The R:GEN Laser System is indicated for use by an ophthalmic physician to produce a wound to the Retinal Pigmented Epithelium (RPE) of the retina of the eye, via focal laser treatment, in the treatment of Clinically Significant Macular Edema (CSME).
Device Story
R:GEN Laser System is a surgical instrument for selective retinal therapy (SRT) in clinic or outpatient settings; operated by ophthalmic physicians. Device utilizes Q-switched Nd:YLF laser (527 nm) to deliver focal laser treatment to retinal pigmented epithelium (RPE). System components include main laser, optical fiber, slit lamp delivery system, footswitch, and LCD touch-screen control panel. Operator sets treatment parameters via GUI; system micro-controller continuously monitors functions during operation. Output is laser energy applied to retina to induce RPE cell regeneration without collateral damage to neural retina or choroid. Clinical benefit includes improved visual acuity and reduced macular thickness in patients with diabetic macular edema.
Clinical Evidence
Two clinical studies evaluated safety and effectiveness for diabetic macular edema (DME). Study 1 (n=23 eyes, Korean subjects): 6-month follow-up showed significant visual acuity improvement (0.35 to 0.23 logMAR, p=0.002) and reduced maximum macular thickness (p=0.006). Study 2 (n=39 eyes, Caucasian subjects): 6-month follow-up showed significant BCVA improvement (43.7 to 46.1 letters, p=0.02). No severe vision loss or treatment-related complications reported in either study.
Technological Characteristics
Q-switched Nd:YLF laser; 527 nm wavelength; 1.7 μs pulse duration; 200 μm spot size; 0.03-0.4 mJ energy. Includes slit lamp delivery system, optical fiber, and LCD touch-screen GUI. Complies with ISO 13485:2003, AAMI/ANSI ES 60601-1, IEC 60601-2-22, and IEC 60825-1. Electrical: 100-240V, 50/60Hz.
Indications for Use
Indicated for ophthalmic physicians to treat Clinically Significant Macular Edema (CSME) in patients via focal laser treatment to the Retinal Pigmented Epithelium (RPE).
Regulatory Classification
Identification
An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.
Predicate Devices
Ellex 2RT Ophthalmic Laser System (Model LR1532) (K122202)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, which is a symbol often associated with healthcare and medicine. The caduceus is depicted with a staff entwined by two snakes and topped with wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 5, 2016
Lutronic, Inc. Jhung Won Vojir, PhD VP of Quality and Regulatory Affairs 6 Neshaminy Interplex, Suite 100 Trevose, PA 19053
Re: K153769
Trade/Device Name: R:GEN Laser System Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HQF Dated: August 26, 2016 Received: August 25, 2016
Dear Dr. Vojir:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA)
, it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name R:GEN Laser System
Indications for Use (Describe)
The R:GEN Laser System is indicated for use by an opthalmic physician to produce a wound to the Retinal Pigmented Epithelium (RPE) of the retina of the eye, via focal laser treatment of Clinically Significant Macular Edema (CSME)
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary Lutronic Corporation R:GEN Laser System
This 510(k) Summary is being submitted in accordance with 21 CFR § 807.92.
# 1. General Information
| Applicant: | Lutronic Corporation<br>219 Sowon-ro<br>Haengsin-dong, Deogyang-gu<br>Goyang-si, Gyeonggi-do 410-220<br>Republic of Korea<br>Tel: (82) 31-908-3440<br>Fax: (82) 31-907-3440 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Jhung Won Vojir, Ph.D.<br>VP of Quality and Regulatory Affairs<br>Lutronic Corporation<br>Six Neshaminy Interplex, Suite 100<br>Trevose, PA 19053<br>Telephone: 215-205-2219<br>Fax: 609-488-6958<br>Email: jvojir@lutronic.com |
| Summary Preparation Date: | October 5, 2016 |
| 2. Names | |
| Trade Name: | R:GEN Laser System |
| Common Name: | Laser, Ophthalmic |
| Classification Name: | Ophthalmic Laser<br>Product Code: HQF<br>21 CFR § 886.4390<br>Panel: Ophthalmic Devices |
{4}------------------------------------------------
### 3. Predicate Device
The 2RT Laser System is a surgical instrument for performing retinal procedures in the treatment of Clinically Significant Macular Edema (CSME).
| 510(K) Number | K122202 |
|------------------------------|--------------------------------------------------------|
| Company Name | Ellex Medical Pty. Ltd. |
| Device Name | Ellex 2RT Ophthalmic<br>Laser System<br>(Model LR1532) |
| Classification<br>Regulation | 21 CFR § 886.4390 |
| Classification Name | Ophthalmic Laser |
| Product Code | HQF |
| Product Code<br>Descriptor | Laser, Ophthalmic |
### 4. Device Description
The R:GEN Laser System is a surgical laser instrument for use by ophthalmic physicians for performing selective retinal therapy (SRT) for treatment of Clinically Significant Macular Edema (CSME) which is a condition secondary to Diabetic Retinopathy. It is designed for use in a clinic or outpatient facility, or in Retinal Specialist's practice.
The R:GEN Laser System is a Q-switched Nd: YLF (Neodymium-doped Yttrium Lithium Fluoride) laser with an emission wavelength of 527 nm (produced by second-harmonic generation). The R:GEN Laser System consists of a main laser, an optical fiber, delivery system including slit lamp, and a footswitch. The graphic user interface (GUI) screen of the control panel is equipped with an LCD (liquid crystal display) touch screen so that operators can easily adjust parameters for optimal settings. The software included provides all the functions, which are necessary to use the device.
The GUI screen of the control panel allows the operator to set the treatment parameters, such as the laser energy for each patient. The GUI screen of the control panel also displays useful information about the system during operation. During operation of the R:GEN Laser System, all system functions are continuously monitored internally by a micro-controller and displayed to the operator.
{5}------------------------------------------------
### 5. Indications for Use
The R:GEN Laser System is indicated for use by an ophthalmic physician to produce a wound to the Retinal Pigmented Epithelium (RPE) of the retina of the eye, via focal laser treatment, in the treatment of Clinically Significant Macular Edema (CSME).
### 6. Substantial Equivalence
The R:GEN Laser System is substantially equivalent (SE) to the legally marketed Ellex 2RT Laser System that is the subject of 510(k) K122202.
| Manufacturer | Lutronic Corporation | Ellex Medical Pty. Ltd. |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | R:GEN Laser System | 2RT (Model number LR1532) Ophthalmic Laser System |
| 510(k) # | K153769 | K122202 |
| Indication for Use | The R:GEN Laser System is indicated for use by an ophthalmic physician to produce a wound to the Retinal Pigmented Epithelium (RPE) of the retina of the eye, via focal laser treatment, in the treatment of:<br>• Clinically Significant Macular Edema (CSME) | The 2RT (LR1532) is indicated for use by a trained ophthalmic physician to produce a wound to the Retinal Pigmented Epithelium (RPE) of the retina of the eye, via focal laser treatment, in the treatment of:<br>• Clinically Significant Macular Edema (CSME) |
| Laser Type | Q-switched Nd:YLF Laser | Q-switched Nd:YAG Laser |
| Laser Wavelengths | 527 nm (green) | 532 nm (green) |
| Laser Energy | 0.03 mJ to 0.4 mJ | 0.1 mJ to 0.6 mJ |
| Pulse Duration | 1.7 μs | 3 ns |
| Spot Size | 200 μm | 400 μm |
| Aiming Beam | 635 nm | 635 nm |
| Type of Delivery System | Slit lamp adapter with optical fiber connected to a separate laser source | Slit lamp adapter with optical fiber connected to a separate laser source |
| Laser Delivery Modes | Single spot | Single spot |
| FDA Classifications | FDCA | Class II (special controls) 21 CFR § 886.4390 |
| | RCHSA | Class III B |
| GUI Screen of the Control Panel | Touch LCD Module | Touch LCD Module |
| Emission Control | Footswitch | Footswitch |
| Dimensions of Laser | 340 mm (W) x 534.5 mm<br>(L) x 619.5 mm (H) | 400 mm (W) x 625 mm (L)<br>x 445 mm (H) |
| Weight of Laser | 52.62 kg | 48 kg |
| Electrical Rating | AC100-240V, 50/60Hz,<br>Power Consumption<br>300VA | AC 100-240V, 50/60 Hz,<br>Power Consumption 60W |
{6}------------------------------------------------
### 7. Performance Data
The Company's Performance Data for the R:GEN Laser System is as follows:
### Bench Testing
Bench testing establish the accuracy of laser spot diameter, pulse duration and energy in laser exposure mode. The R:GEN Laser System bench testing meets all the device specifications and complies with all applicable standards, including ISO 13485:2003, AAMI / ANSI ES 60601-1:2005/(R)2012 and A1:2012, IEC 60601-2-22 Third Edition 2007-05 and IEC 60825-1 Edition 2.0 2007-03.
### Animal Testing
There were two animal studies conducted with R:GEN laser with the objective to evaluate the retinal effects of SRT in two animal models - in vivo mice and ex vivo porcine - and to demonstrate that R:GEN laser SRT is effective in selectively destroying RPE cells without collateral damage to neural retina or choroid with the goal of inducing RPE cell regeneration. Both studies were conducted in full compliance with the good laboratory practices (GLPs).
## Clinical Testing
Two separate clinical studies investigated the safety and effectiveness of the selective retina therapy (SRT) for clinically significant diabetic macular edema (DME).
Twenty-three eyes of 21 Korean subjects with clinically significant DME were treated with SRT and followed for 6 months. The mean subject age was 63.3 years with 14 females and 7 males, with type 2 diabetes and clinically significant DME according to ETDRS criteria. The subjects ranged from 48 to 77 years of age. The number of treatment laser spots was 41.1 (Standard Deviation (SD) = 18.8, range 12 to87 laser spots), and mean treatment pulse energy was 92.1 µJ (SD=8.2, 75 µJ to 130 µJ).
{7}------------------------------------------------
Mean visual acuity improved significantly between baseline and 6 months, from 0.35 logMAR to 0.23 logMAR (p=0.002). After 6 months, 29.4% of SRT-treated eyes gained ≥2 ETDRS lines from baseline, while 41.2% gained 1-2 ETDRS lines, and 17.6% remained stable (<1 ETDRS line change). Conversely, only two eves (11.8%) lost >1 ETDRS lines from baseline. Although there was no significant change in central macular thickness (CMT) (p>0.05), maximum macular thickness (MMT) decreased from 465.8 ± 87.4 um to 434.3 ± 83.9 um (p=0.006), and mean macular sensitivity increased from 20.8 ± 3.4 dB to 22.5 ± 3.5 dB (p=0.02). No eyes experienced severe loss of vision (>2 ETDRS lines) at the 6-month follow-up visit. Individual eyes demonstrated no scotomatous changes in macular sensitivity within the central 10° over 6 months. No treatment-related complications or adverse events were noted in this study.
A study involving thirty-nine eyes of 39 Caucasian subjects with previously untreated non-ischemic DME was performed. The mean subject age was 63.9 year with 16 females and 23 males. The mean number of applied laser spots was 35.2 (SD=24, range 11 to 125 laser spots). The mean treatment pulse energy was 247 uJ (SD=50, range 200 uJ to 325uJ). The mean best-corrected ETDRS visual acuity (BCVA) improved from 43.7 letters (SD=9.1) at baseline to 46.1 letters (SD=10. 5) at the 6-month follow-up (p=0.02). BCVA improved (>5 letters) or remained stable (±5 letters) in 84% of eyes. Thirteen percent (13%) of eyes improved by ≥10 letters, while 16% of eyes lost more than 5 letters. There was no severe loss of vision (≥15 letters). Overall, retinal thickness, hard exudates, and leakage in fundus fluorescein angiography (FFA) did not change significantly (p>0.05), while improvement of BCVA correlated with a reduction of hard exudates (p=0.01) and central retinal thickness (p=0.01). No complications or adverse events were noted during or after treatment, and no patients reported any adverse effects related to SRT or pain during laser application.
#### 8. Conclusion
The R:GEN Laser System and the legally marketed 2RT Laser System cleared under 510(k) number K122202, have the same intended use and Indications for Use statement. While the technological characteristics differ between the two systems, the differences have been established to be minor. Performance testing data established that the R:GEN Laser System is as safe and effective as the legally marketed predicate device and that the R:GEN Laser System does not raise any different questions of safety and effectiveness than the predicate. On this basis and in accordance with 21 CFR § 807.100(b), the R:GEN Laser System is substantially equivalent to the 2RT Laser System and can be legally marketed in the US.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.