X-cera Zirconia Blanks are indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facings, and veneers.
Device Story
X-cera Zirconia Blanks are zirconia-based blocks, discs, or cylinders used as raw material for dental prosthetics. The device is manufactured via uni-axial and iso-static pressing to ensure uniform density. In a dental laboratory setting, a technician uses modern machining equipment to mill the blanks into complex dental shapes, such as crowns, veneers, or denture components. The final prosthetic is then sintered to achieve its final mechanical properties. The device provides a high-strength, biocompatible ceramic material for restorative dentistry, enabling the fabrication of durable, patient-specific dental replacements.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including flexural strength, density, porosity, shrinkage/expansion rates, chemical composition analysis, and biocompatibility testing (cytotoxicity, sensitization, intracutaneous reactivity, acute systemic toxicity, and genotoxicity) in accordance with ISO 6872, ISO 7405, and ISO 10993.
Technological Characteristics
Materials: Zirconia (ZrO2 88-96%, Y2O3 4-10%, Al2O3 <1%, HfO2 <0.05%, SiO2 <0.5%). Form factor: Blocks, discs, cylinders. Processing: Uni-axial and iso-static pressing. Sintering temperature: 1480°C. Mechanical properties: Flexural strength ≥800MPa (ISO 6872), Elastic modulus 210±15 GPa. Standards: ISO 6872 (Type II Class 6), ISO 10993-1, ISO 7405. Non-sterile.
Indications for Use
Indicated for production of artificial teeth in fixed or removable dentures, or for jacket crowns, facings, and veneers. No specific patient population or contraindications described.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three interconnected faces in profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 29, 2016
Shenzhen XiangTong Photoelectricity Technology Co., Ltd. c/o Mr. Mark Job Regulatory Technology Services, LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K153767
Trade/Device Name: X-cera Zirconia Blanks Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain powder for clinical use Regulatory Class: II Product Code: EIH Dated: March 9, 2016 Received: March 15, 2016
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153767
Device Name X-cera Zirconia Blanks
#### Indications for Use (Describe)
X-cera Zirconia Blanks are indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facings, and veneers.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------------------------------------------|-----------------------------------------------|
|-------------------------------------------------------------------------------------|-----------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
for
# X-cera Zirconia Blanks
# 1. 510(k) Number
K153767
#### 2. Submission Sponsor
Shenzhen XiangTong Photoelectricity Technology Co., Ltd. 2~4th floor, 1st Building, South Industrial park of Honghualing, Nanshan District, Shenzhen, China Phone: 0086-755-86001804 Fax: 0086-755-86001486 Contact: Yongqing Chen, Vice-President
#### 3. Submission Correspondent
Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: +86-21-22815850 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Lee Fu, RAC, Senior Consultant, Regulatory Email: project.management@emergogroup.com
#### 4. Date Prepared
03/29/2016
## 5. Device Identification
| Trade/Proprietary Name: | X-cera Zirconia Blanks |
|----------------------------|-----------------------------------|
| Common/Usual Name: | Zirconia Blocks |
| Classification Name: | Porcelain powder for clinical use |
| Classification Regulation: | 872.6660 |
| Product Code: | EIH |
| Device Class: | Class II |
| Classification Panel: | Dental Devices |
#### 6. Legally Marketed Predicate Device(s)
Aidite Zirconia Dental Ceramics (K111291)
{4}------------------------------------------------
#### 7. Device Description
X-cera Zirconia Blanks are provided in either block, disc, cylinder shape and are derived from zirconia powder that has been processed via uni-axial die pressing, followed by iso-static pressing, to achieve final net shapes of uniform density and distribution. These Zirconia Blanks can be fabricated into various prosthetic dental device.
X-cera Zirconia Blanks are offered in ten(10) different product families of shapes and a multitude of different sizes as shown in Figure 1 and Table 1 and are capable of being machined into complex dental shapes using modern machining methods and are similar to the sizes and shapes of the predicate device.
Any difference in the net shapes and dimensions between X-cera Zirconia Blanks and those offered by the manufacturer of the predicate devices are incidental to the performance of the final dental prosthetic devices that will be fabricated from them.
Image /page/4/Figure/4 description: The image shows a variety of Zirconia Blanks. There are 10 different types of Zirconia Blanks, including XS series, XQ series, XW series, XZ series, XC series, XZZ series, XO series, XK series, XA series, and XM series. The Zirconia Blanks are all white, but they come in different shapes and sizes.
Figure 1 X-cera Zirconia Blanks
| Table 2<br>X-cera Zirconia Blanks Product Family(Model) | |
|---------------------------------------------------------|------------------------------------------------------------------|
| Family(Model) | Shapes & Sizes available |
| XQ series | Cylinder: Φ22*25(mm)<br>Disc: Φ98×8(mm) |
| XW series | Discs: Φ100×10 to Φ100×25(mm) (with 3mm protruding step) |
| XO series | Discs: Φ98×10 to 98×25(mm) (with 2mm protruding step) |
| XA series | Discs: Φ100×10 to Φ100×25(mm) (with Φ112×6mm Outer plastic ring) |
| XA series | Discs: Φ98×10 to Φ98×25(mm) (with Ø 115×10mm Outer ring) |
| XA series | Discs: Φ71×10 to Φ71×25(mm) |
| XA series | Blocks: 40×20×16 to 65×30×20(mm) |
| XA series | Blocks: 90×72×10 to 90×72×25(mm)(outer ring) |
| XK series | Discs: Φ98×10 to Φ98×25(mm) |
| XK series | Blocks: 42×16×16 to 60×25×20(mm) |
| XK series | Cylinders: Φ16×16 to Φ20×20(mm) |
| XZ series | Discs: Φ95×10 to Φ95×25(mm) |
| XZ series | Blocks: 43×25×16 to 93×75×25(mm) |
{5}------------------------------------------------
| | Cylinders: $Φ$ 23×16 to $Φ$ 23×22(mm) |
|------------|----------------------------------------------------------------|
| XM series | Blocks: 43×25×16 to 75×42×22(mm) |
| XC series | Cylinders $Φ$ 25×26 to $Φ$ 25×60(mm) (with Outer plastic ring) |
| | Blocks: 26×25×20 to 72×40×25(mm) (with Outer plastic ring) |
| XS series | Blocks: 20×14×15 to 85×40×22(mm) |
| XZZ series | Discs: $Φ$ 95×10 to $Φ$ 95×25(mm) |
Note: The form of size above has two forms, one is shown in D×H and the other is L×W× H, of which, "D"means Diameter, "H"means height, "L" means length, "W" means width.
#### 8. Indication for Use Statement
X-cera Zirconia Blanks are indicated for the production of artificial teeth in fixed or removable dentures, or for jacket crowns, facings, and veneers.
#### 9. Substantial Equivalence Discussion
The following table compares the X-cera Zirconia Blanks to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Manufacturer | Shenzhen XiangTong<br>Photoelectricity<br>Technology Co., ltd. | Qinhuangdao Aidite<br>High-Technical Ceramics<br>Co. | SIGNIFICANT<br>DIFFERENCES |
|---------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Trade Name | X-cera Zirconia Blanks | Aidite Zirconia Dental<br>Ceramics | / |
| 510(k) Number | / | K111291 | / |
| Product Code | EIH | EIH | Same |
| Regulation<br>Number | 21 CFR 872.6660 | 21 CFR 872.6660 | Same |
| Regulation Name | Powder, Porcelain | Powder, Porcelain | Same |
| Indications for Use | are indicated for the<br>production of artificial<br>teeth in fixed or<br>removable dentures, or<br>for jacket crowns,<br>facings, and veneers | are indicated for the<br>production of artificial<br>teeth in fixed or<br>removable dentures, or<br>for jacket crowns,<br>facings, and veneers | Same |
| Directions for Use | Prescription Use | Prescription Use | Same |
| Fundamental<br>Scientific<br>Technology | Iso static pressing | Iso static pressing | Same |
| Main working<br>principle | Be capable of being<br>machined into complex<br>dental shapes using<br>modern machining | Be capable of being<br>machined into complex<br>dental shapes using<br>modern machining | Same |
| Manufacturer | Shenzhen XiangTong<br>Photoelectricity<br>Technology Co., ltd. | Qinhuangdao Aidite<br>High-Technical Ceramics<br>Co. | SIGNIFICANT<br>DIFFERENCES |
| Trade Name | X-cera Zirconia Blanks | Aidite Zirconia Dental<br>Ceramics | / |
| | methods | methods | |
| Applied Standards | ISO 6872<br>ISO 10993-1<br>ISO 7405 | ISO 6872<br>ISO 10993-1<br>ISO 7405 | Same |
| Main Chemical<br>Composition (in<br>mass-%) | ZrO₂ 88% ~ 96%<br>Y₂O₃ 4% ~10%<br>Al₂O₃ <1%<br>HfO₂ <0.05%<br>SiO₂ <0.5% | ZrO₂ <96%<br>Y₂O₃>4%<br>Al₂O₃ <1%<br>HfO₂>1%<br>SiO₂ < 0.02% | Analysis 1 |
| Crystal<br>Morphology | Tetragonal | Tetragonal | Same |
| Color | white | white | Same |
| Sintering<br>temperature | 1480°C | 1480°C | Same |
| Sterile | Non-sterile | Non-sterile | Same |
| Density (post<br>sintering) | 6.0~6.1 g/cm³ | 6.05g/cm³ | Analysis 2 |
| Shrinkage (post<br>sintering) | 19%~22% | 22% | Analysis 3 |
| Categorization<br>According to ISO<br>7405 and ISO<br>10993-1 | 1) Categorization by<br>nature of contact:<br>External communicating<br>devices<br>2) Categorization by<br>duration of contact<br>Permanent contact<br>devices | 1) Categorization by<br>nature of contact:<br>Surface-contacting<br>devices<br>2) Categorization by<br>duration of contact<br>Permanent contact<br>devices | Same |
| Flexural Strength<br>(post sintering) | Comply with ISO 6872 | Comply with ISO 6872 | Same |
| Elastic modulus<br>(post sintering) | 210±15 GPa | 210 GPa | Same |
| Porosity | Less than 0.1% | 0% | Same |
| Type and Class<br>according to ISO<br>6872 | Type II Class 6 | Type II Class 6 | Same |
| Fracture Toughness<br>(after sintered) | Comply with ISO 6872 | Comply with ISO 6872 | Same |
| Coefficient of linear<br>thermal expansion | $10.5±0.5×10^{-6} K^{-1} (25 °C~500°C)$ | $10.5±0.5×10^{-6} K^{-1} (25 °C~500°C)$ | Same |
#### Table 3 – Comparison of Characteristics
{6}------------------------------------------------
## Analysis 1
They are both conformance with the requirement ISO 6872 Type II class 6.
{7}------------------------------------------------
## Analysis 2
The value of Density (post sintering) of Aidite Zirconia Dental Ceramics is a fixed value, and the range of the proposed cover the value of the predicate.
Analysis 3
Proposed device and predicate device are both conformance to the same mechanical and chemical property and biocompatibility standard (ISO 6872, ISO 10993-1, ISO 7405).
## 10. Non-Clinical Performance Data
The following testing has been performed to support substantial equivalence:
- Flexural Strength Test - The test results of mean flexural strength (post sintering) comply with the ISO 6872: 2008 (≥800MPa) and product design requirement (≥900MPa).
- . Density and Apparent Porosity Test - The test results of density and apparent porosity comply with the values specified by manufacturer in the Instruction for Use.
- Shrinkage rate and Expand Rate Test - The test results of shrinkage rate and expand rate comply with the acceptance criteria.
- . Physical and chemical test - The test results comply with the ISO 6872: 2008.
- . Zirconia powder Composition Test - The test results comply with the design specifications and the values specified by manufacturer in the Instruction for Use.
- . Biocompatibility Test – Results comply with ISO 7405 and ISO 10993 including In Vitro Cytotoxicity Test, Sensitization Test, Intracutaneous Reactivity Test, Acute Systemic Test, and Genotoxicity.
The X-cera Zirconia Blanks passed all testing stated above as shown by the acceptable results obtained.
The device passed all the testing in accordance with national and international standards.
## 11. Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device.
{8}------------------------------------------------
#### 12. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
The X-cera Zirconia Blanks, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.