McLaren Proton Therapy System

K160063 · Mclaren Health Care Corporation · LHN · Feb 25, 2016 · Radiology

Device Facts

Record IDK160063
Device NameMcLaren Proton Therapy System
ApplicantMclaren Health Care Corporation
Product CodeLHN · Radiology
Decision DateFeb 25, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The McLaren Proton Therapy System is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Device Story

The McLaren Proton Therapy System is a medical charged-particle radiation therapy system used to treat localized tumors. It consists of a Beam Production System (injector and synchrotron) to generate and accelerate protons, and a Beam Delivery System (gantry, scan/dose, patient positioning, and treatment control subsystems) to shape and direct the beam to the patient. Operated by clinical staff in a treatment facility, the system delivers a controlled proton dose to the target site. The device is used to provide radiation therapy, with the healthcare provider using the system to plan and execute precise dose delivery to the patient's tumor. The primary benefit is the targeted delivery of radiation for tumor treatment.

Clinical Evidence

Bench testing only. Subsystem and full system verification and validation were performed, including beam performance testing under nominal and non-nominal conditions. Testing confirmed accurate beam creation, distribution, and dose delivery. Electrical safety and electromagnetic compatibility were verified per IEC 60601-1 and IEC 60601-1-2. Usability evaluation confirmed user interface functionality.

Technological Characteristics

System comprises Beam Production (injector, synchrotron) and Beam Delivery (scan/dose, gantry, patient positioning, treatment control) subsystems. Operates as a charged-particle radiation therapy system. Electrical safety and EMC testing conducted per IEC 60601-1 and IEC 60601-1-2.

Indications for Use

Indicated for patients with localized tumors and other conditions susceptible to treatment by radiation.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the profile graphic. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 25, 2016 McLaren Health Care Corporation % Daniel Gottlieb, Esq. Attorney Epstein Becker & Green, P.C. 1227 25th Street NW, Suite 700 WASHINGTON DC 20037 Re: K160063 Trade/Device Name: McLaren Proton Therapy System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: LHN Dated: January 11, 2016 Received: January 13, 2016 Dear Mr. Gottlieb: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D.'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K160063 Device Name McLaren Proton Therapy System Indications for Use (Describe) The McLaren Proton Therapy System is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. Type of Use (Select one or both, as applicable) | <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |-------------------------------------------------------------| | <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ #### 510(k) Summary #### McLaren Proton Therapy System # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Epstein Becker Green 1227 25th St. NW Ste. 700 Washington, DC 20037 Telephone: 202-861-1881 Facsimile: 202-861-3581 Contact Person: Daniel Gottlieb Date Prepared: January 11, 2016 ### Name of Device Name/Address of Sponsor McLaren Proton Therapy System McLaren Health Care Corporation G-3235 Beecher Rd., Suite B Flint, MI 48532 ### Common or Usual Name Proton beam therapy systems ### Classification Name Medical Charged-Particle Radiation Therapy System, 21 CFR 892.5050, Product Code LHN ### Predicate Device ProTom Radiance 330™ Proton Beam Therapy System (K134052) Cleared Mar 14, 2014 #### Intended Use The McLaren Proton Therapy System is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. {4}------------------------------------------------ # 510(k) Summary ### McLaren Proton Therapy System ## Technological Characteristics The McLaren Proton Therapy System consists primarily of Beam Delivery and Beam Production systems. These systems are comprised of various components and/or accessories designed to produce and deliver a proton beam appropriate for patient treatment. The system components include: - . Beam Production System. This system produces the proton beam and directs it to the appropriate treatment room and is comprised of the following subsystems: - . Synchrotron subsystem. The accelerator unit is the source of the proton beam and is composed of the injector (which generates the proton beam) and the synchrotron (which accumulates, accelerates, and extracts the proton beam). - . Beam Transport subsystem. This subsystem guides the proton beam extracted from the synchrotron to the treatment room. - Beam Delivery System. This system controls the irradiation dose and shapes the proton beam supplied through the beam transportation system into the configuration required for patient treatment, and directs the beam appropriately. It is comprised of the following subsystems: - . Scan/Dose subsystem - Gantry subsystem . - Patient Positioning subsystem. . - Treatment Delivery Control subsystem # Performance Data Each individual subsystem of the McLaren Proton Therapy System was verified and validated, and full system verification and validation was also performed. Beam performance testing to validate complete system integration under nominal and non-nominal conditions was performed on the full system. Beam delivery testing evaluated the following: - 1. Creation and direction the proton beam appropriately to the patient treatment location; - 2. Production of a transverse and longitudinal distribution appropriate for the patient treatment; and - 3. Delivery of the designated dose to the patient's treatment site. Testing to evaluate electrical safety and electromagnetic compatibility was performed in accordance with IEC 60601-1 and IEC 60601-1-2, and a usability evaluation was conducted to confirm that users can interact with the system user interface to perform treatment with the device system. All testing demonstrated that the system met its specifications for its intended use. # Substantial Equivalence Except for the device name, the McLaren Proton Therapy System has the identical intended use, indications, principles of operation, and technological characteristics as the ProTom Radiance 330'''' Proton Beam Therapy System (K134052). {5}------------------------------------------------ # 510(k) Summary # McLaren Proton Therapy System The McLaren Proton Therapy System is identical to the ProTom Radiance 330™ Proton Beam Therapy System and there are no new questions of safety or effectiveness. Thus, the McLaren Proton Therapy System is substantially equivalent to its predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...