K160021 · Varian Medical Systems, Inc. · JAQ · Apr 13, 2016 · Radiology
Device Facts
Record ID
K160021
Device Name
Varian Sterilization Boxes
Applicant
Varian Medical Systems, Inc.
Product Code
JAQ · Radiology
Decision Date
Apr 13, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5700
Device Class
Class 2
Indications for Use
The Varian Sterilization boxes are intended to securely store the applicator parts during steam sterilization process and subsequent storage.
Device Story
Varian Sterilization Boxes are aluminum containers with hinged lids and fasteners; designed for secure storage and steam sterilization of radiation brachytherapy applicator components. Boxes feature perforated top and bottom surfaces to facilitate steam penetration during pre-vacuum autoclave cycles; internal silicone mats or supports hold applicator parts in place. Used in hospital reprocessing units by sterile processing staff. Boxes do not provide a sterile barrier; must be used with sterilization wrap. Device is non-invasive, non-patient contacting, and reusable for up to 500 steam sterilization cycles. Benefits include organized, secure storage and protection of delicate applicator components during the sterilization process.
Clinical Evidence
No clinical data included. Bench testing only. Testing demonstrated device durability through 500 steam autoclave cycles, compatibility with specified sterilization parameters (132°C/4 min or 134°C/3-5 min), and ability to store/protect applicator components without damage. Usability evaluated per IEC 62366:2007.
Technological Characteristics
Materials: Aluminum alloy, silicone, stainless steel. Design: Rectangular box with hinged lid, latch, and perforated surfaces (17-19% perforation). Energy: Passive (steam autoclave). Connectivity: None. Sterilization: Steam (pre-vacuum autoclave). Non-sterile, reusable (500 cycles). Non-invasive, no patient contact. Dimensions: Various models (e.g., 399-408mm x 259-262mm x 75mm).
Indications for Use
Indicated for secure storage of applicator parts during steam sterilization and subsequent storage. No specific patient population or contraindications as the device is non-invasive and has no patient contact.
Regulatory Classification
Identification
A remote controlled radionuclide applicator system is an electromechanical or pneumatic device intended to enable an operator to apply, by remote control, a radionuclide source into the body or to the surface of the body for radiation therapy. This generic type of device may include patient and equipment supports, component parts, treatment planning computer programs, and accessories.
Predicate Devices
VariSource Endometrial Applicator for VariSource Remote High Dose Rate Afterloader (K980576)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized design featuring three profiles that resemble human faces, stacked on top of each other and connected by a flowing line.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 13, 2016
Varian Medical Systems, Inc. % Mr. Peter Coronado Director, Regulatory Affairs 911 Hansen Way PALO ALTO CA 94304
Re: K160021
Trade/Device Name: Varian Sterilization Boxes Regulation Number: 21 CFR 892.5700 Regulation Name: Remote controlled radionuclide applicator system Regulatory Class: II Product Code: JAO Dated: March 9, 2016 Received: March 11, 2016
Dear Mr. Coronado:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160021
Device Name Varian Sterilization Boxes
Indications for Use (Describe)
The Varian Sterilization boxes are intended to securely store the applicator parts during steam sterilization process and subsequent storage.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Varian Medical Systems. The logo consists of the word "VARIAN" in a serif font, with a stylized "I" that resembles a person. Below the word "VARIAN" is the phrase "medical systems" in a smaller, sans-serif font. The logo is simple and professional, and it conveys the company's focus on medical technology.
Varian Medical Systems, Inc. 3100 Hansen Way Palo Alto, CA 94304-1038 USA Tel +1 650 493 4000 www.varian.com
# Premarket Notification [510(k)] Summary
## Varian Sterilization Boxes
The following information is provided following the format of 21 CFR 807.92
| Submitter's Name: | Varian Medical Systems, Inc.<br>3100 Hansen Way E-110<br>Palo Alto, CA 94304<br><br>Contact Name: Peter J. Coronado<br>Phone: 650.424.6320<br>Fax: 650.646.9200<br><br>Date: February 22, 2016 |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary Name: | Varian Sterilization Boxes |
| Classification Name: | Remote controlled radionuclide applicator system<br>21 CFR 892.5700, JAQ, Class II |
| Common/Usual Name: | Remote controlled radionuclide applicator system |
| Predicate Devices: | [K980576] VariSource Endometrial Applicator for VariSource Remote High Dose<br>Rate Afterloader (predicate is the Sterilization Tray which is a component of the<br>applicator set). |
| Device Description: | The Varian Sterilization boxes consist of an aluminium box with hinged lid and<br>fasteners. There are arrays of holes in the top and bottom surfaces to allow<br>evacuation during pre-vacuum processes. In the Universal Sterilization Box there is<br>a perforated plastic inlay to allow placement of the components or devices to be<br>sterilized. In the device-specific boxes there are supports to hold the components. |
| Intended Use Statement | The Varian Sterilization boxes are intended to securely store the applicator parts<br>during steam sterilization process and subsequent storage. |
| Indications for Use<br>Statement | The Varian Sterilization boxes are indicated to securely store the applicator parts<br>during steam sterilization process and subsequent storage. |
{4}------------------------------------------------
## Technological Characteristics:
| | VARIAN ENDOMETRIAL APPLICATOR FOR<br>VARISOURCE REMOTE HIGH DOSE RATE<br>AFTERLOADER K980576<br><br>Predicate: Sterilization Tray GM11009940<br>(Component of VariSource Endometrial<br>Applicator Set K980576) | Varian Sterilization Boxes<br>GM11010300, GM11010310<br>GM11010320, GM11010340 |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | The Varian VariSource Remote High Dose<br>Rate Afterloader [system, including<br>applicators and accessories] is a device<br>inteded to be used by properly trained<br>and licensed medical personnel to provide<br>radiation brachytherapy. The VariSource<br>Endometrial Applicator which is the<br>subject of this 510(k) is a component of<br>the VariSource system. | The Varian Sterilization boxes are intended<br>to securely store the applicator parts during<br>steam sterilization process and subsequent<br>storage. |
| | The sterilization tray is intended to<br>securely store the applicator parts during<br>steam sterilization process and<br>subsequent storage.<br><br>[The intended use for the sterilization tray<br>isn't specifically cleared however it is a<br>component of the cleared applicator set<br>K980576 and this is the intended use for<br>the component at that time.] | |
| | | |
| Indications for Use | The Varian VariSource Remote High Dose<br>Rate Afterloader [system, including<br>applicators and accessories] is a device<br>inteded to be used by properly trained<br>and licensed medical personnel to provide<br>radiation brachytherapy. The VariSource<br>Endometrial Applicator which is the<br>subject of this 510(k) is a component of<br>the VariSource system. | |
| | The sterilization tray is intended to<br>securely store the applicator parts during<br>steam sterilization process and<br>subsequent storage. | The Varian Sterilization boxes are intended<br>to securely store the applicator parts during<br>steam sterilization process and subsequent<br>storage. |
| | [The indications for use for the<br>sterilization tray aren't specifically cleared<br>however it is a component of the cleared<br>applicator set K980576 and this is the<br>indications for use for the component at<br>that time.] | |
| Design | The box has rectangular base with a latch<br>able lid. The box has perforations on the<br>lid and bottom to allow sterilant<br>penetration. | The box has rectangular base with a latch<br>able lid. The box has perforations on the lid<br>and bottom to allow sterilant penetration. |
| Materials | Aluminum alloy, silicone, stainless steel | Aluminum alloy, silicone, stainless steel |
| Physical Properties | Dimensionally stable<br>No water absorption | Dimensionally stable<br>No water absorption |
| Chemical Properties | N/A, no polymeric composites used as<br>packing materials. | N/A, no polymeric composites used as<br>packing materials. |
| Configuration/<br>Dimension | L400mm x W260mm x H70mm<br>Volume = 7280cm3 | GM11010300:<br>L399mm x W259mm x H75mm<br>Volume = 7751 cm3<br>GM11010310:<br>L408mm x W262mm x H75mm<br>Volume = 8017 cm3<br>GM11010320:<br>L399mm x W259mm x H75mm<br>Volume = 7751 cm3<br>GM11010340:<br>L407mm x W262mm x H75mm.<br>Volume = 7997 cm3 |
| Air permeance | N/A, the box does not provide a sterile<br>barrier. Sterilization wrap has to be used. | N/A, the boxes does not provide a sterile<br>barrier. Sterilization wrap has to be used. |
| Percent of surface perforations | 11% | GM11010300: 18%<br>GM11010310: 17%<br>GM11010320: 18%<br>GM11010340: 19% |
| Method of holding applicators in place | Brackets | Silicone mat upon which items can be placed |
| Sterility | Provided non-sterile | Provided non-sterile |
| Sterilization method | Steam sterilization up to 500 times<br>Pre-vacuum autoclave cycles have been validated for use in the health care facilities as follows:<br>132°C (270°F) exposure temperature set-point, 4 minute exposure time, 25 minute drying time<br>134°C (273°F) exposure temperature set-point, 3 minute exposure time, 25 minute drying time<br>134°C (273°F) exposure temperature set-point, 5 minute exposure time, 25 minute drying time | Steam sterilization up to 500 times<br>Pre-vacuum autoclave cycles have been validated for use in the health care facilities as follows:<br>132°C (270°F) exposure temperature set-point, 4 minute exposure time, 25 minute drying time<br>134°C (273°F) exposure temperature set-point, 3 minute exposure time, 25 minute drying time<br>134°C (273°F) exposure temperature set-point, 5 minute exposure time, 25 minute drying time |
| Microbial Barrier Properties<br>(Packaging Integrity) | The device does not provide a sterile barrier. Sterilization wrap has to be used. Hospital is responsible for wrapping materials, as well as method and shelf life validation. | The device does not provide a sterile barrier. Sterilization wrap has to be used. Hospital is responsible for wrapping materials, as well as method and shelf life validation. |
| Material Compatibility | Compatible to steam and can be used 500 times. | Compatible to steam and can be used 500 times. |
| Toxicological Properties<br>(Biocompatibility, including Sterilant Residue Limits) | N/A, According to ISO 10993-1/FDA G-95-1 the Sterilization Box is a non-invasive product and is not intended to have any patient contact. | N/A, According to ISO 10993-1 / FDA G-95-1 the Sterilization Box is a non-invasive product and are not intended to have any patient contact. |
| Shelf Life | No sterile wrap is supplied with the sterilization box. | No sterile wrap is supplied with the sterilization box. |
| Drying time | 25 minutes | 25 minutes |
| Aeration time | N/A | N/A |
| Where used | Reprocessing units | Reprocessing units |
{5}------------------------------------------------
{6}------------------------------------------------
Non Clinical Tests Bench Testing has been performed to demonstrate that:
- the device can withstand the number of steam autoclave cycles specified in the instructions for use. .
- the device enables the components of the applicators to be stored to give easy access and does not damage them.
- . Sterilization validation has been performed on equivalent devices to demonstrate that the specified applicator components and accessories can be sterilized in the sterilization box.
{7}------------------------------------------------
- Usability was reviewed in comparison to other similar devices to the requirements of IEC 62366:2007.
- Results of Bench Testing showed conformance to applicable requirements and specifications.
Clinical Tests: No clinical tests have been included in this pre-market submission.
Conclusions: All the tests that were performed met the applied pass criteria. Varian considers the device to be safe and effective and to perform as well or better than the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.