# Monopolar single-use HF cable (K160053)

_Olympus Winter & Ibe GmbH · GEI · May 6, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K160053

## Device Facts

- **Applicant:** Olympus Winter & Ibe GmbH
- **Product Code:** [GEI](/productcode/GEI.md)
- **Decision Date:** May 6, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4400
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

HF cable for electrotherapeutical use in endoscopic procedures in conjunction with compatible active accessories and compatible electrosurgical units.

## Device Story

Monopolar single-use HF cable connects electrosurgical generator to resectoscope working element; facilitates delivery of monopolar HF current during endoscopic procedures. Device consists of proximal plug, insulated cord, and distal plug. Used in clinical/endoscopic settings by trained medical professionals. Provides electrical path for cutting/coagulation; enables therapeutic intervention. Single-use, sterile delivery ensures patient safety by eliminating reprocessing requirements.

## Clinical Evidence

No clinical or animal studies performed. Evidence based on bench testing, including electrical safety (AAMI/ANSI ES 60601-1, IEC 60601-2-2, IEC 60601-2-18), EMC (IEC 60601-1-2), and mechanical/durability testing (insertion/retraction force, tensile strength, dielectric strength, fluid ingress, and sterilization validation).

## Technological Characteristics

Monopolar HF cable; proximal/distal connectors with insulated cord. Standards: AAMI/ANSI ES 60601-1, IEC 60601-2-2, IEC 60601-2-18, IEC 60601-1-2. Sterilization: EtO (ISO 11135). Packaging: AAMI ANSI ISO 11607-1. Single-use, sterile. No software.

## Regulatory Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

## Predicate Devices

- HF-cable, monopolar ([K944200](/device/K944200.md)/1)
- Disposable Active Cord ([K890328](/device/K890328.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, serif font. The text is centered on the image and is the only element present.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 6, 2016

Olympus Winter & Ibe GmbH % Mr. Graham Baillie Manager, Regulatory Affairs Gyrus ACMI, Inc. 136 Turnpike Road Southborough, Massachusetts 01772

Re: K160053

Trade/Device Name: Monopolar Single-use HF Cable Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: April 20, 2016 Received: April 21, 2016

Dear Mr. Baillie:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Jennifer R. Stevenson -A

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

K160053

Device Name

Monopolar single-use HF cable

Indications for Use (Describe)

HF cable for electrotherapeutical use in endoscopic procedures in conjunction with compatible active accessories and compatible electrosurgical units.

Type of Use (Select one or both, as applicable)

| <span style="text-decoration: overline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

Monopolar single-use HF cable

# 510(k) PREMARKET NOTIFICATION

## SECTION 2 - 510k Summary

 Olympus Winter & Ibe GmbH Hamburg, Germany

{4}------------------------------------------------

## 510(k) Summary December 07, 2015

# 1. General Information

| Manufacturer:           | Olympus Winter & Ibe GmbH<br>Kuehnstr. 61<br>22045 Hamburg<br>Germany                                                                                                                                                                                         |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                         | Establishment Registration Number: 9610773                                                                                                                                                                                                                    |
| Official Correspondent: | Graham A.L. Baillie<br>Manager, Regulatory Affairs<br>Gyrus ACMI, Inc.<br>136 Turnpike Rd.<br>Southborough, MA 01772-2104<br>Phone: 508.804.2738<br>Fax: 508. 804.2624<br>Email: Graham.baillie@olympus-osta.com<br>Establishment Registration No: 3003790304 |

## 2. Device Identification

| Common Name:            | Electrosurgical, Cutting & Coagulation & Accessories           |
|-------------------------|----------------------------------------------------------------|
| Regulation Number:      | 878.4400                                                       |
| Regulation Description: | Electrosurgical cutting and coagulation device and accessories |
| Device Class:           | II                                                             |
| Product Code:           | GEI                                                            |
| Review Panel:           | General & Plastic Surgery                                      |
| Proprietary/Trade Name: | Monopolar single-use HF cable                                  |

# 3. Predicate Devices

| 510(k) No. | Name                   | Predicate Model No. | Product code /<br>Reg No.        |
|------------|------------------------|---------------------|----------------------------------|
| K944200/1  | HF-cable, monopolar    | A0358               | GCJ / 876.1500<br>FAS / 876.4300 |
| K890328    | Disposable Active Cord | DAC                 |                                  |
|            |                        |                     | GEI / 878.4400<br>HIH / 884.1690 |

{5}------------------------------------------------

The predicates have not been subject to a design-related recall.

No reference devices were used in this submission.

## 4. Product Description

The Olympus monopolar single-use HF cable that is subject to this submission connects a compatible electrosurgical generator with the working element of a resectoscope for the application of monopolar HF current during endoscopic procedures.

The HF cable is delivered in sterile condition and is intended for single use only.

The monopolar single-use HF cable consists of a proximal plug connecting to the HF generator, an insulated cord and a distal plug connecting to the working element of the resectoscope.

## 5. Indications for Use

"HF cable for electrotherapeutical use in endoscopic procedures in conjunction with compatible active accessories and compatible electrosurgical units."

This intended use is essentially the same intended use as that of the predicate devices.

## 6. Comparison of Technological characteristics

The subject and predicate devices are based on the same technological principle with similar elements:

- I HF cable consisting of a proximal plug connecting to the HF generator, an insulated cord and a distal plug connecting to an active instrument
- 트 Predicate and subject devices are designed for use with monopolar current
- I Similar outer dimensions
- I Design changes of the HF cable are minor and do not negatively impact safety or effectiveness of the subject devices
- . There are no patient-contacting materials in predicates and subject device

While one of the predicate devices (A0538) is reusable and can be reprocessed, the other predicate (DAC) and the subject device are provided in sterile condition and are 

{6}------------------------------------------------

intended for single-use only. Accordingly, outside materials differ to suit differing durability needs. Also, material selection of the single-use predicate and subject device differ; however, durability of the new materials has been tested successfully.

#### 7. Performance Data

The following performance data was provided in support of the substantial equivalence determination. All standards applied are FDA recognized international standards.

# Electrical safety and electromagnetic compatibility (EMC)

Electrical Safety was tested according to

| AAMI/ANSI ES 60601-<br>1:2005 + A1:2012, C1:2009<br>and A2:2010 | Medical Electrical Equipment - Part 1.1 General<br>requirements for safety and essential performance.                                                                                                 |
|-----------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| AAMI/ANSI/IEC 60601-2-2<br>2009                                 | Medical Electrical Equipment - Part 2-2: Particular<br>Requirements for the Basic Safety And Essential<br>Performance of High Frequency Surgical Equipment<br>and High Frequency Surgical Accessories |
| IEC 60601-2-18:2009                                             | Medical electrical equipment - Part 2-18: Medical<br>Electrical Equipment - Part 2-18: Particular<br>requirements for the basic safety and essential<br>performance of endoscopic equipment.          |
| IEC 60601-1-2:2007                                              | Medical electrical equipment - Part 1-2: General<br>requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic<br>compatibility - Requirements and tests       |

#### Clinical Evaluation

Clinical and animal studies were not necessary.

A clinical evaluation has been conducted containing a comprehensive literature review.

#### Software

Not applicable as the HF cable does not contain any software.

{7}------------------------------------------------

#### Performance Testing Bench

Conducted tests included the following tests:

- Insertion & Retraction Force tests of connectors -
- -Cable marking durability test
- Surface temperature measurement test -
- Cable insulation HF and Mains Dielectric Strength test -
- -Cable anchorage test
- Fluid ingress test -
- -Continuity test
- Cable tensile strength test -
- Connector grip test -
- Output power measurement -
- -Transport testing

#### 8. Sterilization and Shelf Life

Sterilization is performed according to ISO 11135 and packaging conforms with AAMI ANSI ISO 11607-1:2006. The EtO sterilization cycle has been validated.

A sterility assurance level (SAL) of 106 was reached during validation and will be used for routine sterilization in compliance with regulations in force for sterile medical devices.

The EtO residuals are within the limits after tunnel degassing time.

Shelf Life testing supports a shelf life of 1 year for the monopolar single-use HF cable.

#### 9. Conclusion

The performance data support the safety of the device and demonstrate that the subject device complies with the intended use as specified.

In summary, we believe the monopolar single-use HF cable is substantially equivalent with the predicate devices with respect to the general design approach, function, and the intended use. The HF cable raises no new concerns of safety or effectiveness when compared to the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K160053](https://fda-staging.innolitics.com/device/K160053)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).