Taps are nonpowered hand-held devices intended for bone cutting and drilling on a patient's skull during fracture repair and reconstructive procedures of the cranium. The taps may be used to prepare cranial bone to insert bone fixation screws.
Device Story
Nonpowered, hand-held surgical instruments; used to drill holes and create threads in cranial bone; facilitates insertion of Rapid Resorbable Fixation System bone screws. Operated by surgeons in clinical/OR settings. Input: manual force applied via compatible handles (mini-quick or hex coupling). Output: threaded hole in bone. Benefits: enables secure fixation of cranial plates/screws during reconstructive surgery.
Clinical Evidence
No clinical data. Bench testing only: simulated use/bioskills lab (n=5 users, 4 cadaveric heads), sawbones lab (self-drilling validation), torsional testing (failure vs. stripping torque), axial load testing (stop collar integrity), hex coupling validation, and biocompatibility (cytotoxicity per ISO 10993-5).
Indicated for patients undergoing cranial fracture repair and reconstructive procedures requiring bone fixation screws.
Regulatory Classification
Identification
Manual cranial drills, burrs, trephines, and their accessories are bone cutting and drilling instruments that are used without a power source on a patient's skull.
Predicate Devices
Medtronic Sofamor Danek POWEREASE™ System (K123270)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 18, 2017
Synthes (USA) Products, LLC Ms. Susan Lewandowski Manager, Regulatory Affairs 1301 Goshen Parkway West Chester, Pennsylvania 19380
Re: K153587
Trade/Device Name: Taps for Resorbable Screws Regulation Number: 21 CFR 882.4300 Regulation Name: Manual Cranial Drills, Burrs, Trephines, and Their Accessories Regulatory Class: Class II Product Code: HBG Dated: December 22, 2016 Received: December 23, 2016
Dear Ms. Lewandowski:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153587
Device Name Taps for Resorbable Screws
Indications for Use (Describe)
Taps are nonpowered hand-held devices intended for bone cutting and drilling on a patient's skull during fracture repair and reconstructive procedures of the cranium.
The taps may be used to prepare cranial bone to insert bone fixation screws.
Type of Use (Select one or both, as applicable)
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|--------------------------------------------------------------------------------------|
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Synthes logo. The logo consists of a circular symbol on the left and the word "SYNTHES" in bold, black letters on the right. There is a line underneath the word "SYNTHES".
| Date Prepared: | January 18, 2017 |
|----------------|--------------------------------------------------------------------------------------------------------|
| Submitter: | Synthes USA Products, LLC<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>United States of America |
| Contact: | Susan Lewandowski<br>slewand2@its.jnj.com<br>Telephone 610-719-5852<br>Facsimile 484-356-9682 |
### DEVICE
| Device Name: | Taps for Resorbable Screws |
|----------------------|-----------------------------------------------------------------------------------------|
| Classification Name: | Drills, burrs, trephines, and accessories (Manual) |
| Review Panel: | Neurology |
| Regulatory Class: | II |
| Product Code: | HBG (Manual cranial drills, burrs, trephines, and their accessories)<br>21 CFR 882.4300 |
### PREDICATE DEVICE
Medtronic Sofamor Danek POWEREASE™ System K123270 is the primary predicate.
Synthes Dental Bone Cutting Instruments K150796 is used as a reference device in support of this submission.
#### DEVICE DESCRIPTION
Taps are used to drill a hole and simultaneously create threads in order to accommodate a Rapid Resorbable Fixation System bone screw. The self-drilling fixed-stop taps are manufactured from Stainless Steel 440A which conforms to ASTM F899 Standard Specification for Stainless Steel for Surgical Instruments and ASTM A276 Specification for Stainless Steel Bars and Shapes. The adjustable-length taps (final assembly) are assembled from three components; the adjustable tap (Stainless Steel 440A), the locking collar (Makrolon Rx2530 W/1118 Tint), and the stop collar (Stainless Steel 316L with an aluminum titanium nitride coating).
DePuy Synthes Taps for Resorbable Screws can be used with the following handles: 311.01.98 Handle, with mini quick coupling 311.03 Handle, with mini quick coupling, small 311.005 Screwdriver Handle with hex coupling, small 311.006 Screwdriver Handle with hex coupling, medium 311.007 Screwdriver Handle with hex coupling, large
The DePuy Synthes Taps for Resorbable Screws can be used with the DePuy Synthes Rapidsorb® Rapid Resorbable Fixation System (K062789) screws [1.5mm, 2.0mm, 2.5mm (Emergency) diameters] and plates (1.5 mm and 2.0mm).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Synthes. The logo consists of a circular symbol on the left and the word "SYNTHES" in bold, sans-serif font on the right. The circular symbol appears to be a stylized representation of a medical device or implant. A registered trademark symbol is located to the right of the word "SYNTHES".
## INDICATIONS FOR USE
#### Taps for Resorbable Screws
Taps are nonpowered hand-held devices intended for bone cutting and drilling on a patient's skull during fracture repair and reconstructive procedures of the cranium. The taps may be used to prepare cranial bone to insert bone fixation screws.
#### Medtronic Sofamor Danak POWEREASE System (Primary Predicate)
IPC® System is indicated for the incision/cutting, removal, drilling and sawing of soft and hard tissue and bone, and biomaterials in Neurosurgical (Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures. The IPC® POWEREASE™ System is indicated for drilling, tapping and driving screws and working end attachments during spinal surgery, including open and minimally invasive procedures. It is also used in placement or cutting of screws, posts and rods.
#### Discussion
Both the subject taps and the primary predicate Medtronic Sofamor Danek POWEREASE System K123270 are intended for use in neurosurgical procedures to drill or cut into the skull. There is a difference in the assigned FDA product codes:
- The subject taps are manual instruments, with a corresponding product code of . HBG (Drills, burrs, trephines, and their accessories - manual).
- The primary predicate is a powered instrument with the corresponding product code of HBE (drills, burrs, trephines, and accessories - simple, powered).
Additionally, the subject taps defined herein as Class II product code HBG are the same instruments currently marketed by Synthes as Class I product code LXH (Orthopedic manual surgical instruments). Thus, the subject taps include the intended uses of both product codes (i.e. cranial bone-cutting and general orthopedic bone cutting).
The reference device Synthes Dental Bone Cutting Instruments K150796 does not have the same intended use as the subject taps since the dental bone cutting instruments are intended for trauma, reconstruction, or orthognathic procedures to drill or cut into the upper or lower jaw. The reference device product code is DZJ (Driver, wire, and Bone Drill Manual). Since taps for general orthopedic bone-cutting would typically be considered Class I Exempt instruments, the identified reference device is included to show that taps have been previously cleared within premarket notification submissions.
## COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
#### Primary Predicate
DePuy Synthes considers the subject taps equivalent to the Medtronic Sofamor Danek Sofamor Danek POWEREASE™ System K123270, as well as the Synthes Dental Bone Cutting Instruments K150796.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. A small "R" in a circle is located to the right of the word "SYNTHES", indicating that it is a registered trademark.
### Technological Similarities of Taps to the Primary Predicate
- Both instruments are used for drilling a hole in preparation for a bone screw ●
- . Both are used in the skull
- Both can be used manually
- The Medtronic POWEREASE™ System consists of a hand piece and ● attachments; the subject taps consist of the instruments and handles based on the coupling
### Technological Differences of Taps to Primary Predicate
- The Medtronic POWEREASE™ System is used for drilling, tapping and driving . screws as well as for the placement or cutting of screws, posts and rods; the tap is used to drill a hole and simultaneously create threads in order to receive a bone screw
- The Medtronic POWEREASE™ System can be used manually or with power; the ● subject taps are manual instruments only
- The Medtronic IPC® and POWEREASE™ Systems (K123270) are used in ● Neurosurgical (Cranial, Craniofacial), Orthopedic, Arthroscopic, Spinal, Sternotomy, and General surgical procedures; the subject taps are intended for neurosurgical procedures only
#### Impact on Safety and Effectiveness
There is no impact to safety and effectiveness. The subject devices have a limited intended use compared to the predicate device; the POWEREASE System is used for drilling, tapping, and driving screws while the taps are used to cut bone in preparation of receiving a bone screw. The subject devices are basic handheld instruments compared to the predicate; the tap is connected to a handle (dependent on the coupling feature) while the predicate device consists of a handpiece and multiple attachments. The subject devices are used manually while the predicate can be used either manually or with power.
#### Reference Device
Both the subject taps and the reference Dental Bone Cutting Instruments (K150796) are generally intended to be used during surgery for cutting into bone. The reference device is included to show that taps have been cleared within premarket submissions.
#### Technological Similarities of Taps to the Reference Device
- Both devices are intended to be used to prepare bone for placement of a screw
- Both devices are operated manually ●
- Both are manufactured from stainless steel (440A) .
#### Technological Differences of Taps to the Reference Device
- The subject taps are intended for use in fracture repair and reconstructive procedures of the cranium (covered under a Neurology product code); the Dental Bone Cutting Instruments are intended for use in oral/maxillofacial surgery (covered under a Dental product code).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Synthes logo. The logo consists of a circular symbol on the left and the word "SYNTHES" in bold, sans-serif font on the right. A registered trademark symbol is located to the upper right of the word "SYNTHES". A horizontal line is located underneath the logo.
### Impact on Safety and Effectiveness
There is no impact to safety and effectiveness. The subject taps are intended for use in fracture repair and reconstructive procedures of the cranium (Neurology); the Dental Bone Cutting Instruments are intended for use in oral/maxillofacial surgery (Dental). Regardless of specific intended use, all taps are used to cut bone in preparation of receiving a bone screw. The reference device was included to provide evidence that taps have been cleared within premarket notification submissions.
## PERFORMANCE DATA
### Mechanical Testing
The performance data provided in support of substantial equivalence of the proposed devices are as follows:
- Simulated Use/Bioskills Lab ●
- Sawbones Lab ●
- Torsional Testing of RapidSorb Taps
- . RapidSorb Self-Drilling Tap Axial Load at Strip Out
- Validation of Hex Coupling for RapidSorb Adjustable Taps ●
- Biocompatibility Testing - Cytotoxicity
| Performance Testing - Bench | | |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Test | Test Method Summary | Results |
| Simulated Use/Bioskills Lab -<br>RapidSorb Instruments | A representative range of the<br>subject devices will be selected to<br>represent the entire range of taps.<br>The lab will be conducted to<br>validate various aspects of the<br>device design such as torque<br>resistance, length and screw size<br>diameter. The validation lab will<br>consist of 5 individual<br>participants. There will be 4<br>cadaveric cephalises. The heads<br>(cephalises) will be selected to<br>represent a range of patients (a<br>young male and female, and<br>elderly male and female). The<br>users are independent from the<br>design of the subject device. | The results indicated full<br>validation of the subject device.<br>All acceptance criteria were met. |
| Saw Bones Lab - RapidSorb<br>Instruments | The objective of the lab is to<br>validate that the subject taps are<br>able to self-drill in normal density<br>bone. A representative range of<br>the subject device will be selected<br>to represent the entire range of<br>taps. The range includes all<br>diameters and the longest length<br>of each tap. These represent<br>worst case in regard to torque<br>resistance. | The acceptance criteria was met<br>in that all four users were able to<br>use each of the three subject taps<br>to create three separate tapped<br>holes that will accommodate the<br>corresponding screw (36 total<br>insertions). |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Synthes. The logo consists of a stylized symbol on the left and the word "SYNTHES" in bold, sans-serif font on the right. A registered trademark symbol is located to the right of the word "SYNTHES". A horizontal line is located underneath the logo.
| Performance Testing - Bench | | | | | | |
|-----------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|
| Test | Test Method Summary | Results | | | | |
| Mechanical Test - Torsional<br>Testing of RapidSorb Taps | The objective of this test is to<br>demonstrate that the failure<br>torque of RapidSorb self-drilling<br>taps is above the torque at which<br>the same diameter taps will<br>bottom-out in the testing<br>substrate. Two separate tests will<br>be performed, Failure Torque<br>Test and Stripping Torque Test. | The results of this testing indicate<br>that when the failure torque of all<br>three sizes of RapidSorb Self<br>Drilling Taps (1.5, 2.0, and 2.5<br>mm) are compared to their<br>respective bottoming or stripping<br>torque, the resulting p-values<br>were all 0.000. There is<br>sufficient evidence to reject the<br>null hypothesis in favor of the<br>alternate, that in all cases the<br>RapidSorb failure torque is<br>superior to the Bottoming or<br>Stripping torque (as specified).<br>This satisfies the acceptance<br>criteria. | | | | |
| Mechanical Test - RapidSorb<br>Self-Drilling Tap Axial Load at<br>Strip Out | The objective of this test was to<br>demonstrate that the axial force<br>required to cause the adjustable<br>stop on the adjustable length taps<br>(311.100, 311.101, 311.102,<br>311.110, 311.111, 311.112) to<br>unintentionally move is greater<br>than the axial force that is<br>generated by the strip out torque<br>of each tap. | As can be seen in all comparisons<br>(1.5, 2.0, and 2.5mm) the<br>resulting p-value was 0.000,<br>which is sufficient evidence to<br>reject the null hypothesis in favor<br>of the alternate that the Axial<br>load required to move the<br>adjustable stop was superior to<br>the axial load at strip-out in all<br>cases. This result satisfies the<br>acceptance criteria. | | | | |
| Validation of Hex Coupling for<br>RapidSorb Adjustable Taps | The objective of this test was to<br>validate the compatibility of the<br>Hex coupling on the proximal end<br>of the RapidSorb Adjustable<br>length taps. All tested parts will<br>couple with the designated hex<br>handles and, then be fully inserted<br>into 40 lbs/cf polyurethane foam<br>with no slippage or visible<br>damage (no magnification) of<br>taps proximal coupling. | At the conclusion of the testing,<br>all participants were able to drive<br>all 12 taps with a combination of<br>the 3 specified handles. This was<br>done with no visible damage to<br>the proximal coupling end of the<br>tap. | | | | |
| Biocompatibility (Cytotoxicity) | An in vitro study was conducted<br>to evaluate for potential cytotoxic<br>effects, following guidelines of<br>the ANSI/AAMI/ISO 10993-5,<br>Biological Evaluation of medical<br>devices – Part 5: Test for in vitro<br>cytotoxicity (2009). | At the conclusion of the testing,<br>the test articles extract showed no<br>reactivity of causing cell lysis or<br>toxicity. All test method<br>acceptance criteria were met.<br>The patient contacting<br>components of taps are<br>manufactured from Stainless<br>Steel 440A which conforms to<br>the FDA recognized consensus<br>standard, ASTM F899 - Standard<br>Specification for Stainless Steel<br>for Surgical Instruments,<br>supporting the biocompatibility<br>of this device. | | | | |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. A registered trademark symbol is located to the right of the word.
Performance testing demonstrates that the mechanical performance of the proposed taps is comparable to that of the primary predicate device (taps) and supports substantial equivalence.
## Clinical Studies
Clinical testing was not necessary for the determination of substantial equivalence.
The performance data demonstrate that the mechanical performance of the proposed taps is comparable to that of the primary predicate device and supports substantial equivalence to the primary predicate device for safe and effective bone cutting in the skull.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the logo for SYNTHES. The logo consists of a stylized symbol on the left, followed by the word "SYNTHES" in bold, sans-serif font. A registered trademark symbol is located to the right of the word. The logo is black and white.
# CONCLUSIONS
The proposed devices have the same intended use as the primary predicate device. The mechanical testing included in this submission demonstrates that:
- Any differences in technological characteristics of the primary predicate and/or . reference device do not raise any new questions of safety and effectiveness.
- . The proposed devices are at least as safe and effective as the primary predicate
It is concluded that the information included in this submission supports substantial equivalence.
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "SYNTHES" in a bold, sans-serif font. To the left of the word is a circular logo with a stylized figure inside. A registered trademark symbol is located to the right of the word.
## Device Comparison Table
| | DePuy Synthes Taps for<br>Resorbable Screws (subject<br>device) | Medtronic Sofamor Danek<br>POWEREASE System<br>K123270<br>(primary predicate) | DePuy Synthes Dental Bone<br>Cutting Instruments, e.g. Taps<br>and Countersinks K150796<br>(reference device) |
|--------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | Taps are nonpowered hand-held<br>devices intended for bone<br>cutting and drilling on a patient's<br>skull during fracture repair and<br>reconstructive procedures of the<br>cranium. The taps may be used<br>to prepare cranial bone to insert<br>bone fixation screws. | The IPC System is indicated for<br>the incision/cutting, removal,<br>drilling and sawing of soft and<br>hard tissue and bone, and<br>biomaterials in Neurosurgical<br>(Cranial, Craniofacial),<br>Orthopedic, Arthroscopic,<br>Spinal, Sternotomy, and General<br>surgical procedures.<br>The IPC POWEREASE System<br>is indicated for drilling, tapping<br>and driving screws and working<br>end attachments during spinal<br>surgery, including open and<br>minimally invasive procedures.<br>It is also used in the placement<br>or cutting of screws, posts and<br>rods. | The Dental Bone Cutting<br>Instruments are nonpowered<br>hand-held devices intended for<br>trauma, reconstruction, or<br>orthognathic procedures to drill<br>or cut into the upper or lower<br>jaw and may be used to prepare<br>bone to insert a wire, pin, or<br>screw. |
| FDA Product Code | HBG | HBE, HWE, GWF | DZJ |
| 21 CFR | 882.4300 | 882.4310 | 872.4120 |
| Regulation<br>Description | Manual cranial drills, burrs,<br>trephines, and their accessories | Powered simple cranial drills,<br>burrs, trephines, and their<br>accessories | Bone cutting instrument and<br>accessories |
| Review Panel | Neurology | Neurology | Dental |
| Device Class | 2 | 2 | 2 |
| Device Description | Taps are used to drill a hole and<br>simultaneously create threads in<br>order to accommodate a Rapid<br>Resorbable Fixation System<br>bone screw. The taps are<br>available as self-drilling fixed-<br>stop taps and self-drilling<br>adjustable-stop taps. | The POWEREASETM System<br>Working Ends consists of<br>instruments such as taps, drill<br>bits, screwdrivers, post cutter,<br>set screw break-off tool,<br>reduction nut driver and sleeves.<br>The working ends have a manual<br>alternative. The working ends,<br>listed above, are compatible with<br>the CD HORIZON SOLERA<br>and the TSRH 3Dx Spinal<br>System implants. Of the working<br>ends, only the taps, screwdrivers,<br>drill bits, and sleeves are also<br>compatible with Medtronic's<br>NIM-ECLIPSE Spinal System. | Countersinks are intended to<br>create a countersink feature in<br>bone to provide a contact surface<br>for screw heads and to reduce<br>palpability of the bone screw.<br>Taps are intended to create<br>threads in a pre-drilled hole in<br>order to accomodate a bone<br>screw. |
| Technological<br>characteristics: | | | |
| Principles of<br>function/technology | Taps are designed to be<br>compatible with corresponding<br>Synthes screws and/or drill bits.<br>Taps have cutting threads that<br>run along the length of the<br>instrument. The diameter and | The instrument modifications<br>detailed in this submission have<br>no impact on the technological<br>characteristic of the existing<br>instruments. The working end<br>Taps are intended for tapping | Countersinks feature cutting<br>flutes at the working end of the<br>instrument that continue a<br>distance up the shaft of the<br>instrument and a centering pin at<br>the working end of the |
| | DePuy Synthes Taps for<br>Resorbable Screws (subject<br>device) | Medtronic Sofamor Danek<br>POWEREASE System<br>K123270<br>(primary predicate) | DePuy Synthes Dental Bone<br>Cutting Instruments, e.g. Taps<br>and Countersinks K150796<br>(reference device) |
| | thread profile of the tap is<br>designed to match to the<br>diameter and thread profile of an<br>associated screw. | during spinal surgery, including<br>both open and minimally<br>invasive procedures. The<br>working ends are used to<br>facilitate the placement of the<br>rods. Like the predicate<br>POWEREASE System<br>instruments, the subject<br>instruments are manufactured<br>from stainless steel. | instrument. The user will insert<br>the centering pin into a drilled<br>pilot hole, ensuring that the<br>countersink feature is concentric<br>with the drilled hole.<br>Taps are designed to be<br>compatible with corresponding<br>Synthes screws and/or drill bits.<br>Taps have cutting threads that<br>run along the length of the<br>instrument. The diameter and<br>thread profile of the tap is<br>designed to match to the<br>diameter and thread profile of an<br>associated screw. |
| Dimensions | | | |
| Stop Depth: | Fixed stop taps: 3mm, 4mm,<br>5mm, 6mm, 8mm<br>Adjustable stop taps: 3-8mm<br>(1mm increments) | Unknown | 20-122mm tapping depths |
| Couplings: | Mini-quick coupling<br>Hex coupling | Unknown | Hex coupling |
| Compatible<br>Screws: | Rapidsorb resorbable cortex<br>screws: 1.5mm, 2.0mm, 2.5mm<br>diameter | Unknown | Cortex screws: 1.3-3.0mm<br>diameter |
| Device material(s) | Stainless steel 440A, 316L<br>Polycarbonate (Makrolon<br>Rx2530 W/1118 Tint)<br>Aluminum titanium nitride<br>coating | Stainless Steel | Stainless steel 440A |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the Synthes logo. The logo consists of a circular icon on the left and the word "SYNTHES" in bold, black letters on the right. The icon features a stylized image of a person or figure within a circle, and a registered trademark symbol is located at the upper right of the word "SYNTHES".
(End of summary)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.