← Product Code [NHV](/productcode/NHV) · K153784

# NOSOGARD FILTERS (K153784)

_Evoqua Water Technologies, LLC · NHV · Oct 21, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153784

## Device Facts

- **Applicant:** Evoqua Water Technologies, LLC
- **Product Code:** [NHV](/productcode/NHV.md)
- **Decision Date:** Oct 21, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5665
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The NOSOGARD line of filters are intended to operate on EPA quality drinking water sources as a microbial retention filter. NOSOGARD filters are available in models suitable for the control of bacteria equal to or greater in size than Brevundimons diminuta and models for microbial removal equal to or greater than Legionella pneumophila. The NOSOGARD is suitable for general point of use infection control for procedures such as cleaning, rinsing of equipment, hand washing and bathing where the reduction of such microorganisms in the water is desired. The Nosogard is not intended for use in the production of USP sterile water for use in infusion, injection of fluids for use in dialysis treatments.

## Device Story

Nosogard filters are disposable point-of-use microbial retention devices; installed by facility personnel on faucets or as handheld shower heads. Input is EPA-quality drinking water; output is filtered water with reduced bacterial load. Filters utilize polyethersulfone (PES) pleated membranes within polypropylene housings to physically retain bacteria. Device provides infection control for cleaning, rinsing, hand washing, and bathing. Benefits include reduced risk of waterborne bacterial exposure for patients and staff. No electronic components or software; purely mechanical filtration.

## Clinical Evidence

Bench testing only. Performance validated using ASTM F838 for bacterial retention (Brevundimonas diminuta and Legionella pneumophila) with >10^7 CFU/cm² challenge. Testing included pressure integrity, drop testing, cycle/simulated use testing, and shelf-life verification. Biocompatibility confirmed per USP Class VI and ISO 10993 standards. Sterilization validated per ANSI/AAMI/ISO 11137-2.

## Technological Characteristics

Disposable filter with polyethersulfone (PES) pleated membrane in polypropylene housing. Mechanical filtration principle. No energy source. Various form factors (faucet attachment, handheld shower wand). Sterilized via gamma irradiation. Biocompatibility per ISO 10993 and USP <87>/<88>. Standards: ASTM F838 (bacterial retention), ISO 11607 (packaging).

## Regulatory Identification

A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

## Predicate Devices

- Nephros Inc. – DSU-HTM and SSU-HTM Filters ([K141731](/device/K141731.md))
- PrisMedical - MainStreamTM Water Purification Device ([K012716](/device/K012716.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 21, 2016

Evoqua Water Technologies, LLC Robert Dudek Regulatory Manager 10 Technology Drive Lowell, Massachusetts 01851

Re: K153784

Trade/Device Name: Nosogard Filters Regulation Number: 21 CFR 876.5665 Regulation Name: Water Purification System For Hemodialysis Regulatory Class: Class II Product Code: NHV Dated: September 11, 2016 Received: September 20, 2016

Dear Robert Dudek:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Michael J. Ryan -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K153784

Device Name NOSOGARD FILTERS

#### Indications for Use (Describe)

The NOSOGARD line of filters are intended to operate on EPA quality drinking water sources as a microbial retention filter. NOSOGARD filters are available in models suitable for the control of bacteria equal to or greater in size than Brevundimons diminuta and models for microbial removal equal to or greater than Legionella pneumophila. The NOSOGARD is suitable for general point of use infection control for procedures such as cleaning, rinsing of equipment, hand washing and bathing where the reduction of such microorganisms in the water is desired.

The Nosogard is not intended for use in the production of USP sterile water for use in infusion, injection of fluids for use in dialysis treatments.

Type of Use (Select *one* or *both*, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Evoqua Water Technologies. The logo consists of a stylized water droplet above the company name. The water droplet is light blue and appears to be made of overlapping shapes. The company name, "evoqua," is in a sans-serif font, and the words "WATER TECHNOLOGIES" are in a smaller font below the company name.

# 510k Summary (K153784) of the Nosogard™ Filters

| Submitter:                        | Evoqua Water Technologies, LLC.<br>10 Technology Drive<br>Lowell, MA 01851<br>Establishment Registration - Pending                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                          | Robert Dudek<br>Evoqua Water Technologies, LLC.<br>10 Technology Drive<br>Lowell, MA 01851<br>E-Mail: Robert.Dudek@Evoqua.com<br>Phone: 978-614-7359<br>Fax: 978-441-6037                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Date Prepared                     | September 11, 2016 (amended 10-20-2016)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Trade Name                        | Nosogard Filters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Proposed Class                    | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Classification Name<br>and Number | 21 CFR Part 876.5665 Water Purification Device,<br>General Medical Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Product Code                      | NHV                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Predicate Device 1                | Nephros Inc. – DSU-HTM and SSU-HTM Filters – K141731                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Predicate Device 2                | PrisMedical - MainStreamTM Water Purification Device - K012716                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Device Description                | Nosogard Filters are disposable Polyethersulfone (PES) membrane<br>filters sealed in a polypropylene housing designed to be installed by<br>facility personnel at point of use locations where the control of<br>bacteria in the drinking water supply is desired for device cleaning,<br>handwashing and bathing applications. These filters are designed to<br>be installed as attachments to faucets or as a handheld shower<br>head.                                                                                                                                                                                                                                                                                                                                           |
| Intended Use                      | The NOSOGARD line of filters are intended to operate on EPA quality<br>drinking water sources as a microbial retention filter. NOSOGARD<br>filters are available in models suitable for the control of bacteria equal<br>to or greater in size than Brevundimons diminuta and models for<br>microbial removal equal to or greater than Legionella peumophila. The<br>NOSOGARD is suitable for general point of use infection control for<br>procedures such as cleaning, rinsing of equipment, hand washing and<br>bathing where the reduction of bacterial microorganisms in the water is<br>desired.<br>The Nosogard is not intended for use in the production of USP sterile<br>water for use in infusion, injection or production of fluids for use in<br>dialysis treatments. |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Evoqua Water Technologies. The logo features a stylized water droplet above the company name. The words "WATER TECHNOLOGIES" are written in a smaller font below the company name.

|                               | Evoqua Nosogard Filters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Predicate Device 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|-------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                 | K153784                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | K141731                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | K012716                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Manufacturer<br>Name          | Evoqua Water<br>Technologies, LLC.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Nephros Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | PrisMedical                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Indications<br>for Use        | The Nosogard Filters are<br>intended to operate on EPA<br>quality drinking water<br>sources as a microbial<br>retention filter and are<br>suitable for general point of<br>use infection control for<br>procedures such as<br>cleaning, rinsing of<br>equipment, hand washing<br>and bathing where the<br>reduction of bacterial<br>microorganisms in the water<br>is desired<br>The Nosogard is not<br>intended for use in the<br>production of USP sterile<br>water for use in infusion,<br>injection or production of<br>fluids for use in dialysis<br>treatments. | The DSU-H and<br>SSU-H Ultrafilters are<br>intended to be used<br>to filter EPA quality<br>drinking water. The<br>filters retain bacteria,<br>viruses and<br>endotoxin. By<br>providing ultrapure<br>water for washing<br>and drinking, the<br>filters aid in infection<br>control. The filters<br>produce water that is<br>suitable for wound<br>cleansing, cleaning of<br>equipment used in<br>medical procedures<br>and washing of<br>surgeon's hands. The<br>filters are not<br>intended to provide<br>water that can be<br>used as a substitute<br>for USP sterile water. | To produce from<br>EPA grade drinking<br>water, sterile<br>purified water to be<br>used within 24 hours<br>of collection that is<br>suitable for: -<br>Cleaning and rinsing<br>open wounds -<br>Infection control<br>(cleaning equipment<br>used in medical<br>procedures, medical<br>personnel's hands) -<br>Use as a diluent for<br>enteral, nutritional,<br>oral vaccine, or oral<br>drug preparations -<br>All other uses of<br>sterile purified water<br>the practitioner or<br>clinician deems<br>necessary -Not for<br>parenteral<br>administration |
| Feed Water<br>Source          | Drinking water plumbing at<br>point of use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | In-Line plumbing                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Feed Container<br>(tank)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                               | Evoqua Nosogard Filters<br>K153784                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate Device 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Predicate Device 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Feed Water<br>Quality         | Potable Water as defined<br>by the EPA National<br>Primary Drinking Water<br>Regulations (EPA Quality<br>Water)                                                                                                                                                                                                                                                                                                                                                                                                                                                       | EPA Quality Water                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | EPA Grade Water                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Useful Life                   | Up to 7 days, 31 days or 62<br>days depending on model                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | up to 3 Mo. (SSU-H)<br>up to 6 Mo. (DSU-H)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | 3 Liters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Maximum Inlet<br>Pressure     | 89.9 psi (pounds / square<br>inch)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 75 psi (SSU-H)<br>100 psi (DSU-H)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Gravity feed (< 10 psi)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Sterile Device                | Shipped in sterile condition<br>and package                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Casing                        | Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | ABS Plastic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Polycarbonate<br>Plastic                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Bacteria Filter<br>Element(s) | Polyethersulfone (PES)<br>pleated membrane                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Polysulfone<br>Ultrafiltration Hollow<br>Fiber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Polyethersulfone flat<br>membrane                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Performance<br>Claims         | Evoqua Nosogard Filters                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device 1                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Predicate Device 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Bacteria<br>Reduction         | Greater than 107                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Greater than 1011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Greater than 107                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Virus Reduction               | No performance claims                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Greater than 108                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Greater than 104                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Endotoxin<br>Reduction        | No performance claims                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Greater than 105                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Greater than 104                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Organic<br>Reduction          | No performance claims                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | TOC reduced to less<br>than 1 ppm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Ion Reduction                 | None                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Greater than 103<br>dissociable ions                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Evoqua Water Technologies. The logo features a stylized water droplet in a light blue color above the company name. The company name, "evoqua," is written in a dark blue sans-serif font, with the words "WATER TECHNOLOGIES" in a smaller, lighter blue font below it.

Table 1 - Comparison to predicate devices

### Device Description & Summary of Technological Characteristics:

The Nosogard Filter are disposable Polyethersulfone (PES) membrane filters sealed in a polypropylene housing designed to be installed by facility personnel at point of use locations where the control of bacteria in the EPA quality drinking water supply is desired for device cleaning, handwashing and bathing applications. These filters are designed to be installed as attachments to faucets or as a handheld shower head attached to a water outlet. Further details of the technology may be located in the 510k submission Section 12, Materials and Design. The filter is not intended or marketed as a device to reduce endotoxin or virus in general washing applications and the difference to predicate devices is not applicable. Biocompatibility testing (located in the 510k submission Section 16, Biocompatibility) demonstrates that all materials are safe for use in this

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Evoqua Water Technologies. The logo consists of a stylized water droplet above the company name. The water droplet is light blue and appears to be formed by two curved shapes. The company name, "evoqua," is written in a dark teal sans-serif font, and below it, in smaller light teal letters, are the words "WATER TECHNOLOGIES."

application and are free of harmful extractable or leachable chemicals. These materials of construction are commonly used in other medical devices that have been previously approved for use in products regulated under 21CFR876.5665

The product models are based on the anticipated useful life of the product in specific roles. These are identified as filters suitable for maximum use periods of 7 days (type Nosoqard 7/31 Filter), 31 days (Nosoqard 31 Day Filter) or 62 days (Nosoqard 62 Day Filter) depending on the size of the bacterial component targeted for removal and other particulate matter in the source water. Internal differences in membrane arrangement, density of pleating and pre-filtration elements within the cartridge determine the maximum anticipated useful life of the filter. The Nosogard line of filters are offered with a choice of style of connection to water source, outlet flow type and both a large and normal size 7/31 model. In addition handheld shower wands are available in 31 and 62 day models. Each model is identified by a unique part number and information regarding application is also identified on the product labeling.

| Part<br>Number | Description                         | Inlet Connection | Outlet Flow | Microbial Performance                |                                       |
|----------------|-------------------------------------|------------------|-------------|--------------------------------------|---------------------------------------|
|                |                                     |                  |             | Brevundimons diminuta<br>(or larger) | Legionella pneumophila<br>(or larger) |
| W2T42138       | Nosogard 7/31 Filter (1.25" size)   | Quick Connect    | Stream      | Not to exceed 7 days                 | Not to exceed 31 days                 |
| W2T42134       | Nosogard 7/31 Filter (1.25" size)   | Quick Connect    | Shower      | Not to exceed 7 days                 | Not to exceed 31 days                 |
| W2T42138       | Nosogard 7/31 Filter (2.5" size)    | Quick Connect    | Stream      | Not to exceed 7 days                 | Not to exceed 31 days                 |
| W2T42138       | Nosogard 7/31 Filter (2.5" size)    | Quick Connect    | Shower      | Not to exceed 7 days                 | Not to exceed 31 days                 |
| W2T42138       | Nosogard 7/31 Filter (2.5" size)    | Quick Connect    | 1/2" NPT    | Not to exceed 7 days                 | Not to exceed 31 days                 |
| W2T82022       | Nosogard 31 Day Filter (1.25" size) | Quick Connect    | Shower      | Not to exceed 31 days                | Not to exceed 31 days                 |
| W2T82022       | Nosogard 31 Day Filter (1.25" size) | Quick Connect    | Stream      | Not to exceed 31 days                | Not to exceed 31 days                 |
| W2T82022       | Nosogard 62 Day Filter (1.25" size) | Quick Connect    | Shower      | Not to exceed 62 days                | Not to exceed 62 days                 |
| W2T82022       | Nosogard 62 Day Filter (1.25" size) | Quick Connect    | Stream      | Not to exceed 62 days                | Not to exceed 62 days                 |
| W2T82022       | Nosogard 31 Day Shower Wand         | 1/2" NPT         | Shower      | Not applicable                       | Not to exceed 31 days                 |
| W2T82023       | Nosogard 31 Day Shower Wand         | 1/2" NPT         | Shower      | Not to exceed 31 days                | Not to exceed 31 days                 |
| W2T82023       | Nosogard 62 Day Shower Wand         | 1/2" NPT         | Shower      | Not to exceed 62 days                | Not to exceed 62 days                 |

#### NOSOGARD® FILTER MODELS

Table 2 – Nosoqard Models & Part Numbers

# Non-Clinical Performance Testing

Results reflected in summary Table 3, Nosogard Performance Testing are contained in the submission sections titled Biocompatibility and Bench Testing in full detail.

All filter designs were placed in test assemblies and operated continuously for a period at least equal to the useful life of the model with repetitive inoculations of the subject microbe. Testing for bacterial reduction rates of 10 to the 7th were demonstrated successful at the time of installation and remained effective throughout the use period stated for each model using the ASTM F838 Standard Test Method for Determining Bacterial Retention of Membrane Filters Utilized for Liguid Filtration. This reduction rate testing and ASTM F838 standard was also utilized to demonstrate bacterial grow through of the membrane did not occur during the stated use period for each model.

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for Evoqua Water Technologies. The logo features a stylized water droplet above the company name. The droplet is light blue and appears to be formed by two curved lines. Below the company name, in smaller font, are the words "WATER TECHNOLOGIES".

All filter designs and housings were tested to pressures equal to or greater than stated operating limits to demonstrate the Nosogard filter retains integrity at the stated operating limits. Integrity testing and pressure testing was conducted using the protocols in ASTM F838 Standard Test Method for Determining Bacterial Retention of Membrane Filters Utilized for Liquid Filtration.

Drop testing was conducted to demonstrate that the filter was not compromised at the stated maximum operating pressure and maximum operating temperatures after being subjected to a 6 foot unprotected drop. Integrity was verified with membrane diffusional flow testing.

Simulated use testing (cycle testing) demonstrates the Nosogard filter successfully retains integrity after exposure to repeated pressurization, flow and depressurization cycles at maximum stated temperatures and pressure. Integrity was verified with diffusional flow testing.

Sterilization testing conducted in accordance with ANSI/AAMI/ISO 11137-2:2012 Sterilization of health care products was conducted post gamma irradiation and demonstrates that the dosage and exposure utilized is adequate to assure a sterile product prior to release.

Testing of the packaging utilizing ISO 11607:2009 Packaging for terminally sterilized medical devices demonstrated the packaging is capable of maintaining a sterile barrier for a period of three years.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Evoqua Water Technologies. The logo features a stylized water droplet in a light blue color above the company name. The company name, "evoQUA," is written in a sans-serif font, with the "Q" stylized to resemble a water wave. Below the company name, the words "WATER TECHNOLOGIES" are written in a smaller, sans-serif font.

| Nosogard Performance Testing |                                    |                                                                                                                                                       |                    |                                                        |                                                               |                                                                                         |                                                                                       |
|------------------------------|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------|--------------------------------------------------------|---------------------------------------------------------------|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Nosogard<br>Model            | Parameter                          | Minimum Objective of Non-<br>Clinical Testing & Materials for<br>Nosogard Brand Filters                                                               | Passing<br>Result? | Nosogard<br>Model                                      | Parameter                                                     | Minimum Objective of Non-<br>Clinical Testing & Materials for<br>Nosogard Brand Filters | Passing<br>Result?                                                                    |
| Nosogard 7/31<br>Filter      | 7day, B.dim grow<br>through.       | No grow through ≤7days when<br>challenge with >107 cfu/cm²<br>Brewundimonas diminuta                                                                  | Passed             | NOSOGARD 31<br>Day shower wand<br>(handheld<br>shower) | 31day, legionella<br>grow through.                            | No grow through ≤31days when<br>challenge with >107 cfu/cm²<br>Legionella pneumophila   | Passed                                                                                |
|                              | 31day, legionella<br>grow through. | No grow through ≤31days when<br>challenge with >107 cfu/cm²<br>Legionella pneumophila                                                                 | Passed             |                                                        | Flow Rate                                                     | 0.79ft³/min at 44psi inlet pressure                                                     | Passed                                                                                |
|                              | Flow rate                          | 1.25" = 0.22ft³/min at 44psi inlet<br>pressure<br>2.5" = 0.40ft³/min at 29psi inlet<br>pressure                                                       | Passed             |                                                        | Housing burst<br>pressure                                     | 10Barg at 40°C                                                                          | Passed                                                                                |
|                              | Housing burst<br>pressure          | 10Barg at 40°C                                                                                                                                        | Passed             |                                                        | Pulse testing                                                 | Integral post 124 cycles x 7mins x 0-<br>5Barg pulses at 40°C                           | Passed                                                                                |
|                              | Pulse testing                      | 7day:<br>Integral post 2000 cycles x 5mins x<br>0-6.2Barg pulses at 66°C<br>31day:<br>Integral post 2000 cycles x 5mins x<br>0-6.2Barg pulses at 66°C | Passed             |                                                        | Shelf life                                                    | Packaging forms a sterile barrier for<br>at least 3 years.                              | Passed                                                                                |
|                              | Shelf life                         | Packaging forms a sterile barrier for<br>at least 3 years.                                                                                            | Passed             |                                                        | Biocompatibility                                              | Meets requirements in FDA<br>memorandum #G95-1                                          | Passed                                                                                |
|                              | Biocompatibility                   | Meets requirements in FDA<br>memorandum #G95-1                                                                                                        | Passed             | NOSOGARD 31<br>Day shower wand<br>(handheld<br>shower) | 31day, B.dim<br>grow through.                                 | No grow through ≤31days when<br>challenge with >107 cfu/cm²<br>Brewundimonas diminuta   | Passed                                                                                |
|                              |                                    |                                                                                                                                                       |                    |                                                        | Flow Rate                                                     | 0.32ft³/min at 44psi inlet pressure                                                     | Passed                                                                                |
| NOSOGARD 31<br>Day Filter    | 31day, B.dim<br>grow through.      | No grow through ≤31days when<br>challenge with >107 cfu/cm²<br>Brewundimonas diminuta                                                                 | Passed             |                                                        | Housing burst<br>pressure                                     | 10Barg at 40°C                                                                          | Passed                                                                                |
|                              | Flow Rate                          | 0.19ft³/min at 44psi inlet pressure                                                                                                                   | Passed             |                                                        | Pulse testing                                                 | Integral post 124 cycles x 7mins x 0-<br>5Barg pulses at 40°C                           | Passed                                                                                |
|                              | Housing burst<br>pressure          | 10Barg at 40°C                                                                                                                                        | Passed             |                                                        | Shelf life                                                    | Packaging forms a sterile barrier for<br>at least 3 years.                              | Passed                                                                                |
|                              | Pulse testing                      | Integral post 2000 cycles x 5mins x<br>0-6.2Barg pulses at 66°C                                                                                       | Passed             |                                                        | Biocompatibility                                              | Meets requirements in FDA<br>memorandum #G95-1                                          | Passed                                                                                |
|                              | Shelf life                         | Packaging forms a sterile barrier for<br>at least 3 years.                                                                                            | Passed             |                                                        | NOSOGARD 62<br>Day shower wand<br>(handheld<br>shower)        | 62day, B.dim<br>grow through.                                                           | No grow through ≤62days when<br>challenge with >107 cfu/cm2<br>Brewundimonas diminuta |
|                              | Biocompatibility                   | Meets requirements in FDA<br>memorandum #G95-1                                                                                                        | Passed             |                                                        |                                                               | Flow Rate                                                                               | 0.39ft³/min at 44psi inlet pressure                                                   |
| NOSOGARD 62<br>Day Filter    | 62day, B.dim<br>grow through.      | No grow through ≤62days when<br>challenge with >107 cfu/cm²<br>Brewundimonas diminuta                                                                 | Passed             | Housing burst<br>pressure                              |                                                               | 10Barg at 40°C                                                                          | Passed                                                                                |
|                              | Flow Rate                          | 0.25ft³/min at 44psi inlet pressure                                                                                                                   | Passed             | Pulse testing                                          | Integral post 248 cycles x 7mins x 0-<br>5Barg pulses at 40°C | Passed                                                                                  |                                                                                       |
|                              | Housing burst<br>pressure          | 10Barg at 40°C                                                                                                                                        | Passed             | Shelf life                                             | Packaging forms a sterile barrier for<br>at least 3 years.    | Passed                                                                                  |                                                                                       |
|                              | Pulse testing                      | Integral post 4000 cycles x 5mins x<br>0-6.2Barg pulses at 66°C                                                                                       | Passed             | Biocompatibility                                       | Meets requirements in FDA<br>memorandum #G95-1                | Passed                                                                                  |                                                                                       |
|                              | Shelf life                         | Packaging forms a sterile barrier for<br>at least 3 years.                                                                                            | Passed             |                                                        |                                                               |                                                                                         |                                                                                       |
|                              | Biocompatibility                   | Meets requirements in FDA<br>memorandum #G95-1                                                                                                        | Passed             |                                                        |                                                               |                                                                                         |                                                                                       |

# Biocompatibility Testing Results

Biocompatibility of Nosogard Filters was tested in the form of fully assembled products and passed the following tests for biocompatibility:

- 1. General Chapter <88> of the United States Pharmacopeia and National Formulary, Section USP Class VI Plastics
	- a. Intracutaneous Reactivity Testing
	- b. Acute Systemic Injection Testing

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Evoqua Water Technologies. The logo features a stylized water droplet in shades of blue above the company name. The company name is written in a modern, sans-serif font, with "evoqua" in a darker blue and "WATER TECHNOLOGIES" in a lighter blue below it.

- c. Intramuscular Implant Testing
- 2. General Chapter <87> of the United States Pharmacopeia, Biological Reactivity Tests, In Vitro
	- a. Minimal Essential Media Elution Test for Cytotoxicity
- 3. ISO 10993:2010 Biological evaluation of medical devices Part 5:
	- a. Part 5: Tests for in vitro cytotoxicity
	- b. Part 10: Tests for irritation and Skin Sensitization

## Summary of Standards Utilized in the Design and Performance Evaluation of the Nosogard

- General Chapter <88> of the United States Pharmacopeia and National ● Formulary, Section USP Class VI Plastics
- General Chapter <87> of the United States Pharmacopeia, Biological . Reactivity Tests, In Vitro
- . ISO 10993:2010 Biological evaluation of medical devices
- ASTM F838 Standard Test Method for Determining Bacterial Retention of ● Membrane Filters Utilized for Liquid Filtration
- ANSI/AAMI/ISO 11137-2:2012 Sterilization of health care products ●
- ISO 11607:2009 Packaging for terminally sterilized medical devices ●
- Blue Book Memorandum #G95-1, Use of International Standard ISO-. 10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing

### Summary of Substantial Equivalence:

Based on non-clinical performance testing, the Nosogard filters have been found to be substantially equivalent to the predicate PrisMedical MainStream Filter (K012716) and Nephros DSU & SSU filters (K141731) for the intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K153784](https://fda-staging.innolitics.com/device/K153784)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).