← Product Code [DXN](/productcode/DXN) · K153600

# Consult Station TS (K153600)

_H4D Sas · DXN · Mar 17, 2017 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153600

## Device Facts

- **Applicant:** H4D Sas
- **Product Code:** [DXN](/productcode/DXN.md)
- **Decision Date:** Mar 17, 2017
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Consult Station TS is intended to be used by the general public to perform a health check, including the measurement of certain vital signs (temperature, weight and BMI, blood pressure and pulse rate). The lay user autonomously self-performs all the measurements listed above. Using an audio-video conferencing feature incorporated in the Consult Station TS to connect with a healthcare professional remotely, it is possible to perform additional examinations: ENT (Ear, Nose and Throat) and skin examinations, auscultation sounds and blood oxygen level. The lay user and the healthcare professional are able to see and to interact with each other. The Consult Station TS is not intended for diagnostic use. Lay users should consult their personal physician if they have concerns regarding the results of their health check. The Consult Station TS is intended for lay users of fourteen (14) years old or older. The Consult Station TS is not intended for children under the age of fourteen (14). The Consult Station TS can be deployed in indoor public areas, such as: pharmacies, senior living communities, shopping centers, corporations, etc.

## Device Story

Consult Station TS is an automated health booth for public use; collects vital signs (weight, BMI, temperature, BP, pulse) via integrated sensors. Operates in two modes: Unsupervised (lay user self-performs measurements) and Supervised (audio-video link to remote physician for additional exams: ENT, skin, auscultation, blood oxygen). Booth contains computer, touch screen, camera, speakers, printer, and medical instruments. Software manages MMI, guides user via video/voice, and facilitates remote connection. Physician guides user to place instruments (otoscope, dermatoscope, stethoscope, pulse oximeter) correctly. Results printed on ticket; data accessible via secure web app. Benefits include remote access to professional health guidance and reliable data collection for physician consultation.

## Clinical Evidence

Bench testing and usability study conducted. Usability study included 48 U.S. participants across 3 groups (lay users >14-21, lay users >21, healthcare professionals) over 96 sessions. Evaluated system usability, functionality, and human factors. Individual components (BP monitor, thermometer, pulse oximeter) validated against standards (ISO 80601-2-30, ASTM E1965-98, ISO 80601-2-56, ISO 80601-2-61) and original manufacturer clinical data.

## Technological Characteristics

Booth structure: resin/glass fiber with gelcoat. Instruments: electronic scale, infrared non-contact thermometer, electronic sphygmomanometer, otoscope, dermatoscope, stethoscope, pulse oximeter. Connectivity: serial/USB interfaces, networked web application. Software: .NET C# (WPF, WCF, ASP.NET MVC), MS SQL Server 2012, IIS 7.5. Hardware: Intel Atom N455/D525 processor, Windows 7. Sterilization: non-sterile, cleaning/disinfection instructions provided.

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- CSI Model 9K Managed Health System ([K101198](/device/K101198.md))
- Omron Hem-705CP Automated Monitor System ([K903134](/device/K903134.md))
- Infrared Thermometer (Model: LX-26E, LX-260TE, PRO LX-261E, LX-360, LX-361T, BW-CX10) ([K160816](/device/K160816.md))

## Reference Devices

- Pulse oximeter ([K140785](/device/K140785.md))
- Stethoscope ([K102893](/device/K102893.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a human figure with three faces in profile.

March 17, 2017

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

H4D SAS % Esin Yesilalan Senior Regulatory Scientist Voisin Consulting Life Sciences 222 Third Street Suite 3121 Cambridge, Massachusetts 02142

Re: K153600

Trade/Device Name: Consult Station TS Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II Product Code: DXN, FLL, DQA Dated: February 27, 2017 Received: February 28, 2017

Dear Esin Yesilalan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply 

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Muda Jellun
for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K153600

Device Name Consult Station TS

### Indications for Use (Describe)

The Consult Station TS is intended to be used by the general public to perform a health check including the measurement of certain vital signs (weight, BMI, temperature, blood pressure and pulse rate). The lay user autonomously self-performs all the measurements listed above.

Using an audio-video conferencing feature incorporated in the Consult Station TS to connect with a healthcare professional remotely. it is possible to perform additional examinations: ENT (Ear. Nose. Throat) and skin examinations. auscultations sounds and blood oxygen level. The lay user and the healthcare professional are able to see and to interact each other.

The Consult Station TS is not intended for diagnostic use. Lay users should consult their personal physician if they have concerns regarding the results of their health check.

The Consult Station TS is intended for lay users of fourteen (14) years old or older. The Consult Station TS is not intended for children under the age of fourteen.

The Consult Station can be deployed in indoor public areas, such as: pharmacies, senior living communities, shopping centers, corporations, etc.

Type of Use (Select one or both, as applicable)

| <div> <input checked="true" type="checkbox"/>     Prescription Use (Part 21 CFR 801 Subpart D)    </div> | <div> <input checked="true" type="checkbox"/>     Over-The-Counter Use (21 CFR 801 Subpart C)    </div> |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDEDThis section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

EF

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for H4D, which stands for Health for Development. The logo features the letters H4D in a bold, sans-serif font, with the "H" and "4" in blue and the "D" in gray. To the left of the letters is a blue graphic of a grid of squares. To the right of the letters are the words "HEALTH FOR DEVELOPMENT" in a smaller, sans-serif font, with "HEALTH" in gray and "FOR DEVELOPMENT" in blue.

# SECTION 5 510 (k) SUMMARY

## CONTENTS

| <b>5</b> | <b>510(K) SUMMARY</b>                                     |  | <b>5-2</b> |
|----------|-----------------------------------------------------------|--|------------|
| 5.1      | Submitter Information                                     |  | 5-2        |
| 5.2      | Device Identification                                     |  | 5-2        |
| 5.3      | Identification of Predicate Devices                       |  | 5-2        |
| 5.4      | Device Description                                        |  | 5-3        |
| 5.5      | Indications for Use                                       |  | 5-4        |
| 5.6      | Comparison to Predicate Devices                           |  | 5-5        |
|          | 5.6.1 Summary of Technological Characteristics            |  | 5-6        |
|          | 5.6.2 Electromagnetic Compatibility and Electrical Safety |  | 5-7        |
|          | 5.6.3 Performance and Clinical Testing                    |  | 5-8        |
|          | 5.6.4 Clinical Studies                                    |  | 5-9        |
|          | 5.6.5 Biocompatibility Testing                            |  | 5-10       |
|          | 5.6.6 Comparison with the Predicates and Conclusion       |  | 5-10       |
| 5.7      | Device Comparison Table                                   |  | 5-10       |
| 5.8      | Conclusion                                                |  | 5-22       |

### List of Tables

| Table 5-1 Table Predicate Devices for Consult Station TS®                                                                             | 5-5  |
|---------------------------------------------------------------------------------------------------------------------------------------|------|
| Table 5-2 Table Predicate Devices for the blood pressure module.                                                                      | 5-5  |
| Table 5-3 Table Predicate Devices for the thermometer                                                                                 | 5-6  |
| Table 5-4 Device Comparison Table                                                                                                     | 5-11 |
| Table 5-5 Comparison of the Blood Pressure Module with Omron Hem-705CP                                                                | 5-19 |
| Table 5-6 Comparison of the Thermometer IR with Infrared Thermometer (Model: LX-26E, LX-260TE, PRO LX-261E, LX-360, LX-361T, BW-CX10) | 5-21 |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for H4D Health for Development. The logo features a blue, stylized globe made of small squares on the left. To the right of the globe is the text "H4D" in a bold, blue font, followed by a vertical line. Next to the line is the text "HEALTH FOR DEVELOPMENT" in a smaller, gray font. 

### 5 510(K) SUMMARY

#### 5.1 Submitter Information

| Company Name:    | H4D                                   |
|------------------|---------------------------------------|
| Company Address: | 92 avenue Kleber, 75016 Paris, FRANCE |
| Company Phone:   | +33 (0) 1 46 24 20 64                 |
| Company Fax:     | 33 (0) 9 58 29 67 66                  |
| Contact Person:  | Marta Rivas                           |
| Date of Summary: | February 24, 2017                     |

#### 5.2 Device Identification

| Trade Name:          | Consult Station TS ®                           |
|----------------------|------------------------------------------------|
| Common Name:         | Health booth                                   |
| Classification Name: | Non-invasive blood pressure measurement system |
| Product Code:        | DXN, FLL, DQA                                  |
| Regulation Number:   | 870.1130                                       |
| Device Class:        | Class II                                       |

#### 5.3 Identification of Predicate Devices

H4D's Consult Station TS® is substantially equivalent to the following predicate device:

- CSI Model 9K Managed Health System, manufactured by Computerized Screening, Inc. and cleared for commercial distribution under 510(k) K101198;
Among the FDA Class II device components of the Consult Station, the non-invasive blood pressure module and the thermometer carry a CE approval in the EU; however, they are not cleared in the US. Nevertheless, this blood pressure module is substantially equivalent to Omron Hem-705CP Automated Monitor System (510(k) K903134) and the temperature module is equivalent to Infrared Thermometer (Model: LX-26E, LX-260TE, PRO LX-261E, LX-361T, BW-CX10) (510k number K160816).

- . Omron Hem-705CP Automated Monitor System 510(k) K903134;
- Infrared Thermometer (Model: LX-26E, LX-260TE, PRO LX-261E, LX-360, LX-361T, BW-. CX10) (510k number K160816).

{5}------------------------------------------------

#### 5.4 Device Description

The Consult Station TS is an automated non-invasive screening device intended to be used by the general public (lay users). Equipped with sensors and instruments, the Consult Station TS presents itself in the shape of a health booth and allows to perform health check-ups in order to collect reliable medical data.

The Consult Station TS is not a diagnostic device and only provides data in order for lay users to consult with their personal physician. Lay users should consult with their physician if they have concerns regarding the results of their health check-up.

Two scenarios for the use of the Consult Station TS are available:

- A scenario where the lay user will autonomously self-perform the measurements of his/her vital 1. signs ("Unsupervised Mode");
This use does not require assistance of a healthcare professional. The lay user autonomously self-performs a health check allowing him/her to collect a number of his/her vital signs: weight and BMI, systolic and diastolic blood pressure, pulse rate and temperature. An user interface guides the lay user with video tutorials, a series of interactive screens and voice instructions through the various steps of the health check cycle.

- 2. An alternative scenario where, at first, the lay user will autonomously self-perform measurements of his/her vital signs ("Unsupervised Mode"), and then be remotely connected with a healthcare professional through the audio-video conferencing feature to perform additional examinations ("Supervised Mode").
This use that requires assistance of a healthcare professional to ensure that data is appropriately collected and interpreted. After performing a health check (see scenario above), the lay user is connected through the audio-video conferencing feature with a healthcare professional, the lay user and the healthcare professional being able to see and to interact with each other.

The healthcare professional can then remotely guide the lay user to use as needed:

- · an otoscope for ENT examinations (inspection of the external ear canal),
- · a dermatoscope for skin surface examinations,
- · a stethoscope for auscultation of sounds associated with the heart and lungs,
- · a pulse oximeter for the measurement of the blood oxygen level.

The otoscope (ENT examination), the dermatoscope (skin examination), the stethoscope (auscultations sounds) and the pulse oximeter (blood oxygen level) are only used in the Supervised Mode. The use of the pulse oximeter, dermatoscope, otoscope and the stethoscope are restricted on the order of a licensed practitioner.

In both scenarios, at the end of cycle, a ticket is printed inside the booth which summarizes the results of the measurement as well as user access codes for subsequent online access.

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the logo for H4D, which stands for Health for Development. The logo features the letters "H4D" in a bold, sans-serif font, with the "H" and "4" in a darker shade of blue and the "D" in a lighter shade. To the left of the letters is a graphic of blue squares arranged in a grid-like pattern. To the right of the letters is the text "HEALTH FOR DEVELOPMENT" in a smaller, sans-serif font.

Lay users may manually enter blood glucose level measurements. The Consult Station TS may accept loaded data from an external blood glucometer (if such device is FDA cleared).

Consult Station is intended for users of 14 years old or older. The Consult Station TS is not intended for children under the age of fourteen (14).

Consult Station TS system consists of:

- A health booth,
- A software embedded in the health booth,
- . A Consult Access software containing an audio-video conferencing feature,
- . A web application which provides secure access by patients and health professionals to their anonymized data.

The health booth is an enclosed space, accessible through a door and containing a seat, sensors and instruments as well as communication tools (i.e. screens, camera, speakers, printer).

The software embedded in the health booth is responsible for the management of the man-machine interface (MMI) used to provide instructions to the patient, to collect patient data and to print them via the printer on a ticket. The software is also responsible for starting, stopping and repeating measurements. These functionalities are implemented by communicating via the serial or USB interfaces of the connected devices involved (sending commands followed by receiving of results).

The Consult Access is a software that allows through an audio-video conferencing connecting a healthcare professional with a patient seated inside the Consult Station. The patient and the healthcare professional will be able to see and to interact with each other. The healthcare professional can remotely guide the patient throughout all the measurements and examinations performed. The otoscope, the dermatoscope and the oximeter are only available while the video conferencing feature is in use.

The web application provides secure access by patients and healthcare professionals (only with the patient's agreement) to the measurements performed via the Consult Station TS.

#### ર.5 Indications for Use

The Consult Station TS is intended to be used by the general public to perform a health check, including the measurement of certain vital signs (temperature, weight and BMI, blood pressure and pulse rate). The lay user autonomously self-performs all the measurements listed above.

Using an audio-video conferencing feature incorporated in the Consult Station TS to connect with a healthcare professional remotely, it is possible to perform additional examinations: ENT (Ear, Nose and Throat) and skin examinations, auscultation sounds and blood oxygen level. The lay user and the healthcare professional are able to see and to interact with each other.

{7}------------------------------------------------

The Consult Station TS is not intended for diagnostic use. Lay users should consult their personal physician if they have concerns regarding the results of their health check.

The Consult Station TS is intended for lay users of fourteen (14) years old or older. The Consult Station TS is not intended for children under the age of fourteen (14).

The Consult Station TS can be deployed in indoor public areas, such as: pharmacies, senior living communities, shopping centers, corporations, etc.

#### 5.6 Comparison to Predicate Devices

H4D is claiming substantial equivalence for Consult Station TS with the CSI Model 9K Managed Health System, which is currently commercially available in the U.S. Relevant information on the predicate devices is shown in Table 5-1.

| Table 5-1 | Table Predicate Devices for Consult Station TS® |
|-----------|-------------------------------------------------|
|-----------|-------------------------------------------------|

| Device Name      | CSI Model 9K Managed Health System |
|------------------|------------------------------------|
| Manufacturer     | Computerized Screening, Inc.       |
| 510(k) Number    | K101198                            |
| Regulatory Class | Class II                           |
| Common Name      | Automated Blood Pressure Monitor   |
| Clearance Date   | 04 Aug 2010                        |

H4D is claiming substantial equivalence for the blood pressure module with the Omron Hem-705CP, which is currently commercially available in the U.S. Relevant information on the predicate devices is shown in Table 5-1

| Table Predicate Devices for the blood pressure module<br>Table 5-2 |
|--------------------------------------------------------------------|
|--------------------------------------------------------------------|

| Device Name      | Omron Auto Oscillometric DIG BP<br>MONT/HEM-704C 705C |
|------------------|-------------------------------------------------------|
| Manufacturer     | Omron Marshall Products, INC.                         |
| 510(k) Number    | K903134                                               |
| Regulatory Class | Class II                                              |
| Common Name      | Noninvasive blood pressure<br>measurement system      |
| Clearance Date   | 01/08/1991                                            |

H4D is claiming substantial equivalence for the temperature with the Infrared thermometer (Model: LX-26E, LX-260TE, PRO LX-261E, LX-360, LX-361T, BW-CX10) which are currently commercially available in the U.S. Relevant information on the predicate devices is shown in Table 5-3.

{8}------------------------------------------------

| Device Name      | Infrared Thermometer (Model: LX-26E,<br>LX-260TE, PRO LX-261E, LX-360,<br>LX-361T, BW-CX10) |
|------------------|---------------------------------------------------------------------------------------------|
| Manufacturer     | Nexmed Technology Co., Ltd                                                                  |
| 510(k) Number    | K160816                                                                                     |
| Regulatory Class | Class II                                                                                    |
| Regulation Name  | Clinical Electronic Thermometer                                                             |
| Clearance Date   | 09/21/2016                                                                                  |

#### Table 5-3 Table Predicate Devices for the thermometer

#### 5.6.1 Summary of Technological Characteristics

The physiological parameters measurements available in Unsupervised Mode (the scenario where the lay user autonomously self-performs all measurements of his/her vital signs) are :

- The weight and BMI calculation,
- The temperature,
- . The blood pressure and pulse rate.

The associated instruments for physiological parameters measurement consist of the following:

- . An electronic scale,
- . An infrared non-contact forehead thermometer,
- . An electronic sphygmomanometer.

The exams available in Supervised Mode (the scenario where the lay user remotely connected with a healthcare professional through the audio-video conferencing feature to perform additional examinations) are:

- . ENT examinations (inspection of the external ear canal) using an otoscope,
- Skin surface examinations using a dermatoscope,
- . Auscultation of sounds associated with the heart and lungs using a stethoscope,
- Blood oxygen level measurement using a pulse oximeter.

The Hardware computer components consist of the following:

- . Height measuring device using ultrasonic;
- . A touch screen which allows to enter data or to answer questions and the start and stop the measurements :
- . An embedded computer with interface connections for sensors and instruments;
- . Environmental sensors (temperature, humidity percentage) ;
- A printer ensuring the editing of a ticket which summarizes the measurement results and the ● patient identifiers.

The Remote Monitoring components consist of the following:

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image contains a logo for an organization called H4D, which stands for Health for Development. The logo consists of two parts: a graphic element on the left and the text "H4D | HEALTH FOR DEVELOPMENT" on the right. The graphic element is a stylized representation of interconnected squares, forming a spherical shape. The text is arranged in two lines, with "H4D" in a larger font size and "HEALTH FOR DEVELOPMENT" in a smaller font size.

- A camera and loudspeakers ;
- A touch screen used as input device for the lay user. Measurements can be started and stopped bylay users using the controls displayed on the touch screen;
- . A LCD 19" screen for man-machine interface. The LCD screen is used for displaying interactive screens, videos instructions and measurement results. The functionality to gain remote access to a qualified physician is available via a LCD screen. Once measurements are done (note: the proprietary software of the instruments take over the Consult Station TS during that time) the results are displayed on a LCD screen.

### The Software solution consists of the following:

- A software embedded in the Consult Station for starting, stopping and repeating measurements, management of the MMI used to give instructions to the patient, to collect patient data and to print them on the printer;
- . An audio-video conference feature through which users are remotely connected with a healthcare professional:
- . A website, on which users and physicians can consult the medical data collected, through the use of a secure user ID and password (found on the ticket printed inside the booth).

Software requires the following technologies/environments:

- .NET C# Language (Visual Studio 2013, framework 4.5) with WPF, WCF, Entity framework and ASP.NET MVC 2
- . MS SOL Server 2012 database server
- . IIS 7.5 web server"

Hardware of the Consult Station TS includes a computer with Microsoft Windows 7 Operating system with a Single Board Industrial Computer PCM-9363 with Embedded Intel Atom N455 1.66 GHz/D525 1.8 GHz Processor.

#### 5.6.2 Electromagnetic Compatibility and Electrical Safety

The Consult Station TS has been subject to electromagnetic compatibility and electrical safety and essential performance testing and is compliant with the standards as listed below:

- IEC 60601-1 :2005 + A1 :2012 and AAMI ANSI ES60601-1:2005 and A1:2012 and A2:2010 Electrical equipment - Part 1: General requirements for basic safety and essential performance;
- IEC 60601-1-11 Edition 1.0 2010-04 Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in homecare environment;

{10}------------------------------------------------

Image /page/10/Picture/1 description: The image contains a logo for "H4D | HEALTH FOR DEVELOPMENT". The logo consists of the letters "H4D" in a bold, sans-serif font, with the "H" and "4" in a darker shade of blue and the "D" in a lighter shade. To the right of "H4D" is the text "HEALTH FOR DEVELOPMENT", with "HEALTH" in a smaller font size and "FOR DEVELOPMENT" in a larger font size. The text is also in a blue color, but slightly lighter than the "H4D" letters. To the left of the letters is a graphic of blue squares.

- Consult Station TS meets requirements of compliance with the standard IEC 60601-1-2 Edition ● 3: 2007-03; Electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests and by equivalence with the standard IEC 60601-1-2 Edition 3: 2007
- . Acceptance tests for the home environment and US deviations have been completed according to ANSVAAMI/IEC 60601-1-:2014 Electrical Safety Medical electrical equipment Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetidisturbances -- Requirements and test.

#### Performance and Clinical Testing 5.6.3

The Consult Station TS has been subject to bench and clinical evaluation and meets requirements for the basic safety and essential performance of all its components as listed below:

### Automated non-invasive sphygmomanometers

- . The automated sphygmomanometer has been subject to bench tests and meets requirements of compliancewith the standard ANSI AAMI ISO 80601-2-30:2009 & A1:2013, Particular Requirements for Safety, including essential performance, of automatic cycling non-invasive blood pressure measuring equipment;
The automated sphygmomanometer has been also subject to clinical tests by their original manufacturer before their integration in the Consult Station TS and meets requirements for Clinical and Performance testing according to EN 1060-4: 2004 Non-invasive sphygmomanometers - Part 4 (Procedures for determining the accuracy of the entire automated non-invasive blood pressure system was completed by the supplier of the blood pressure module). The study was completed according to EN ISO 1060-4, as per procedure "Sequential blood pressure measurement" (N3). The methodology, criteria and results of the test meet the compliance with the standard ANSVAAMI/ISO 81060-2 as per procedure "Same arm sequential method".

### Infrared thermometer

- . The IR thermometer has been tested on bench and is compliant with the standard ASTM E1965-98 (Reapproved 2009) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature:
- The IR thermometer has been subject to clinical investigation and is compliant with the standard ISO 80601-2-56 First Edition 2009-10-01 Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement.

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for H4D, which stands for Health for Development. The logo features a stylized globe made of blue squares on the left side. To the right of the globe, the text "H4D" is written in a bold, sans-serif font, with a vertical line separating it from the words "HEALTH FOR DEVELOPMENT" written in a smaller, sans-serif font. The overall design is clean and modern, suggesting a focus on global health initiatives.

### Pulse oximeter

- The pulse oximeter has been subject to clinical tests by their original manufacturer before their . integration in the Consult Station TS and meets requirements for Clinical testing according to ISO 80601-2-61 First edition 2011-04-01 Medical electrical equipment - Part 2-61 (Particular requirements for basic safety and essential performance of pulse oximeter equipment was completed according to its original 510(k) (K140785)
### Software

Software validation has been satisfactorily completed according to "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" (May 11, 2005) for a device of Moderate level of concern. Software Requirements Specification, Architecture Design Chart, Software Design Specifications, Traceability Analysis, Software Development Environment Description, Verification and Validation and Revision Level History were completed in accordance with relevant FDA guidance documents and with the international standard IEC 62304: 2006 Medical device software - Lifecycle software process.

#### 5.6.4 Clinical Studies

An Usability Study, containing 48 U.S. participants in the human factors testing, was conducted by a third part Connected Healthcare Solutions. The H4D Human Factors validation protocol considered and tested various Use Cases, Risks, Known Use Errors and Expert Analysis of hazards and risks associated with use of the devices on forty-eight (48) distinct test subjects in three (3) distinct User groups in a total of ninety-six (96) Unsupervised sessions over a period of approximately two (2) weeks of live, observed and documented testing.

The study poplation was allocated in 3 user groups:

- . Lay user patients: Male and female ages > 14 - 21 years of age o Patients under the age of 18 had parental consent.
- . Lay user patients: Male and female ages 21 years of age and older
- Health Care Professionals ●

The primary study objective was to evaluate the usability of the Consult Station TS:

- System usability by patients was assessed using a quantitative usability instrument and semi-● structured one-on-one interviews.
- System functionality was assessed through semi-structured one-on-one interviews and ● Health Care Provider's feedback in terms of successful vital signs collections.
- . Patient compliance using the system was assessed through number of prescribed actions versus actual actions.

The secondary study objective was to evaluate the human factors utility of the Consult Station TS:

- . System human factors was assessed using a qualitative, subjective instrument and one-onone interviews.

{12}------------------------------------------------

#### 5.6.5 Biocompatibility Testing

The Consult Station TS has been evaluated in accordance with international and FDA recognized standard, ISO 10993-1 as well as FDA's 510(k) Memorandum G95-1. The Consult Station TS medical components are considered surface devices with skin contact and limited duration. Accordingly, the following biocompatibility tests have been conducted:

- . Cytotoxicity according to ISO 10993-5
- . Sensitization according to ISO 10993-10
- Skin Irritation according to ISO 10993-10

All tests were performed on sterilized finished product in accordance with Good Laboratory Practices (GLP) with no deviations. These regulations set forth the minimum basic requirements for study conduct, personnel, facilities, equipment, written protocols, operating procedures, study reports, and a system of quality assurance oversight for each study to help assure the safety of FDA regulated products.

#### 5.6.6 Comparison with the Predicates and Conclusion

This paragraph describes the substantial equivalence between the Consult Station TS and the predicates CSI Model 9K Managed Health System (K101198).

Among the FDA Class II device components of the Consult Station, the non-invasive blood pressure module is substantially equivalent to Omron Hem-705CP Automated Monitor System (510(k) K903134). The table 5-5 compares the module with Omron Hem-705CP.

Similarly, thermometer component of the Consult Station TS is manufactured by H4D is substantially equivalent to Infrared Thermometer (Model: LX-260TE, PRO LX-261E, LX-360, LX-361T, BW-CX10) (510k number K160816). The table 5.6 compares the thermometer module with Infrared Thermometer (Model: LX-26E, LX-260TE, PRO LX-261E, LX-360, LX-361T, BW-CX10) (510k number K160816).

#### 5.7 Device Comparison Table

Table 5-4 compares the similarities and differences between Consult Station TS device and the predicate CSI Model 9K Managed Health System (K101198).

{13}------------------------------------------------

| Table 5-4                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Device Comparison Table                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |            |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Consult Station TS®<br>(subject 510(k))                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | CSI Model 9K<br>Managed Health<br>System (K101198)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |            |
| FDA Product<br>Code                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | DXN, FLL, DQA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | DXN                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same (unchanged) CSI Model<br>9K                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |            |
| Regulation<br>No.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 21 CFR 870.1130<br>(Non-invasive blood pressure<br>measurement system)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | 21 CFR 870.1130<br>(Non-invasive blood<br>pressure measurement<br>system)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Same regulation number<br>(unchanged) CSI Model 9K                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |            |
| Device Class                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Same device class (unchanged)<br>CSI Model 9K                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |            |
| Intended Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Consult Station TS is an automated<br>non-invasive screening device<br>intended to be used by the general<br>public.<br>It is not a diagnostic device and<br>only furnishes data so users can<br>consult their personal physician.<br>User should consult their personal<br>physician if they have concerns<br>regarding their measurement.<br>The Consult Station TS covers two<br>primary functions:<br>• It allows a lay user to<br>autonomously self-perform the<br>measurement of his/her vital<br>signs (temperature, weight and<br>BMI calculation, blood pressure,<br>pulse rate).<br>• It offers a lay user the possibility<br>to connect via an audio-video<br>conferencing feature with a<br>healthcare professional remotely<br>connected, the lay user and the<br>healthcare professional being<br>able to see and to interact with<br>each other. The healthcare<br>professional will then be able to<br>conduct additional examinations:<br>ENT (Ear, Nose and Throat) and<br>skin examinations, auscultation<br>of heart and lungs sounds and<br>measurement of blood oxygen<br>level and guide the lay user to<br>use the instruments for his/her<br>examination. The healthcare<br>professional will also be able to<br>analyze the readings from these                           | Unsupervised means for<br>measuring and tracking<br>an individual's blood<br>pressure, pulse rate, and<br>if optionally equipped,<br>weight and temperature.<br>Users may manually<br>enter or download blood<br>glucose level.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Unsupervised Mode:<br>The Consult Station TS has the<br>same intended use (i.e. furnishes<br>medical data so lay users can<br>consult their personal physician)<br>and follows the same principle<br>of operation (i.e. self-operation).<br>performing health check-ups via<br>an automated non-invasive<br>device in the form of a health<br>booth with panels, windows,<br>door and seat) as the predicates.<br>Both the Consult Station TS and<br>CSI Model 9K measure systolic<br>and diastolic blood pressure,<br>pulse rate and (if optionally equipped for CSI Model 9K)<br>weight and temperature.<br>Lay<br>users may also manually enter<br>or download blood glucose level<br>with both devices.<br>Supervised Mode:<br>When using the audio-video conferencing feature, the<br>Consult Station TS integrates<br>additional instruments (pulse oximeter, stethoscope, otoscope and dermatoscope) to perform<br>specific examinations (ENT and<br>skin examinations, auscultation sounds and blood oxygen level)   |            |
| Consult Station TS®<br>(subject 510(k))                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | CSI Model 9K<br>Managed Health<br>System (K101198)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |            |
| examinations.<br>Lay users may also download blood<br>glucose level provided by an<br>external glucometer (not integrated<br>in the booth, (if such device is FDA<br>cleared).). Upon completion, data<br>may be stored and accessed by the<br>user via a website.<br>Consult Station is intended for<br>users fourteen years or older. The<br>Consult Station TS is not intended<br>for children under the age of<br>fourteen.<br>The use of the pulse oximeter,<br>dermatoscope, otoscope and the<br>stethoscope are restricted on the<br>order of a licensed practitioner.<br>Users may manually enter or<br>download blood glucose level. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | predicates. Use of these<br>supplementary functionalities<br>requires guidance of a physician<br>via audio-conferencing to ensure<br>that medical devices are<br>properly placed on the lay user<br>and data is appropriately<br>collected for interpretation by<br>the physician.<br>The selected pulse oximeter<br>(K140785) and stethoscope<br>(K102893) are FDA cleared for<br>"remote use by a physician<br>using a computer" as they are<br>intended to be used when<br>integrated into the Consult<br>Station TS.<br>Addition of the oximeter does<br>not affect the safety and<br>effectiveness of the Consult<br>Station TS.<br>The stethoscope is FDA cleared,<br>510(k), K102893. The otoscope/<br>dermatoscope are MD class I,<br>FDA exempt. The addition of<br>the stethoscope electronic<br>otoscope/dermatoscope does not<br>raise new questions in terms of<br>safety /effectiveness of the<br>Consult Station TS.<br>Consult Station TS has<br>additional (non-MD regulated)<br>functionality of remote<br>consultation with a professional<br>health.<br>The data in the Consult Station<br>is HIPAA compliant; therefore<br>this function does not raise new<br>questions in terms of<br>safety/effectiveness in<br>comparison to predicates. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |            |
|                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Consult Station TS®<br>(subject 510(k))                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | CSI Model 9K<br>Managed Health<br>System (K101198)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Comparison                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |            |
| Indications<br>for Use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | The Consult Station TS is intended<br>to be used by the general public to<br>perform a health check, including<br>the measurement of certain vital<br>signs (temperature, weight and<br>BMI, blood pressure and pulse<br>rate). The lay user autonomously<br>self-performs all the measurements<br>listed above.<br>Using an audio-video conferencing<br>feature incorporated in the Consult<br>Station TS to connect with a<br>healthcare professional remotely, it<br>is possible to perform additional<br>examinations: ENT (Ear, Nose and<br>Throat) and skin examinations,<br>auscultation sounds and blood<br>oxygen level. The lay user and the<br>healthcare professional are able to<br>see and to interact with each other.<br>The Consult Station TS is not<br>intended for diagnostic use. Lay<br>users should consult their personal<br>physician if they have concerns<br>regarding the results of their health<br>check.<br>The Consult Station TS is intended<br>for lay users of fourteen (14) years<br>old or older. The Consult Station<br>TS is not intended for children<br>under the age of fourteen (14).<br>The Consult Station TS can be<br>deployed in indoor public areas,<br>such as: pharmacies, senior living<br>communities, shopping centers,<br>corporations, etc. | CSI Model 9K Managed<br>Health System Kiosk is<br>an automated non-<br>invasive screening device<br>intended for voluntary<br>use by the general public.<br>The device measures<br>systolic and diastolic<br>blood pressure, pulse and<br>if optionally equipped,<br>weight and temperature.<br>Users may also manually<br>enter or download blood<br>glucose level. Data from<br>the Model 9K may be<br>stored by the user for<br>tracking purposes.<br>It is not a diagnostic<br>device, and only<br>furnishes data so users<br>can consult their personal<br>physicians.<br>The device is intended<br>for users fourteen years<br>or older. Children under<br>the age of fourteen must<br>be accompanied by an<br>adult.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Similar general intended use as<br>CSI Model 9K.<br>The addition of new exams in<br>audio-video conferencing<br>feature: inspection of the<br>external ear canal, body surface<br>examination and auscultation of<br>sounds associated with internal<br>organs and blood oxygen level<br>measurement does not change<br>the general intended use of the<br>Consult Station TS compared<br>with the CSI Model 9K.…

---

**Source:** [https://fda-staging.innolitics.com/device/K153600](https://fda-staging.innolitics.com/device/K153600)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com