INTRAcel Premium Fractional RF Micro Needle (FRM) System
K153727 · Jeisys Medical, Inc. · GEI · Jul 21, 2016 · General, Plastic Surgery
Device Facts
Record ID
K153727
Device Name
INTRAcel Premium Fractional RF Micro Needle (FRM) System
Applicant
Jeisys Medical, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Jul 21, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The INTRAcel Premium device is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.
Device Story
INTRAcel Premium is an electrosurgical system using bipolar radiofrequency (RF) energy delivered via micro-needle electrodes. The device consists of a main frame, handpiece, and disposable tips. It is operated by a clinician in a clinical setting. The user programs treatment parameters (energy levels, depth) via a touch-screen interface. The device delivers RF energy to the skin to induce thermal coagulation of collagen, facilitating hemostasis and tissue remodeling. The clinician monitors the procedure via the display and manual interface. The system includes safety features such as emergency switches, key switches, and audible/visual indicators. Benefits include controlled tissue coagulation for dermatological and surgical applications.
Clinical Evidence
No clinical testing was required for the 510(k) clearance. Evidence includes bench testing (electrical safety, EMC, dimensional accuracy), biocompatibility (cytotoxicity, sensitization, reactivity, systemic toxicity), and sterilization validation. Preclinical performance was supported by an in-vivo porcine model evaluating wound healing and collagen response, and an ex-vivo animal tissue study comparing the subject device to the predicate, demonstrating similar conical diamond-shaped tissue coagulation columns.
Indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three overlapping profiles of human faces, creating a sense of community and connection. The logo is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 21, 2016
Jeisys Medical Incorporated c/o Ms. Priscilla Chung LK Consulting Group USA, Inc. 2651 E. Chapman Ave., Ste 110 Fullerton. CA 92831
Re: K153727
Trade/Device Name: INTRAcel Premium Fractional RF Micro Needle (FRM) System Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical, cutting and coagulation device & accessories Regulatory Class: Class II Product Code: GEI Dated: June 7, 2016 Received: June 17, 2016
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Christopher J. Ronk -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K153727
Device Name
INTRAcel Premium Fractional RF Micro Needle (FRM) System
Indications for Use (Describe)
The INTRAcel Premium Fractional RF Micro Needle is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.
| Type of Use (Select one or both, as applicable) | <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
EF
PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
# 510(k) Summary (K153727)
#### 1. Submission Sponsor
Jeisys Medical, Inc. 307 Daeryung Techno Town 8th Gamasan-ro 96, Geumcheon-Gu, Seoul 153-775 Korea Phone: (82) 2.2603.6417 Fax: (82) 2.2603.6447 Contact: Hyo Seok Lee, RA Manager
#### 2. Submission Correspondent
Priscilla Chung LK Consulting Group USA, Inc. 2651 E Chapman Ave Ste 110, Fullerton CA 92831 Tel: 714-202-5789 Email: juhee.c@LKconsultinggroup.com
#### 3. Date Prepared
July 19, 2016
#### 4. Device Identification
| Trade/Proprietary Name: | INTRAcel Premium Fractional RF Micro Needle (FRM) System |
|----------------------------|----------------------------------------------------------------|
| Common/Usual Name: | Electrosurgical coagulation device and accessories |
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Classification Regulation: | 878.4400 |
| Product Code: | GEI |
| Device Class: | Class II |
| Classification Panel: | General & Plastic Surgery |
#### 5. Predicate Devices
Lutronic — INFINI, K121481
{4}------------------------------------------------
#### 6. Device Description
The INTRAcel Premium consists of the following components:
- INTRAcel Premium (Main Frame)
- Active Accessory
- : 10xINTRAcelTip 1011 1x Power cable 1x Foot switch 1xHandpiece 1xHandpiece hanger 1x User's manual
#### 7. Intended Use
The INTRAcel Premium device is indicated for use in dermatological and general surgical procedures for electrocoagulation and hemostasis.
#### 8. Comparison of Technological Characteristics
The INTRAcel Premium device shares the same or similar indications for use, device operation, overalltechnical and functional capabilities, and therefore is substantially equivalent to the predicate devices. The INTRAcel Premium device is similar in design and function to the predicate devices for the modes of operation and use.
The devices have a control unit that can be programmed utilized for the patient parameters. The devices are equipped with an input device either a stylus or touch screen to program the parameters. In addition, they are equipped with manual interface, enabling to operate the device during therapy, and a display, which shows the set and indicated parameters. During operation the devices have an applicator instrument attached to the main unit.
These devices all have the same intended use and indications for use as the INTRAcel Premium device. Devices use RF energy delivered through micro needle electrodes to provide treatment controlled by a user controlled interface.
#### 9. Substantial Equivalence Discussion
The following table compares the INTRAcel Premium device to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
{5}------------------------------------------------
| | | | | Table 5A - Comparison of Characteristics |
|--|--|--|--|------------------------------------------|
|--|--|--|--|------------------------------------------|
| | Manufacturer<br>Trade Name | Lutronic<br>Predicate Device<br>INFINI | Jeisys Medical, Inc.<br>Subject Device<br>INTRAcel Premium |
|-----------------------------|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | | K121481 | K153727 |
| Intended Use | | The device is intended for use in<br>dermatologic and general surgical<br>procedures for electrocoagulation<br>and hemostasis and the<br>percutaneous treatment of facial<br>wrinkles. | The INTRAcel Premium is<br>intended for use in dermatologic<br>and general surgical procedures<br>for electrocoagulation and<br>hemostasis. |
| Design | | The device uses Bipolar<br>radiofrequency to apply heat therapy<br>to the skin, resulting skin resurfacing<br>and collagen remodeling for<br>treatment of wrinkles. | The device uses Bipolar radio<br>frequency to apply therapy for<br>use in Dermatologic and General<br>Surgical procedures<br>electrocoagulation and<br>hemostasis. |
| Materials | | Plastic and metal enclosure with<br>hand-piece for application | Plastic and metal enclosure with<br>hand-piece for application |
| Construction | | Constructed of materials that<br>conform to safety standards and<br>requirements | Constructed of materials that<br>conform to safety standards and<br>requirements |
| Interface | | Touch screen user applied interface<br>to program and setthe controls for<br>the patient application; there is a<br>hand-piece utilized to deliver the<br>treatment | Touch screen user applied<br>interface to program and set the<br>controls for the patient<br>application; there is a hand-piece<br>utilized to deliver the treatment |
| Electr<br>osurgical<br>Unit | Energy Source<br>Mode of Operation<br>Power Source<br>Frequency<br>Nominal Operating<br>Power | Radio Frequency<br>Bipolar<br>Unknown<br>1 MHz<br>50W (Up to 20 Level) | Radio Frequency<br>Bipolar<br>120 VAC, 50/60 Hz<br>1 MHz<br>12.5W (Level 1 -4),<br>32W (Level 5)<br>40.5 W(Level 6)<br>50 W(Level 7) |
| | Maximum power<br>delivered to the<br>patient | 50W | 50W |
| | Maximum power of<br>per pin delivered to<br>the patient | 50W | 50W |
| | Impedance | | 200 ohm |
| | Treatment<br>temperature range | 36 - 43°C | 39°C - 42°C |
| | Treatment levels | 20 Level | 1 – 7 levels (1 low, 7 high) |
| | Dimensions | 362 mm (W) x 40 mm (L) x 108 mm<br>(H) | 350 mm (W) x 400 mm (L) x 108<br>mm (H) |
| | Weight | 28 kg | 63 kg |
{6}------------------------------------------------
| | Safety features | | Safety interlocks, removable key<br>switches, hand-piece triggering,<br>visual and audible indicators, cut-off<br>temperature switches, emergency<br>off button. | There are an emergency switch<br>and a key switch so that the<br>operator can control by touch<br>screen based on window O/S<br>program. |
|-------------------------|-------------------------------|---------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Active<br>Access<br>ory | Dimensions | Depth of<br>skin ablation | 0.5-3.5mm | 0.5/0.8/1.5/2.0mm |
| | | Pin<br>thickness | 0.25mm | 0.25mm |
| | | Needle<br>Quantity | 49ea | 49ea |
| | | Pin<br>length | 7.6mm | 11mm |
| | | The<br>distance<br>between<br>the pin | 1.3mm | 1.3mm |
| | Material | | Unknown+SUS304 | ABS+SUS304 |
| | RF treatment area | | Spot Size (Treated area) : 1cm X 1cm | Spot Size (Treated area) : 1cm X<br>1cm |
| | Recommended<br>Treatment Time | | 10 min | 10min~15min |
#### 10. Non-Clinical Performance Data
The device's hardware and software development, verification, and validation have been carried out in accordance with FDA guidelines. The software was tested against the established Software Design Specifications for each of the test plans to assure the deviceperforms as intended. The Device Hazard analysis was completed and risk control implemented to mitigate identified hazards. The testing result supports that all the hardware specifications and software specifications have met the acceptance criteria of each module and interaction of processes. The INTRAcel Premium device passed all testing and supports the claims of substantial equivalence and safe operation.
The INTRAcel Premium device complies with the applicable voluntary standards for Electromagnetic Compatibility and Safety. The device passed all the electrical and safety testing according tonational and international standards.
| Testing Type | Test Description | Test Result |
|-------------------------------------------------------------------|--------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety and<br>Electromagnetic<br>Compatibility Testing | • IEC 60601-1:2012<br>• IEC 60601-2-2:2009<br>• IEC 60601-1-2:2007 | The INTRAcel Premium met<br>all acceptance criteria in<br>accordance with IEC 60601-<br>1:2009, IEC 60601-2-2:2009<br>and IEC 60601-1-2:2007. |
| Performance Testing | • Dimensional Inspection<br>• Accuracy of Measurement | The INTRAcel Premium met<br>all acceptance criteria. |
{7}------------------------------------------------
| Biocompatibility Testing | Cytotoxicity Sensitization Intracutaneous Reactivity Acute Systemic Toxicity | The active electrodes are<br>biocompatible. |
|-----------------------------------------|-----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Sterilization and Shelf<br>Life Testing | Ethylene Oxide Residues Bioburden Sterility Bacteriostasis/Fungistasis Shelf Life | The INTRAcel Premium met<br>all acceptance criteria. |
| Software design Testing | IEC62304:2006 | The INTRAcel Premium met<br>all acceptance criteria in<br>accordance with IEC<br>62304:2006 |
The preclinical performance test by in-vivo micropig model was performed to investigate the wound healing response in porcine subjects after the Fractional RF Micro Needle (FRM) treatment using histopathological examination.The treatment was performed at the energy level, 0.5J, 1.25J and 3.0J and in depth ofmicroneedling0.5mm, 0.8mm, 1.5mm and 2.0mm. Histologic evaluation was done by H&E and Trichrome for immediately after, 7 days after, 14 days after, 28 days after and 10 weeks after procedure.Histologically, the thermally coagulated collagen after treatment was replaced by new collagen.
Additional preclinical performance test was performed to investigate ex-vivo animal tissue testing comparing the subject device and predicate device power delivery characteristics effect on tissue.The treatment was performed at the energy level, 0.5J, 1.25J and 3.0J in depth of microneedling0.5mmand 2.0mm. Histologic evaluation was done by H&E and Trichrome for immediately after procedure. Histologically, both Subject device and predicate device created conical diamond shaped tissue coagulation inthe dermis and show similar coagulated column. The INTRAcel Premium Fractional RF Micro Needle (FRM) System was found to be substantially equivalent to the predicate device foruseindermatological and general surgical procedures for electrocoagulation and hemostasis.
### 11. Clinical Testing
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate devices. However, we did clinical testing to validate safety and efficacy. It was published in the surgical corner of Journal of Drugs in Dermatology. The verification and validation testing of the device software and electrical safety and EMC testing of the device was found to acceptable and supports the claims of substantial equivalence.
#### 12. Statement of Substantial Equivalence
It has been shown in this 510(k) submission that the difference between the INTRAcel Premium system and the predicate device do not raise any questions regarding its safety and effectiveness. The INTRAcel Premium system, as designed and manufactured, is determined to be substantially equivalent to the referenced predicate device.
{8}------------------------------------------------
#### 13. Conclusion
The INTRAcel Premium device has similar intended use and technological characteristics as the predicate devices.
The information provided in this submission supports the substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.