K153620 · Oreon Technologies, Inc. · IYO · May 5, 2016 · Radiology
Device Facts
Record ID
K153620
Device Name
Enovare Ultrasound System
Applicant
Oreon Technologies, Inc.
Product Code
IYO · Radiology
Decision Date
May 5, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Indications for Use
The Enovare system's ultrasound scanner connected with a probe enables real time diagnostic ultrasound imaging for B-mode, Tissue and Contrast Harmonic Imaging, PWD, Power Doppler and combined mode imaging. Signal Analysis and display. Guidance of biopsy and therapy needles, geometrical measurements and calculation of parameters. The system offers optional probes each type optimized to image structure and orientation of tissues during specific clinical applications. These include general Abdominal, Intraoperative, Cardiac, Neurosurgery, Transrectal, Small organs, Urological, Transvaginal, Musculoskeletal.
Device Story
Enovare is a diagnostic ultrasound system consisting of a cart, keyboard, LCD monitor, and various probes. It captures ultrasound signals via piezoelectric transducers; processes these signals to generate B-mode, Tissue/Contrast Harmonic, PWD, Power Doppler, and Color Doppler images. Used in clinical settings (e.g., OR, clinic) by healthcare professionals for real-time imaging, needle guidance, and geometric measurements/calculations. The system supports 3D reconstruction of 2D images. Output is displayed on the integrated monitor to assist clinicians in anatomical visualization and diagnostic decision-making. Benefits include non-invasive tissue assessment and procedural guidance.
Clinical Evidence
Bench testing only. The device was tested for electrical safety (IEC 60601-1 Edition 3.1), biocompatibility (ISO 10993), and acoustic output (Track III requirements). No clinical data was provided.
Technological Characteristics
System includes cart, keyboard, LCD monitor, and piezoelectric transducers. Operates as an ultrasonic pulsed echo imaging system. Supports B-mode, PWD, CFM, Power Doppler, and Tissue Harmonic Imaging. Connectivity includes optional 3D reconstruction unit. Sterilization via Steris. Safety compliance: IEC 60601-1 Edition 3.1 (Class II, Type B/BF).
Indications for Use
Indicated for diagnostic ultrasound imaging or fluid flow analysis of the human body, including general abdominal, intraoperative, cardiac, neurosurgery, transrectal, small organs, urological, transvaginal, and musculoskeletal applications.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 5, 2016
Oreon Technologies Inc. % Mr. Howard Fidel President 520 White Plains Road TARRYTOWN NY 10591
Re: K153620
Trade/Device Name: Enovare Ultrasound System Regulation Number: 21 CFR 892.1560 Regulation Name: Ultrasonic pulsed echo imaging system Regulatory Class: II Product Code: IYO, IYN, IYX Dated: December 18, 2015 Received: February 22, 2016
Dear Mr. Fidel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153620
Device Name Enovare Ultrasound System
#### Indications for Use (Describe)
The Enovare system's ultrasound scanner connected with a probe enables real time diagnostic ultrasound imaging for Bmode, Tissue and Contrast Harmonic Imaging, PWD, Power Doppler and combined mode imaging. Signal Analysis and display. Guidance of biopsy and therapy needles, geometrical measurements and calculation of parameters. The system offers optional probes each type optimized to image structure and orientation of tissues during specific clinical applications. These include general Abdominal, Intraoperative, Cardiac, Neurosurgery, Transrectal, Small organs, Urological, Transvaginal, Musculoskeletal.
Type of Use (Select one or both, as applicable)
| | Prescription Use (Part 21 CFR 801 Subpart D) |
|--|----------------------------------------------|
| | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
System: Enovare
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Modes of Operation | | | | |
|-----------------------------|--------------------|-----|-------------------|---------------|----------|
| Track 3 | B | PWD | Color Doppler + B | Color + PWD+B | Harmonic |
| Abdominal | p | n | n | n | n |
| Intra-operative (abdominal) | n | n | n | n | n |
| Laparoscopic | n | n | n | n | n |
| Trans-rectal | p | n | n | n | n |
| Trans-vaginal (not fetal) | p | n | n | n | n |
| Small Parts | p | n | n | n | n |
n= new indication, p = previously cleared by FDA
### (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH
{4}------------------------------------------------
Transducer: Intraoperative 2001114 System: Enovare Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application Modes of Operation | | | | | |
|-----------------------------------------|---|-----|-------------------|---------------|----------|
| Track 3 | B | PWD | Color Doppler + B | Color + PWD+B | Harmonic |
| Abdominal | | | | | |
| Intra-operative (abdominal) | n | n | n | n | n |
| Laparoscopic | n | n | n | n | n |
| Trans-rectal | | | | | |
| Trans-vaginal (not fetal) | | | | | |
| Small Parts | | | | | |
n= new indication, p = previously cleared by FDA
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH
{5}------------------------------------------------
System: Enovare _____________________________________________________________________________________________________________________________________________________________ follows: follows.
Clinical Applicatio Modes of Operatio
| Clinical Application | Modes of Operation | | | | |
|-----------------------------|--------------------|-----|-------------------|---------------|----------|
| Track 3 | B | PWD | Color Doppler + B | Color + PWD+B | Harmonic |
| Abdominal | | | | | |
| Intra-operative (abdominal) | | | | | |
| Laparoscopic | | | | | |
| Trans-rectal | p | n | n | n | n |
| Trans-vaginal (not fetal) | p | n | n | n | n |
| Small Parts | | | | | |
n= new indication, p = previously cleared by FDA
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH
{6}------------------------------------------------
Transducer: Linear 2001021 System: _Enovare Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows: lonows.
Clinical Applicatio Mode s of Operatio
| Clinical Application | Modes of Operation | | | | |
|-----------------------------|--------------------|-----|-------------------|---------------|----------|
| Track 3 | B | PWD | Color Doppler + B | Color + PWD+B | Harmonic |
| Abdominal | p | n | n | n | n |
| Intra-operative (abdominal) | | | | | |
| Laparoscopic | | | | | |
| Trans-rectal | | | | | |
| Trans-vaginal (not fetal) | | | | | |
| Small Parts | p | n | n | n | n |
n= new indication, p = previously cleared by FDA
### (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH
{7}------------------------------------------------
# 510(k) Summary
### I. SUBMITTER
Oreon Technologies Inc. Suite 500 520 White Plains Road Tarrytown, NY 10591
Phone: 914 413 0087 Email: howard@oreonetch.com
Contact Person: Howard Fidel Date Prepared: October 25, 2015
### II. DEVICE
Name of Device: Enovare Common or Usual Name: Diagnostic Ultrasound System Classification Names: Ultrasonic Pulsed Echo Imaging System (90 IYO CFR 892.1560) Ultrasonic Pulsed Doppler Imaging System (90 IYN, CFR 892.1560) Diagnostic Ultrasound Transducer (90 ITX, CFR 892.1570) Regulatory Class: II
### III. PREDICATE DEVICES
B & K Medical Flex Focus 1202, K123254
3G Ultrasound Sonalis K043189
# IV. DEVICE DESCRIPTION
The Enovare supports the following scanning modes and combinations thereof: B-mode (incl. Tissue Harmonic Imaging), PWD mode, CFM mode, Amplitude (Power) Doppler mode.
The system consists of:
- system cart
- keyboard ●
- ultrasound probe(s) ●
- LCD monitor ●
{8}------------------------------------------------
The system can perform simple geometric measurements and perform calculations in the vascular and urology areas.
# V. INDICATIONS FOR USE
The Enovare system's ultrasound scanner connected with a probe enables real time diagnostic ultrasound imaging for B-mode, Tissue and Contrast Harmonic Imaging, PWD. Power Doppler. Color Doppler and combined mode imaging. Signal Analysis and display. Guidance of biopsy and therapy needles, geometrical measurements and calculation of parameters. An optional 3-D unit can reconstruct a series of 2-D images into a single 3-D volume and display this on the screen.
The system offers optional probes each type optimized to image structure and orientation of tissues during specific clinical applications. These include general Abdominal, Intraoperative, Cardiac, Neurosurgery, Transrectal, Small organs, Urological, Transvaginal, Musculoskeletal.
# VI. COMPARISON OF TECHNICAL CHARACTERISTICS
This device operates identically to the predicate devices in that the piezoelectric material in the transducer used as an ultrasound source to transmit sound waves into the body. Sound waves are reflected back to the transducer and converted to electrical signals that are processed and displayed as images.
# VII SAFETY AND EFFECTIVENESS
The Enovare is compliant to IEC 60601-1 Edition 3.1, International Electrotechnical Commission, Safety of Medical Electrical Equipment, Class II, Type B. Transducer is Type BF.
# VIII BIOCOMATIBILITY
Enovare was tested in accordance with the testing requirements of the ISO 10993 recognized standards and found to be safe for its intended use.
# IX CONCULUSION
The Enovare Ultrasound System is the same device as the predicate Sonalis system with additional added features found in the B&K Medical Flex Focus predicate. Otherwise the Enovare system also has updated components from the Sonalis that do not affect safety and efficacy. The Enovare has the same intended uses as found in the predicate devices and uses the same technological characteristics.
{9}------------------------------------------------
| Indications for use | Enovare | Sonalis<br>K043189 | B & K Medical Flex Focus 1202<br>K123254 | | Thermal, Mechanical,<br>and Electrical Safety | 7. Patient Contact<br>Materials | 8. Cleaning,<br>Disinfection,<br>Sterilization |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | General purpose ultrasound device for obtaining diagnostic medical images. Optional probes are available for general abdominal, small organ, gynecological, intraoperative, peripheral vascular muskulo skeletal and urological anatomy. | General purpose ultrasound device for obtaining diagnostic medical images. Optional probes are available for general abdominal, small organ, gynecological and urological anatomy. | Abdominal, Cardiac, Fetal, Intraoperative, Transurethral, Neurosurgery, Pediatrics, Transrectal, Small Parts, Transvaginal, Peripheral Vascular, Muskulo-skeletal | Enovare | Designed to comply with IEC 60601-1 Edition<br>3.1, and/or prior editions. | The materials of construction of the patient<br>contact portions of probes, components or<br>accessories are identical to those used in<br>similar probes, components or accessories of<br>the Sonalis System and other predicate<br>ultrasound systems. | Cleaning and disinfection are the same as with<br>the predicate Sonalis, and is described on page<br>40 of the user's manual. Sterilization is by<br>Steris, the same as the B&K predicate and is<br>described on page 41 of the user's manual. |
| Scan Modes | B, PWD, CFM and combinations, Tissue Harmonic Imaging | Full Screen 2-D, Four Quadrant B-mode, B-mode, B-mode. | B, M, PWD, CFM and combinations, Tissue Harmonic Imaging | Sonalis<br>K043189 | IEC 60601-1-2 | The materials of construction of the patient<br>contact portions of 5 probes, components or<br>accessories are identical to those used in similar<br>probes, components or accessories of the<br>Sonada predicate ultrasound system. | Cleaning and disinfection is by industry<br>standard methods described the user's manual. |
| Transducers | Peripheral Vascular<br>Endo-Rectal<br>Intraoperative | Peripheral Vascular<br>Endo-Rectal | Peripheral Vascular<br>Endo-Rectal<br>Intraoperative | B & K Medical Flex Focus 1202<br>K123254 | IEC 60601-1-2 | | Sterilization by Steris. |
| Acoustic Output | All transducers are within Track III requirements. | All transducers are within Track I requirements. | 1. All transducers are within Track III requirements. | | | | |
| Biometric<br>Measurements | Same as Sonalis predicate. See page 12 of the system's User's Manual included in Section 4 of this submission. | Distance 6% - 14%<br>Area 10%<br>Ellipse 6%-23% | | | | | |
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
{10}------------------------------------------------
omparison of Oreon Technologies Enovare to the Predicate Device
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.