K153552 · Ageless Health Industrial Co., Ltd. · DXN · Jun 1, 2016 · Cardiovascular
Device Facts
Record ID
K153552
Device Name
AGE Automatic Upper Arm Blood Pressure Monitor
Applicant
Ageless Health Industrial Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
Jun 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
AGE Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual.
Device Story
Battery-powered automatic non-invasive blood pressure monitor; uses oscillometric technique to measure systolic/diastolic blood pressure and pulse rate. Input: pressure signals from inflatable upper arm cuff. Processing: automated inflation/deflation and signal analysis to derive BP and pulse. Output: digital display of systolic/diastolic BP, pulse rate, date, time, WHO BP classification bar, and memory records. Used by medical professionals or patients at home. Healthcare providers use output for routine BP monitoring and clinical decision-making. Benefits: enables convenient, non-invasive, self-administered or clinical BP tracking.
Clinical Evidence
Bench testing only. Evaluated per IEC 60601-1 (safety), IEC 60601-1-2 (EMC), ISO 10993-5 (cytotoxicity), ISO 10993-10 (irritation/sensitization), and ISO 81060-2 (clinical validation of automated measurement). No clinical trial data provided.
Technological Characteristics
Battery-powered (4x AA) oscillometric monitor. Materials: patient-contacting materials tested per ISO 10993-5/10. Connectivity: standalone. Software: embedded firmware. Standards: IEC 60601-1, IEC 60601-1-2, ISO 81060-2.
Indications for Use
Indicated for adult individuals to measure diastolic and systolic blood pressure and pulse rate using a non-invasive oscillometric technique with an inflatable upper arm cuff. Available in six cuff sizes.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
AGE Automatic Upper Arm Blood Pressure Monitor (K123882)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, a symbol often associated with medicine and healthcare.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 1, 2016
Ageless Health Industrial Co., Ltd % Cecelia Ceng Regulatory Manager Guangzhou GLOMED Biological Technology Co., Ltd. Suite 306, Kecheng Mansion, No.121 Science Road Guangzhou Science Park Guangzhou, 510663 CN
Re: K153552
Trade/Device Name: AGE Automatic Upper Arm Blood Pressure Monitor, Model: BA-801X, BA-802X, BA-803X, BA-805X, BA-806X, BA-811X, BA-812X, BA-813X, BA-821X, BA-822X, BA-823X, BA-826X (X can be A, B, C, D, E, F). Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: November 25, 2015 Received: December 11, 2015
Dear Cecelia Ceng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours. Shawn W. Forrest -S 2016.06.01 22:58:44 -04'00'
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
#
510(k) Number (if known) K153552
### Device Name
AGE Automatic Upper Arm Blood Pressure Monitor Model: BA-801X, BA-802X, BA-805X, BA-806X, BA-811X, BA-812X, BA-813X, BA-821X, BA-822X, BA-823X, BA-826X (X can be A, B, C, D, E, F).
### Indications for Use (Describe)
AGE Automatic Upper Arm Blood Pressure Monitor is for use by medical professionals or at home and is a non-invasive blood pressure measurement system intended to measure the diastolic blood pressures and pulse rate of an adult individual by using a non-invasive technique in which an inflatable cuff is wrapped around the upper arm. The cuff circumference is six sizes.
| Type of Use (Select one or both, as applicable) | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <table style="border:none"><tr><td><span style="font-size: 12px;">☐ Registration Use (Part 21 CFR 601 Subpart D)</span></td><td><span style="font-size: 12px;">☒ One-Time Central Use (21 CFR 601 Subpart G)</span></td></tr></table> | <span style="font-size: 12px;">☐ Registration Use (Part 21 CFR 601 Subpart D)</span> | <span style="font-size: 12px;">☒ One-Time Central Use (21 CFR 601 Subpart G)</span> |
| <span style="font-size: 12px;">☐ Registration Use (Part 21 CFR 601 Subpart D)</span> | <span style="font-size: 12px;">☒ One-Time Central Use (21 CFR 601 Subpart G)</span> | |
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Chapter 6. 510(k) Summary
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 870.1130.
#### Submitter Information 1.
Sponsor Name: Dongguan Ageless Health Industrial Co., Ltd Address: 3/F, A1 Bldg, Dongshen Sima Industrial Area, No.33 Shenbei Road, Sima Village, Changping Town, Dongquan City, Guangdong Province, China Contact Person: Victor Wan (Vice-president) Phone: +86-769-81158038 Fax: +86-769-82289331 E-mail: victor@agelh.com
### Application Correspondent:
Contact Person: Ms. Cecilia Ceng / Mr. Tim Wong Guangzhou GLOMED Biological Technology Co., Ltd. Tel: +86-20-61099984 Email: requlatory@glomed-info.com
### 2. Subject Device Information
| Type of 510(k): | Traditional |
|----------------------|------------------------------------------------|
| Common Name: | Noninvasive blood pressure measurement systems |
| Trade Name: | AGE Automatic Upper Arm Blood Pressure Monitor |
| Classification Name: | Noninvasive blood pressure measurement system |
| Review Panel: | Cardiovascular |
| Product Code: | DXN |
| Regulation Number: | 21 CFR 870.1130 |
| Regulation Class: | 2 |
{4}------------------------------------------------
| Sponsor: | Dongguan Ageless Health Industrial Co., Ltd |
|--------------------|------------------------------------------------|
| Common Name: | Noninvasive blood pressure measurement systems |
| Trade Name: | AGE Automatic Upper Arm Blood Pressure Monitor |
| 510(k) number: | K123882 |
| Review Panel: | Cardiovascular |
| Product Code: | DXN |
| Regulation Number: | 21 CFR 870.1130 |
| Regulation Class: | 2 |
#### 4. Device Description
AGE Automatic Upper Arm Blood Pressure Monitor is a battery driven automatic non-invasive blood pressure meter. It can automatically conduct the inflation and measurement, which can measure systolic and diastolic blood pressure as well as the pulse rate of adult at arm within its claimed range and accuracy via the Oscillometry technique. The device also has low voltage indication, which will be triggered when the battery is low.
#### 5. Intended Use
AGE Automatic Upper Arm Blood Pressure Monitor is intended for use by medical professionals or at home to monitor and display diastolic, systolic blood pressure and pulse rate on adult each time, with an air cuff buckled around one's arm according to the instruction in the user's guide manual.
#### Test Summary 6.
AGE Automatic Upper Arm Blood Pressure Monitor has been evaluated the safety and performance by lab bench testing according to the following standards:
- IEC 60601-1, Medical Electrical Equipment Part 1: General Requirements for Safety, 2005 + � A1:2012
- � IEC 60601-1-2, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic Disturbances - Requirements and tests, 2014
- � ISO 10993-5, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity, 2009
- � ISO 10993-10, Biological evaluation of medical devices -- Part 10: Tests for irritation and skin sensitization, 2010
- � AAMI / ANSI / ISO 81060-2 Second Edition, Non-Invasive Sphygmomanometers - Part 2: Clinical Validation Of Automated Measurement Type. (Cardiovascular)
{5}------------------------------------------------
### 7. Comparison to Predicate Device
Compare with predicate device, the subject device is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device and predicate device do not raise and new questions of safety or effectiveness.
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
|--------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| Product Name | AGE Automatic Upper Arm Blood<br>Pressure Monitor | AGE Automatic Upper Arm Blood<br>Pressure Monitor | -- |
| Intended Use and Indications for Use | | | |
| Intended Use | AGE Automatic Upper Arm Blood<br>Pressure Monitor is intended for<br>use by medical professionals or at<br>home to monitor and display<br>diastolic, systolic blood pressure<br>and pulse rate on adult each time,<br>with the cuff around the left upper<br>arm according to the instruction in<br>the user's guide manual. | AGE Automatic Upper Arm Blood<br>Pressure Monitor is intended for<br>use by medical professionals or at<br>home to monitor and display<br>diastolic, systolic blood pressure<br>and pulse rate on adult each time,<br>with the cuff around the left upper<br>arm according to the instruction in<br>the user's guide manual. | SE |
| Indications for<br>Use | AGE Automatic Upper Arm Blood<br>Pressure Monitor is for use by<br>medical professionals or at home<br>and is a non-invasive blood<br>pressure measurement system<br>intended to measure the diastolic<br>and systolic blood pressures and<br>pulse rate of an adult individual by<br>using a non-invasive technique in<br>which an inflatable cuff is wrapped<br>around the upper arm. The cuff<br>circumference is six sizes. | AGE Automatic Upper Arm Blood<br>Pressure Monitor is for use by<br>medical professionals or at home<br>and is a non-invasive blood<br>pressure measurement system<br>intended to measure the diastolic<br>and systolic blood pressures and<br>pulse rate of an adult individual by<br>using a non-invasive technique in<br>which an inflatable cuff is wrapped<br>around the upper arm. The cuff<br>circumference is limited to 24~34<br>cm. | SE<br>Note 1 |
| ELECTRICAL REQUIREMENT | | | |
| Power Supply | DC 6V (4 X AA 1.5V alkline<br>batteries) | 6Vdc (4 "AA" batteries) | SE |
| PERFORMANCE SPECIFICATION | | | |
| Measuring<br>Method | Oscillometry | Oscillometry | SE |
| Measurement<br>range for<br>pressure | 0~280 mmHg | 0~280 mmHg | SE |
| Operating<br>Range for<br>Pressure (Cuff<br>Pressure) | 0~280 mmHg | 0~280 mmHg | SE |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
| Cuff Display<br>Range | 0~280 mmHg | 0~280 mmHg | SE |
| Pressure<br>resolution | 1mmHg or 0.1kPa | 1mmHg or 0.1kPa | SE |
| Accuracy | Pressure: ±3 mmHg<br>Pulse: ±5% | Pressure: ±3mmHg<br>Pulse: ±5% | SE |
| Patient<br>Population | Adult | Adult | SE |
| Measurement<br>Site of Body | Upper Arm | Upper Arm | SE |
| Inflation and<br>Deflation | Automatic | Automatic | SE |
| Memory Size | 2 x 90 sets record (BA-822X is 90<br>sets ) | 2 x 90 sets record | SE<br>Note 2 |
| Software<br>Version | V01 | V1.1 | SE<br>Note 3 |
| Indicators | Blood Pressure (Systolic and<br>Diastolic), Pulse, Date, Time, WHO<br>BP Classification Indicating Bar,<br>Low Battery Icon, Heart Icon,<br>Memory Record Number | Blood Pressure (Systolic and<br>Diastolic), Pulse, Date, Time, WHO<br>BP Classification Indicating Bar,<br>Low Battery Icon, Heart Icon,<br>Memory Record Number | SE |
| Physical<br>Dimensions<br>Weight | Please refer to "Table 1 Physical Dimensions and weight for each model"<br>of "Chapter 6" | | SE<br>Note 4 |
| Cuff<br>Circumference | size A: 17cm--22cm (SMALL<br>ADULT CUFF)<br>size B: 22cm--30cm (ADULT<br>CUFF-1)<br>size C: 24cm--34cm (ADULT<br>CUFF-2)<br>size D: 22cm--42cm (L-LARGE<br>ADULT CUFF)<br>size E: 30cm--42cm (LARGE<br>ADULT CUFF)<br>size F: 42cm--50cm (EXTRA<br>LARGE ADULT CUFF) | 24~34 cm | SE<br>Note 1 |
| OPERATING & STORAGE CONDITIONS | | | |
| Storage<br>Environment | Temperature: -20°C ~ +65°C<br>Humidity: 10~95%RH<br>Atmospheric Pressure:86 kPa~106<br>kPa | Temperature: -20°C ~ +65°C<br>Humidity: 10~95%RH<br>Atmospheric Pressure:86 kPa~106<br>kPa | SE |
| Working<br>Environment | Temperature: 5°C ~ 40°C<br>Humidity: 15~90%RH<br>Atmospheric Pressure:86 kPa~106 | Temperature: 5°C ~ 40°C<br>Humidity: 10~90%RH<br>Atmospheric Pressure:86 kPa~106<br>kPa | SE<br>Note 2 |
| Elements of<br>Comparison | Subject Device | Predicate Device | Verdict |
| | kPa | | |
| COMPLIANCE STANDARDS | | | |
| Electrical,<br>Mechanical and<br>Thermal<br>Evaluation | IEC 60601-1<br>IEC 60601-1-2 | IEC 60601-1<br>IEC 60601-1-2 | SE |
| Biocompatibility<br>Evaluation | All the patient contracting materials<br>are evaluated by the<br>biocompatibility standard ISO<br>10993 -5, -10. | All the patient contracting materials<br>are evaluated by the<br>biocompatibility standard ISO<br>10993 -5, -10. | SE |
| Performance | ISO 81060-2 | AAMI SP10 | SE<br>Note 5 |
{6}------------------------------------------------
{7}------------------------------------------------
### Note 1
Although there is difference for measurement cuff circumference of subject device and predicate device, both of them are complied with ISO 81060-2. This difference does not affect the safety and effectiveness.
### Note 2
Although some specifications of operating & storage conditions, memory size are different for subject device and predicate device, they are both complied with IEC 60601-1. The differences do not affect the safety and effectiveness. The storage and operating pressure ranges of subject device and predicate device are same, and the storage and operating pressure ranges of predicate device was provided by the manufacturer.
### Note 3
The software version of subject device is updated based on the predicate device, there is no different in PWB control circuit and software among models. The software is same as K123882.
### Note 4
These differences of physical dimensions and weight have been verified and validated to demonstrate that it does not affect the safety and effectiveness of subject device.
### Note 5
Although the standards of performance have updated and substituted from AAMI SP10 to ISO 81060-2, all of the requirements in these three standards are intended for blood pressure monitor. Therefore,
{8}------------------------------------------------
both of them met the requirements. The differences do not affect the safety and effectiveness.
{9}------------------------------------------------
| Subject Device | | | Predicate Device | | |
|----------------|-------------------------------|----------------------------------------------|------------------|-------------------------------|----------------------------------------------|
| Model<br>Name | Appearance Picture | Size and<br>Weight | Model Name | Appearance Picture | Size and<br>Weight |
| BA-801X | Image: Blood pressure monitor | 115mm (L)<br>103mm (W)<br>67.5mm (H)<br>460g | BA-801 | Image: Blood pressure monitor | 115mm (L)<br>103mm (W)<br>67.5mm (H)<br>460g |
| BA-802X | Image: Blood pressure monitor | 115mm (L)<br>103mm (W)<br>67.5mm (H)<br>460g | BA-802 | Image: Blood pressure monitor | 115mm (L)<br>103mm (W)<br>67.5mm (H)<br>460g |
{10}------------------------------------------------
| Subject Device | | | Predicate Device | | |
|----------------|-------------------------------|------------------------------------------------|------------------|-------------------------------|------------------------------------------------|
| Model<br>Name | Appearance Picture | Size and<br>Weight | Model Name | Appearance Picture | Size and<br>Weight |
| BA-803X | Image: Blood pressure monitor | 119mm (L)<br>100mm (W)<br>63.2mm (H)<br>456g | BA-803 | Image: Blood pressure monitor | 119mm (L)<br>100mm (W)<br>63.2mm (H)<br>456g |
| BA-805X | Image: Blood pressure monitor | 118.3mm (L)<br>100mm (W)<br>62.9mm (H)<br>465g | BA-805 | Image: Blood pressure monitor | 118.3mm (L)<br>100mm (W)<br>62.9mm (H)<br>465g |
{11}------------------------------------------------
| Subject Device | | | Predicate Device | | |
|----------------|-----------------------------------------------------------------------------------------------------|--------------------------------------------|------------------|-----------------------------------------------------------------------------------------------------|--------------------------------------------|
| Model<br>Name | Appearance Picture | Size and<br>Weight | Model Name | Appearance Picture | Size and<br>Weight |
| BA-806X | Image: Blood pressure monitor with arm cuff, LCD screen, and buttons labeled Memory, Power, and Set | 144mm (L)<br>112mm (W)<br>70mm (H)<br>472g | BA-806 | Image: Blood pressure monitor with arm cuff, LCD screen, and buttons labeled Memory, Power, and Set | 144mm (L)<br>112mm (W)<br>70mm (H)<br>472g |
| BA-811X | Image: Blood pressure monitor with arm cuff, LCD screen, and buttons labeled Memory, Power, and Set | 144mm (L)<br>112mm (W)<br>67mm (H)<br>468g | BA-811 | Image: Blood pressure monitor with arm cuff, LCD screen, and buttons labeled Memory, Power, and Set | 144mm (L)<br>112mm (W)<br>67mm (H)<br>468g |
LCD
{12}------------------------------------------------
| Subject Device | | | Predicate Device | | |
|----------------|-------------------------------|---------------------------------------------|------------------|-------------------------------|---------------------------------------------|
| Model<br>Name | Appearance Picture | Size and<br>Weight | Model Name | Appearance Picture | Size and<br>Weight |
| BA-812X | Image: Blood pressure monitor | 113.5mm (L)<br>92mm (W)<br>48mm (H)<br>418g | BA-812 | Image: Blood pressure monitor | 113.5mm (L)<br>92mm (W)<br>48mm (H)<br>418g |
| BA-813X | Image: Blood pressure monitor | 115.3mm (L)<br>92mm (W)<br>48mm (H)<br>412g | BA-813 | Image: Blood pressure monitor | 115.3mm (L)<br>92mm (W)<br>48mm (H)<br>412g |
{13}------------------------------------------------
| Subject Device | | | Predicate Device | | |
|----------------|-----------------------------------------------------------------------------------------------------------------------|-------------------------------------------|------------------|--------------------|--------------------|
| Model<br>Name | Appearance Picture | Size and<br>Weight | Model Name | Appearance Picture | Size and<br>Weight |
| BA-821X | Image: Blood pressure monitor with LCD screen showing readings, with labels for 'Set', 'Power', and 'Memory' buttons. | 150mm (L)<br>100mm(W)<br>70mm (H)<br>610g | | | |
| BA-822X | Image: Blood pressure monitor with LCD screen, with labels for 'LCD', 'Memory', and 'Power' buttons. | 153mm (L)<br>90mm(W)<br>58mm (H)<br>380g | | | |
{14}------------------------------------------------
| Subject Device | | | Predicate Device | | |
|----------------|---------------------------------------------------------------------------------------------------------------------|-------------------------------------------|------------------|--------------------|--------------------|
| Model<br>Name | Appearance Picture | Size and<br>Weight | Model Name | Appearance Picture | Size and<br>Weight |
| BA-823X | Image: Blood pressure monitor with arm cuff. The display reads 120/90. Labels point to LCD, Set, Power, and Memory. | 140mm (L)<br>95mm(W)<br>70mm (H)<br>420g | | | |
| BA-826X | Image: Blood pressure monitor with arm cuff. The display reads 120/80. Labels point to LCD, Set, Memory, and Power. | 140mm (L)<br>115mm(W)<br>60mm (H)<br>500g | | | |
{15}------------------------------------------------
### 8. Conclusion
The subject device AGE Automatic Upper Arm Blood Pressure Monitor has all features of the predicate device. The few differences do not affect the safety and effectiveness of the subject device. Thus, the subject device is substantially equivalent to the predicate device.
# 9. Summary Prepared Date 26 April 2016
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.