K153748 · Care Fusion · CAF · May 17, 2017 · Anesthesiology
Device Facts
Record ID
K153748
Device Name
AirLife Misty Finity Nebulizer
Applicant
Care Fusion
Product Code
CAF · Anesthesiology
Decision Date
May 17, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5630
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Misty Finity is a nebulizer designed for administration of aerosolized drug to the respiratory system. Misty Finity may be used with pediatric (ages 2 years and above) and adult patients. The product is single patient use device, nonsterile and used in professional healthcare environments under a doctor's supervision and by skilled clinician.
Device Story
AirLife™ Misty Finity™ Nebulizer is a small volume pneumatic jet nebulizer for delivering aerosolized drugs to the respiratory system. Device consists of three components: top, one-piece jet, and bottom reservoir. Liquid medication is placed in the reservoir; connected to an external compressed air or oxygen source (2-4 LPM) via flexible tubing. Principle of operation: gas stream shatters liquid into particles (15-500 microns); particles impact a baffle to reduce size (<15 microns); larger particles settle via gravity/inertia to repeat nebulization; particles <5 microns are inhaled by patient. Used in professional healthcare environments by skilled clinicians under physician supervision. Single-patient use device; supports multiple treatments. Provides continuous aerosol delivery. Benefits patient by enabling targeted respiratory drug administration.
Clinical Evidence
Bench testing only. Aerosol characterization performed using 7-stage Cascade Impaction (NGI) per USP 38 <601> with three medications (Albuterol, Budesonide, Cromolyn) at 2 and 4 LPM. Metrics included MMAD, GSD, and mass distribution. Biocompatibility testing conducted per AAMI/ANSI/ISO 10993 series (cytotoxicity, sensitization, irritation, implantation, genotoxicity, extractables/leachables) for prolonged contact.
Technological Characteristics
Pneumatic jet nebulizer; 10 ml fill capacity; 2-4 LPM flow rate. Materials evaluated per AAMI/ANSI/ISO 10993-1:2009. Non-sterile, single-patient use. Mechanical operation; no software or electronic components.
Indications for Use
Indicated for pediatric (ages 2+) and adult patients requiring administration of aerosolized medication to the respiratory system.
Regulatory Classification
Identification
A nebulizer is a device intended to spray liquids in aerosol form into gases that are delivered directly to the patient for breathing. Heated, ultrasonic, gas, venturi, and refillable nebulizers are included in this generic type of device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, composed of three intertwined figures. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus. The seal is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 17, 2017
Carefusion/Vyaire Medical, Inc. Colleen O'Keeffe Acting Director, Regulatory Affairs 26125 Riverwoods Blvd Mettawa, IL 60045
Re: K153748
Trade/Device Name: AirLife™ Misty Finity™ Nebulizer Regulation Number: 21 CFR 868.5630 Regulation Name: Nebulizer Regulatory Class: Class II Product Code: CAF Dated: April 12, 2017 Received: April 17, 2017
Dear Colleen Okeeffe:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153748
Device Name AirLife™ Misty FinityTM Nebulizer
#### Indications for Use (Describe)
The Misty Finity is a nebulizer designed for administration of aerosolized drug to the respiratory system. Misty Finity may be used with pediatric (ages 2 years and above) and adult patients. The product is single patient use device, nonsterile and used in professional healthcare environments under a doctor's supervision and by skilled clinician.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|
| <span style="text-decoration: overline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Vyaire Medical. The logo is primarily blue, with the word "Vyaire" in a stylized, sans-serif font. Below "Vyaire" is the word "MEDICAL" in a smaller, sans-serif font. Above the word "Vyaire" are three curved, light blue lines that give the impression of a wave or airflow.
ne: (224) 706 -6818
# 510K Summary
#### 1. Submitter
Carefusion/Vyaire Medical, Inc. 26125 Riverwoods Blvd Mettawa, IL 60045 Phone: (224) 706-6818
Contact Person: Colleen O'Keeffe Acting Director, Regulatory Affairs
Date Prepared: May 16, 2017
#### 2. Device
| Trade name | AirLife™ Misty Finity™ Nebulizer |
|----------------------|----------------------------------|
| Common Name: | Small Volume Nebulizer |
| Classification Name: | Nebulizer 21 CFR 868.5630 |
| Regulatory Class: | II |
| Product Code: | CAF |
#### 3. Predicate Device
Uni-HEART K943600 cleared on 10/13/1994.
#### 4. Device Description
The Misty Finity nebulizer is a small volume nebulizer designed to deliver continuous aerosolized drugs to the respiratory system over extended periods of time. The Misty Finity is a single patient use device, and may be used for multiple treatments. The nebulizer is filled with a liquid, typically respiratory medication and connected to the air source via flexible tubing operating between 2 LPM and 4 LPM. The nebulizer is disassembled and reassembled after adding the medication in the bottom (reservoir), and then the device is reassemble in a snap fit. The Misty Finity consists of three components: nebulizer top, one-piece jet, and nebulizer bottom. The Nebulizer is marketed with oxygen tubing and an adult aerosol mask. Both tubing and mask are single-use accessories. The pediatric aerosol mask can be ordered online/from the company website.
#### 5. Principal of Operation
The nebulizer works by having the fluid come into contact with the stream of gas. The gas shatters the liquid into small particles (approximately 15-500 microns in size). These particles then impact a baffle that further reduces the size of the particles (< 15 microns). The majority of larger particles settle inside the nebulizer as a result of gravity and inertia, returning the mist to liquid to repeat the nebulization process. The smaller
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Vyaire Medical. The logo is in blue and features the word "Vyaire" in a stylized font. Above the word "Vyaire" are three curved lines in a lighter shade of blue. Below the word "Vyaire" is the word "MEDICAL" in a smaller, sans-serif font.
particles (<5 microns) are then administered as the patient inhales. The treatment is completed when the majority of fluid is nebulized.
#### 6. Indication for use
The Misty Finity is a nebulizer designed for administration of aerosolized drug to the respiratory system. Misty Finity may be used with pediatric (ages 2 years and above) and adult patients. The product is single patient use device, non-sterile and used in professional healthcare environments under a doctor's supervision and by skilled clinician.
## 7. Comparison of technological characteristics with the predicate device
| Element of<br>comparison | Proposed Device<br>K153748 | Predicate Device<br>K943600 | Comparison Assessment |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The Misty Finity is a nebulizer<br>designed for administration of<br>aerosolized drug to the<br>respiratory system. Misty Finity<br>may be used with pediatric (ages<br>2 years and above) and adult<br>patients. The product is single<br>patient use device, non-sterile<br>and used in professional<br>healthcare environments under<br>a doctor's supervision and by<br>skilled clinician. | The proposed device is intended<br>to spray liquids in aerosol form<br>into gases that are delivered<br>directly to the patient for<br>breathing for all patients'<br>populations requiring nebulized<br>medications. | Although worded differently, the<br>indications for use (IFU)<br>statements identify the same: 1)<br>principle of operation (i.e., both<br>devices aerosolize drugs so they<br>can be delivered to the<br>respiratory system) and 2) both<br>devices are intended for all<br>populations requiring nebulized<br>medications (i.e., ages 2 years<br>and above) and adults).<br>Although not stated in the<br>predicate device IFU, both<br>devices are: 1) single use (see<br>discussion below), 2) are<br>provided non-sterile and 3) are<br>prescription products which<br>require that they are used in<br>professional healthcare<br>environments under a doctor's<br>supervision and by skilled<br>clinician. The difference in<br>verbiage in the IFU for the Misty<br>Finity Nebulizer and the<br>predicate device does not<br>present any new questions of<br>safety and effectiveness. |
| Principle of Operation | Pneumatic Jet | Pneumatic Jet | Same |
| Characteristics | | | |
| Flow rates | 2-4 LPM | 2-4 LPM | Same |
| Maximum Fill Capacity<br>(capacity of medication<br>cup) | 10 ml | 10 ml | Same |
| Shelf life | 11 months | unknown | The Shelf-life and the Useful Life |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Vyaire Medical. The logo is in blue and features the word "Vyaire" in a stylized font. Below the word "Vyaire" is the word "MEDICAL" in a smaller, sans-serif font. Above the word "Vyaire" are three curved lines in a lighter shade of blue.
26125 Riverwoods Blvd Mettawa, IL 60045 Phone: (224) 706 -6818 carefusion.com
| Element of<br>comparison | Proposed Device<br>K153748 | Predicate Device<br>K943600 | Comparison Assessment |
|--------------------------|-------------------------------------------------|-----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Useful life | 48 hours | unknown | of the predicate device has not<br>been published. The real-time<br>aging and Biocompatibility<br>testing of the subject device<br>support the indicated Shelf-life<br>and Useful life and do not<br>present any new questions of<br>safety and effectiveness. |
| Type of the device | Disposable<br>Single patient use<br>Non sterile | Disposable<br>Single use<br>Non sterile | The predicate device labeling<br>indicates that it is single use (i.e.,<br>used once). Misty Finity is a<br>single patient use device (i.e.,<br>the same patient may use it<br>more than once). The cleaning<br>validations of the subject device<br>support substantial equivalence<br>without posing any new<br>questions of safety and<br>effectiveness. |
| Type of gas source | Compressed air or oxygen | Compressed air or oxygen | Same |
## 8. Performance Data
The proposed device was tested to ensure compliance to the following standards:
#### 8.1 Performance Testing
| Performance Characteristic | Standard |
|---------------------------------------------------|--------------|
| Products for Nebulization – Characterization Test | USP 38 <601> |
Aerosol Characterization using 7- stage Cascade Impaction (NGI) for three common aerosolized medications 1
| | Aerosol Characteristics | 2 lpm | | 4 lpm | |
|---------------------------------------|--------------------------------------|------------|-----------------|------------|-----------------|
| Drug<br>Description | | | (mean ± 95% CI) | | (mean ± 95% CI) |
| Albuterol<br>Sulfate:<br>2.5 mg/ 3 ml | MMAD (µm) 2 | 3.4 ± 0.3 | | 2.7 ± 0.3 | |
| | GSD 3 | 2.8 ± 0.2 | | 2.6 ± 0.1 | |
| | Units | % (mass) 4 | µg 5 | % (mass) 4 | µg 5 |
| | Particles ( ≤ 1 µm) | 15.7 ± 2.9 | 523 ± 96 | 18.4 ± 2.7 | 1531 ± 184 |
| | Particles ( 1-5 um) | 46.6 ± 2.9 | 1553 ± 96 | 52.8 ± 2.0 | 4400 ± 168 |
| | Particles ( ≤ 5 µm) | 62.3 ± 3.3 | 2076 ± 109 | 71.2 ± 3.2 | 5931 ± 263 |
| | Particles ( ≥ 5 µm) | 37.7 ± 3.3 | 1258 ± 109 | 28.8 ± 3.2 | 2402 ± 263 |
| | Total Mass of Drug in Nebulizer (µg) | 3333 | | 8333 | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Vyaire Medical. The logo is in blue and consists of the word "Vyaire" in a stylized font, with the word "MEDICAL" in smaller letters underneath. There is a light blue swoosh above the word "Vyaire".
| | Calculated Time to Deliver 100% of Dose<br>(minutes) 7,8 | 94 ± 12 | 104 ± 10 |
|-------------------------------------------|----------------------------------------------------------|------------|--------------|
| Budesonide<br>Suspension:<br>0.5 mg/ 2 ml | MMAD (µm) | 3.8 ± 0.3 | 4.2 ± 0.2 |
| | GSD | 3.1 ± 0.2 | 2.4 ± 0.2 |
| | Units | % (mass) 4 | µg 5 |
| | % Particles (≤ 1 µm) | 16.2 ± 1.4 | 162 ± 14 |
| | % Particles (1-5 μm) | 38.7 ± 1.3 | 387 ± 13 |
| | % Particles (≤ 5 µm) | 54.9 ± 1.6 | 549 ± 16 |
| | % Particles (≥ 5 μm) | 45.1 ± 1.6 | 451 ± 16 |
| | Total Mass of Drug in Nebulizer (µg) 6 | 1000 | 2500 |
| | Calculated Time to Deliver 100% of Dose<br>(minutes) 7 | 73 ± 8 | 91 ± 13 |
| Cromolyn<br>Sodium:<br>20 mg/ 2 ml | MMAD (µm) | 2.3 ± 0.3 | 3.6 ± 0.4 |
| | GSD | 3.5 ± 0.3 | 2.5 ± 0.1 |
| | Units | % (mass) 4 | µg 5 |
| | % Particles (≤ 1 µm) | 16.7 ± 2.3 | 6693 ± 924 |
| | % Particles (1-5 μm) | 51.9 ± 2.9 | 20751 ± 1152 |
| | % Particles (≤ 5 µm) | 68.6 ± 4.9 | 27444 ± 1956 |
| | % Particles (≥ 5 µm) | 31.4 ± 4.9 | 12556 ± 1956 |
| | Total Mass of Drug in Nebulizer (µg) 6 | 40000 | 100000 |
| | Calculated Time to Deliver 100% of Dose<br>(minutes) 7 | 117 ± 16 | 108 ± 10 |
| | | 14.5 ± 1 | 361 ± 24 |
| | | 40.6 ± 2.1 | 1015 ± 53 |
| | | 55.1 ± 2 | 1377 ± 50 |
| | | 44.9 ± 2 | 1123 ± 50 |
| | | 23 ± 2.4 | 22953 ± 2386 |
| | | 38.5 ± 2 | 38545 ± 1618 |
| | | 61.5 ± 3.5 | 61498 ± 3522 |
| | | 38.5 ± 3.5 | 38502 ± 3522 |
1 Performance data using Cascade Impactor (NGI) per USP 35 <1601> chilled to 5 °C with an extraction flow of 15 \/min for three commonly aerosolized medications. Values represent the Mean and 95% confidence interval of the mean.
2 MMAD is Mass Median Aerodynamic Diameter.
3 GSD is Geometric Standard Deviation.
4 Values are the percent (by mass) of drug substance delivered in the specified range of aerosol particles.
్ Values are the total drug mass in nebulizer, multiplied by the % (mass) for each micron range. This assumes 100% dose delivery and represents the theoretical maximum drug mass for a given micron range.
otal Mass of Drug in Nebulizer is the mass of drug substance placed in the nebulizer cap for dose listed. At the drug concentrations specified, these drug masses equate to a 4 ml fill volume for 2 lpm, and a 10 ml fill volume for 4 lpm.
7 Calculated by dividing the Total Mass of Drug in Nebulizer, by the Drug Delivery Rate. This is a theoretical calculation which assumes 100% dose delivery.
8 Calculated time to deliver 100% of 3 ml dose of Albuterol sulfate at 2.5 mg/3 ml concentration:
- 68 minutes @ 2 lpm ●
- 30 minutes @ 4 lpm .
The aerosol characterization testing with the adult and pediatric face mask demonstrated that the proposed device performance is equivalent to the predicate device.
#### 8.2 Biocompatibility
Tests for an externally communicating, tissue by way of gas path and direct mucosal contact with prolonged contact (greater than 24 hours but less than 30 days): Cytotoxicity, Sensitization, Irritation, Muscle Implantation, Genotoxicity and Extractables/Leachables.
| Performance Characteristic | Standard |
|----------------------------------------------------------------------------|----------------------------|
| Biological Evaluation of Medical Devices—Part 1:<br>Evaluation and Testing | AAMI/ANSI/ISO 10993-1:2009 |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Vyaire Medical. The logo is blue and features the word "Vyaire" in a stylized font. Below the word "Vyaire" is the word "MEDICAL" in a smaller, sans-serif font. Above the word "Vyaire" are three curved lines in a lighter shade of blue.
| Performance Characteristic | Standard |
|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------|
| Biological evaluation of medical devices – Part 3: Tests<br>for genotoxicity, carcinogenicity and reproductive<br>toxicity | AAMI/ANSI/ISO 10993-3:2009 (R2014) |
| Biological Evaluation of Medical Devices-Part 5: Tests<br>for In Vitro Cytotoxicity | AAMI/ANSI/ISO 10993-5:2009 (R2014) |
| Biological evaluation of medical devices - Part 6: Tests<br>for local effects after implantation | AAMI/ANSI/ISO 10993-6:2009 (R2014) |
| Biological Evaluation of Medical Devices-Part 10: Tests<br>for Irritation and Skin Sensitization. | AAMI/ANSI/ISO 10993-10:2010<br>(R2014) |
| Biological evaluation of medical devices – Part 11:<br>Tests for systemic toxicity | AAMI/ANSI/ISO 10993-11:2006<br>(R2010) |
| Biological evaluation of medical devices – Part 12:<br>Sample preparation and reference materials | AAMI/ANSI/ISO 10993-12: 2012 |
| Biological Evaluation of Medical Devices Part 17:<br>Establishment of allowable limits for leachable<br>substances | AAMI/ANSI/ISO 10993-17:2002 |
| Biological Evaluation of Medical Devices Part 18:<br>Chemical characterization of materials | AAMI/ANSI/ISO 10993-18:2005 |
#### 9. Conclusion
The non-clinical data demonstrate that the Misty Finity nebulizer is as safe and as effective as the predicate and therefore substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.