# PleuraFlow System (K153681)

_Clearflow, Inc. · OTK · Jan 20, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153681

## Device Facts

- **Applicant:** Clearflow, Inc.
- **Product Code:** [OTK](/productcode/OTK.md)
- **Decision Date:** Jan 20, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4780
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Real-World Evidence, Pediatric

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K153681 · Jan 20, 2016 | PleuraFlow System | Clearflow, Inc. | Published clinical literature (Sirch et al., 2015) reporting on clinical use of the device | The sponsor cites post-market clinical data from a published study to support the safety and effectiveness of the PleuraFlow System. | Post-market clinical data; Clinical literature; Thoracic surgery |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Active Clearance of Chest Drainage Catheters Reduces Retained Blood (Sirch et al., 2015); Clinical study (published literature) | Patients undergoing cardiothoracic surgery | Not applicable for this study | Reduction of retained blood |

## Indications for Use

The PleuraFlow System is indicated for use during cardiothoracic surgical procedures and chest trauma. Its Active Clearance Technology® proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings.

## Device Story

PleuraFlow System consists of silicone chest tube and clearance apparatus; used in clinical settings for cardiothoracic surgery or chest trauma. Device inputs: blood/fluid accumulation in chest tube. Principle of operation: Active Clearance Technology uses PTFE-coated clearance wire with distal loop, advanced into chest tube via magnetic shuttle; wire/loop manually advanced/retracted to break up clots/obstructions. Output: patent chest tube enabling fluid evacuation. Healthcare provider (surgeon) uses device to maintain tube patency, reducing retained blood. Benefits: prevention of chest tube occlusion, improved drainage, reduced retained blood complications.

## Clinical Evidence

No new clinical data generated for this specific 510(k). Safety and effectiveness previously established via design validation/verification (K093565, K150042). Biocompatibility per ISO 10993-1. Mechanical testing (flow rate, tensile strength) performed per BS EN 1617:1997 and BS EN 1618:1997. Post-market clinical evidence for the PleuraFlow system cited from Sirch et al. (2015) in the Journal of Thoracic and Cardiovascular Surgery.

## Technological Characteristics

Silicone chest tube; PTFE-coated clearance wire with distal loop; magnetic shuttle. 20FR size with 2-inch Effective Drainage Length (4 eyelets). Barium stripe for visualization. Mechanical actuation via manual advancement/retraction of wire. Biocompatibility per ISO 10993-1. Testing per BS EN 1617:1997 and BS EN 1618:1997.

## Regulatory Identification

A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.

## Predicate Devices

- PleuraFlow System ([K150042](/device/K150042.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an emblem featuring a stylized depiction of three human profiles forming a wing-like shape.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 20, 2016

Clearflow Incorporated Mr. Dov Gal Vice President Regulatory Affairs and Quality Assurance 1630 South Sunkist St, Suite E Anaheim, California 92806

Re: K153681

Trade/Device Name: PleuraFlow® System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OTK, GBX Dated: September 17, 2015 Received: September 18, 2015

Dear Mr. Gal:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you: however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Section 4: Indications for Use Statement

K153681 510(k) Number:

PleuraFlow® System Device Name:

Indications for Use: The PleuraFlow System is indicated for use during cardiothoracic surgical procedures and chest trauma. Its Active Clearance Technology® proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings.

| Prescription Use _X_        | AND/OR | Over-The-Counter Use _ |
|-----------------------------|--------|------------------------|
| (Part 21 CFR 801 Subpart D) |        | (21 CFR 807 Subpart C) |

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

{3}------------------------------------------------

## Section 5: 510(k) Summary

The following information is provided as required by 21 CFR § 807.87 for PleuraFlow® System with Short Effective Drainage Length (SEDL) Chest Tube 510(k) premarket notification. In response to the Safe Medical Devices Act of 1990, the following is a summary of the safety and effectiveness information upon which the substantial equivalence determination is based.

| Date of Submission:      | December 21, 2015                                                                                                                                               |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant:               | ClearFlow, Inc.<br>1630 S. Sunkist St., Suite E<br>Anaheim CA 92806                                                                                             |
| Primary Contact Person:  | Dov Gal, DVM<br>ClearFlow, Inc.<br>1630 S. Sunkist St., Suite E<br>Anaheim CA 92806<br>Phone: 714-916-5014<br>Fax: 714-916-5019<br>Email: dgal@clearlflow.com   |
| CEO/President:           | Paul Molloy<br>ClearFlow, Inc.<br>1630 S. Sunkist St., Suite E<br>Anaheim CA 92806<br>Phone: 714-905-5271<br>Fax: 714-916-5019<br>Email: pmolloy@clearlflow.com |
| Device Proprietary Name: | PleuraFlow® System                                                                                                                                              |

{4}------------------------------------------------

Introduction/drainage; wound drain catheter system. Device Common Name:

Regulatory Class and Name: Class II, Powered Suction Pump

Product Codes: OTK and GBX

Indication For Use: The PleuraFlow System is indicated for use during cardiothoracic surgical procedures and chest trauma. Its Active Clearance Technology proactively removes clots formed inside the chest tube to prevent or minimize chest tube occlusion with clot. A patent chest tube enables evacuation of blood and fluid from the operative site after closure of the surgical wound and reduces retained blood. The product is indicated for adult and pediatric patients including infant, preadolescent and adolescent patients under clinical settings.

Predicate Device: Predicate device is the PleuraFlow System (K150042) by ClearFlow, Inc.

Device Description: The PleuraFlow System is comprised of a silicone Chest Tube and a Clearance Apparatus. Currently the PleuraFlow Chest Tube is available in four (4) models with standard sizes (20FR, 24FR, 28FR and 32FR). The Chest Tubes of currently marketed and above-references PleuraFlow System models include six (6) eyelets distributed along an Effective Drainage Length of 4 inches (10.2 cm). The Effective Drainage Length is defined as the length of the Chest Tube having eyelets for the influx of fluid. The current Chest Tube models are available with a Cut Length of 19 inches (48.3 cm) with graduated measurements in centimeters from the distal eyelet. Each Chest Tube has a barium stripe to facilitate visualization. The Chest Tube is connected to a Clearance Apparatus, which is connected to the tubing from the drainage canister.

> The new PleuraFlow System model is similar to the current models except for the Chest Tube, which has four eyelets instead of 6 eyelets. The short Effective Drainage length is only available with a 20FR Chest Tube identifies as: Short Effective Drainage Length (SEDL). This

{5}------------------------------------------------

model has four (4) eyelets distributed along an Effective Drainage Length of 2 inches (5.1 cm). This Chest Tube model has a Cut Length of 19 inches (48.3 cm) with graduated measurements in centimeters from the distal eyelet and a barium stripe to facilitate visualization, as all of the currently marketed Chest Tubes have. Same as the marketed models, the 20FR SEDL Chest Tube is connected to a Clearance Apparatus, which is connected to the tubing from the drainage canister. The Clearance Apparatus that is part of the PleuraFlow System with 20FR SEDL Chest Tube is same as the predicate. It consists of a Guide Tube and a PTFE-coated Clearance Wire with a Loop set on its distal end, bent at a 105-degree angle. The Clearance Apparatus is advanced into the PleuraFlow Chest Tube using a magnetic Shuttle. When indicated, the Clearance Wire and Loop is advanced and retracted within the PleuraFlow Chest Tube to proactively prevent or break up and clear any tube obstructions or clogging to keep the tube open.

The PleuraFlow System with 20FR Short Effective Drainage Length (SEDL) Chest Tube can be used when the surgeon determines that a chest tube with a shorter length of drainage holes, 2" (5.1 cm) is needed. This includes patients having smaller dimension chests (length of torso) that cannot accommodate Chest Tubes with longer EDLs.

- Performance Data: The safety and effectiveness of the PleuraFlow System has been previously demonstrated through design validation and verification that were cleared under 510(k) premarket notification K093565 and K150042. The biocompatibility evaluation for the PleuraFlow System was conducted under predicate 510(k) K093565, in accordance with the FDA Blue Book Memorandum #G95-1 and International Standard ISO 10993-1.
Mechanical testing was performed to establish substantial equivalence to the predicate 510(k) K093565. Testing was performed in accordance with BS EN 1617:1997 and BS EN 1618:1997. These included flow rate and tensile strength testing.

Device performance has been further demonstrated through post-

{6}------------------------------------------------

market data published in the Journal of Thoracic and Cardiovascular Surgery [Sirch J, Ledwon M, Püski T, Grossmann I, Boyle EM, Pfeiffer S, Fischlein T. Active Clearance of Chest Drainage Catheters Reduces Retained Blood. J Thorac Cardiovasc Surg. 2015 Oct 22. pii: S0022-5223(15)01970-4. doi: 10.1016/j.jtcvs.2015.10.015].

Conclusion: The device Indication for Use is same as the predicate. The evaluation of the PleuraFlow System does not raise any additional concerns regarding safety and effectiveness and may be considered substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K153681](https://fda-staging.innolitics.com/device/K153681)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).