← Product Code [JAQ](/productcode/JAQ) · K160045

# Heyman Packing Applicator Set (K160045)

_Varian Medical Systems, Inc. · JAQ · Nov 17, 2016 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K160045

## Device Facts

- **Applicant:** Varian Medical Systems, Inc.
- **Product Code:** [JAQ](/productcode/JAQ.md)
- **Decision Date:** Nov 17, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5700
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The Heyman Packing Applicator Set is indicated for cancer treatment of the endometrium using HDR or PDR brachytherapy.

## Device Story

Heyman Packing Applicator Set consists of various sized Heyman-capsules and accessories (cleaning caps, leak stop buttons, channel markers) used in intracavitary brachytherapy. Capsules connect to compatible remote afterloaders via guide tubes to deliver radioactive sources to target endometrial tissue. Used in healthcare facilities by trained personnel. Applicator guides radioactive source to precise treatment location. Device is MR conditional, reusable, and requires steam sterilization by user. No software, electronics, or biologics included.

## Clinical Evidence

No clinical data or animal studies were included. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (cytotoxicity, sensitization, irritation, systemic/subacute toxicity, implantation), sterilization validation, and mechanical/MRI safety analysis.

## Technological Characteristics

Intracavitary brachytherapy applicator set. Components: Heyman-capsules (various diameters) and accessories. Materials: Biocompatible, MR conditional. Energy source: External (remote afterloader). Connectivity: None (mechanical interface). Sterilization: Steam (pre-vacuum). Software: None.

## Regulatory Identification

A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.

## Predicate Devices

- GammaMed Plus High Dose Rate Remote Afterloading System ([K983436](/device/K983436.md))

## Reference Devices

- Varian Cleaning Caps, Varian Leak Stop Buttons, Varian Leak Stop Channel Marker Sets ([K152018](/device/K152018.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 17, 2016

Varian Medical Systems, Inc. % Mr. Peter J. Coronado Director, Regulatory Affairs 3100 Hansen Way PALO ALTO CA 94304

Re: K160045

Trade/Device Name: Hevman Packing Applicator Set Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: JAQ Dated: November 3, 2016 Received: November 4, 2016

Dear Mr. Coronado:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael O'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K160045

Device Name Heyman Packing Applicator Set

Indications for Use (Describe)

Heyman Packing Applicator Set is indicated for use for treating endometrial cancer using HDR or PDR brachytherapy.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

A partner for life

Varian Medical Systems, Inc. Corporate Headquarters 3100 Hansen Way Palo Alto, CA 94304-1038

Telephone: 1.650.493.4000 Toll Free: 800.544.4636 www.varian.com

# 510k Summary

# 510(k) Submission for Heyman Packing Applicator Set

As required by 21 CFR 807.92, Reference: FDA's Guidance Document "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]" (June 2014).

#### l. SUBMITTER

| Submitter's Name: | Varian Medical Systems<br>3100 Hansen Way, m/s E-110<br>Palo Alto CA 94304-1038 |
|-------------------|---------------------------------------------------------------------------------|
| Contact Name:     | Peter J. Coronado                                                               |
| Position:         | Director, Regulatory Affairs                                                    |
| Phone:            | 1.650.424.6230                                                                  |
| Fax:              | 1.650.646.9200                                                                  |
| Email:            | submissions.support@varian.com                                                  |
| Date Prepared:    | Thursday January 07, 2016                                                       |

#### II. DEVICE

| Name of Device:    | Heyman Packing Applicator Set                                         |
|--------------------|-----------------------------------------------------------------------|
| Common/Usual Name: | System, Applicator, Radionuclide,<br>Remote-controlled                |
| Regulation Name:   | Remote controlled radionuclide applicator<br>system (21 CFR 892.5700) |
| Regulatory Class:  | Class II                                                              |
| Product Code:      | JAQ                                                                   |

{4}------------------------------------------------

#### PREDICATE DEVICE .

Name of Predicate: GammaMed Plus High Dose Rate Remote Afterloading System 510k Number: K983436

This predicate has not been subject to a design-related recall. K152018 (Varian Cleaning Caps, Varian Leak Stop Buttons, Varian Leak Stop Channel Marker Sets) is used as a reference device for this submission.

#### IV. DEVICE DESCRIPTION

The Heyman Packing Applicator Set consists of various sized Heyman-capsules which may be connected to compatible brachytherapy afterloaders via guide tubes. The set also includes a number of associated accessories for sterilization (Cleaning Cap, Leak Stop Button, and Leak Stop Channel Marker Sets). The Heyman Packing Applicator Set is intended for cancer treatment of the endometrium using HDR or PDR brachytherapy. The Heyman Packing Applicator Set is an applicator for intracavitary Brachytherapy. The applicator acts to quide the radioactive source to the correct location or locations for treatment.

The device does not contain or consist of software/firmware. The device does not contain any biologics or drug components. The device has patient-contacting materials. The device is designed for repeated use. No parts of the system are provided sterile. The device can be steam sterilized with common parameters using pre-vacuum sterilization. This device is used on female patients.

The Heyman Packing Applicator Set is intended to be used in a healthcare or treatment facility by trained and qualified personnel.

{5}------------------------------------------------

| Product Number | Component List                                                | Associated<br>510(k) Number |
|----------------|---------------------------------------------------------------|-----------------------------|
| GM11005280     | Heyman-capsule, 4mm diameter x<br>17.5mm length, with mandrin | This 510(k)                 |
| GM11005110     | Heyman-capsule, 6mm diameter x<br>17.5mm length, with mandrin |                             |
| GM11005040     | Heyman-capsule, 8mm diameter x<br>17.5mm length, with mandrin |                             |
| GM11010510     | Leak Stop Channel Marker Set, 1 - 12                          | Cleared under<br>K152018    |
| GM11003810     | Leak stop button, pack of 50                                  |                             |
| GM11010770     | Cleaning Cap, dark blue (pack of 25)                          |                             |

The associated accessories include:

#### INDICATIONS FOR USE V.

Indications for Use Statement: The Heyman Packing Applicator Set is indicated for cancer treatment of the endometrium using HDR or PDR brachytherapy.

### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

Both the subject and predicate devices are intended for use in the treatment of cancer through intracavitary brachytherapy. The subject device is based on a subset of applicator accessories from the predicate device.

{6}------------------------------------------------

At a high level, the subject and predicate devices are based on the following same technological elements:

- Biocompatible Heyman-capsules of various sizes .
- Application in intracavitary brachytherapy treatments ●
- . Same anatomical treatment site/application
- Sterilized by user prior to use (and reuses) .

The following main differences exist between the subject and predicate devices:

- Material change in the probe connect of the Heyman-capsule
- . Validation of machine cleaning and the associated accessories
- . Heyman capsules for the subject device are MR conditional
- Added VariSource afterloader compatibility .

#### VII. PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

# Non-Clinical Tests:

## Biocompatibility Testing:

The biocompatibility evaluation for the subject device was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995, and International Standard ISO 10993-1 "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Management Process," as recognized by FDA. This included the following tests:

- . Cytotoxicity
- Sensitization ●

{7}------------------------------------------------

- Irritation
- Systemic Toxicity ●
- Subacute Toxicity ●
- . Implantation

The Heyman Packing Applicator Set is a surface device in contact with mucosal membrane and prolonged contact duration of 24 hours up to 30 days.

## Sterilization Testing:

Sterilization testing was performed for the subject device was conducted to assess the effectiveness of the provided cleaning, disinfection, and sterilization procedures for the device. Furthermore, the components of the subject set were subjected to sterilization cycles up to the stated use life and evaluated for performance and any damage that might affect safety or effectiveness.

- . Validation/Efficacy Testing of Cleaning, Disinfection, Sterilization Cycles
- . Verification of the Suitability of Applicator Design and Material for the Stated Use Life and Reprocessing Cycles

## Electrical Safety and Electromagnetic Compatibility (EMC):

This item is not applicable to the subject device. No electrical safety and electromaqnetic compatibility tests have been included in this submission in support of the substantial equivalence determination.

## Software Verification and Validation Testing:

This item is not applicable to the subject device; the device does not contain or consist of software/firmware. No software verification and validation testing has been included in this submission in support of the substantial equivalence determination.

## Mechanical and Acoustic Testing:

- MRI Safety Tests and Analysis ●

{8}------------------------------------------------

- CT Compatibility Tests and Analysis ●
- Validation with the VariSource afterloader series

# Animal Study / Clinical Tests:

No animal studies or clinical tests have been included in this submission in support of the substantial equivalence determination.

# VIII. CONCLUSIONS

The results of the non-clinical tests support the safety and effectiveness of the device under the specified use conditions. Varian believes that the validation and verification testing demonstrates that the subject device performs comparably to the predicate device, meeting the intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K160045](https://fda-staging.innolitics.com/device/K160045)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com