ebb Complete Tamponade System

K153591 · Clinical Innovations, LLC · OQY · Jan 14, 2016 · Obstetrics/Gynecology

Device Facts

Record IDK153591
Device Nameebb Complete Tamponade System
ApplicantClinical Innovations, LLC
Product CodeOQY · Obstetrics/Gynecology
Decision DateJan 14, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 884.4530
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ebb Complete Tamponade System is indicated for use in providing temporary control or reduction of postpartum uterine bleeding. Inflation of the Vaginal Balloon anchors the Uterine Balloon and provides vaginal tamponade if vaginal bleeding is present. The ebb Complete Tamponade System should only be used in the setting of post-partum uterine bleeding when conservative management is warranted.

Device Story

Disposable, multiple-lumen catheter with two inflatable balloons; uterine balloon (750 mL capacity) for intrauterine tamponade; vaginal balloon (350 mL capacity) for anchoring and vaginal tamponade. Operated manually by clinicians in clinical settings; inflation via isotonic IV fluid (saline/Ringers Lactate) using spike or Luer assembly. Includes drainage and irrigation lumens. Device retained for up to 24 hours. Modification adds UV bands to balloon ends. Provides mechanical pressure to control hemorrhage; benefits patient by reducing blood loss during postpartum complications.

Clinical Evidence

Bench testing only. Device subjected to cyclic physiologic pressures under conditions exceeding those in published literature; no failures reported. No clinical data provided.

Technological Characteristics

Disposable, multiple-lumen catheter; uterine balloon (750 mL) and vaginal balloon (350 mL); spike/Luer fill assemblies; drainage/irrigation lumens. Sterilized via Ethylene Oxide (SAL 10^-6). 3-year shelf life. UV bands added to balloon ends.

Indications for Use

Indicated for patients experiencing postpartum uterine bleeding where conservative management is warranted; provides temporary control or reduction of bleeding via uterine and vaginal tamponade.

Regulatory Classification

Identification

An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.

Special Controls

*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 14, 2016 Clinical Innovations, LLC Tom Haueter Director, QA/RA 747 West 4170 South Murray, UT 84123 Re: K153591 Trade/Device Name: ebb Complete Tamponade System Regulation Number: 21 CFR 884.4530 Regulation Name: Obstetric-Gynecologic Specialized Manual Instrument Regulatory Class: Class II Product Code: OQY Dated: December 14, 2015 Received: December 16, 2015 Dear Tom Haueter, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply {1}------------------------------------------------ with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Herbert P. Lerner -S Benjamin R. Fisher, Ph.D. for Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K153591 Device Name ebb Complete Tamponade System #### Indications for Use (Describe) The ebb Complete Tamponade System is indicated for use in providing temporary control or reduction of postpartum uterine bleeding. Inflation of the Vaginal Balloon anchors the Uterine Balloon and provides vaginal tamponade if vaginal bleeding is present. The ebb Complete Tamponade System should only be used in the setting of post-partum uterine bleeding when conservative management is warranted. Type of Use (Select one or both, as applicable) | <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | |--------------------------------------------------------------------------------------| | <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Clinical Innovations. The logo consists of a circular design with four curved segments in shades of blue and green. To the right of the logo is the text "CLINICAL INNOVATIONS" in gray and green. Below the company name is the tagline "forMOM. forBABY. forLIFE." ### 510(k) SUMMARY ## Clinical Innovations' ebb Complete Tamponade System Clinical Innovations, LLC 747 W. 4170 S. Murray, UT 84123 Phone: 801-260-6100 Fax: 801-266-7373 Contact Person: Tom Haueter Date Prepared: December 14, 2015 #### Name of Device and Name/Address of Sponsor ebb Complete Tamponade System Clinical Innovations, LLC 747 W. 4170 S. Murray, UT 84123 Common or Usual Name: Intrauterine tamponade balloon Classification Name: Class II per 21 CFR 884.4530 Product Code: OQY Predicate Device: K150573 ebb Complete Tamponade System #### Purpose of the Special 510(k) notice: The subject ebb Complete Tamponade System is a modification to the previous ebb Complete Tamponade System (K150573). This modification is the addition of a UV Band around the proximal and distal ends of the vaginal and uterine balloons. {4}------------------------------------------------ #### Device Description: The ebb Complete Tamponade System is a disposable, multiple lumen catheter attached to an inflatable balloon system designed to provide tamponade for controlling hemorrhage from the uterus and vagina. The device consists of two inflatable balloons: The upper uterine balloon is inflated inside the uterus and the lower vaginal balloon is inflated inside the vagina. Inflation is accomplished with isotonic intravenous fluid such as normal saline or Ringers Lactate. The uterine balloon catheter has separate lumens to enable inflation/deflation, irrigation and drainage. The vaginal balloon catheter has a lumen to enable inflation/deflation. The uterine and vaginal balloons are permanently assembled and are not to be separated. The device may be retained in position for up to 24 hours in the post-operative mode of treatment. The ebb Complete Tamponade System is supplied sterile in peel open pouches for one time use to a single patient. #### Indications for Use: The ebb Complete Tamponade System is indicated for use in providing temporary control or reduction of postpartum uterine bleeding. Inflation of the Vaginal Balloon anchors the Uterine Balloon and provides vaginal tamponade if vaginal bleeding is present. The ebb Complete Tamponade System should only be used in the setting of post-partum uterine bleeding when conservative management is warranted. | | Ebb Complete Tamponade System<br>Device | Ebb Complete Tamponade System<br>(K150573) | Comparison | |--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------| | Intended Use<br>/Indications For Use | The ebb Complete Tamponade<br>System is indicated for use in<br>providing temporary control or<br>reduction of postpartum uterine<br>bleeding. Inflation of the Vaginal<br>Balloon anchors the Uterine Balloon<br>and provides vaginal tamponade if<br>vaginal bleeding is present. The ebb<br>Complete Tamponade System<br>should only be used in the setting of<br>post-partum uterine bleeding when<br>conservative management is<br>warranted. | The ebb Complete Tamponade<br>System is indicated for use in<br>providing temporary control or<br>reduction of postpartum uterine<br>bleeding. Inflation of the Vaginal<br>Balloon anchors the Uterine Balloon<br>and provides vaginal tamponade if<br>vaginal bleeding is present. The ebb<br>Complete Tamponade System<br>should only be used in the setting of<br>post-partum uterine bleeding when<br>conservative management is<br>warranted. | Same | | Technology | Tamponade achieved by inflation of<br>balloon with saline solution | Tamponade achieved by inflation of<br>balloon with saline solution | Same | | Operating Principle | Manual Operation | Manual Operation | Same | #### Technological Characteristics: {5}------------------------------------------------ | Components | - Uterine balloon to create tamponade, recommended fill volume of 750 ml<br>- Vaginal balloon to assist in positioning, recommended fill volume of 350 ml<br>- Spike assembly to allow for rapid fluid fill from IV bag<br>- Leuer assembly for a measured fill<br>- Drainage lumen<br>- Separate irrigation lumen | - Uterine balloon to create tamponade, recommended fill volume of 750 ml<br>- Vaginal balloon to assist in positioning, recommended fill volume of 350 ml<br>- Spike assembly to allow for rapid fluid fill from IV bag<br>- Leuer assembly for a measured fill<br>- Drainage lumen<br>- Separate irrigation lumen | Same | |-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------| | Safety Features | Labels on Lumens for immediate identification | Labels on Lumens for immediate identification | Same | | Sterility | Ethylene Oxide, SAL of 10-6 | Ethylene Oxide, SAL of 10-6 | Same | | Reuse | Single Use | Single Use | Same | | Shelf Life | 3 Years | 3 Years | Same | | Dimensions | - Recommended uterine balloon fill volume of 750 ml<br>- Recommended vaginal balloon fill volume of 350 ml | - Recommended uterine balloon fill volume of 750 ml<br>- Recommended vaginal balloon fill volume of 350 ml | Same | # Performance Data Testing performed on the modified devices demonstrated significant improvement in strength over the previous version. The testing consisted of subjecting the device to cyclic physiologic pressures under more challenging conditions than indicated in published literature. None of the modified devices failed under those conditions. # Substantial Equivalence The ebb Complete Tamponade System has the same intended use and similar indications, principles of operation, and technological characteristics as the previous ebb Complete Tamponade System. The minor differences is the addition of a UV Rings at the distal end and proximal ends of the uterine and vaginal balloons, does not raise any new questions of safety or effectiveness. Performance data demonstrates that the ebb Complete Tamponade System is as safe and effective as the ebb Complete Tamponade System. Thus, the ebb Complete Tamponade System is substantially equivalent to its predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...