinus dressing, including lack of toxicity, biocompatibility, and non-dislocation after application. In addition, comparative bench testing has established substantial equivalence of SinuBand to the predicate device. MeroGel Nasal Dressing and Sinus Stent, in several tests critical to the function of an effective nasal/sinus dressing, including maximum sustained load, flexibility, ability to absorb water, pH to support a wound healing environment, ease of placement and removal, and ability to conform to nasal anatomy. SinuBand and MeroGel also have a comparable residence time in the sinus cavity of approximately 2 weeks. The nonclinical and bench testing data demonstrate that SinuBand is substantially equivalent to MeroGel and the other identified predicate devices that are currently marketed for the same intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.