Juliet® Ti PO, OL & TL lumbar interbody devices are indicated for intervertebral body fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. Juliet® Ti lumbar interbody devices are to be used with supplemental fixation that has been cleared for use in the lumbosacral spine. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.
Device Story
Juliet® Ti is a range of intervertebral body spacers designed for lumbar fusion procedures via open posterior or transforaminal approaches. Implants are additively manufactured from titanium alloy, featuring both solid and porous structures to facilitate bone ingrowth. The device is used by surgeons in an operating room setting to restore disc height and provide stability in conjunction with supplemental fixation and bone graft (autogenous or allogenic). The variety of lengths, widths, heights, and lordotic angles allows for anatomical adaptation to individual patient conditions. The device is supplied sterile for single use, while associated surgical instruments are reusable and provided non-sterile. Clinical benefit is derived from the stabilization of the spinal segment to promote fusion in patients with degenerative disc disease.
Clinical Evidence
Bench testing only. Mechanical performance validated via static and dynamic axial compression (ASTM F2077), static and dynamic shear compression (ASTM F2077), and subsidence (ASTM F2267). Porous structure characterization performed via static shear strength (ASTM F1044), fatigue shear strength (ASTM F1160), static tensile strength (ASTM F1147), and abrasion resistance (ASTM F1978). Material properties characterized per ASTM F3001 and ASTM E8/E8M.
Technological Characteristics
Intervertebral body fusion device made of Ti6Al4V ELI titanium alloy (ASTM F136). Additively manufactured with solid and porous structures. Dimensions vary by length, height, width, and lordotic angle. Gamma sterilization. Standalone implant used with supplemental fixation.
Indications for Use
Indicated for skeletally mature patients with degenerative disc disease (DDD) of the lumbar spine (L2-S1) at one or two contiguous levels, including those with up to Grade 1 spondylolisthesis or retrolisthesis, who have failed at least 6 months of non-operative treatment.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of human profiles facing right, arranged in a cascading manner. The profiles are rendered in a dark color, creating a sense of depth and connection. Encircling the profiles is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 15, 2016
SPINEART Mr. Franck Pennesi Director of Industry & Quality International Center Cointrin 20, route de Pré-Bois - CP1813 1215 Geneva. SWITZERLAND
Re: K153621
Trade/Device Name: Juliet® Ti Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: June 13, 2016 Received: June 15, 2016
Dear Mr. Pennesi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
K153621
Page 1 of 1
510(k) Number (if known) K153621
Device Name Juliet® Ti
#### Indications for Use (Describe)
Juliet® Ti PO, OL & TL lumbar interbody devices are indicated for intervertebral body fusion procedures in sketally mature patients with degenerative disc disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1. DDD is defined as discogenic back pain with degeneration of the disc confirmed by patient history and radiographic studies. These DDD patients may also have up to Grade 1 spondylolisthesis at the involved level(s). These spinal implants are to be used with autogenous bone graft and/or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft. Juliet® Ti lumbar interbody devices are to be used with supplemental fixation that has been cleared for use in the lumbosacral spine. Patients should have at least six (6) months of non-operative treatment prior to treatment with an intervertebral cage.
#### Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Zane W. Wyatt Division of Orthopedic Devices
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Traditional 510k JULIET * Ti
Image /page/3/Picture/1 description: The image shows the logo for Spineart. The logo consists of the word "spineart" in a stylized font, with a butterfly above the word. The butterfly is purple, and the word "spineart" is in a dark color, with the "a" and "r" in a lighter shade of purple.
### 510(k) SUMMARY
| Submitted by | SPINEART |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | International Center Cointrin |
| | 20 route de pré-bois |
| | CP1813 |
| | 1215 GENEVA 15 |
| | SWITZERLAND |
| Contacts | Franck PENNESI Director of Industry & Quality |
| | Phone : +41 22 570 1246 Fax : +41 22 799 4026 |
| | Mail : fpennesi@spineart.com |
| | Regulatory contact : Dr Isabelle DRUBAIX (Idée Consulting) |
| | Mail : idrubaix@nordnet.fr |
| Date Prepared | June 13, 2016 |
| Common Name | Intervertebral body fusion device |
| Trade Name | JULIET ® Ti |
| Classification Name | Intervertebral Fusion Device With Bone Graft, Lumbar |
| Class | II |
| Product Code | MAX |
| CFR section | 888.3080 |
| Device panel | ORTHOPEDIC |
| Legally marketed predicate devices | Primary predicate: Cascadia Interbody System (K150481) manufactured by K2M, |
| | Inc |
| | Additional predicates: Dynamik intervertebral body fusion device (renamed |
| | Juliet) (K081888) manufactured by Spineart; JULIET PO (K142277) manufactured |
| | by Spineart; Capstone Control Spinal System (K120368) manufactured by |
| | Medtronic Sofamor Danek Usa, In; Capstone Spinal System (K133650) |
| | manufactured by Medtronic Sofamor Danek Usa, In; LUCENT (K071724) |
| | manufactured by Spinal Elements, Inc |
| Indications for use | Juliet® Ti PO, OL & TL lumbar interbody devices are indicated for intervertebral<br>body fusion procedures in skeletally mature patients with degenerative disc<br>disease (DDD) of the lumbar spine at one or two contiguous levels from L2-S1.<br>DDD is defined as discogenic back pain with degeneration of the disc confirmed<br>by patient history and radiographic studies. These DDD patients may also have up<br>to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). These spinal<br>implants are to be used with autogenous bone graft and/or allogenic bone graft<br>comprised of cancellous and/or corticocancellous bone graft. Juliet® Ti lumbar<br>interbody devices are to be used with supplemental fixation that has been<br>cleared for use in the lumbosacral spine. Patients should have at least six (6)<br>months of non-operative treatment prior to treatment with an intervertebral<br>cage. |
| Description of the device | Spineart Juliet® Ti PO, OL & TL spinal implants consist of a range of intervertebral<br>body spacers with various shapes and designs so as to be implanted via an open<br>posterior or transforaminal approach and to adapt different patient's conditions.<br>Spineart JULIET® Ti cage is offered in a variety of lengths, heights, widths and<br>lordotic angles to adapt to a variety of patient anatomies.<br>Juliet® Ti PO, OL & TL spinal implants are made of Titanium alloy Ti6Al4V ELI<br>conforming to ASTM F136. Juliet® Ti PO, OL & TL spinal implants are additively<br>manufactured and present both solid and porous structures.<br>The Juliet Ti cages are delivered sterile (gamma sterilization) and supplied with<br>surgical instruments (reusable - provided non-sterile). |
| of<br>Summary<br>Technological<br>Characteristics | The subject JULIET Ti device is identical in indications for use, surgical technique,<br>manufacturing method and raw material to the predicate device cleared in<br>K150481. The subject JULIET Ti device shares similar design features, comparable<br>heights, widths, depths, lordotic angles and surgical instrumentation as the<br>predicate devices (K081888, K142277, K120368, K133650, K071724). |
| Discussion of Testing | Testing in compliance with: FDA's "Class II Special Controls Guidance Document:<br>Intervertebral Body Fusion Device" was performed for the JULIETⓇ Ti Cages and<br>demonstrated substantially equivalent performance to the identified predicate<br>devices. The following mechanical tests were performed: Static and dynamic axial<br>compression (per ASTM F2077), Static and dynamic Shear compression (per<br>ASTM F2077), Subsidence (per ASTM F2267)<br>The characterization of the chemical, physical and mechanical properties of the<br>material was performed in accordance with ASTM F3001 and ASTM E8/E8M.<br>The porous structure featured on titanium alloy implants additively<br>manufactured was performed based on the FDA's Guidance for industry on the<br>testing of metallic plasma sprayed coatings on orthopedic implants to support<br>reconsideration of postmarket surveillance requirements. The following tests<br>were performed: Static shear strength (per ASTM F1044), Fatigue shear strength<br>(per ASTM F1160) Static tensile strength (per ASTM F1147) and abrasion<br>resistance (per ASTM F1978). |
| Conclusion | Based on the design features, the use of established well known materials,<br>feature comparisons, indications for use, and results of the mechanical testing,<br>the JULIET Ti has demonstrated substantial equivalence to the identified<br>predicate devices. |
{4}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.