K153731 · 410 Medical Innovation, LLC · FPA · Aug 25, 2016 · General Hospital
Device Facts
Record ID
K153731
Device Name
LifeFlow Rapid Infusion Device (LifeFlow Device)
Applicant
410 Medical Innovation, LLC
Product Code
FPA · General Hospital
Decision Date
Aug 25, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The LifeFlow™ Rapid Infusion Device is an intravenous administration set intended for rapid* delivery of fluids from a container into a patient's vascular system. The device is intended to deliver only crystalloid and colloid resuscitative fluids. These devices may be used for any pediatric or adult patient population with consideration given to adequacy of vascular anatomy, appropriateness for the solution being infused and duration of therapy. *Capable of rates greater than 150 mL/min through a 20G needle.
Device Story
LifeFlow™ Rapid Infusion Device is a single-use, disposable IV administration set with a levered hand pump (Handle). Input: IV fluid from standard bags. Operation: Tubing set placed into Handle; manual compression of lever actuates internal syringe to deliver fluid; automatic refill upon release. Output: Rapid fluid delivery (>150 mL/min through 20G needle). Used in ambulance, ER, and ICU by clinicians. Benefits: Enables rapid resuscitation in emergency settings. Healthcare providers view fluid delivery via clear canopy on Handle. Device is non-sterile Handle with gamma-sterilized IV tubing set.
Clinical Evidence
Bench testing only. Testing included reliability, pressure, tactile feel, flow rate evaluations, and user validations. Biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility, pyrogenicity) and extractables/leachables studies performed per ISO 10993-1:2009. Sterilization validated per ANSI/AAMI/ISO 11137-2:2013.
Technological Characteristics
Manual levered hand pump; IV administration set with check valve, tubing, syringe, bag spike, thumb clamp, Luer access device, and male Luer connection. Materials tested per ISO 10993-1. Sterilization: Gamma radiation (ISO 11137-2). Connectivity: None. Form factor: Disposable, single-use.
Indications for Use
Indicated for rapid delivery of crystalloid and colloid resuscitative fluids into the vascular system of pediatric or adult patients, contingent upon adequate vascular anatomy and appropriateness of the solution and therapy duration.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of a human figure with three faces in profile, one behind the other, with flowing lines representing the body.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 25, 2016
410 Medical Innovation, LLC % Dawn Reilly-O'Dell Principal Consultant Full Circle Regulatory Consulting, LLC 201 West Main Street, Suite 305 Durham, North Carolina 27701
Re: K153731
Trade/Device Name: LifeFlow™ Rapid Infusion Device (LifeFlow™ Device) Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: II Product Code: FPA Dated: July 22, 2016 Received: July 25, 2016
Dear Dawn Reilly-O'Dell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Ryan -S
for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Food and Drug Administration
510(k) Number (if known)
K153731
Device Name
LifeFlow™ Rapid Infusion Device (LifeFlow™ Device)
Indications for Use (Describe)
The LifeFlow™ Rapid Infusion Device is an intravenous administration set with Handle intended for rapid* delivery of fluids from a container into a patient's vascular system. The device is intended to deliver only crystalloid and colloid resuscitative fluids. These devices may be used for any pediation with consideration with consideration given to adequacy of vascular anatomy, appropriateness for the solution being infused and duration of therapy.
*Capable of rates greater than 150 mL/min through a 20G needle.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration
Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMBnumber."
{3}------------------------------------------------
# 510(K) SUMMARY
| I. SUBMITTER<br>[807.92(a)(1)] | 410 Medical Innovation, LLC.<br>201 West Main St, Suite 207, Durham, NC 27701<br>Phone: (844) 410-0410<br><br>Contact: Galen Robertson, C.O.O.<br>Date Prepared: August 22, 2016 |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| II. DEVICE NAME<br>[807.92(a)(2)] | Trade: LifeFlow™ Rapid Infusion Device (LifeFlow™ Device)<br><br>Common: IV Administration Set with Hand Pump<br><br>Classification: IV Administration Set<br>(21 CFR 880.5440, Class II, Product Code FPA) |
| III. PREDICATE DEVICE<br>[807.92(a)(3)] | IV Administration Set with Hand Pump<br>(21 CFR 880.5440, Class II, Product Code FPA)<br>B. Braun, Inc., K140838<br><br>This predicate has not been subject to a design-related recall. No reference devices were used in this submission. |
# IV. DEVICE DESCRIPTION [807.92(A)(4)]
Each LifeFlow™ Rapid Infusion Device (LifeFlow™ Device) is a disposable, single-use device that includes an intravenous administration set (Tubing Set) and a Handle and is used to deliver up to 4 Liters of fluid from a container into a patient's vascular system rapidly through manual compression of the levered Handle.
The Tubing Set is sterile, compatible with standard IVs and IV fluid bags and comprised of various generic components, such as a check valve, tubing, syringe, bag spike, thumb clamp, Luer access device, and Luer connections. Each of the needleless components are 510(k) cleared.
The addition of a hand pressure pump, the Handle, provides the capability for rapidly delivering IV fluids. The end of the Tubing Set is placed into the Handle. The clear canopy, through which the graduations and contents of the syringe can be viewed during use, is closed. The Handle is manually compressed to actuate the syringe, which delivers fluid to the patient and then automatically refills when the Handle is released.
{4}------------------------------------------------
## Use Environments: Ambulance, Emergency Room/Department, ICU
Duration of Use: Within 24 hours of starting initial fluid resuscitation
# V. INDICATIONS FOR USE [807.92(A)(5)]
The LifeFlow™ Rapid Infusion Device is an intravenous administration set intended for rapid* delivery of fluids from a container into a patient's vascular system. The device is intended to deliver only crystalloid and colloid resuscitative fluids. These devices may be used for any pediatric or adult patient population with consideration given to adequacy of vascular anatomy, appropriateness for the solution being infused and duration of therapy.
*Capable of rates greater than 150 mL/min through a 20G needle.
# VI. Explanation of Differences in Indications for Use Compared to Predicate:
The indications for use for both subject and predicate devices are substantially equivalent. The differences include 4 minor wording modifications numbered and bolded in the table below; a description of each follows.
| SUBJECT DEVICE | PREDICATE DEVICE |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. The LifeFlow™ Rapid Infusion device<br>is an intravenous administration set | 1. The IV Administration Sets with Hand<br>Pump are intravenous administration sets |
| 2. is intended for rapid* delivery of fluids<br>from a container into a patient's vascular<br>system.<br>*Capable of rates greater than 150<br>mL/min through a 20G needle. | 2. intended for delivery of fluids from a<br>container into a patient's vascular system. |
| 3. These devices may be used for any<br>pediatric or adult patient population with<br>consideration given to adequacy of<br>vascular anatomy, appropriateness for the<br>solution being infused and duration of<br>therapy. | 3. These devices may be used for any<br>patient population with consideration<br>given to adequacy of vascular anatomy<br>and appropriateness for the solution being<br>infused and duration of therapy. |
| 4. The device is intended to deliver only<br>crystalloid and colloid resuscitative fluids. | 4. When the hand pump is activated, the<br>device is intended to deliver only<br>crystalloid and colloid resuscitative fluids. |
1. The first modification reflects the differences in naming of the two devices.
2. The second modification is the addition of the word "rapid". This description of fluid delivery, although not included in the predicate's indications for use, is included in the
{5}------------------------------------------------
Device Description published in the predicate's 510(k) Summary. A definition of rapid flow rate has been added for clarity and is based upon performance testing of the subject device.
3. Both predicate and subject devices are indicated for use with any patient population as long as proper consideration has been given to adequacy of vascular anatomy, appropriateness of the solution being infused and duration of therapy. The subject wording further clarifies "any patient population".
4. It was unclear from the predicate indications for use if the statement "when the hand pump is activated" allows for the syringe and administration tubing to be used without the hand pump to deliver other drugs or fluids. Therefore, the statement was removed for clarity and to align with the supporting performance testing of the subject device.
These modifications do not change the intended use in comparison to the predicate device and do not affect the safety and effectiveness of the device when used as intended.
# VII. TECHNOLOGICAL CHARACTERISTICS [807.92(A)(6)]
The LifeFlow™ Rapid Infusion Device has the same indication for use as the B. Braun IV Administration Sets with Hand Pump in that they are both indicated for infusion therapy for the delivery of fluids to any patient through an increased flow rate provided by use of a manually activated hand pump. In the case of the proposed device, this hand pump is a levered Handle; in the case of the B. Braun IV Administration Sets, this hand pump is a flexible cylindrical chamber. The tubing length of the predicate device could not be confirmed; however, measurements of simulated use environments supported that the tubing length of the LifeFlow Device is sufficient for its intended use environments and therefore substantially equivalent.
| Technical Characteristics | Subject<br>LifeFlow Device | Predicate<br>IV Administration<br>Set with Hand Pump | Substantial<br>Equivalence |
|--------------------------------|----------------------------|------------------------------------------------------|----------------------------|
| Manual hand pump | Yes | Yes | Equivalent |
| Resuscitative fluids | Crystalloid and<br>colloid | Crystalloid and<br>colloid | Equivalent |
| Sterile IV Administration Set | Yes | Yes | Equivalent |
| Single Use | Yes | Yes | Equivalent |
| Needle-free | Yes | Yes | Equivalent |
| Cannula/catheter<br>connection | Male Luer | Male Luer | Equivalent |
#### Substantial Equivalence Table
{6}------------------------------------------------
| Technical Characteristics | Subject<br>LifeFlow Device | Predicate<br>IV Administration<br>Set with Hand Pump | Substantial<br>Equivalence |
|---------------------------------------------|----------------------------------|------------------------------------------------------|----------------------------|
| Fluid container connector | IV spike | IV spike | Equivalent |
| Prevents backflow with dual-<br>check valve | Yes | Yes | Equivalent |
| Injection site / Luer access<br>device | Yes | Yes | Equivalent |
| Tubing set length | 70 in. (50.8 cm)<br>(approx.) | Similar | Functionally<br>Equivalent |
| Rapid fluid delivery | Yes (flow rates<br>> 150 mL/min) | Yes | Equivalent |
## VIII. PERFORMANCE DATA [807.92(B)(1)]
The following FDA recognized performance standards and guidance were utilized in evaluating the functionality of the LifeFlow™ Device:
- ISO 594-1:1986, Conical fittings with a 6% (Luer) taper for syringes, needles and . certain other medical equipment - Part 1: General requirements
- ISO 594-2:1998, Conical fittings with a 6% (Luer) taper for syringes, needles and ● certain other medical equipment - Part 2: Lock fittings
- ISO 8536-4:2010, Infusion equipment for medical use – Part 4: Infusion sets for single use, gravity feed
- ISO 10993-1:2009, Biological evaluation of medical devices - part 1: Evaluation and testing within a risk management process
- . ANSI/AAMI/ISO 11137-2:2013, Sterilization of health care products - Radiation -Part 2: Requirements for development, validation, and routine control of a sterilization process for medical devices
- AAMI/ANSI HE75:2009/(R)2013, Human Factors Engineering - Design of Medical Devices
- . Guidance for Industry and FDA Staff - Intravascular Administration Sets Premarket Notification Submissions [510(k)], July 11, 2008
Functional performance testing was completed with the proposed LifeFlow Device to demonstrate that the device performs as intended. Testing included reliability, pressure, tactile feel, flow rate evaluations and user validations. Results of testing evidence that the proposed device performs in a manner that is substantially equivalent to the predicate.
{7}------------------------------------------------
### Biocompatibility
The materials of construction of the fully assembled LifeFlow Device's IV set were tested according to ISO 10093-1:2009, as recognized by FDA. Testing for cytotoxicity, sensitization, irritation, systemic toxicity, hemocompatibility and pyrogenicity (LAL and material mediated tests) was completed. Chemical characterization, including extractables and leachables studies, and subsequent toxicology risk assessment, were also conducted.
Biocompatibility test results verify that the LifeFlow Device materials of construction are safe for their clinical application.
#### Sterilization
The Handle is not sterile, but the IV set is gamma sterilized. Sterilization and shelf-life testing were completed according to the FDA recognized ANSI/AAMI/ISO 11137-2:2013, Sterilization of healthcare products - Radiation - Part 2: Establishing the Sterilization Dose, and ISO 11607, respectively.
### IV. CONCLUSIONS [807.92(B)(3)]
Biocompatibility, sterilization and non-clinical functional testing support the safety of the device and demonstrate that the LifeFlow Device meets its intended performance requirements. Results documented in user validation confirm the LifeFlow Device is as safe and as effective as the predicate device currently marketed for the same intended use. Therefore, it can be concluded based on the nonclinical tests discussed above that the LifeFlow Device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.