StageOne Disposable Cement Spacer Molds for Temporary Hemi-Shoulder Prosthesis
K160071 · Biomet, Inc. · MBB · Jun 16, 2016 · Orthopedic
Device Facts
Record ID
K160071
Device Name
StageOne Disposable Cement Spacer Molds for Temporary Hemi-Shoulder Prosthesis
Applicant
Biomet, Inc.
Product Code
MBB · Orthopedic
Decision Date
Jun 16, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3027
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Disposable cement spacer molds are indicated for use to mold a temporary hemi-shoulder replacement for skeletally mature patients undergoing a two-stage revision procedure due to a septic process. The temporary prosthesis is molded using Palacos® R+G Bone Cement or Refobacin® Bone Cement R and inserted into the humeral medullary canal and glenoidal cavity following removal of the existing total shoulder replacement implants and debridement. The device is intended for use in conjunction with systemic antibiotic therapy (standard treatment approach to an infection). The hemi-shoulder prosthesis made from the StageOne™ disposable cement molds is not intended for use more than 180 days, at which time it must be explanted and permanent devices implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion, etc.) Due to the inherent mechanical limitations of the hemi-shoulder prosthesis material (Palacos® R-G Bone Cement and Refobacin® Bone Cement R), the temporary hemi-shoulder prosthesis is only indicated for patients who will consistently follow activity limitations throughout the implant period.
Device Story
Device consists of sterile, single-use medical-grade silicone molds. Clinician fills mold with Palacos® R+G or Refobacin® Bone Cement R using a dispenser/gun. After curing, the molded temporary hemi-shoulder prosthesis is removed from the mold and implanted into the humeral medullary canal and glenoidal cavity following debridement of infected total shoulder implants. Used in clinical/surgical settings during the first stage of a two-stage revision procedure. Device remains in situ for up to 180 days to maintain joint space and deliver local antibiotic therapy while systemic treatment addresses infection. Benefits include temporary joint stabilization and infection management prior to permanent reconstruction.
Clinical Evidence
No clinical data was required to establish substantial equivalence. Evidence is based on non-clinical bench testing, including mechanical performance and gentamicin elution testing comparing the subject device to the predicate.
Technological Characteristics
Single-use, sterile silicone molds. Dimensions and form factor unchanged from predicate. Energy source: None (manual). Connectivity: None. Sterilization: Unchanged. Materials: Medical-grade silicone.
Indications for Use
Indicated for skeletally mature patients undergoing two-stage revision shoulder arthroplasty due to septic process. Contraindicated for patients unable to follow activity limitations during the 180-day maximum implant period.
Regulatory Classification
Identification
Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”
Predicate Devices
StageOne™ Disposable Cement Spacer Molds for Temporary Hemi-Shoulder Prosthesis (K131135)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing to the right, arranged in a way that they appear to be interconnected or overlapping. The profiles are rendered in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile graphic.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 16, 2016
Biomet, Incorporated Ms. Heidi Busz Global Regulatory Affairs Associate 56 East Bell Drive Warsaw, Indiana 46581
Re: K160071
Trade/Device Name: StageOne™ Disposable Cement Spacer Molds for Temporary Hemi-Shoulder Prosthesis Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: MBB, HSD, KWS Dated: April 15, 2016 Received: April 18, 2016
Dear Ms. Busz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K160071
#### Device Name
StageOne™ Disposable Cement Spacer Molds for Temporary Hemi-Shoulder Prosthesis
#### Indications for Use (Describe)
Disposable cement spacer molds are indicated for use to mold a temporary hemi-shoulder replacement for skeletally mature patients undergoing a two-stage revision procedure due to a septic process. The temporary prosthesis is molded using Palacos® R+G Bone Cement or Refobacin® Bone Cement R and inserted into the humeral medullary canal and glenoidal cavity following removal of the existing total shoulder replacement implants and debridement. The device is intended for use in conjunction with systemic antibiotic therapy (standard treatment approach to an infection).
The hemi-shoulder prosthesis made from the StageOne™ disposable cement molds is not intended for use more than 180 days, at which time it must be explanted and permanent devices implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion, etc.)
Due to the inherent mechanical limitations of the hemi-shoulder prosthesis material (Palacos® R-G Bone Cement and Refobacin® Bone Cement R), the temporary hemi-shoulder prosthesis is only indicated for patients who will consistently follow activity limitations throughout the implant period.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;">\[X]</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: sans-serif;">[]</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "BIOMET" in a bold, sans-serif font. The letters are all capitalized and evenly spaced. The color of the text is a dark gray, contrasting with the white background. The overall impression is clean and professional.
# 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the StageOne™ Shoulder Spacer Molds 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005
| Sponsor: | Biomet Inc.<br>56 East Bell Drive<br>PO Box 587<br>Warsaw, IN 46581<br>Establishment Registration Number: 1825034 |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Heidi Busz<br>Regulatory Affairs Associate |
| Phone: | 574-372-4249 |
| Fax: | 574-372-4710 |
| Date: | May 31, 2016 |
| Subject Device: | Trade Name: StageOne™ Disposable Cement Spacer Molds for<br>Temporary Hemi-Shoulder Prosthesis<br>Common Name: Bone Cement Shoulder Spacer Mold; Disposable<br>Cement Spacer Molds for Temporary Hemi-Shoulder Prosthesis;<br>StageOne™ Shoulder Spacer Mold |
| Classification Name: | MBB-Polymethylmethacrylate (PMMA) bone cement (21 CFR 888.3027) HSD- Shoulder joint humeral (hemi-shoulder) metallic uncemented prosthesis (21 CFR 888.3690) KWS-Shoulder joint metal/polymer semi-constrained cemented prosthesis (21 CFR 888.3660) |
# Legally marketed devices to which substantial equivalence is claimed:
- K131135-StageOne™ Disposable Cement Spacer Molds for Temporary Hemi-Shoulder ● Prosthesis
- K112983-Tecres S.p.A. InterSpace® Shoulder
# Device Description
The single-use cement spacer molds are sterile disposables made of medical grade silicone. They are intended to be filled with Palacos® R+G Bone Cement or Refobacin® Bone Cement R by injecting with a dispenser/gun into the mold. After the cement cures, the hemi-shoulder prosthesis is to be removed from the mold and placed into the joint space. The hemi-shoulder prosthesis remains in place (180 days or less) until the second stage of the two-stage procedure is performed to implant a conventional shoulder joint prosthesis.
# Indications for Use
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "BIOMET" in a sans-serif font. The letters are all capitalized and appear to be a dark gray color. The letters are evenly spaced and the word is centered in the image. The overall impression is clean and professional.
Disposable cement spacer molds are indicated for use to mold a temporary hemi-shoulder replacement for skeletally mature patients undergoing a two-stage revision procedure due to a septic process. The temporary prosthesis is molded using Palacos® R+G Bone Cement or Refobacin® Bone Cement R and inserted into the humeral medullary canal and glenoidal cavity following removal of the existing total shoulder replacement implants and debridement. The device is intended for use in conjunction with systemic antimicrobial antibiotic therapy (standard treatment approach to an infection).
The hemi-shoulder prosthesis made from the StageOne™ disposable cement molds is not intended for use more than 180 days, at which time it must be explanted and permanent devices implanted or another appropriate treatment performed (e.g. resection arthroplasty, fusion, etc.)
Due to the inherent mechanical limitations of the hemi-shoulder prosthesis material (Palacos® R+G Bone Cement and Refobacin® Bone Cement R), the temporary hemi-shoulder prosthesis is only indicated for patients who will consistently follow activity limitations throughout the implant period.
# Summary of Technological Characteristics
The subject modifications include the removal Cobalt™ G-HV Bone Cement from the Instructions for Use and the addition of Palacos® R+G Bone Cement and Refobacin® Bone Cement R for use with the subject device.
The subject modification does not result in any changes to the StageOne™ Shoulder Spacer Mold design features, materials, sizes, sterilization method, manufacturing process, sterility assurance level, or shelf life of the StageOne™ Shoulder Spacer Molds.
### Summary of Performance Data (Nonclinical and/or Clinical)
- . Non-Clinical Tests
- Mechanical performance testing and Gentamicin Elution testing of StageOne™ o Shoulder Spacers fabricated with Palacos® R+G Bone Cement and Refobacin® Bone Cement R were compared to that of the predicate. The results demonstrated that StageOne™ Shoulder Spacers fabricated with Palacos® R+G Bone Cement and Refobacin Bone Cement R possess mechanical and elution characteristics comparable to those of the predicate device. Design validation test results confirm that the StageOne™ Shoulder Spacer Molds utilized with Palacos® R+G Bone Cement and Refobacin® Bone Cement R meet design requirements for use.
- Clinical Tests
- o Clinical data was not required to establish substantial equivalence between the subject StageOne™ Shoulder Spacer Molds and the predicate device.
### Substantial Equivalence Conclusion
Based on the similarities in design, function, indications for use and fundamental scientific technology, the devices that are the subject of this submission are similar to the predicate devices and do not introduce any new risks of safety or efficacy. Therefore, Biomet concludes that the subject devices are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.