Contact Lens Case is a lens care product to be used by the contact lens wearer or practitioner for storing soft(hydrophilic) contact lenses while not being worn. Not designed for heat disinfecting system. Use only with chemical disinfection.
Device Story
Lens storage container for soft (hydrophilic) contact lenses; used by wearers or practitioners. Device consists of two adjoining wells with either screw-top or integral-hinged caps. Input: contact lenses and chemical disinfection solution. Output: storage environment for lenses. Operation: manual placement of lenses into wells filled with chemical disinfectant. Benefit: provides secure, clean storage environment preventing lens dehydration and contamination. Used in home or clinical settings.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including leakage testing and biocompatibility evaluation (ISO 10993-5, ISO 10993-10, ISO 10993-11).
Technological Characteristics
Material: Polypropylene (PP). Design: Two adjoining wells with screw-top or integral-hinged caps. Capacity: >1.5 ml per well. Inner well height: >8.5 mm. Biocompatibility: Tested per ISO 10993-5, 10993-10, 10993-11. Non-sterile.
Indications for Use
Indicated for storage of soft (hydrophilic) contact lenses while not being worn by contact lens wearers or practitioners. Not for use with heat disinfection systems; use only with chemical disinfection.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or fabric. The profiles are arranged in a staggered formation, with the first profile being the largest and the third being the smallest. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 7, 2016
Danyang Chanxin Glasses Cases Factory c/o Mr. Mike Gu Regulatory Affairs Manager OSMUNDA Medical Device Consulting Co., Ltd Level 7, Jin Gui Business Center 982 Cunyun Road, Baiyun District, Guangzhou Guangdong, China 510420
Re: K160005
Trade/Device Name: Contact Lens Case (multiple brand names) Regulation Number: 21 CFR 886.5928 Regulation Name: Contact Lens Case Regulatory Class: Class II Product Code: LRX Dated: April 25, 2016 Received: April 28, 2016
Dear Mr. Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
Page 2 - Mr. Mike Gu
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR: Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR
regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or 301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K160005
Device Name Contact Lens Case
Indications for Use (Describe)
Contact Lens Case is a lens care product to be used by the contact lens wearer or practitioner for storing soft (hydrophilic) contact lenses while not being worn. Not designed for heat disinfecting system. Use only with chemical disinfection.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo with a stylized flower-like design on the left and the text "诚欣 Chan Xin" on the right. The flower-like design is composed of overlapping petals in different colors, including blue, red, yellow, and cyan. The text "诚欣 Chan Xin" is written in a simple, sans-serif font, with the Chinese characters "诚欣" appearing above the English text "Chan Xin".
# 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided:
### l. SUBMITTER
DANYANG CHANXIN GLASSES CASES FACTORY
HUADIAN CHUANGYE PARK, DANGUI ROAD, DEVELOPING ZONE, DANYANG CITY,
JIANGSUPROVINCE, CHINA
Phone: +86-511-86903193
Fax: +86-511-86903193
| Primary Contact Person: | Mike Gu |
|---------------------------|--------------------------------------------|
| | Regulatory Affairs Manager |
| | OSMUNDA Medical Device Consulting Co., Ltd |
| | Tel: (+86) 20-6232 1333 |
| | Fax: (+86) 20-8633 0253 |
| Secondary Contact Person: | Lemon Chen |
| | CEO |
| | DANYANG CHANXIN GLASSES CASES FACTORY |
| | Tel: +86-0511-86903193 |
| | Fax: +86- 0511-86903193 |
| | Email: lemon@jschanxin.com |
| Preparing date: | 30 Dec 2015 |
| II. | |
| DEVICE | |
| Name of Device: | Contact Lens Case |
| Common/Usual Name: | Contact Lens Case |
| Classification Names: | Case, Contact Lens |
| Regulation Class: | II |
| Product Code: | LRX |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains a logo with a stylized flower-like design on the left and the text "诚欣 Chan Xin" on the right. The flower-like design is composed of overlapping curved shapes in different colors, including blue, yellow, red, and black. The text "诚欣" is in Chinese characters, and "Chan Xin" is the corresponding English translation or transliteration, written in a smaller font size below the Chinese characters.
#### III. PREDICATE DEVICE
## Contact Lens Case(K071081)
This predicate has not been subject to a design-related recall. No reference devices were used in this submission.
### IV. DEVICE DESCRIPTION
Contact Lens Case is a lens care product to be used by the contact lens wearer or practitioner for storing soft(hydrophilic) contact lenses while not being worn. This device is not sterile and not for heat-disinfection. Use only with chemical disinfection.
There are two main series includes the variant model DC-2001, DC-3001 and DC-6001, the other includes DC-7001 and DC-8001. All these five variant models are made of Polypropylene, which follow the same design principle with the same intended use.
All the five variant models of this device have a capacity of over 1.5 ml in each case well. And the inner height of the all wells exceeds 8.5 mm. With regard to the Center Thickness of the normal hydrophilic and hydrophobic contact lens will not outnumber 8.5 mm, the capacity is sufficient for contact lens to be fully immersed under use condition.
### V. INDICATION FOR USE
Contact Lens Case is a lens care product to be used by the contact lens wearer or practitioner for storing soft(hydrophilic) contact lenses while not being worn. Not designed for heat disinfecting system. Use only with chemical disinfection.
### VI. SUBSTANTIAL EQUIVALENCES
- Technology: The proposed device Contact lens case has similar product design as the predicate device. The major differences are due to the different design for adjoining wells with integral hinged caps or screw-top caps, the different dimension and appearance, made of the PP material with different color additives. These are not relating to the safety or effectiveness aspects. Thus they are substantially equivalent.
Summary of Non-Clinical Leakage Testing: Production units of Contact Lens Cases have
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo with a stylized flower-like design on the left and Chinese characters along with the English words "Chan Xin" on the right. The flower-like design is composed of curved lines in various colors, including yellow, blue, red, and black. The Chinese characters are positioned above the English words, which are written in a simple, sans-serif font.
undergone leakage testing. None of the tested lens cases Tests: showed any leakage and all of the Leakage Tests were passed successfully.
Summary of Clinical Tests: The subject of this premarket submission, Contact lens case did not require clinical studies to support substantial equivalence.
The following comparison table identifies the similarities and differences of the proposed device Contact Lens Case to the legally marketed predicate device Contact Lens Case (k071081) to which substantial equivalency is claimed.
| Items | Predicate Device | Proposed Device | Discussion of<br>Differences |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| | CONTACT LENS CASE | CONTACT LENS CASE | |
| | K071081 | | |
| Manufacturer | Ningbo Kaida Rubber &<br>Plastic Technology Co.,Ltd. | DANYANG<br>CHANXIN<br>GLASSES CASES FACTORY | --- |
| Device name | Multiple Brand Names | Contact Lens Case | --- |
| Classification Name | Contact Lens Case | Contact Lens Case | Identical |
| Product Code | LRX | LRX | Identical |
| Intended Use | The contact lens case is a<br>lens care product to be<br>used by the contact lens<br>wearer or practitioner for<br>storing contact lenses<br>while not being worn. The<br>device is not designed for<br>heat disinfecting system.<br>Use only with chemical<br>disinfection. | Contact Lens Case is a lens<br>care product to be used by<br>the contact lens wearer or<br>practitioner for storing<br>soft(hydrophilic) contact<br>lenses while not being<br>worn. Not designed for<br>heat disinfecting system.<br>Use only with chemical<br>disinfection. | Equivalent |
| Indications for Use | Storage and Disinfection of<br>Soft, Rigid Gas permeable<br>or Hard contact lenses<br>while not being worn | Storage and Disinfection of<br>Soft contact lenses while<br>not being worn | Equivalent |
| Disinfection Type | Chemical Disinfection<br>Not Heat-Disinfection | Chemical Disinfection<br>Not Heat-Disinfection | Identical |
| Items | Predicate Device | Proposed Device | Discussion of |
| | CONTACT LENS CASE | CONTACT LENS CASE | Differences |
| | K071081 | | |
| Design | Two adjoining Wells with<br>Screw-Top Caps | Two adjoining Wells with<br>Screw-Top Caps: DC-2001,<br>DC-3001 and DC-6001<br>Two adjoining Wells with<br>Integral-Hinged<br>Caps:<br>DC-7001 and DC-8001 | Equivalent |
| Main Materials | Polypropylene(PP) | Polypropylene(PP) | Equivalent |
| Dimension | Each case well has a<br>capacity of over 1.5 ml.<br>The inner height of the<br>wells exceeds 8.5mm. | Each case well has a<br>capacity of over 1.5 ml.<br>The inner height of the<br>wells exceeds 8.5mm. | Equivalent |
| Color | Purple | colorless, transparent<br>blue, transparent violet,<br>blue, violet, and white | Different,<br>the colors<br>of proposed<br>device have<br>been<br>demonstrat<br>ed to be<br>biocompati<br>ble safe. |
| Effectiveness | The capacity is sufficient<br>for contact lens to be fully<br>immersed<br>under<br>use<br>condition. | The capacity is sufficient<br>for contact lens to be fully<br>immersed<br>under<br>use<br>condition. | Identical |
| Safety | ISO 10993-5 Biological<br>evaluation of medical devices<br>-- Part 5: Tests for in vitro<br>cytotoxicity;<br>ISO 10993-10 Biological<br>evaluation of medical devices<br>-- Part 10: Tests for irritation<br>and skin sensitization<br>ISO 10993-11 Biological<br>evaluation of medical devices<br>-- Part 11: Tests for systemic<br>Toxicity. | ISO 10993-5 Biological<br>evaluation of medical devices<br>-- Part 5: Tests for in vitro<br>cytotoxicity;<br>ISO 10993-10 Biological<br>evaluation of medical devices<br>-- Part 10: Tests for irritation<br>and skin sensitization<br>ISO 10993-11 Biological<br>evaluation of medical devices<br>-- Part 11: Tests for systemic<br>Toxicity. | Identical |
| Sterility | No | No | Identical |
Comparison Table of Predicate device and Proposed Device
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains a logo with a stylized flower-like design on the left and the text "Chan Xin" in both Chinese and English on the right. The flower-like design is composed of four curved shapes in different colors: blue, yellow, red, and black. The text "Chan Xin" is written in a simple, sans-serif font, with the Chinese characters above the English text.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains a logo with a stylized flower-like design on the left and the text "Chan Xin" on the right. The flower-like design is composed of overlapping curved lines in various colors, including blue, yellow, red, and teal. The text "Chan Xin" is written in a serif font, with the first word in a larger, bolder font than the second word.
### VII. CONCLUSION
DANYANG CHANXIN GLASSES CASES FACTORY considers the Contact lens case to be as safe, as effective, and performance is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.