K160091 · Medtronic, Inc. · FLL · May 20, 2016 · General Hospital
Device Facts
Record ID
K160091
Device Name
Temperature Probe, Affinity Temperature Probe
Applicant
Medtronic, Inc.
Product Code
FLL · General Hospital
Decision Date
May 20, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Indications for Use
The temperature probe is intended for use for continuous blood temperature monitoring as measured at a temperature monitoring adapter located within a Medtronic extracorporeal circulation device as specified in the device's instructions for use. The temperature probe is designed for use with a YSI tele-thermometer to monitor and display temperature.
Device Story
Temperature probe consists of a thermistor sensor housed in a stainless steel sheath, connected to a 3-meter cable with a 6.35-mm phono plug. Device operates by measuring electrical resistance changes in the thermistor, which are equated to temperature changes. Used in clinical settings during extracorporeal circulation; operated by healthcare professionals. Input is thermal energy from blood; output is an electrical signal processed by a YSI tele-thermometer to display temperature. Benefits include continuous, real-time blood temperature monitoring to assist clinical decision-making during perfusion procedures.
Clinical Evidence
Bench testing only. No clinical data required. Performance testing included IEC 60601 electrical safety and EMC, temperature range accuracy, precision/repeatability, response time, life testing, and liquid ingress/chemical exposure testing. All tests passed.
Indicated for continuous blood temperature monitoring in patients undergoing extracorporeal circulation using Medtronic devices.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized symbol that resembles three human profiles facing to the right, stacked on top of each other, with a wing-like shape extending above them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 20, 2016
Medtronic, Inc. Choua Thao Regulatory Affairs Specialist 7611 Northland Dr. Brooklyn Park, Minnesota 55428
Re: K160091
Trade/Device Name: Temperature Probe Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: April 19, 2016 Received: April 20, 2016
Dear Choua Thao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K160091
Device Name Temperature Probes
#### Indications for Use (Describe)
The temperature probe is intended for use for continuous blood temperature monitoring as measured at a temperature monitoring adapter located within a Medtronic extracorporeal circulation device as specified in the device's instructions for use. The temperature probe is designed for use with a YSI tele-thermometer to monitor and display temperature.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"> </span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Image /page/2/Picture/18 description: The image shows the text "KXXXXXXXX" on the first line and "1-42" on the second line. The text is in a simple, sans-serif font. The text is black and the background is white.
{3}------------------------------------------------
# 510(k) Summary K160091
| Date Prepared: | May 12, 2016 |
|-----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428 |
| | Establishment Registration Number: 2184009 |
| Contact Persons: | Choua Thao<br>Regulatory Affairs Specialist<br>Medtronic Perfusion Systems<br>Phone: 763.514.9842<br>Fax: 763.367.8360<br>Email: choua.thao@medtronic.com<br><br>Alternate Contact:<br>Susan C. Fidler<br>Senior Regulatory Affairs Manager<br>Medtronic Perfusion Systems<br>Phone: 763.514.9839<br>Fax: 763.367.8360<br>Email: susan.c.fidler@medtronic.com |
| Device Name and Classification<br>Trade Name: | Temperature Probe |
| Common Name: | Clinical Electronic Thermometer |
| Regulation Number: | 880.2910 |
| Product Code: | FLL |
| Product Classification: | Class II |
| Predicate Device<br>K100645 | Affinity Temperature Probe |
|-----------------------------|----------------------------|
| Reference Device<br>K831528 | Temperature Probe |
{4}------------------------------------------------
### Device Description
The temperature probe devices are intended for use with the temperature monitoring adapter of compatible Medtronic devices and the YSI™ tele-thermometer 1. The probe has a thermistor sensor housed in a stainless steel sheath connected to a 3-m (10-ft) cable, terminating with a 6.35-mm (1/4-in) phono plug.
The temperature probes meet the requirements of IEC 60601-1 when connected to a temperature monitor Class I or Class II device of type "CF."
## Indications for Use
The temperature probe is intended for use for continuous blood temperature monitoring as measured at a temperature monitoring adapter located within a Medtronic extracorporeal circulation device as specified in the device's instructions for use. The temperature probe is designed for use with a YSI tele-thermometer to monitor and display temperature.
## Comparison to Predicate and Reference Devices
A comparison of the proposed temperature probes to the currently marketed predicate devices (K831528 and K100645) indicates the following similarities:
- Same operating principle
- Same fundamental technological characteristics .
- o Substantially equivalent overall device design
- Substantially equivalent materials o
- o Same energy source
- Substantially equivalent intended use/indications .
The following technological differences exist between the subject and predicate devices. The subject devices contain the following:
| Differences | Impact Discussion |
|---------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Use of lead free solder | Testing demonstrated that the change in solder had no impact<br>on operation of the device. |
| Use of PVC for cable | Testing demonstrated that the change in cable material had no<br>impact on the operation of the device. |
| Use of RTV (room temperature<br>vulcanization silicone) | Testing demonstrated that the change in adhesive had no<br>impact on the operation of the device. |
| Use of additional insulation sleeve | Testing demonstrated that the additional insulation sleeve did<br>not negatively impact the operation of the device. |
<sup>1</sup> YSI is a trademark of YSI Incorporated.
{5}------------------------------------------------
The following is a comparison table of the proposed temperature probes to the currently marketed predicate devices. The temperature probe devices are substantially equivalent.
| | Reference I –<br>K831528 | Reference II –<br>K831528 | Predicate - K100645 | Current Submission | |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Element of<br>Comparison | Temperature Probe | Temperature Probe | Affinity Temperature<br>Probe | Temperature Probe | Substantially<br>Equivalent? |
| | FDA Clearance<br>30June1983 | FDA Clearance<br>30June1983 | FDA Clearance<br>14September2012 | | |
| Models | 1384 | TP | ATP210 | 1384, TP, ATP210 | |
| Indications for Use | Reusable temperature<br>probes are designed for<br>continuous temperature<br>measurement and<br>control with monitors<br>specifically designed for<br>use with the 400 Series<br>temperature. | The SciMed Model TP<br>Thermistor Probe is<br>intended to be used with<br>the SciMed Model TMA<br>Temperature Monitoring<br>Adapter and YSI Series<br>400 equipment. The<br>thermistor is housed in a<br>stainless steel probe, and<br>is connected to a ten foot<br>long shielded cable that is<br>terminated with a ¼ inch<br>phone plug. | The Affinity Temperature<br>Probe is intended for use<br>for continuous blood<br>temperature monitoring as<br>measured at a<br>temperature monitoring<br>adapter located within a<br>Medtronic extracorporeal<br>circulation device as<br>specified in the device's<br>Instructions for Use. The<br>Temperature Probe is<br>designed for use with a<br>YSI™ Tele-thermometer<br>to monitor and display<br>temperature. | The temperature probe is<br>intended for use for<br>continuous blood temperature<br>monitoring as measured at a<br>temperature monitoring<br>adapter located within a<br>Medtronic extracorporeal<br>circulation device as specified<br>in the device's instructions for<br>use. The temperature probe<br>is designed for use with a YSI<br>tele-thermometer to monitor<br>and display temperature. | Yes |
| Operating<br>Principle | A device that measures<br>differences in resistance<br>and equates that to<br>changes in temperature | A device that measures<br>differences in resistance<br>and equates that to<br>changes in temperature | A device that measures<br>differences in resistance<br>and equates that to<br>changes in temperature | A device that measures<br>differences in resistance and<br>equates that to changes in<br>temperature | Yes |
| Components | Reusable | Reusable | Reusable | Reusable | Yes |
| Element of<br>Comparison | Reference I –<br>K831528 | Reference II –<br>K831528 | Predicate - K100645 | Current Submission | Substantially<br>Equivalent? |
| | Temperature Probe | Temperature Probe | Affinity Temperature<br>Probe | Temperature Probe | |
| | FDA Clearance<br>30June1983 | FDA Clearance<br>30June1983 | FDA Clearance<br>14September2012 | | |
| Models | 1384 | TP | ATP210 | 1384, TP, ATP210 | |
| Sensor | Thermistor Sensor Tip | Thermistor Sensor Tip | Thermistor Sensor Tip | Thermistor Sensor Tip | Yes |
| Materials | • Thermistor<br>• Sensor Tip -<br>Stainless Steel,<br>Brass, Ceramic<br>• Cable - TPE (C-<br>Flex) and Copper<br>• Phono Plug -<br>Brass | • Thermistor Sensor<br>Tip - Stainless Steel,<br>Brass, Ceramic<br>• Cable - TPE (C-<br>Flex) and Copper<br>• Phono Plug - Brass | • Probe tip - Stainless<br>Steel, Brass,<br>Ceramic<br>• Cable - TPE (C-<br>Flex) and Copper<br>• Phono Plug - Brass | • Probe tip –<br>Stainless Steel and Brass<br>• Cable - PVC and Copper<br>• Phono Plug - Brass | Yes |
| Thermistor | Negative<br>Temperature<br>Coefficient (NTC)<br>Thermistor | Negative Temperature<br>Coefficient (NTC)<br>Thermistor | Negative Temperature<br>Coefficient (NTC)<br>Thermistor | Negative Temperature<br>Coefficient (NTC) Thermistor | Yes |
| Cable | PVC Insulated 2<br>conductor twisted<br>pair2 | PVC Insulated 2<br>conductor twisted pair | PVC Insulated 2<br>conductor twisted pair | PVC Insulated 2 conductor<br>twisted pair | Yes |
| Phono Plug | 6.35-mm (1/4-in)<br>phono plug | 6.35-mm (1/4-in) phono<br>plug | 6.35-mm (1/4-in) phono<br>plug | 6.35-mm (1/4-in) phono plug | Yes |
Comparison Table – Indications for Use and Physical Specification
{6}------------------------------------------------
² Two 24 gauge wires twisted around each other throughout the length of the cable.
{7}------------------------------------------------
## Summary of Performance Data
Bench testing was used to demonstrate the performance characteristics of the temperature probe devices. Clinical testing was not required to establish substantial equivalence.
The following performance tests were conducted:
| Test | Description | Result |
|----------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------|
| IEC 60601 | Ensures the temperature probes meet the requirements of IEC 60601 electrical safety and Electromagnetic compatibility (EMC). | Pass |
| Temperature Range | Ensures the temperature probes meet the requirements of the system at the extreme temperature ranges. | Pass |
| Ambient Temperature Environment | Ensures the temperature probes meet the requirements of the system at the normal temperature ranges. | Pass |
| Accuracy | Accuracy over the entire temperature range specified for the device. | Pass |
| Precision and Repeatability | Precision and repeatability of measurements over the temperature range specified with the effects of air currents, over the entire temperature range specified. | Pass |
| Time | Indicate the time required for the device to obtain a steady state reading. | Pass |
| Life Testing | Ensures the temperature probes meet the requirements of the expected life of the probe. | Pass |
| Liquid Ingress and Chemical Exposure Testing | Ensures the temperature probes meet the Ingress Protection and cleaning requirements. | Pass |
#### Conclusion
Medtronic has determined that the temperature probe devices described in this submission has shown to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.