K153739 · Mipm Mammendorfer Institut Fur Physik Und Medizin · MWI · Jun 28, 2016 · Cardiovascular
Device Facts
Record ID
K153739
Device Name
MRI Patient Monitor TeslaDUO
Applicant
Mipm Mammendorfer Institut Fur Physik Und Medizin
Product Code
MWI · Cardiovascular
Decision Date
Jun 28, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The MRI Patient Monitor TeslaDUO is intended for monitoring of vital signs during MRI examinations (MRI procedures) of patients. The TeslaDUO is intended for use in the Adult and Pediatric populations for continuous monitoring of Pulse Oximetry (SpO2). The TeslaDUO is intended for use in the Adult, Pediatric and Neonatal populations for continuous monitoring of Non-Invasive Blood Pressure (NIBP). The TeslaDUO is intended for use by health care professionals.
Device Story
MRI-compatible patient monitor for use in MRI suites; monitors vital signs during MRI procedures. Inputs: SpO2 sensors and NIBP cuffs. Transforms physiological signals into graphic and numeric data displayed on-screen. Operated by healthcare professionals. Provides real-time monitoring to assist clinical decision-making during imaging. Benefits patient by enabling continuous vital sign tracking in high-magnetic-field environments. Device is magnet-resistant and HF-proof; battery or AC powered.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety (IEC 60601-1), electromagnetic compatibility (IEC 60601-1-2), alarm performance (IEC 60601-1-8), biocompatibility (ISO 10993-1), and MRI-specific environmental testing (magnetostatic forces, magnetic/MRI immunity, and RF/gradient field-induced heating).
Technological Characteristics
Housing: coated aluminum, antimagnetic stainless steel, plastics. Power: Li-Ion battery or 100-240 VAC. Connectivity: Standalone monitor. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-49, ISO 10993-1, ISO 14971, IEC 62304, ISO 80601-2-61 (SpO2), IEC 80601-2-30 (NIBP).
Indications for Use
Indicated for continuous monitoring of SpO2 in adult and pediatric patients, and NIBP in adult, pediatric, and neonatal patients during MRI procedures.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The symbol is placed to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the symbol.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 28, 2016
MIPM Mammendorfer Institut Fur Physik Und Medizin GmbH % Andre Kindsvater Senior Consultant RA & OA Emergo Group 16 Congress Avenue Suite 1400 Austin, Texas 78701
Re: K153739
Trade/Device Name: MRI Patient Monitor TeslaDUO Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: May 22, 2016 Received: May 26, 2016
Dear Andre Kindsvater:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153739
Device Name MRI Patient Monitor TeslaDUO
Indications for Use (Describe)
The MRI Patient Monitor TeslaDUO is intended for monitoring of vital signs during MRI examinations (MRI procedures) of patients.
The TeslaDUO is intended for use in the Adult and Pediatric populations for continuous monitoring of Pulse Oximetry (SpO2).
The TeslaDUO is intended for use in the Adult, Pediatric and Neonatal populations for continuous monitoring of Non-Invasive Blood Pressure (NIBP).
The TeslaDUO is intended for use by health care professionals.
Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (301) 443-6740
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
# Section 5 – 510(k) Summary
# MRI Patient Monitor TeslaDUO
# K153739
#### 1. Submission Sponsor
MIPM Mammendorfer Institut für Physik und Medizin GmbH Oskar-von-Miller-Str. 6 Mammendorf 82291 Germany Phone: (+49) 8145 9209-12 Fax: (+49) 8145 9209-13 Contact: Hans Schwaiger Title: General Manager
#### 2. Submission Correspondent
Emergo Global Consulting, LLC 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: +41 78 843 7077 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: André Kindsvater, Senior Consultant, RA/QA Email: project.management@emergogroup.com
#### 3. Date Prepared
April 28th, 2016
### 4. Device Identification
| Trade/Proprietary Name: | MRI Patient Monitor TeslaDUO |
|-------------------------|--------------------------------------------------------------------------|
| Common/Usual Name: | Physiological Patient Monitor |
| Classification Name: | monitor, physiological, patient (without arrhythmia detection or alarms) |
| Regulation Number: | 870.2300 |
| Product Code: | MWI |
| Device Class: | Class II |
| Classification Panel: | Cardiovascular |
{4}------------------------------------------------
## 5. Legally Marketed Predicate Device(s)
K142032, MRI Patient Monitoring System Tesla M3 MIPM Mammendorfer Institut für Physik und Medizin GmbH
#### 6. Device Description
The MRI Patient Monitor TeslaDUO ("TeslaDUO") is a magnetic resonance compatible patient monitoring device for use in MRI rooms (HF-proof, magnet-resistant) to monitor and display data from vital signs sensors during MRI examinations (MRI procedures) of patients.
It is capable of continuously monitoring and displaying data in graphic and numeric form from the following sensors/measurement modules:
- Pulse Oximetry (SpO2 and Pulse Rate)
- Non-Invasive Blood Pressure (NIBP and PR)
### 7. Indication for Use Statement
The MRI Patient Monitor TeslaDUO is intended for monitoring of vital signs during MRI examinations (MRI procedures) of patients.
The TeslaDUO is intended for use in the Adult and Pediatric populations for continuous monitoring of Pulse Oximetry (SpO2).
The TeslaDUO is intended for use in the Adult, Pediatric and Neonatal populations for continuous monitoring of Non-Invasive Blood Pressure (NIBP).
The TeslaDUO is intended for use by health care professionals.
### 8. Substantial Equivalence Discussion
The following table compares the MRI Patient Monitor TeslaDUO to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance testing. The comparison of the devices provides more detailed information regarding the basis for the determination of substantial equivalence. The subject device does not raise any new issues of safety or effectiveness based on the similarities to the predicate device.
{5}------------------------------------------------
| Manufacturer | MIPM | MIPM | Significant<br>Differences |
|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Trade Name | MRI Patient Monitor<br>TeslaDUO | MRI Patient Monitoring<br>System Tesla M3 | |
| 510(k) Number | TBD | K142032 | N/A |
| Product Code | MWI | MWI | Equivalent |
| Regulation Number | 870.2300 | 870.2300 | Equivalent |
| Regulation Name | Monitor, Physiological, Patient<br>(without arrhythmia detector<br>or alarms) | Monitor, Physiological,<br>Patient (without arrhythmia<br>detector or alarms) | Equivalent |
| Indications for Use | The MRI Patient Monitor<br>TeslaDUO is intended for<br>monitoring of vital signs<br>during MRI examinations (MRI<br>procedures) of patients.<br>The TeslaDUO is intended for<br>use in the Adult and Pediatric<br>populations for continuous<br>monitoring of Pulse Oximetry<br>(SpO2) | The MRI Patient Monitoring<br>System Tesla M3 is intended<br>for monitoring of vital signs<br>during MRI examinations (MRI<br>procedures) of patients.<br>The Tesla M3 is intended for<br>use in the Adult, Pediatric and<br>Neonatal populations for the<br>continuous monitoring of<br>Electrocardiogram (ECG),<br>Non-Invasive Blood Pressure<br>(NIBP), Invasive Blood<br>Pressure (IBP), Temperature,<br>Respiration, Capnography<br>(etCO2), Oxygen and<br>Anesthetic Agents.<br>The Tesla M3 is intended for<br>use in the Adult and Pediatric<br>populations for the<br>continuous monitoring of<br>Pulse Oximetry (SpO2). | Equivalent<br>TeslaDUO does not<br>have ECG, IBP, CO2,<br>Temperature,<br>Oxygen and<br>Anesthetic Agents<br>options |
| | The TeslaDUO is intended for<br>use in the Adult, Pediatric and<br>Neonatal populations for<br>continuous monitoring of<br>Non-Invasive Blood Pressure<br>(NIBP).<br>The TeslaDUO is intended for<br>use by health care<br>professionals. | The Tesla M3 is intended for<br>use by health care<br>professionals. | |
| Intended Population | Adult, Pediatric, Neonatal | Adult, Pediatric, Neonatal | Equivalent |
| Mechanism of<br>Action | Data Acquisition | Data Acquisition | Equivalent |
| Technology<br>Overview | PEMS and Software | PEMS and Software | Equivalent |
| Pulse Oximetry | Waveform and Numeric | Waveform and Numeric | Equivalent |
| NIBP | Numeric | Numeric | Equivalent |
| Manufacturer | MIPM | MIPM | Significant<br>Differences |
| Trade Name | MRI Patient Monitor<br>TeslaDUO | MRI Patient Monitoring<br>System Tesla M3 | |
| Material | Housing mainly coated<br>aluminium,<br>antimagnetic stainless steel,<br>plastics | Housing mainly coated<br>aluminium,<br>antimagnetic stainless steel,<br>plastics | Equivalent |
| Latex Free | Yes | Yes | Equivalent |
| Sterile | No | No | Equivalent |
| Single-Use | No | No | Equivalent |
| Shelf Life | N/A | N/A | Equivalent |
| MRI safe | Yes | Yes | Equivalent |
| Mode of operation | Continuous | Continuous | Equivalent |
| Energy Output | none | none | Equivalent |
| Software | Yes | Yes | Equivalent |
| Battery Operated | Li-lon | Li-lon | Equivalent |
| AC Powered | 100 to 240 VAC, 50/60 Hz | 100 to 240 VAC, 50/60 Hz | Equivalent |
| Complies with ISO<br>10993-1 | Yes | Yes | Equivalent |
| Electrical Safety<br>Testing Passed | Yes | Yes | Equivalent |
# Table 5A – Comparison of Characteristics
{6}------------------------------------------------
## 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of TeslaDUO and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, MIPM completed a number of nonclinical performance tests against applicable standards.
| | Test | Pass / fail criteria | Results |
|---|----------------------------------|------------------------------------------------------------------------------|---------|
| 1 | Electrical safety | Compliance to IEC 60601-1:2012 | Passed |
| 2 | Electromagnetic<br>compatibility | Compliance to EN/IEC 60601-1-2: 2007<br>Compliance to EN/IEC 60601-1-2: 2014 | Passed |
| 3 | Multifunction<br>Patient Monitor | Compliance to IEC 60601-2-49: 2011-02 | Passed |
| 4 | Alarms | Compliance to IEC 60601-1-8:2006+A1:2012-<br>11 | Passed |
| 5 | Biocompatibility | Compliance to ISO 10993-1 | Passed |
| 6 | Risk<br>Management | Compliance to ISO 14971:2007 | Passed |
| 7 | Software | Compliance to IEC 62304:2006 | Passed |
| Table 5B – Performance Standards Testing Summary | | |
|--------------------------------------------------|--|--|
| | | |
{7}------------------------------------------------
| | Test | Pass / fail criteria | Results |
|----|----------------|----------------------------------------------------------------------|---------|
| 8 | Pulse Oximeter | Compliance to ISO 80601-2-61: 2011 | Passed |
| 9 | NIBP | Compliance to IEC 80601-2-30:2009-<br>01 (ed.1.0) | Passed |
| 10 | Batteries | IEC 62133:2012 | Passed |
| 11 | Usability | IEC 60601-1-6:2010-01 + A1:2013<br>IEC 62366:2007 (Ed. 1) + A1: 2014 | Passed |
The TeslaDUO passed all the testing in accordance with internal requirements, national standards, and international standards shown below to support substantial equivalence of the subject device:
Extensive tests were performed to verify the proper functioning of the TeslaDUO and the accessories in the intended Magnetic Resonance (MR) environment. Both, the influence of the TeslaDUO on the functioning of the intended MRI Scanners and the influence of the intended MRI Scanners on the functioning of the TeslaDUO were tested.
Further tests were undertaken to demonstrate the performance of the device and its accessories in the intended MRI environment, such as:
- Measurement of Magnetostatic Forces,
- Magnetic Immunity,
- MRI Immunity, and
- Radio Frequency field induced and Gradient field induced Heating
All tests passed.
#### 10. Clinical Performance Data
There was no human clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with proven safety and efficacy for the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
### 11. Statement of Substantial Equivalence
lt has been shown in this 510(k) submission that the difference between the MRI Patient Monitor TeslaDUO and the predicate devices MRI Patient Monitoring System Tesla M3 do not raise any questions regarding its safety and performance.
The TeslaDUO has the similar intended use and the same technological characteristics as the previously cleared predicate device Tesla M3.
The TeslaDUO as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device(s).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.