Quickie® and Zippie® power wheelchairs are battery-operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position. The Zippie power wheelchairs are specifically for people who are slightly smaller in stature—including children.
Device Story
Battery-powered, user-controlled electric wheelchairs; Quickie Pulse for medium/large individuals; Zippie ZM-310 for smaller individuals/children. Optional Bluetooth Mouse Module (BMM) added to control bus; allows users to utilize existing wheelchair joystick/input device to control pointer functions on external Bluetooth-enabled devices (e.g., tablets/PCs). Feature only active when wheelchair is stationary. Used in homes, healthcare facilities, and outdoors. Provides convenience for disabled users to interact with external electronics without leaving the chair. BMM operates in 2.4-2.485 GHz ISM band; frequency-hopping spread-spectrum. Safety ensured by ensuring BMM operation does not interfere with wheelchair braking or electronic control systems.
Clinical Evidence
No clinical or animal data provided. Substantial equivalence supported by bench testing, including electromagnetic compatibility (EMC) testing per ANSI/RESNA WC2 standards (extended to 3.0 GHz) and operational verification of the wireless pointer function.
Technological Characteristics
Battery-powered electric wheelchair; BMM adds Bluetooth 2.4-2.485 GHz transmitter (Class 1, 100mW max, frequency-hopping). Connectivity: Bluetooth wireless to external devices. EMC tested to ANSI/RESNA WC2. Software: Embedded firmware for BMM pointer control logic.
Indications for Use
Indicated for persons restricted to a sitting position requiring mobility assistance, including children and individuals of smaller stature for Zippie models.
Regulatory Classification
Identification
A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.
Predicate Devices
Quickie Pulse (Models 6BC, 6SC, 6CC, 5BC, 5CC, 6MPC) and Zippie ZM-310 (Models BC & SC) (K142457)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a series of human profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 14, 2016
Sunrise Medical (US) LLC Laurie Roberts Director, Regulatory Affairs 2842 Business Park Av Fresno, California 93727
Re: K160031
Trade/Device Name: Quickie Pulse (Models 6BC, 6SC, 6CC, 5BC, 5CC, 6MPC) and Zippie ZM-310 (Models BC, SC) Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: Class II Product Code: ITI Dated: June 9, 2016 Received: June 9, 2016
Dear Laurie Roberts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J哥offmann -A
Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K160031
Device Name
Quickie Pulse (Models 6BC, 6SC, 6CC, 5BC, 5CC, 6MPC) and Zippie ZM-310 (Models BC, SC)
Indications for Use (Describe)
Quickie® and Zippie® power wheelchairs are battery-operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position. The Zippie power wheelchairs are specifically for people who are slightly smaller in stature-including children.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) SUMMARY
## Submitter
| Company | Sunrise Medical (US) LLC<br>2842 Business Park Avenue<br>Fresno, CA 93727<br>Tel: 800-333-4000<br>Fax: 559-294-2872 |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact | Laurie Roberts, MS, RAC<br>Director, Regulatory Affairs |
| Device | |
| Trade name | Quickie Pulse (Models 6BC, 6SC, 6CC, 5BC, 5CC, 6MPC) and Zippie<br>ZM-310 (Models BC, SC) |
| Class name | Wheelchair, powered (21 CFR 890.3860) |
| Product code | ITI |
| Class | 2 |
| Predicate | |
| Devices | Quickie Pulse: Models 6BC, 6SC, 6CC, 5BC, 5CC, 6MPC and Zippie ZM-<br>310: Models BC & SC; legally marketed per 510(k): K142457 |
| Recalls | These predicate devices have not been subject to a design-related recall. |
| Device Description | |
| Overview | Quickie and Zippie wheelchairs are battery-powered, user-controlled, electric<br>wheelchairs. All models of these wheelchairs will be offered with an optional<br>wireless pointer control feature. |
| Wireless<br>pointer<br>control<br>feature | This option entails adding a Bluetooth Mouse Module (BMM)-a small<br>Bluetooth transmitter-mounted to the chair and wired into its control bus. |
| | With BMM in place, the wheelchair's controller can then use the pointer-<br>control feature built into the BMM hardware and software. |
| | Pointer control can only be activated when the wheelchair is stationary and<br>not actively performing wheelchair functions. |
| Purpose of<br>feature | The BMM lets users use the chair's joystick, or other input device, to control<br>the pointer functions of Bluetooth-enabled devices (such as personal or<br>tablet computers) while the chair is not in operation. |
| Model<br>distinctions | The wireless pointer control feature will be offered on all models of the<br>Quickie Pulse and Zippie ZM-310 series chairs.<br><br>Technologically, all of these chair designs are identical. Model designations<br>relate to widths and heights of their seats and various components: such as<br>seating, wheels, and other options available. Quickie Pulse chairs are sized to fit medium to large size individuals Zippie ZM-310 chairs fit individuals of smaller stature—including children |
| Use<br>environment | These chairs, along with the BMM, are meant for use in healthcare facilities,<br>in the home, outdoors, and other places of individual activity for the disabled. |
Continued on next page
{4}------------------------------------------------
### Device Description (continued)
#### Intended Use
| Indications<br>for Use | Quickie® and Zippie® power wheelchairs are battery-operated devices with wheels that are intended for medical purposes to provide mobility to persons restricted to a sitting position. The Zippie power wheelchairs are specifically for people who are slightly smaller in stature—including children. |
|------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature fact | The Indications for Use for the Quickie and Zippie power wheelchairs with wireless pointer control are identical to those for the chairs that do not have the feature. The feature is strictly an add-on for the convenience of the users. |
## Technological Characteristics and Comparison
| Design | Technological characteristics—design, materials, motors, electronics, power source, etc.—are identical to the predicate devices with the exception of the addition of the BMM. |
|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| BMM option | By adding the BMM, wireless pointer control can be engaged.<br>The BMM adds The ability to use the chair's joystick, or other input device, to control the pointer functions of Bluetooth-enabled electronic devices Some extra wiring to the electronic system raising the issue of continued electromagnetic compatibility for the chairs A low-power radio-frequency transmitter in the 2 to 2.5 GHz Bluetooth frequency range raising the issue of electromagnetic emissions interfering with chair brake function (chair is not in motion at the time of transmission but may be braking to hold chair on an incline) |
{5}------------------------------------------------
### Technological Characteristics and Comparison (continued)
| Parameters | K160031 (subject device) | K142457 (predicate device) |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Models | Quickie Pulse® (Models: 6BC,<br>6SC, 6CC, 5BC, 5CC, 6MPC)<br>and Zippie® ZM-310 (Models:<br>BC & SC) | Same |
| Intended Use | To provide mobility to persons<br>limited to a seating position<br>that have the capability of<br>operating a powered<br>wheelchair | Same |
| Indications for Use | Quickie® and Zippie® power<br>wheelchairs are battery-<br>operated devices with wheels<br>that are intended for medical<br>purposes to provide mobility<br>to persons restricted to a<br>sitting position. The Zippie<br>power wheelchairs are<br>specifically for people who are<br>slightly smaller in stature-<br>including children. | Same |
| Bluetooth feature available | YES<br>Bluetooth technology operates<br>in unlicensed industrial,<br>scientific and medical (ISM)<br>band at 2.4 to 2.485 GHz,<br>using a spread-spectrum,<br>frequency-hopping, full-duplex<br>signal at a nominal rate of<br>1600 hops/sec.<br>Class 1 output - used<br>primarily in industrial use<br>cases - 100 meters Power is<br>100mW max for Class 1 | NO |
| Other technical features | Same | Same |
| Evaluation methods | Same | Same |
{6}------------------------------------------------
#### Performance Data
The following non-clinical performance data were provided in support of the substantial equivalence determination.
| Performance | Operational performance of wireless pointer function without detriment to<br>wheelchair function or features<br>Chair's electrical braking continued to hold on a maximum slope during<br>BMM operation Wireless pointer function properly executed on typical external Bluetooth-<br>enabled electronic device from wheelchair inputs |
|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EMC | Electromagnetic compatibility testing carried out to cover functional<br>verification and device performance to ANSI/RESNA WC2 testing plus over<br>extended frequency range through 3.0 GHz with BMM mounted in its<br>closest proximity (worst-case) position to the wheelchair controller<br>demonstrated Immunity of total chair system with BMM mounted and connected No effect on chair during BMM activation |
| Software<br>qualification | Wireless pointer function demonstrated while chair not in operation,<br>qualifying the BMM and its software by successful device performance. |
No animal or clinical data necessary to demonstrate the performance of this option.
#### Conclusion
The performance data support the electromagnetic compatibility of the BMM to the Quickie and Zippie power wheelchairs and their operation. The operational verification and validation of the wireless pointer control feature demonstrate that the feature performs as intended in the specified use conditions. Thus, the optional feature added to these chairs produce comparable results to the predicate devices for the same intended use. The devices are substantially equivalent to their predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.