The Nester® and Tornado® Embolization Coils are intended for arterial and venous embolization in the peripheral vasculature.
Device Story
Nester® and Tornado® Embolization Coils are passive implantable devices used for arterial and venous embolization in peripheral vasculature. Constructed from platinum wire and synthetic fibers, these coils are delivered via a wire guide through a standard delivery catheter. The Nester coils feature a helical shape, while Tornado coils feature a tapered vortex shape. The device is operated by a physician in a clinical setting. Upon deployment, the coils physically obstruct blood flow to the target vessel, facilitating embolization. The device is MR Conditional. Clinical benefit is derived from the mechanical occlusion of vessels to treat vascular conditions.
Clinical Evidence
Bench, MRI, and animal testing performed. Bench testing included biocompatibility (cytotoxicity, sensitization, reactivity, toxicity, genotoxicity, implantation, hemocompatibility, pyrogenicity), wire/coil tensile strength, torque strength, delivery friction, delivery fatigue, and fiber security. MRI testing evaluated displacement force, torque, artifact, and RF-induced heating. Animal testing involved 1, 3, and 6-month safety evaluations in swine, showing no substantially adverse biological response. All tests met predetermined acceptance criteria.
Technological Characteristics
Materials: platinum wire, synthetic fibers. Form factor: helical (Nester) or tapered vortex (Tornado) coils. Dimensions: 0.018, 0.035, 0.038 inch catheter compatibility; 2-20 cm lengths; 2-20 mm diameters. Energy source: none (passive implant). Connectivity: none. Sterilization: not specified. MR Conditional.
Indications for Use
Indicated for arterial and venous embolization in the peripheral vasculature for patients requiring occlusion of blood vessels.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three human profiles facing to the right, stacked on top of each other, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 19, 2016
Cook Incorporated David Lehr, RAC Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404
Re: K153778
Trade/Device Name: Nester® and Tornado® Embolization Coils Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: May 11, 2016 Received: May 12, 2016
Dear David Lehr:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153778
#### Device Name
Nester® and Tornado® Embolization Coils
Indications for Use (Describe)
The Nester® and Tornado® Embolization Coils are intended for arterial and venous embolization in the peripheral vasculature.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. 812.339.2235 TOLLEREE: 800.457.4500 WWW.COOKMEDICAL.COM
Image /page/3/Picture/2 description: The image shows the Cook Medical logo. The logo consists of the word "COOK" in white, with a registered trademark symbol next to it. Below the word "COOK" is the word "MEDICAL" in white, set against a red background. The red background is a rectangle that extends slightly beyond the width of the word "COOK".
#### 2.0 510(k) SUMMARY
#### As required by 21 CFR §807.92 Date Prepared: April 1, 2016
#### I. SUBMITTER
| Submission: | Traditional 510(k) Premarket Notification |
|-----------------------|---------------------------------------------------------------|
| Applicant: | Cook Incorporated |
| Contact: | David Lehr, RAC |
| Applicant Address: | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone Number: | (812) 335-3575 ext. 102309 |
| Contact Fax Number: | (812) 332-0281 |
#### II. DEVICE
| Trade Name: | Nester® and Tornado® Embolization Coils |
|----------------------|----------------------------------------------|
| Common Name: | Vascular Embolization Device |
| Classification Name: | Device, Vascular, For Promoting Embolization |
| Regulation/Class: | 21 CFR §870.3300/Class II |
| Product Code | KRD |
## III. PREDICATE DEVICE
The device subject of this submission is considered substantially equivalent to the predicate device, the Cook Retracta® Detachable Embolization Coil (K123712/K151676). This predicate has never been subject to a design-related recall.
## IV. DEVICE DESCRIPTION
The Nester® and Tornado® Embolization Coils are constructed from coiled platinum wire and synthetic fibers. The wire forms primary coil diameters that can be delivered through catheters with end hole diameters of 0.018, 0.035, and 0.038 inch. For the Nester® Embolization Coils, the extended embolus lengths range from 3 to 20 cm, and upon deployment the coiled embolus diameters range from 2 to 20 mm. For the Tornado Embolization Coils, the extended embolus lengths range from 2 to 14.2 cm, and upon deployment the coiled embolus tapering diameters range from 3/2 to 10/5 mm.
## V. INDICATIONS FOR USE
The Nester® and Tornado® Embolization Coils are intended for arterial and venous embolization in the peripheral vasculature.
{4}------------------------------------------------
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The subject devices, the Nester® and Tornado® Embolization Coils, and the predicate device, the Retracta® Detachable Embolization Coil (K123712/151676), are substantially equivalent in that these devices have identical intended uses and similar technological characteristics. The predicate device is a combined embolization coil and delivery wire. Its embolization coil (made of platinum wire and nylon fibers) is delivered by a detachment mechanism. The subject devices are pushable embolization coils (also made of platinum wire and nylon fibers) delivered with a wire guide through a delivery catheter. The predicate coils (intended for delivery through catheters with an end hole size of 0.035 inch) have a helical shape, and are available in lengths of 7 cm or 14 cm and diameters ranging from 4 to 20 mm. The subject Nester coils (intended for delivery through catheters with end hole sizes of 0.018, 0.035, and 0.038 inch) also have a helical shape and are available in lengths ranging from 2 to 20 cm and in diameters ranging from 2 to 20 mm. The subject Tornado coils (also intended for delivery through catheters with end hole sizes of 0.018, 0.035, and 0.038 inch) have a tapered vortex (or tornado) shape and are available in lengths ranging from 2.0 to 14.2 cm and tapering diameters ranging from 3/2 to 10/5 mm. Additionally, the subject devices, like the predicate device, are labeled as MR Conditional. However, there are differences in the MR scanning conditions to reflect the testing performed on the subject devices. Based on the comparison of the design, intended use, materials, fundamental technology, and principle of operation, the subject devices are considered to be substantially equivalent to the currently marketed predicate device.
#### VII. PERFORMANCE DATA
The subject devices underwent the applicable testing listed below to ensure reliable design and performance under the testing parameters. Performance (bench, MRI, and animal) and biocompatibility testing was conducted in accordance with applicable FDA guidance documents to confirm the reliable performance of critical device characteristics.
- . Biocompatibility Testing - Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, Subchronic Toxicity, Genotoxicity, Implantation, Hemocompatibility, and Pyrogenicity tests were performed on the implantable embolization coils. Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity, and Hemocompatibility tests were performed on the device's
{5}------------------------------------------------
loading stylet and loading cartridge. All tests results met the acceptance criteria, where applicable, or demonstrated that the device is biocompatible.
- . Wire Tensile Testing - Testing shows the raw wire has a peak load value greater than or equal to the minimum tensile strength requirements. The predetermined acceptance criteria were met.
- . Coil Tensile Testing - Testing characterized the embolization coils' uniaxial tensile strength.
- . Rotations to Failure Testing - Testing characterized the embolization coils' torque strength.
- . Delivery Friction Testing - Testing shows that the embolization coils fully deploy into the portion of the target artery in an anatomical model. The predetermined acceptance criteria were met.
- . Delivery Fatigue Testing – Testing shows that the embolization coils do not have any visual defect after delivery. The predetermined acceptance criteria were met.
- Fiber Security Testing Testing shows that an entire fiber is not released from the . coils during simulated delivery conditions. The predetermined acceptance criteria were met.
- MRI Testing MRI compatibility was assessed by evaluating magnetic field . interactions (displacement force and torque), artifact, and RF-induced heating in accordance with FDA guidance titled, Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment.
- . Animal Testing - One-month, three-month, and six-month safety evaluations in swine were performed on platinum embolization coils. Testing shows that the target arteries do not indicate a substantially adverse biological response. The predetermined acceptance criteria were met.
## VIII. CONCLUSIONS
The data included in this submission indicate that the subject devices do not raise new questions of safety or effectiveness compared to the predicate device. This supports a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.