K153669 · Arthocare Corporation · MBI · Jan 19, 2016 · Orthopedic
Device Facts
Record ID
K153669
Device Name
MultiFIX S Ultra Knotless Fixation System
Applicant
Arthocare Corporation
Product Code
MBI · Orthopedic
Decision Date
Jan 19, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The MultiFIX S Ultra Knotless Fixation Device is indicated for use in fixation of soft tissue to bone. Examples of such procedures include: Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis, and deltoid repair Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction Foot: Hallux valgus reconstruction Elbow: Tennis elbow repair, biceps tendon reattachment Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction. ITB tenodesis: patellar ligament and tendon avulsions
Device Story
Implantable bone anchor system for arthroscopic/orthopedic soft tissue-to-bone fixation; eliminates manual surgical knot tying. System comprises PEEK anchor preloaded on disposable inserter handle. Surgeon inserts anchor into bone; utilizes interference fit for bone locking and plug/cylinder compression for suture locking. Used in shoulder, ankle, foot, elbow, and knee procedures. Provides secure fixation to facilitate tissue healing. Single-use, sterile, irradiation-sterilized.
Clinical Evidence
Bench testing only. In vitro testing per FDA Guidance (April 1996) included static and cyclic fatigue testing in simulated bone substrate. Design verification and simulated use testing confirmed performance specifications met.
Indicated for soft tissue to bone fixation in orthopedic/arthroscopic procedures including shoulder, ankle, foot, elbow, and knee repairs. Applicable to patients requiring ligament, tendon, or capsule reconstruction/reattachment.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, with three profiles overlapping to create a sense of depth and connection. The profiles are rendered in black and white. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 19, 2016
ArthoCare Corporation Ms. Laura Kasperowicz Principle Regulatory Affairs Specialist 15285 Alton Parkway, Suite 200 Irvine, California 92618
Re: K153669
Trade/Device Name: MultiFIX® S Ultra Knotless Fixation System Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: December 18, 2015 Received: December 21, 2015
Dear Ms. Kasperowicz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153669
Device Name MultiFIX S Ultra Knotless Fixation System
Indications for Use (Describe)
The MultiFIX S Ultra Knotless Fixation Device is indicated for use in fixation of soft tissue to bone. Examples of such procedures include:
Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shiftleapsulelabral reconstruction, biceps tenodesis, and deltoid repair
Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction
Foot: Hallux valgus reconstruction
Elbow: Tennis elbow repair, biceps tendon reattachment
Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction. ITB tenodesis: patellar ligament and tendon avulsions
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for ArthroCare. The logo consists of a stylized, circular graphic on the left and the company name "ArthroCare" on the right. The graphic is made up of several curved lines in a maroon color, arranged to create a sense of movement or flow. The text "ArthroCare" is written in a dark gray, sans-serif font, with the registered trademark symbol "®" appearing in superscript next to the word "Care".
# 510(k) Summary
# ArthroCare® Corporation
MultiFIX® S Ultra Knotless Fixation System
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### General Information
| Submitter Name: | ArthroCare Corporation, a Smith & Nephew Company | |
|----------------------|----------------------------------------------------------------------------------------------------------|--|
| Address | 15285 Alton Parkway, Suite 200<br>Irvine, CA. 92618 | |
| Contact Person: | Laura Kasperowicz<br>Principle Regulatory Affairs Specialist<br>Phone: 949-585-2406<br>Fax: 949-585-2401 | |
| Date Prepared: | December 18, 2015 | |
| Device Name | | |
| Proprietary Name: | MultiFIX® S Ultra Knotless Fixation System | |
| Common Name: | Bone Anchor | |
| Classification Name: | Smooth or threaded metallic bone fixation fastener | |
| Device Class: | Class II | |
| Product Code: | MBI | |
| CFR Section: | 21 CFR 888.3040 | |
## Predicate Device
MultiFIX® S Knotless Fixation System: K151660 (cleared July 16, 2015)
#### Descrintion
The MultiFIX S Ultra Knotless Fixation System (MultiFIX S Ultra) is an implantable bone anchor with inserter handle designed for use in arthroscopic and orthopedic procedures. The MultiFIX S Ultra is a knotless fixation device, meaning that manually tying surgical knots is not necessary for the fixation of suture to tissue.
The MultiFIX S Ultra consists of two primary parts: an implantable bone anchor and an anchor inserter, which is preloaded with the anchor. The anchor inserter is a disposable tool.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for ArthroCare. The logo consists of a stylized, abstract symbol on the left and the word "ArthroCare" on the right. The symbol is made up of several curved lines that are arranged in a circular shape. The word "ArthroCare" is written in a sans-serif font and is in a dark gray color. There is a registered trademark symbol to the right of the word "ArthroCare".
The entire product is packaged in a tray with a Tyvek® lid, and the finished product is sterilized by irradiation. Both the anchor and inserter are designed for single use only.
### Intended Use/Indications For Use
The MultiFIX S Ultra Knotless Fixation Device is indicated for use in fixation of soft tissue to bone.
Examples of such procedures include:
Shoulder: Bankart Repair, SLAP lesion repair, acromio-clavicular separation, rotator cuff repair, capsule shift/capsule-labral reconstruction, biceps tenodesis, and deltoid repair
Ankle: Lateral instability, medial instability, Achilles tendon repair/reconstruction, and midfoot reconstruction
Foot: Hallux valgus reconstruction
Elbow: Tennis elbow repair, biceps tendon reattachment
Knee: Extra-capsular repairs; reattachment of: medial collateral ligament, posterior oblique ligament or joint capsule closure to anterior proximal tibia; extra capsular reconstruction, ITB tenodesis; patellar ligament and tendon avulsions
### Non-Clinical Data
Bench testing was performed on both the proposed and predicate devices in accordance with the FDA Guidance Document, Testing Bone Anchors, April 1996. This in vitro testing involved insertion of the anchors into a simulated human bone substrate followed by both static and cyclic fatigue testing. Design Verification testing was performed in order to verify the design outputs meet the design inputs. Simulated use testing was performed on the device and accessory devices which will be promoted for use with the MultiFIX S Ultra system in order to affirm safety and effectiveness.
The test results demonstrate that the proposed MultiFIX S Ultra meets its design, performance, and safety specifications. Based on the test results, the proposed device performs as intended and mechanical properties are substantially equivalent to the predicate device when used in accordance with labeling.
### Clinical Data
No clinical or animal data are included in this submission.
### Summary
All testing demonstrates that the proposed MultiFIX S Ultra performs as intended and has acceptable mechanical properties when used in accordance with the labeling.
As the intended use, operating principle, materials and technological characteristics are comparable to the predicate device, the proposed MultiFIX S Ultra Knotless Fixation System is substantially equivalent. The minor differences between the proposed MultiFIX S Ultra and predicate device do not raise any new questions of safety or effectiveness.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for ArthroCare. The logo consists of a circular graphic on the left and the word "ArthroCare" on the right. The circular graphic is made up of several curved red lines. The word "ArthroCare" is written in a dark gray sans-serif font, and there is a registered trademark symbol after the word.
| Comparison of Technological Characteristics | | |
|---------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Characteristics | Predicate Device<br>MultiFIX S<br>(K151660) | Proposed Device<br>MultiFIX S Ultra |
| Intended Use | Fixation of soft tissue to bone | Same |
| Delivery Method | Arthroscopic and Limited Access | Same |
| How Supplied | Sterile | Same |
| Brand Name of Qualified<br>Suture | #2 UHMWPE MagnumWire® Suture Thread<br>and/or UltraBraid® Suture Thread and/or<br>UltraTape® | Same |
| # of Suture Legs | 2, 3 or 4 Suture Thread Legs<br>or a maximum of 2 Suture Thread Legs<br>and 2 Suture Tape Legs | 2, 3 or 4 Suture Thread Legs<br>or a maximum of 4 Suture Tape Legs |
| Suture Snare | Single Loop | Dual Loop |
| Anchor Material | Invibio PEEK Optima® with laser marking | Same |
| Design Technology | Pound in Anchor with screw | Same |
| Bone Locking Mechanism | Interference Fit<br>(threaded screw) | Same |
| Suture Locking Mechanism | Plug/Cylinder Compression | Same |
| Diameter of Cortical Lock | 5.5 mm & 6.5 mm | Same |
| Anchor Deployed Length | 20 to 23 mm | Same |
| Device Length | 291mm | Same |
| Sterilization Method | Irradiation | Same |
| Packaging | Sterile / Thermoform Tray with Tyvek Lid | Same |
| Accessory Devices<br>Promoted for use with the<br>MultiFIX S System | Insertion Guide with Obturator, Bone<br>Punch, Anchor Extraction Tool, Reusable<br>FirstPass Suture Passer, SmartStitch Suture<br>Passer, PerfectPasser Suture Passer | Same and FirstPass® ST Suture Passer |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.