MoMe Kardia Wireless Ambulatory ECG Monitoring and Detection System
K160064 · Infobionic, Inc. · QYX · Mar 11, 2016 · Cardiovascular
Device Facts
Record ID
K160064
Device Name
MoMe Kardia Wireless Ambulatory ECG Monitoring and Detection System
Applicant
Infobionic, Inc.
Product Code
QYX · Cardiovascular
Decision Date
Mar 11, 2016
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
MoMe® Kardia is intended to be used for: 1. Patients who experience transient symptoms that may suggest cardiac arrhythmia. 2. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation) 3. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath) 4. Patients recovering from cardiac surgery or interventional procedures who are indicated for outpatient arrhythmia monitoring. 5. ECG data recorded by the device can be analyzed by other processing systems to provide Holter style reports.
Device Story
MoMe Kardia is a wearable, single-unit ambulatory ECG monitor; acquires ECG and accelerometer data; transmits data via cellular technology to a web-based remote server (MoMe Software System). System uses proprietary algorithms to analyze data; flags events for physician review. Device supports Holter, Event, and MCT monitoring modes. Used in physician practices, clinics, and research institutions; operated by healthcare providers. Patient intervention not required for capture, but device includes a patient event trigger. Output allows physicians to review ECG data and arrhythmia detections to aid in diagnosis and clinical decision-making; benefits patients by enabling remote monitoring of cardiac symptoms.
Clinical Evidence
Bench testing only. The device was tested for compliance with IEC 60601-1, IEC 60601-1-2, and ANSI/AAMI/IEC 60601-2-47. No clinical trial data was provided.
Technological Characteristics
Single-unit body-worn device; 3 electrodes; 2-channel ECG acquisition; accelerometer for motion data. Connectivity: cellular modem for data transmission to cloud-based server. Power: Li-ion rechargeable battery. Software: proprietary server-side algorithm for arrhythmia detection. Standards: IEC 60601-1, IEC 60601-1-2, ANSI/AAMI/IEC 60601-2-47.
Indications for Use
Indicated for patients experiencing transient symptoms suggesting cardiac arrhythmia, requiring monitoring of drug effects on ventricular rate in atrial arrhythmias, or recovering from cardiac surgery/interventional procedures. Contraindicated for patients requiring attended, in-hospital monitoring for life-threatening arrhythmias.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
TruVue™ Wireless Ambulatory ECG Monitoring System (K100155)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 15, 2023
Infobionic, Inc. Carrie Neuberger Regulatory Consultant To Infobionic Inc. 600 Suffolk Street Lowell, Massachusetts 01854
Re: K160064
Trade/Device Name: MoMe Kardia Wireless Ambulatory ECG Monitoring and Detection System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: Class II Product Code: QYX, DSI
Dear Carrie Neuberger:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated March 11, 2016. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Kozen, OHT2: Office of Cardiovascular Devices, 301-796-5813, jennifer.kozen@fda.hhs.gov.
Sincerely,
Jennifer W. Shih -S
Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Public Health Service
Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol is a depiction of an abstract human figure embracing or protecting another figure, representing the department's mission of protecting the health of all Americans and providing essential human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 11, 2016
Infobionic, Inc. Ms. Carrie Neuberger Regulatory Consultant to Infobionic Inc. 600 Suffolk Street Lowell, Massachusetts 01854
Re: K160064 Trade/Device Name: MoMe Kardia Wireless Ambulatory ECG Monitoring And Detection System Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including ST-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: DSI Dated: January 12, 2016 Received: January 13, 2016
Dear Ms. Carrie Neuberger,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{2}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours
FDA
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
#### 510(k) Number (if known) K160064
Device Name
MoMe® Kardia Wireless Ambulatory ECG Monitoring and Detection System
Indications for Use (Describe) MoMe® Kardia is intended to be used for:
1. Patients who experience transient symptoms that may suggest cardiac arrhythmia.
2. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation)
3. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath)
4. Patients recovering from cardiac surgery or interventional procedures who are indicated for outpatient arrhythmia monitoring.
5. ECG data recorded by the device can be analyzed by other processing systems to provide Holter style reports.
MoMe® Kardia is contraindicated for:
1. MoMe® Kardia is contraindicated for those patients requiring attended, in-hospital monitoring for life threatening arrhythmias.
Note: MoMe® Kardia does not provide interpretive statements. Interpretation and diagnosis is the responsibility of a physician.
Type of Use (Select one or both, as applicable)
| <span></span> | <span></span> |
|---------------|------------------------------------------------------------|
| <div></div> | <span></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| <span></span> | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Infobionic. The logo consists of the word "INFOBIONIC" in a combination of blue and green colors. The "INFO" part of the word is in blue, while the "BIONIC" part is in green. There is a blue heart symbol in the middle of the word. To the left of the word, there is a cluster of blue and green circles arranged in a semi-circular pattern.
MoMe® Kardia
# K160064 Page 1 of 5
# Section 5 510(K) SUMMARY COMPLYING WITH 21 CFR 807.92
| Date Prepared: | January 12, 2015 |
|----------------------|---------------------------------------------------------------------------------------------------|
| Submitter Name: | InfoBionic, Inc. |
| Submitter Address: | 600 Suffolk Street, Lowell, MA, 01854 |
| Contact Person: | Carrie Neuberger, Regulatory Consultant |
| Phone Number: | 415.640.3377 |
| Device Trade Name: | MoMe® Kardia Wireless Ambulatory ECG Monitoring<br>and Detection System |
| Device Common Name: | Continuous ECG monitor and Arrhythmia Detection |
| Classification Name: | Arrhythmia Detector and alarm, 21 CFR 870.1025,<br>DSI |
| Predicate Device: | K100155, Biomedical Systems Corporation's<br>TruVue™ Wireless Ambulatory ECG Monitoring<br>System |
## Device Description:
MoMe® Kardia is a wireless, remote monitoring system designed to aid physicians in their diagnosis of cardiac arrhythmias in patients with a demonstrated need for cardiac monitoring.
MoMe® Kardia includes the wearable MoMe® Kardia Device that acquires and stores ECG and motion (accelerometer) data and transmits that data via cellular technology to the MoMe® Software System (K152491), a web-based remote server software with proprietary algorithms for analysis, using the MoMe® Device Communications Protocol. MoMe® Software System analyzes the data via the embedded algorithm and when indicated, data identified by the algorithm is flagged for physician review. MoMe® Kardia requires no patient intervention to capture or analyze data however does provide a patient event trigger.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Infobionic. The logo consists of the word "INFOBIONIC" in blue and green letters. The "INFO" part of the word is in blue, and the "BIONIC" part is in green. There is a blue heart in the middle of the word. To the left of the word is a cluster of blue and green dots.
MoMe® Kardia
K160064
Page 2 of 5
MoMe® Kardia supports three cardiac monitoring modes:
| 1 | Holter |
|---|--------------------------------|
| 2 | Event Monitoring |
| 3 | Mobile Cardiac Telemetry (MCT) |
MoMe® Kardia:
- Is non-invasive and poses no significant safety issues ●
- Uses existing electrode and ECG technology ●
- ls used in an adjunctive fashion, where physicians also use patient ● symptoms and other tests, in the diagnosis or monitoring of patients with cardiac arrhythmias.
MoMe Kardia is not an emergency service. If the patient is experiencing symptoms that he/she is concerned about, the patient needs to seek immediate medical attention.
## Indications for Use:
MoMe® Kardia is indicated for:
- 1. Patients who experience transient symptoms that may suggest cardiac arrhythmia.
- 2. Patients who require monitoring of effect of drugs to control ventricular rate in various atrial arrhythmias (e.g. atrial fibrillation)
- 3. Patients with symptoms that may be due to cardiac arrhythmias. These may include but are not limited to symptoms such as: a) dizziness or lightheadedness; b) syncope of unknown etiology in which arrhythmias are suspected or need to be excluded; and c) dyspnea (shortness of breath)
- Patients recovering from cardiac surgery or interventional procedures 4. who are indicated for outpatient arrhythmia monitoring.
- ECG data recorded by the device can be analyzed by other processing 5. systems to provide Holter style reports.
MoMe® Kardia is contraindicated for:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Infobionic. The logo is made up of the word "INFOBIONIC" in blue and green letters. The letters "INFO" are in blue, and the letters "BIONIC" are in green. There is a blue heart in the middle of the word. To the left of the word is a cluster of blue and green dots.
MoMe® Kardia
K160064
Page 3 of 5
- 1. MoMe® Kardia is contraindicated for those patients requiring attended, inhospital monitoring for life threatening arrhythmias.
Note: MoMe® Kardia does not provide interpretive statements. Interpretation and diagnosis is the responsibility of a physician.
## Performance Data and Standards Compliance:
The MoMe® Kardia submission was written according to and in conformance with FDA Guidance "Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm" released on October 2003. The test reports in the submission demonstrate the MoMe® Kardia meets its intended use and design requirements.
The MoMe® Kardia was tested and conforms to following voluntary FDA recognized standards,
- IEC 60601-1:2005/R(2012) and A1:2012: Medical 1. electrical equipment - Part 1: General requirements for basic safety and essential performance
- 2. IEC 60601-1-2:2007/R(2012): Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests (Edition 3)
- 3. ANSI/AAMI/IEC 60601-2-47: 2012 Medical electrical equipment- Part 2-47: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems
## Technological Characteristics:
MoMe® Kardia differs from the predicate device mainly in one area of the device technology and design. The MoMe® Kardia is a single unit body worn device used for acquisition and transmission of ECG data to the monitoring center while the predicate device is a two unit device comprising of a sensor unit to acquire ECG data and a handheld unit which transmits the data to the monitoring center. In predicate device, the sensor unit communicates with cellular modem using Bluetooth technology.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Infobionic. The logo is made up of the word "INFOBIONIC" in blue and green letters. The letters "INFO" are in blue, and the letters "BIONIC" are in green. There is a small blue heart in the middle of the "O" in "BIONIC". To the left of the word is a cluster of blue and green circles.
MoMe® Kardia
K160064
Page 4 of 5
Both the devices, MoMe® Kardia and the predicate device, transmit the recorded ECG data using cellular modem to remote Monitoring centers where the ECG data is analyzed using proprietary software algorithm.
## Rationale for Substantial Equivalence:
| Parameter | MoMe® Kardia | TruVue™ Wireless<br>Ambulatory ECG<br>Monitoring System |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|
| 510(k) Reference<br>Number | Not applicable, this 510(k)<br>submission | K100155 |
| FDA Classification | Class II | Class II |
| Product Code/<br>Classification Code/<br>Common Name | DSI<br>21 CFR 870.1025<br>Arrhythmia detector and<br>alarm | DSI<br>21CFR 870.1025<br>Arrhythmia detector and<br>alarm |
| Indications for Use | Indications for use are<br>identical to predicate device.<br>Please refer to Indications for<br>Use in Section 12.2 below. | Refer to Indications for Use in<br>Section 12.2 below |
| Intended Use | For ECG reporting and<br>arrhythmia detection in<br>patients with non-life<br>threatening arrhythmias | For ECG reporting and<br>arrhythmia detection in<br>patients with non-life<br>threatening arrhythmias |
| Number of<br>electrodes | 3 | 3 |
| Number of ECG<br>channels | 2 channels | 2 channels |
| Ambulatory ECG<br>Performance<br>Standards | IEC60601-2-47 | EC 38 |
| ECG Acquisition | Single unit body worn (sensor<br>+ cellular module) | Body worn sensor, handheld<br>device with cellular module) |
| ECG Transmission<br>to Cellular | On board (single unit) | Bluetooth |
| ECG Transmission<br>to Monitoring<br>Center | Cellular | Cellular |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Infobionic. The logo is made up of the word "INFOBIONIC" in blue and green letters. To the left of the word is a cluster of blue and green dots. The "O" in "BIONIC" is replaced with a blue heart.
MoMe® Kardia
| Parameter | MoMe® Kardia | TruVue™ Wireless<br>Ambulatory ECG<br>Monitoring System |
|--------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| User Event Trigger | Device User Interface | Handheld User Interface |
| Battery Type | Li lon rechargeable | Sensor - 1 AAA<br>Monitor - Li Ion rechargeable |
| Environments<br>where device may<br>be used | Physician practices, clinics,<br>research institutions | Physician practices, clinics,<br>research institutions |
| Environment where<br>device data storage<br>and reports are<br>generated | Remote (cloud-based) server | Remote (cloud-based) server |
| Prescription Use | Yes, intended to be used by<br>physicians and health care<br>providers only, not for use by<br>patients | Yes, intended to be used by<br>physicians and health care<br>providers only, not for use by<br>patients |
| Physician access to<br>patient<br>physiological and<br>event information | Yes | Yes |
| Arrhythmia<br>detection algorithm | Proprietary/Server side | Proprietary/Server side |
## Conclusion:
The MoMe® Kardia Indications for Use are the same as the predicate device Indications for Use. The MoMe® Kardia utilizes equivalent operating principles and technology as compared to the predicate device. The descriptive information and performance testing in this submission demonstrate that the MoMe® Kardia meets the expected performance requirements, does not raise new issues of safety or effectiveness, and is therefore substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.