K160050 · Stryker Corporation · GEI · May 3, 2016 · General, Plastic Surgery
Device Facts
Record ID
K160050
Device Name
SERFAS 90-S Electrosurgical Probe
Applicant
Stryker Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
May 3, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Stryker SERFAS 90-S Electrosurgical Probe is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in orthopedic and arthroscopic procedures of joints such as the knee, shoulder, elbow, hip, ankle, and wrist.
Device Story
The SERFAS 90-S Electrosurgical Probe is a disposable, single-use accessory for the SERFAS Energy System; used in arthroscopic orthopedic procedures. The device delivers radiofrequency (RF) energy to soft tissue for resection, ablation, and coagulation; also provides hemostasis for blood vessels. Operated by surgeons in clinical settings; the probe connects to a compatible Stryker RF generator console. The device features a redesigned ceramic tip, electrode face suction path, and handle to improve mechanical robustness and prevent unintentional activation. Surgeons use the probe to remove tissue and control bleeding during joint surgery; the device benefits patients by facilitating minimally invasive arthroscopic interventions. The probe is provided sterile via ethylene oxide.
Clinical Evidence
Bench testing only. Testing included aggressive use, tip cantilever strength, electrode bend testing, thermal expansion, leak testing, and system compatibility with Stryker consoles. Biocompatibility was assessed per ISO 10993-1:2009. Sterilization validation followed ISO 11135:2014 and residuals were assessed per ISO 10993-7. All tests passed.
Technological Characteristics
RF electrosurgical probe; disposable; ethylene oxide sterilized. Materials: ceramic and handle components (biocompatible per ISO 10993-1). Features: suction path, electrode face. Energy source: RF generator (external). Connectivity: compatible with Stryker RF consoles.
Indications for Use
Indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in orthopedic and arthroscopic joint procedures (knee, shoulder, elbow, hip, ankle, wrist).
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, resembling a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 3, 2016
Stryker Corporation Somi Ekwealor Regulatory Affairs Analyst 5900 Optical Court San Jose, California 95138
Re: K160050
Trade/Device Name: SERFAS 90-S Electrosurgical Probe Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: April 1, 2016 Received: April 4, 2016
Dear Somi Ekwealor:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| K160050 | DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Page 1 of 1 | Page 19 of 681 |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|----------------|
| | <b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. | |
| 510(k) Number (if known) | K160050 | | |
| Device Name | Stryker® SERFAS 90-S Electrosurgical Probe | | |
| Indications for Use (Describe) | The Stryker SERFAS 90-S Electrosurgical Probe is indicated for resection, ablation, and coagulation of soft tissue and<br>hemostasis of blood vessels in orthopedic and arthroscopic procedures of joints such as the knee, shoulder, elbow, hip,<br>ankle, and wrist. | | |
| Type of Use (Select one or both, as applicable) | | | |
| | <div> <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </div> | <div> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </div> | |
| | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | |
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | Page 1 of 1 | PSC Publishing Services (301) 443-6740 | EF |
|----------------------|-------------|----------------------------------------|----|
|----------------------|-------------|----------------------------------------|----|
189 ງຸງ 10 6 ໂ ລອງປີ 1
{3}------------------------------------------------
## 1. General Information
| 510(k) Sponsor | Stryker Corporation |
|-------------------------|-----------------------------------------------------------------------------------|
| Address | 5900 Optical Court<br>San Jose, CA 95138 |
| FDA Registration Number | 2936485 |
| Correspondence Person | Mr. Somi Ekwealor, MSRS, RAC<br>Regulatory Affairs Analyst<br>Stryker Corporation |
| Contact Information | Email: somi.ekwealor@stryker.com<br>Phone: (408) 754-2356<br>Fax: (408) 754-2507 |
| Date Prepared | 08 Jan 2016 |
# 2. Device Identification
Proposed Device:
| Proprietary Name | Stryker® SERFAS 90-S Electrosurgical Probe |
|---------------------|-------------------------------------------------------------------|
| Common Name | RF Probe |
| Classification Name | Electrosurgical Cutting and Coagulation Device<br>and Accessories |
| Regulation Number | 21 CFR 878.4400 |
| Product Code | GEI |
| Regulatory Class | II |
### Predicate Device:
| Proprietary Name | SERFAS Energy System (probe only*) |
|------------------------|----------------------------------------------------------------|
| Common Name | RF System, RF Generator, RF Probe, RF Footswitch |
| Premarket Notification | K041810 |
| Classification Name | Electrosurgical Cutting and Coagulation Device and Accessories |
| Regulation Number | 21 CFR 878.4400 |
| Product Code | GEI |
| Regulatory Class | II |
* The predicate device includes the entire energy system and accessories while the Proposed device includes the electrosurgical probe (90-S Redesign) only. This premarket notification refers to the SERFAS 90-S Redesign unless Original SERFAS 90-S is stated.
{4}------------------------------------------------
### Device Description
The Stryker SERFAS 90-S Electrosurgical Probe (hereafter referred to as "Proposed device") is an accessory to the SERFAS Energy System, marketed through K041810, which is intended for resection, ablation, and coagulation of soft tissue via radiofrequency (RF) ablation. RF ablation probes are the main tool used in most arthroscopic procedures for the removal of tissue and the coagulation of bleeding vessels. The Proposed device is a disposable single-use electrosurgical device provided sterile via Ethylene Oxide sterilization.
### 3. Indication for Use
The Stryker SERFAS 90-S electrosurgical probe is indicated for resection, ablation, and coagulation of soft tissue and hemostasis of blood vessels in orthopedic and arthroscopic procedures of joints such as the knee, shoulder, elbow, hip, ankle, and wrist.
### 4. Comparison of Technological Characteristics with the Predicate Device
The Stryker SERFAS 90-S electrosurgical probe and predicate device are similar in technological characteristics and design. The predicate device includes the entire SERFAS Energy system while the Proposed device includes the probe only.
The differences between the Proposed device and the predicate are minor and raise no new questions of safety and effectiveness, therefore supporting that the SERFAS 90-S electrosurgical probe is substantially equivalent to the predicate device currently on the market. The following differences exist between the Proposed and predicate device:
- 트 Ceramic and Electrode Key Feature Design
- Electrode Face Suction Path
- 트 Handle Design
### Ceramic and Electrode Key Feature Design
Both the Proposed device and predicate device ceramic raw materials are identical but the devices have slightly different dimensional specifications. The intent of this change is to increase the mechanical robustness of the probe's tip and thereby address the failure mode of tip breakage. This difference is not critical to the intended surgical use of the device and does not impact the effectiveness of the device when used as labeled, however the safety of this product will be improved as a result of this change given the reduction in occurrence of tip breakage.
{5}------------------------------------------------
This minor difference in the Proposed device does not introduce additional risks and raises no new questions of safety or effectiveness and is substantially equivalent to the predicate device.
#### Electrode Face Suction Path
The Proposed device contains minor electrode face suction path design differences compared to the predicate device. Both the Proposed device and predicate device electrode raw materials are identical but the devices have slightly different dimensional specifications.
This minor difference in the Proposed device does not introduce additional risks and raises no new questions of safety or effectiveness and is substantially equivalent to the predicate device.
#### Handle Design
The patient contacting raw materials of the Proposed device and predicate device handle are identical. The Proposed device has a slightly different design specification for improved safety. This difference is not critical to the intended surgical use of the device and does not affect the effectiveness of the device when used as labeled. however the safety of this product will be improved as a result of this change because the occurrence of unintentional activation will be decreased.
This minor difference in the Proposed device does not introduce additional risks and raises no new questions of safety or effectiveness and is substantially equivalent to the predicate device.
#### 5. Performance Testing
The Stryker SERFAS 90-S Electrosurgical probe was tested for performance in accordance with internal design specification and with the applicable performance standards, which established the substantial equivalence determination. The following non-clinical tests were conducted and are summarized in this premarket notification:
| Test Name | Description | Results |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Aggressive Use | Determines if a probe can last its full lifetime without failure<br>while being subjected to an extended activation period on<br>tissue at the highest power setting. | Pass |
| Tip Cantilever | Determines if the probe tip can withstand the normal, side,<br>and back force required to permanently deform the lumen | Pass |
{6}------------------------------------------------
| Test Name | Description | Results |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Electrode Bend | Verifies the electrode can withstand a 7lb force for a worst<br>case scenario. | Pass |
| Thermal<br>Expansion (Heat)<br>Test | Verifies that the SERFAS probe tip assembly does not<br>experience thermal expansion capable of causing catastrophic<br>failure when subject to excessive heat | Pass |
| Leak | Verifies the condition of the ceramic-outer lumen assembly<br>and the condition of the suction clamp of all styles of Stryker<br>and non-Stryker RF ablation probes. | Pass |
| System<br>Compatibility | Verifies that the electrosurgical probe is compatible with<br>Stryker's legally marketed consoles | Pass |
Biocompatibility was assessed in accordance with ISO 10993-1:2009 "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process" and related collateral standards for patient contacting materials (see Section 15 -Biocompatibility).
Sterilization was assessed in accordance with ISO 11135:2014 - Sterilization of health-care products - Ethylene Oxide - Requirements for the development, validation and routine control of a sterilization process for medical devices. Sterilization residuals were assessed in accordance with ISO 10993-7 - Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals (see Section 14 - Sterilization and Shelf Life).
Bench performance testing was conducted to ensure that the device functioned as intended and met design specifications and acceptance criteria. Test results obtained verified the safety and effectiveness of the device per design specifications and applicable standards (see Section 18 - Performance Testing - Bench).
#### 6. Conclusion
Based on the intended use, technological characteristics, performance testing, and nonclinical testing in comparison to the predicate device, the Stryker SERFAS 90-S Electrosurgical Probe does not introduce additional risks and raises no new questions of safety and effectiveness, thus demonstrating substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.