← Product Code [QYX](/productcode/QYX) · K160064

# MoMe Kardia Wireless Ambulatory ECG Monitoring and Detection System (K160064)

_Infobionic, Inc. · QYX · Mar 11, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K160064

## Device Facts

- **Applicant:** Infobionic, Inc.
- **Product Code:** [QYX](/productcode/QYX.md)
- **Decision Date:** Mar 11, 2016
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

MoMe® Kardia is intended to be used for: 1. Patients who experience transient symptoms that may suggest cardiac arrhythmia. 2. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation) 3. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath) 4. Patients recovering from cardiac surgery or interventional procedures who are indicated for outpatient arrhythmia monitoring. 5. ECG data recorded by the device can be analyzed by other processing systems to provide Holter style reports.

## Device Story

MoMe Kardia is a wearable, single-unit ambulatory ECG monitor; acquires ECG and accelerometer data; transmits data via cellular technology to a web-based remote server (MoMe Software System). System uses proprietary algorithms to analyze data; flags events for physician review. Device supports Holter, Event, and MCT monitoring modes. Used in physician practices, clinics, and research institutions; operated by healthcare providers. Patient intervention not required for capture, but device includes a patient event trigger. Output allows physicians to review ECG data and arrhythmia detections to aid in diagnosis and clinical decision-making; benefits patients by enabling remote monitoring of cardiac symptoms.

## Clinical Evidence

Bench testing only. The device was tested for compliance with IEC 60601-1, IEC 60601-1-2, and ANSI/AAMI/IEC 60601-2-47. No clinical trial data was provided.

## Technological Characteristics

Single-unit body-worn device; 3 electrodes; 2-channel ECG acquisition; accelerometer for motion data. Connectivity: cellular modem for data transmission to cloud-based server. Power: Li-ion rechargeable battery. Software: proprietary server-side algorithm for arrhythmia detection. Standards: IEC 60601-1, IEC 60601-1-2, ANSI/AAMI/IEC 60601-2-47.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- TruVue™ Wireless Ambulatory ECG Monitoring System ([K100155](/device/K100155.md))

## Reference Devices

- MoMe® Software System ([K152491](/device/K152491.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

November 15, 2023

Infobionic, Inc. Carrie Neuberger Regulatory Consultant To Infobionic Inc. 600 Suffolk Street Lowell, Massachusetts 01854

Re: K160064

Trade/Device Name: MoMe Kardia Wireless Ambulatory ECG Monitoring and Detection System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: QYX, DSI

Dear Carrie Neuberger:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated March 11, 2016. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Kozen, OHT2: Office of Cardiovascular Devices, 301-796-5813, jennifer.kozen@fda.hhs.gov.

Sincerely,

Jennifer W. Shih -S

Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Public Health Service

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol is a depiction of an abstract human figure embracing or protecting another figure, representing the department's mission of protecting the health of all Americans and providing essential human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 11, 2016

Infobionic, Inc. Ms. Carrie Neuberger Regulatory Consultant to Infobionic Inc. 600 Suffolk Street Lowell, Massachusetts 01854

Re: K160064 Trade/Device Name: MoMe Kardia Wireless Ambulatory ECG Monitoring And Detection System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: DSI Dated: January 12, 2016 Received: January 13, 2016

Dear Ms. Carrie Neuberger,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{2}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours

FDA

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

#### 510(k) Number (if known) K160064

Device Name

MoMe® Kardia Wireless Ambulatory ECG Monitoring and Detection System

Indications for Use (Describe) MoMe® Kardia is intended to be used for:

1. Patients who experience transient symptoms that may suggest cardiac arrhythmia.

2. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation)

3. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath)

4. Patients recovering from cardiac surgery or interventional procedures who are indicated for outpatient arrhythmia monitoring.

5. ECG data recorded by the device can be analyzed by other processing systems to provide Holter style reports.

MoMe® Kardia is contraindicated for:

1. MoMe® Kardia is contraindicated for those patients requiring attended, in-hospital monitoring for life threatening arrhythmias.

Note: MoMe® Kardia does not provide interpretive statements. Interpretation and diagnosis is the responsibility of a physician.

Type of Use (Select one or both, as applicable)

| <span></span> | <span></span>                                              |
|---------------|------------------------------------------------------------|
| <div></div>   | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span></span> | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Infobionic. The logo consists of the word "INFOBIONIC" in a combination of blue and green colors. The "INFO" part of the word is in blue, while the "BIONIC" part is in green. There is a blue heart symbol in the middle of the word. To the left of the word, there is a cluster of blue and green circles arranged in a semi-circular pattern.

MoMe® Kardia

# K160064 Page 1 of 5

# Section 5 510(K) SUMMARY COMPLYING WITH 21 CFR 807.92

| Date Prepared:       | January 12, 2015                                                                                  |
|----------------------|---------------------------------------------------------------------------------------------------|
| Submitter Name:      | InfoBionic, Inc.                                                                                  |
| Submitter Address:   | 600 Suffolk Street, Lowell, MA, 01854                                                             |
| Contact Person:      | Carrie Neuberger, Regulatory Consultant                                                           |
| Phone Number:        | 415.640.3377                                                                                      |
| Device Trade Name:   | MoMe® Kardia Wireless Ambulatory ECG Monitoring<br>and Detection System                           |
| Device Common Name:  | Continuous ECG monitor and Arrhythmia Detection                                                   |
| Classification Name: | Arrhythmia Detector and alarm, 21 CFR 870.1025,<br>DSI                                            |
| Predicate Device:    | K100155, Biomedical Systems Corporation's<br>TruVue™ Wireless Ambulatory ECG Monitoring<br>System |

## Device Description:

MoMe® Kardia is a wireless, remote monitoring system designed to aid physicians in their diagnosis of cardiac arrhythmias in patients with a demonstrated need for cardiac monitoring.

MoMe® Kardia includes the wearable MoMe® Kardia Device that acquires and stores ECG and motion (accelerometer) data and transmits that data via cellular technology to the MoMe® Software System (K152491), a web-based remote server software with proprietary algorithms for analysis, using the MoMe® Device Communications Protocol. MoMe® Software System analyzes the data via the embedded algorithm and when indicated, data identified by the algorithm is flagged for physician review. MoMe® Kardia requires no patient intervention to capture or analyze data however does provide a patient event trigger.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Infobionic. The logo consists of the word "INFOBIONIC" in blue and green letters. The "INFO" part of the word is in blue, and the "BIONIC" part is in green. There is a blue heart in the middle of the word. To the left of the word is a cluster of blue and green dots.

MoMe® Kardia

K160064

Page 2 of 5

MoMe® Kardia supports three cardiac monitoring modes:

| 1 | Holter                         |
|---|--------------------------------|
| 2 | Event Monitoring               |
| 3 | Mobile Cardiac Telemetry (MCT) |

MoMe® Kardia:

- Is non-invasive and poses no significant safety issues ●
- Uses existing electrode and ECG technology ●
- ls used in an adjunctive fashion, where physicians also use patient ● symptoms and other tests, in the diagnosis or monitoring of patients with cardiac arrhythmias.

MoMe Kardia is not an emergency service. If the patient is experiencing symptoms that he/she is concerned about, the patient needs to seek immediate medical attention.

## Indications for Use:

MoMe® Kardia is indicated for:

- 1. Patients who experience transient symptoms that may suggest cardiac arrhythmia.
- 2. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation)
- 3. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath)
- Patients recovering from cardiac surgery or interventional procedures 4. who are indicated for outpatient arrhythmia monitoring.
- ECG data recorded by the device can be analyzed by other processing 5. systems to provide Holter style reports.

MoMe® Kardia is contraindicated for:

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Infobionic. The logo is made up of the word "INFOBIONIC" in blue and green letters. The letters "INFO" are in blue, and the letters "BIONIC" are in green. There is a blue heart in the middle of the word. To the left of the word is a cluster of blue and green dots.

MoMe® Kardia

K160064

Page 3 of 5

- 1. MoMe® Kardia is contraindicated for those patients requiring attended, inhospital monitoring for life threatening arrhythmias.
Note: MoMe® Kardia does not provide interpretive statements. Interpretation and diagnosis is the responsibility of a physician.

## Performance Data and Standards Compliance:

The MoMe® Kardia submission was written according to and in conformance with FDA Guidance "Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm" released on October 2003. The test reports in the submission demonstrate the MoMe® Kardia meets its intended use and design requirements.

The MoMe® Kardia was tested and conforms to following voluntary FDA recognized standards,

- IEC 60601-1:2005/R(2012) and A1:2012: Medical 1. electrical equipment - Part 1: General requirements for basic safety and essential performance
- 2. IEC 60601-1-2:2007/R(2012): Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests (Edition 3)
- 3. ANSI/AAMI/IEC 60601-2-47: 2012 Medical electrical equipment- Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems

## Technological Characteristics:

MoMe® Kardia differs from the predicate device mainly in one area of the device technology and design. The MoMe® Kardia is a single unit body worn device used for acquisition and transmission of ECG data to the monitoring center while the predicate device is a two unit device comprising of a sensor unit to acquire ECG data and a handheld unit which transmits the data to the monitoring center. In predicate device, the sensor unit communicates with cellular modem using Bluetooth technology.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Infobionic. The logo is made up of the word "INFOBIONIC" in blue and green letters. The letters "INFO" are in blue, and the letters "BIONIC" are in green. There is a small blue heart in the middle of the "O" in "BIONIC". To the left of the word is a cluster of blue and green circles.

MoMe® Kardia

K160064

Page 4 of 5

Both the devices, MoMe® Kardia and the predicate device, transmit the recorded ECG data using cellular modem to remote Monitoring centers where the ECG data is analyzed using proprietary software algorithm.

## Rationale for Substantial Equivalence:

| Parameter                                            | MoMe® Kardia                                                                                                               | TruVue™ Wireless<br>Ambulatory ECG<br>Monitoring System                                               |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| 510(k) Reference<br>Number                           | Not applicable, this 510(k)<br>submission                                                                                  | K100155                                                                                               |
| FDA Classification                                   | Class II                                                                                                                   | Class II                                                                                              |
| Product Code/<br>Classification Code/<br>Common Name | DSI<br>21 CFR 870.1025<br>Arrhythmia detector and<br>alarm                                                                 | DSI<br>21CFR 870.1025<br>Arrhythmia detector and<br>alarm                                             |
| Indications for Use                                  | Indications for use are<br>identical to predicate device.<br>Please refer to Indications for<br>Use in Section 12.2 below. | Refer to Indications for Use in<br>Section 12.2 below                                                 |
| Intended Use                                         | For ECG reporting and<br>arrhythmia detection in<br>patients with non-life<br>threatening arrhythmias                      | For ECG reporting and<br>arrhythmia detection in<br>patients with non-life<br>threatening arrhythmias |
| Number of<br>electrodes                              | 3                                                                                                                          | 3                                                                                                     |
| Number of ECG<br>channels                            | 2 channels                                                                                                                 | 2 channels                                                                                            |
| Ambulatory ECG<br>Performance<br>Standards           | IEC60601-2-47                                                                                                              | EC 38                                                                                                 |
| ECG Acquisition                                      | Single unit body worn (sensor<br>+ cellular module)                                                                        | Body worn sensor, handheld<br>device with cellular module)                                            |
| ECG Transmission<br>to Cellular                      | On board (single unit)                                                                                                     | Bluetooth                                                                                             |
| ECG Transmission<br>to Monitoring<br>Center          | Cellular                                                                                                                   | Cellular                                                                                              |

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Infobionic. The logo is made up of the word "INFOBIONIC" in blue and green letters. To the left of the word is a cluster of blue and green dots. The "O" in "BIONIC" is replaced with a blue heart.

MoMe® Kardia

| Parameter                                                                | MoMe® Kardia                                                                                            | TruVue™ Wireless<br>Ambulatory ECG<br>Monitoring System                                                 |
|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| User Event Trigger                                                       | Device User Interface                                                                                   | Handheld User Interface                                                                                 |
| Battery Type                                                             | Li lon rechargeable                                                                                     | Sensor - 1 AAA<br>Monitor - Li Ion rechargeable                                                         |
| Environments<br>where device may<br>be used                              | Physician practices, clinics,<br>research institutions                                                  | Physician practices, clinics,<br>research institutions                                                  |
| Environment where<br>device data storage<br>and reports are<br>generated | Remote (cloud-based) server                                                                             | Remote (cloud-based) server                                                                             |
| Prescription Use                                                         | Yes, intended to be used by<br>physicians and health care<br>providers only, not for use by<br>patients | Yes, intended to be used by<br>physicians and health care<br>providers only, not for use by<br>patients |
| Physician access to<br>patient<br>physiological and<br>event information | Yes                                                                                                     | Yes                                                                                                     |
| Arrhythmia<br>detection algorithm                                        | Proprietary/Server side                                                                                 | Proprietary/Server side                                                                                 |

## Conclusion:

The MoMe® Kardia Indications for Use are the same as the predicate device Indications for Use. The MoMe® Kardia utilizes equivalent operating principles and technology as compared to the predicate device. The descriptive information and performance testing in this submission demonstrate that the MoMe® Kardia meets the expected performance requirements, does not raise new issues of safety or effectiveness, and is therefore substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K160064](https://fda-staging.innolitics.com/device/K160064)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
