The proposed device is intended for use as the disposable, conductive adhesive interface between the patient's skin and the electrical stimulator to apply electrical stimulation current, and is intended to be used with marketed electrical stimulators, i.e. TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). It's for OTC use.
Device Story
TENS Electrodes function as disposable, conductive adhesive interfaces between patient skin and external electrical stimulators (TENS/EMS). Device consists of electrode pad and wire; available in various shapes (rectangular, round, butterfly). Applied by patient (OTC use) to skin; adhesive hydrogel secures device without external materials. Device conducts electrical current from stimulator to patient tissue. Benefits include facilitating TENS/EMS therapy for pain management or muscle stimulation. No active processing or electronic components; passive conductive interface.
Clinical Evidence
Bench testing only. Evaluated via electrical impedance, adhesive performance, tensile strength, conformability, impedance distribution uniformity, biocompatibility (ISO 10993), stability, and reuse testing. Safety of lead wires/cables verified per IEC 60601-1 and IEC 60601-2-10.
Technological Characteristics
Non-sterile, flexible laminated structure. Materials: non-woven fabric, adhesive, PVC-coated wire/terminal, hydrogel, PET fabric, carbon fiber, reinforcing fiber, poly-isobutylene, carbon black, silicon-coated PET liner. Passive conductive interface. Dimensions vary by model. Conforms to IEC 60601-1 and IEC 60601-2-10.
Indications for Use
Indicated for use as a disposable, conductive adhesive interface between patient skin and electrical stimulators (TENS/EMS) for OTC application of electrical stimulation current. No specific patient population, age, or gender restrictions.
Regulatory Classification
Identification
A cutaneous electrode is an electrode that is applied directly to a patient's skin either to record physiological signals (e.g., the electroencephalogram) or to apply electrical stimulation.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 19, 2016
Cathay Manufacturing Corp. % Mr. Fan, Tao Consultant Shanghai YunTao Management Consultant Room 202, No. 14, Lane 986, Gaosi Road Pudong District Shanghai, 201203 China
Re: K160081
Trade/Device Name: TENS Electrodes Regulation Number: 21 CFR 882.1320 Regulation Name: Cutaneous Electrode Regulatory Class: Class II Product Code: GXY Dated: September 3, 2016 Received: September 19, 2016
Dear Mr. Fan. Tao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807): labeling (21 CFR Part 801 and Part 809): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801and Part 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michaelリ. Hoffmann -A
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160081
Device Name TENS Electrodes
Indications for Use (Describe)
The proposed device is intended for use as the disposable, conductive interface between the patient's skin and the electrical stimulator to apply electrical stimulation current, and is intended to be used with marketed electrical stimulators, i.e. TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation). It's for OTC use.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitter: | CATHAY MANUFACTURING CORP.<br>No. 328, Xishe Road, Maogang Town, Songjiang Area<br>Shanghai, CN 201607 |
|----------------------|--------------------------------------------------------------------------------------------------------|
| Contact: | Mr. Fan, Tao<br>Phone: (86)18616806984<br>Fax: (86)-21-57876338<br>fant01@uwellmed.com |
| Date Prepared: | Oct. 18, 2016 |
| Device Trade Name: | TENS Electrodes |
| Device Common Name: | Neurostimulation Electrodes |
| Classification name: | Cutaneous Electrodes |
| Class: | II |
| Regulation: | 882.1320 |
| Product Code: | GXY |
# Primary Predicate Device:
| Device | Company | Product Code | 510(k) Number |
|-----------------------------|---------------------------------------------------|--------------|---------------|
| SELF-ADHESIVE<br>ELECTRODES | CATHAY Healthcare<br>Equipment Manufacturing Inc. | GXY | K080276 |
# A Reference Device Chosen:
| Top-Rank Adhesive | Top-Rank Health Care Co., | GXY | K132588 |
|-------------------|---------------------------|-----|---------|
| Electrodes | Ltd | | |
{4}------------------------------------------------
#### 1. Indication for Use:
The proposed device is intended for use as the disposable, conductive adhesive interface between the patient's skin and the electrical stimulator to apply electrical stimulation current, and is intended to be used with marketed electrical stimulators. i.e. TENS (Transcutaneous Electrical Nerve Stimulation) and EMS (Electrical Muscular Stimulation).
It's for OTC use.
## 2. Product Description:
The proposed TENS Electrodes are non-sterile laminated, flexible structures composed of materials commonly used in this application:
The electrodes are designed for single-patient application use. Because of the adhesive nature of the biocompatible hydrogel, no securing materials are required to secure the device to the patient's skin.
The proposed device mainly consists of electrode pad and wire. The electrode pad is available in rectangular shape, round shape, and butterfly shape, etc.
| Shape | Product Model | Shape | Product Model |
|--------------|-------------------------|--------------|-----------------------|
| Image: Shape | ☐ CM2040FC ☐ CM5570FC | Image: Shape | ☐ CM2222FC ☐ CM6060FC |
| | ☐ CM3050FC ☐ CM5575FC | | ☐ CM3232FC ☐ CM6565FC |
| | ☐ CM3060FC ☐ CM5590FC | | ☐ CM4040FC ☐ CM7070FC |
| | ☐ CM4050FC ☐ CM6070FC | | ☐ CM4545FC ☐ CM7575FC |
| | ☐ CM4060FC ☐ CM6080FC | | ☐ CM4848FC ☐ CM8080FC |
| | ☐ CM4070FC ☐ CM6090FC | | ☐ CM5050FC ☐ CM8585FC |
| | ☐ CM4080FC ☐ CM60100FC | | ☐ CM5555FC ☐ CM9090RC |
| | ☐ CM4090FC ☐ CM7080FC | Image: Shape | ☐ CM22RC ☐ CM60RC |
| | ☐ CM40100FC ☐ CM70100FC | | ☐ CM25RC ☐ CM65RC |
| | ☐ CM4580FC ☐ CM70120FC | | ☐ CM32RC ☐ CM70RC |
| | ☐ CM4590FC ☐ CM75125FC | | ☐ CM40RC ☐ CM75RC |
| | ☐ CM45100FC ☐ CM80100FC | | ☐ CM45RC ☐ CM85RC |
| | ☐ CM5060FC ☐ CM80110FC | | ☐ CM50RC ☐ CM90RC |
| | ☐ CM5070FC ☐ CM80120FC | Image: Shape | ☐ CM55RC |
| | ☐ CM5090FC ☐ CM80130FC | | ☐ CM4050TC ☐ CM5580SC |
| | ☐ CM50100FC ☐ CM90120FC | Image: Shape | ☐ CM4560TC ☐ CM7090TC |
| | ☐ CM50130FC ☐ CM90130FC | | ☐ CM4570SC |
| | ☐ CM5565FC ☐ CM100180FC | | ☐ CM2542YC |
| | | | |
wire: _ 0 2. Omn Packaging : 22PCS 04PCS 6PCS 88PCS
{5}------------------------------------------------
| ELEMENT<br>OF<br>COMPARISON | SUBJECT DEVICE | Primary Predicate Device | A Reference Device chosen | |
|-----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | K080276 | K132588 | |
| | | SELF-ADHESIVE<br>ELECTRODES | Top-Rank Adhesive Electrodes | |
| Indications for<br>Use | The proposed device is<br>intended for use as the<br>disposable, conductive<br>adhesive interface between<br>the patient's skin and the<br>electrical stimulator to apply<br>electrical stimulation<br>current, and is intended to<br>be used with marketed<br>electrical stimulators, i.e.<br>TENS (Transcutaneous<br>Electrical Nerve<br>Stimulation) and EMS<br>(Electrical Muscular<br>Stimulation).<br>It's for OTC use | The electrotherapy<br>electrodes are<br>intended to be used to<br>apply electrical<br>stimulation current to<br>the patient's skin or<br>record physiological<br>signals | Adhesive electrodes are<br>intended for use as the<br>disposable, conductive<br>adhesive interface between<br>the patient's skin and the<br>electrical stimulator. Top-Rank<br>Adhesive Electrodes are<br>intended to be used with<br>marketed Electrical<br>Stimulators, i.e. TENS<br>(Transcutaneous Electrical<br>Nerve Stimulation) and<br>EMS(Electrical Muscular<br>Stimulation). It's for OTC use. | |
| OTC or<br>Prescription | For OTC Use | For Prescription use | For OTC use | |
| Sterility | Non-sterile | Non-sterile | Non-sterile | |
| Design Features | Substrate / Wire/ Hydro-Gel | Substrate / Wire/ Hydro-Gel | Substrate /Wire/Hydro-Gel | |
| | Scrim / Conductive Fiber | Scrim / Conductive Fiber | Scrim/Conductive Fiber | |
| | Carbon Conductive Film / Liner | Carbon Conductive Film / Liner | Carbon Conductive Film / Liner | |
| Materials:<br>Substrate;<br>Wire:<br>Hydro-gel;<br>Scrim;<br>Conductive<br>fiber;<br>Carbon<br>conductive<br>film;<br>Liner | Non-woven fabric+ adhesive | Non-woven fabric+ adhesive | Non-woven fabric+ adhesive | |
| | Wire & Terminal coated PVC | Wire & Terminal coated PVC | Wire & Terminal coated PVC | |
| | Hydro-gel | Hydro-gel | Hydro-gel | |
| | PET fabric | PET fabric | PET fabric | |
| | Carbon fiber + reinforcing<br>fiber | Carbon fiber + reinforcing fiber | Carbon fiber + reinforcing fiber | |
| | Poly-isobutylene, Carbon<br>Black | Poly-isobutylene, Carbon Black | Poly-isobutylene, Carbon Black | |
| | Transparent PET liner<br>coated silicon | Transparent PET liner coated<br>silicon | Transparent PET liner coated silicon | |
| Biocompatibility | Conforms to<br>ISO10993 | Conforms to<br>ISO10993 | Conforms to<br>ISO10993 | |
| Labeling | confirm to 21 CFR Part 801 | confirm to 21 CFR Part 801 | confirm to 21 CFR Part 801 | |
| Performance | Conforms to IEC 60601-1<br>and<br>IEC 60601-2-10:2012,<br>Part 2 Particular<br>requirements for the basic<br>safety and essential<br>performance of nerve and<br>muscle stimulators. | Conforms to IEC 60601-1<br>and<br>IEC 60601-2-10:2012,<br>Part 2 Particular<br>requirements for the basic<br>safety and essential<br>performance of nerve and<br>muscle stimulators. | Conforms to IEC 60601-1<br>and<br>IEC 60601-2-10:2012,<br>Part 2 Particular requirements for the<br>basic safety and essential<br>performance of nerve and muscle<br>stimulators. | |
| Electrical<br>impedance | Individual pad impedance<br>below 200 ohms @ 60<br>kHz | By linking the connector to<br>gel using a multi-meter. Set<br>the range at x1k ohm.<br>Pass if the needle of multi-<br>meter moves | By linking the connector to gel using<br>a multi-meter. Set the range at x1k<br>ohm.<br>Pass if the needle of multi-meter<br>moves | |
| | ELEMENT<br>OF<br>COMPARISON | SUBJECT DEVICE | Primary Predicate Device | A Reference Device chosen |
| | | | K080276 | K132588 |
| | | | SELF-ADHESIVE<br>ELECTRODES | Top-Rank Adhesive Electrodes |
| | Electrode types,<br>shapes, and<br>sizes used for<br>testing | Largest model:<br>CM100180FC (Rectangle<br>shape: 100 x 180 mm);<br>Smallest model:<br>CM2222FC(square<br>shape: 22 x 22 mm);<br>Irregular representative:<br>CM2542YC(crescent ) | Largest and irregular model:<br>CM100200 (butterfly shape:<br>100 x 200 mm)<br>Smallest model:<br>CM2222 (Square shape: 22<br>x 22 mm) | A most common used sample:<br>AF5050WP(Square Shape: 50 x50<br>mm) |
| To | Adhesive<br>performance on | >= No.14 steel ball<br>(Diameter: 11.112 mm)<br>following AS07-003 /<br>GB/T4852-2002 | >= No.14 steel ball<br>(Diameter: 11.112 mm)<br>following AS07-003 /<br>GB/T4852-2002 | >= No.14 steel ball (Diameter:<br>11.112 mm) following AS07-003 /<br>GB/T4852-2002 |
| | Adhesive<br>holding strength<br>performance | A sample pad with 25mm<br>length and 25 mm width;<br>Hang a weight of 200g<br>without detach within 30s | A sample pad with 25mm<br>length and 25 mm width;<br>Hang a weight of 200g<br>without detach within 30s | A sample pad with 25mm length and<br>25 mm width:<br>Hang a weight of 200g without<br>detach within 30s |
| | Tensile Strength | Electrode is held up with a<br>weight of 500g hung on<br>the wire, and can hold on<br>at least 1 minutes | Electrode is held up with a<br>weight of 500g hung on the<br>wire, and can hold on at least<br>1 minutes | Electrode is held up with a weight of<br>500g hung on the wire, and can hold<br>on at least 1 minutes |
| Test | Conformability | No more than 10% of the<br>adhesive area of the<br>device shall have<br>separated from the skin<br>surface at 1 h after<br>application. | No more than 10% of the<br>adhesive area of the device<br>shall have separated from<br>the skin surface at 1 h after<br>application. | No more than 10% of the adhesive<br>area of the device shall have<br>separated from the skin surface at 1<br>h after application. |
| | Impedance<br>Distribution<br>uniformity test | 1) No significant deviation<br>between resistance<br>values measured for each<br>combination mode.<br>2) The maximum value<br>and the minimum value<br>should be within ±10% of<br>average value | 1) No significant deviation<br>between resistance values<br>measured for each<br>combination mode.<br>2) The maximum value and<br>the minimum value should be<br>within ±10% of average<br>value | 1) No significant deviation between<br>resistance values measured for each<br>combination mode.<br>2) The maximum value and the<br>minimum value should be within<br>±10% of average value |
| | Stability and<br>shelf life | 2 year | 2 year | 2 year |
| | Re-usable | For single patient | For single patient | For single patient |
| | Intended<br>population | No specified population | No specified population | No specified population |
# 4. Summary of Technology Characteristics and Performance Testing (Comparison to the Primary Predicate Device and Reference Device)
{6}------------------------------------------------
{7}------------------------------------------------
| Comparison<br>Summary | Compare to primary predicate device specified in K080276, our device and the<br>predicate device are same in Essential Components, raw materials, physical<br>features, and same manufacturing processes. One difference is that our proposed<br>device does not claim for recording physiological signals as the predicate device<br>does. Another difference is predicate device is intended for prescription use, and our<br>proposed product is for Over-the-Counter (OTC) use.<br>Our product and the predicate device are the same in intended for use, essential<br>components, physical and general safety features based on ISO international<br>standards. The hydro-gel used in K080276 is not different from what is used in<br>CATHAY TENS Electrodes.<br>Compare to both primary predicate device and reference device, the proposed<br>device has different test method for electrical impedance. However, this test method<br>is a more sensitive & improved test method. Previously, proposed device test<br>method for electrical impedance was exactly the same as predicate device and<br>reference device, and all tests can pass previous test method also. |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Compare to the primary predicate device and reference device, the subject device<br>is constructed in the same manner,<br>is constructed of same or similar materials,<br>have same or similar indications for use, and<br>have similar performance characteristics<br>Any minor visual, dimensional or labeling differences between the primary<br>predicate device and reference device and the proposed electrodes do not pose risk<br>to their performance and usage. |
## 5. Bench Testing (non-clinical):
Compare to primary predicate product specified in K080276, our device and the predicate device are same in Essential Components, raw materials, physical features, and same manufacturing processes. The biocompatibility performance equivalence evidence of proposed electrode can be demonstrated.
The proposed TENS Electrodes evaluated through Electrode Performance testing (Electrical Impedance Performance, Adhesive Performance, and Tensile Strenght), Biocompatibility testing, Stability testing, and Reuse testing.
The comparison tests conclusively prove that the proposed electrode is similar / equivalence as performance efficient at electrical impedance, adhesive performances, tensile strength, conformability, and impedance distribution uniformity performance as the primary predicate device and reference device.
For lead wires & cables used in proposed device, the safety performances are demonstrated by the Third party, TUV Rheiland, through testing following IEC
{8}------------------------------------------------
60601-1:2005+CORR. 1(2006) + CORR.2(2007) + AM1(2012), Part 1General requirements for basic safety and essential performance, and IEC 60601-2-10:2012, Part 2 Particular requirements for the basic safety and essential performance of nerve and muscle stimulators.
The results of bench testing & type testing provide reasonable assurance that the proposed device has been designed and validated to assure conformance to the requirements for its indication for use.
## 6. Conclusion:
Based on the indication for use, technology characteristics, and performance testing, the proposed product, TENS Electrodes, has been shown to be appropriate for its indication for use and is considered to be substantially equivalent to the primary predicate device and reference device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.