K153547 · Hitachi Medical Systems America, Inc. · LNH · Mar 31, 2016 · Radiology
Device Facts
Record ID
K153547
Device Name
ECHELON Oval V 5.1 MRI System
Applicant
Hitachi Medical Systems America, Inc.
Product Code
LNH · Radiology
Decision Date
Mar 31, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The ECHELON Oval MRI System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (TI), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
Device Story
ECHELON Oval V5.1 is a 1.5 Tesla superconducting MRI system. It uses RF excitation and magnetic field gradients to manipulate hydrogen nuclei (protons) in the body, measuring relaxation times (T1, T2) and proton density to reconstruct cross-sectional images. Operated by trained clinicians in a clinical setting, the system provides anatomical and physiological information for diagnostic interpretation. New features in V5.1 include 32-channel RF receiver support, upgraded Xeon 3.5GHz CPU, FARCON reconstruction hardware, and 'Soft Sound' pulse sequences to reduce acoustic noise. The device also supports ASL perfusion (pCASL), NCC function for coil reconstruction, and 'Vivid' image processing. These enhancements improve image quality and workflow without altering the fundamental MRI principle of operation. The output is used by physicians to aid in clinical diagnosis.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including acoustic noise measurements for the 'Soft Sound' feature and verification of compliance with NEMA MS 1-5, MS 8, and IEC 60601 series standards.
Indicated for diagnostic imaging of the head, body, spine, and extremities using T1, T2, proton density weighted imaging, diffusion weighted imaging, MR angiography, and spectroscopy. Applicable for whole-body imaging.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 31, 2016
Hitachi Medical Systems America, Inc. % Mr. Doug Thistlethwaite Manager of Regulatory Affairs 1959 Summit Commerce Park TWINSBURG OH 44087
Re: K153547
Trade/Device Name: ECHELON Oval V 5.1 MRI System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: March 2, 2016 Received: March 3, 2016
Dear Mr. Thistlethwaite:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Robert Ocks
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
4-1
### 510(k) Number (if known) K153547
Device Name ECHELON OVAL V5.1 MRI system
#### Indications for Use (Describe)
The ECHELON Oval MRI System is an imaging device, and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (TI), spin-spin relaxation time (T2), and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
| Anatomical Region: | Head, Body, Spine, Extremities |
|--------------------|-----------------------------------------|
| Nucleus excited: | Proton |
| Diagnostic uses: | T1, T2, proton density weighted imaging |
| | Diffusion weighted imaging |
| | MR Angiography |
| | Image processing |
| | Spectroscopy |
| | Whole Body |
Type of Use (Select one or both, as applicable)
[X] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5
510(k) Summary
{4}------------------------------------------------
| Submitter: | Hitachi Medical Systems America, Inc.<br>1959 Summit Commerce Park<br>Twinsburg, Ohio 44087-2371 |
|-------------------------|--------------------------------------------------------------------------------------------------|
| Contact: | Douglas J. Thistlethwaite |
| Telephone number: | 330-425-1313 |
| Telephone number: | 330-963-0749 |
| E-mail: | thistlethwaited@hitachimed.com |
| Date: | November 30, 2015 |
| Device Name | |
| Regulation Number: | 892.1000 |
| Regulation Name: | Magnetic resonance diagnostic device |
| Product Code | LNH |
| Class | 2 |
| Panel | Radiology |
| Trade/Proprietary Name: | ECHELON Oval V5.1 MRI System |
# Submitter Information
# Device Intended Use
The ECHELON Oval V5.1 System is an imaging device and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR system produces transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. The images produced by the MR system reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in diagnosis determination.
### Anatomical Region: Head, Body, Spine, Extremities
Nucleus excited: Proton
Diagnostic uses:
- T1, T2, proton density weighted imaging
- Diffusion weighted imaging ●
- MR Angiography ●
- Image processing
- Spectroscopy
- . Whole Body
# Device Description
### Function
The ECHELON OVAL V5.1 is a Magnetic Resonance Imaging System that utilizes a 1.5 Tesla superconducting magnet in a gantry design. The design was based on the ECHELON MRI system. The ECHELON OVAL has been designed to enhance clinical utility as compared to the ECHELON by taking advantage of open architecture.
{5}------------------------------------------------
## Scientific Concepts
Magnetic Resonance imaging (MRI) is based on the fact that certain atomic nuclei have electromagnetic properties that cause them to act as small spinning bar magnets. The most ubiquitous of these nuclei is hydrogen, which makes it the primary nuclei currently used in maqnetic resonance imaging. When placed in a static maqnetic field, these nuclei assume a net orientation or alignment with the magnetic field, referred to as a net magnetization vector. The introduction of a short burst of radiofrequency (RF) excitation of a wavelength specific to the magnetic field strength and to the atomic nuclei under consideration can cause a re-orientation of the net magnetization vector. When the RF excitation is removed, the protons relax and return to their original vector. The rate of relaxation is exponential and varies with the character of the proton and its adjacent molecular environment. This re-orientation process is characterized by two exponential relaxation times, called T1 and T2. A RF emission or echo that can be measured accompanies these relaxation events.
The emissions are used to develop a representation of the relaxation events in a three dimensional matrix. Spatial localization is encoded into the echoes by varying the RF excitation, applying appropriate magnetic field gradients in the x, y, and z directions, and changing the direction and strength of these gradients. Images depicting the spatial distribution of the NMR characteristics can be reconstructed by using image processing techniques similar to those used in computed tomography.
### Physical and Performance Characteristics
MRI is capable of producing high quality anatomical images without the associated risks of ionizing radiation. The biological properties that contribute to MR image contrast are different from those responsible for x-ray image contrast. In MR imaging, difference in proton density, blood flow, and T1 and T2 relaxation times can all contribute to image contrast. By varying the pulse sequence characteristics, the resulting images can emphasize T1, T2, proton density, or the molecular diffusion of water or other proton containing molecules. And MR system has the Function of measuring spectroscopy.
### Performance Evaluation
There is no change from the previous predicate device that would affect performance.
The ECHELON OVAL V5.1 MRL is equivalent to the ECHELON OVAL V5.0 MRJ (K151015) with the following exceptions:
- 32ch in RF receiver channel is added to product specification.
- Xeon 3.5GHz in CPU platform is added to product specification.
- Reconstruction configuration hardware is changed to from the TIP to the FARCON. ●
- Coil mode of Spine coil is added for over 16ch receiver channels.
- The Soft Sound function in pulse sequences is added to functional specification.
{6}------------------------------------------------
# Device Technological Characteristics
The control and image processing hardware and the base elements of the system software are identical to the predicate device. The ECHELON OVAL includes V5.1 software is substantially equivalent to the ECHELON OVAL V5.0 (K151015). See tables below.
The technological characteristics in regards to hardware of the ECHELON Oval MRI System and the predicate are listed in Table 1.
| | ITEM | ECHELON OVAL V5.0 (K151015)<br>PREDICATE | ECHELON OVAL V5.1 | DIFFERENCE<br>ANALYSIS |
|----------------------|----------------------------------------------------------|------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|------------------------|
| System | Standards Met | NEMA: MS 1, MS 2, MS 3, MS 4,<br>MS 5, MS 8,<br>IEC: 60601-1, 60601-1-2, 60601-2-<br>33, 62304 | NEMA: MS 1, MS 2, MS 3, MS 4,<br>MS 5, MS 8,<br>IEC: 60601-1, 60601-1-2, 60601-2-<br>33, 62304 | No |
| Magnet and<br>Gantry | Type and Field<br>Strength<br>Resonant Frequency | Super-conducting magnet,<br>horizontal bore, 1.5 Tesla<br>63.86 MHz | Super-conducting magnet,<br>horizontal bore, 1.5 Tesla<br>63.86MHz | No<br>No |
| Gradient<br>System | Gradient Strength<br>Slew Rate<br>Rise Time | 34mT/m<br>150 T/m/sec<br>227µsec to 34mT/m | 34mT/m<br>150 T/m/sec<br>227µsec to 34mT/m | No<br>No<br>No |
| | Audible Noise (MCAN)<br>Ambient<br>Lpeak<br>Leq | 58 dBA<br>125 dBA<br>117 dB | 58 dBA<br>125 dBA<br>117 dBA | No<br>No<br>No |
| RF System | Transmitter channels<br>Peak Envelop Power<br>Duty Cycle | 2<br>40 kW<br>100% (Gating max), 12.5% at full<br>power | 2<br>40 kW<br>100% (Gating max), 12.5% at full<br>power | No<br>No<br>No |
| | RF receiver channel | 16 | 16, 32 | See Table 2 |
#### Table 1 Comparison: Hardware
The hardware differences from the ECHELON Oval MRI System to the predicate device are analyzed in Table 2.
#### Table 2 Hardware Comparison Analysis
| FDA<br>Requirements | Analyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does not<br>constitute a new intended use; and any differences in technological characteristics are accompanied by information that<br>demonstrates the device is as safe and effective as the predicate and do not raise different questions of safety and<br>effectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (see<br>section 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)). | | | |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------------|----------------------------------|
| Device<br>Modification<br>Summary | 32ch in RF receiver channel is added to product specification.<br>Xeon 3.5GHz in CPU platform is added to product specification.<br>Reconstruction configuration hardware is changed to from the TIP and 1 PC to the FARCON and 2 PCs. | | | |
| Significant<br>Changes | □ Manufacturing Process | □ Labeling | □ Technology | □ Performance |
| | □ Engineering | □ Materials | □ Others | ☑ None (See rationale statement) |
| HITACHI<br>Rationale<br>Statement | Modified specification doesn't constitute a new intended use. There are no significant changes in technological characteristics.<br>For safety, gradient system and RF system is controlled according to same regulation as ECHELON Oval V5.0 (K151015). So, safety and effectively of the device are same as ECHELON Oval V5.0 (K151015) | | | |
{7}------------------------------------------------
The technological characteristics in regards to coils of the ECHELON Oval MRI System and the predicate are listed in Table 3.
| Table 3 Comparison: RF Coils | | | | |
|------------------------------|----------------|---------------------------------------------------------------|---------------------------------------------------------------|------------------------|
| ITEM | | ECHELON OVAL V5.0 (K151015)<br>PREDICATE | ECHELON OVAL | DIFFERENCE<br>ANALYSIS |
| RF Coils | Transmit Coil | T/R Body | T/R Body | No |
| | Receiver Coils | WIT Posterior Head/Neck coil, WIT<br>Anterior Head attachment | WIT Posterior Head/Neck coil,<br>WIT Anterior Head attachment | No |
| | | WIT Torso coil | WIT Torso coil | No |
| | | Extremity coil (Knee) | Extremity coil (Knee) | No |
| | | WIT Anterior Neck attachment | WIT Anterior Neck attachment | No |
| | | Hand/Wrist coil | Hand/Wrist coil | No |
| | | WIT Anterior NV attachment | WIT Anterior NV attachment | No |
| | | Breast | Breast | No |
| | | MP coil 140A, B | MP coil 140A, B | No |
| | | Shoulder | Shoulder | No |
| | | WIT Spine coil 12 | WIT Spine coil 12 | No |
| | | WIT Spine coil 8 | WIT Spine coil 8 | No |
| | | Foot/Ankle | Foot/Ankle | No |
| | | Flexible Extremity (Long Bone) | Flexible Extremity (Long Bone) | No |
| | | WIT Cardiac | WIT Cardiac | No |
| | | PV | PV | No |
The coil differences from the ECHELON Oval MRI System to the predicate device are analyzed
| | | Table 4 Coil Comparison Analysis | | |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|--------------|----------------------------------|
| FDA<br>Requirements | Analyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does not constitute a new intended use; and any differences in technological characteristics are accompanied by information that demonstrates the device is as safe and effective as the predicate and do not raise different questions of safety and effectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (see section 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)). | | | |
| Device<br>Modification<br>Summary | The performance and technological characteristics of the coils are the same as ECHELON Oval V5.0 (K151015). | | | |
| Significant<br>Changes | ☐ Manufacturing Process | ☐ Labeling | ☐ Technology | ☐ Performance |
| | ☐ Engineering | ☐ Materials | ☐ Others | ☑ None (See rationale statement) |
{8}------------------------------------------------
The technological characteristics in regards to changes in functionality of the ECHELON Oval MRI System as compared to the predicate are listed in Table 5.
| ITEM | DIFFERENCES | ANALYSIS |
|-----------------------|-----------------------------------------------------------------------|-------------|
| Operating System | None | No |
| CPU Platform | Core i3 3.5GHz in CPU platform is also available on ECHELON Oval V5.1 | See Table 6 |
| Application Software | Going from V5.0A to V5.1A | See Table 6 |
| Scan Tasks | None | No |
| 2D Processing Tasks | Vivid Image is available | See Table 6 |
| 3D Processing Tasks | None | No |
| Analysis Tasks | None | No |
| Maintenance Tasks | None | No |
| Viewport Tools | None | No |
| Film, Archive Tools | None | No |
| Network Tools | None | No |
| Protocol Enhancements | ASL perfusion (pCASL) is available, NCC function is available. | See Table 6 |
| Pulse Sequences | 2D Soft FSE, 2D Soft FIR, 2D Soft SE, 3D Soft TOF is available | See Table 6 |
Table 5 Comparison: Functionality
The functionality differences from the ECHELON Oval MRI System to the predicate device are analyzed in
| FDA<br>Requirements | Analyze why any differences between the subject device and predicate(s) do not render the device NSE (e.g., does not<br>constitute a new intended use; and any differences in technological characteristics are accompanied by information that<br>demonstrates the device is as safe and effective as the predicate and do not raise different questions of safety and<br>effectiveness than the predicate ), affect safety or effectiveness, or raise different questions of safety and effectiveness (see<br>section 513(i)(1)(A) of the FD&C Act and 21 CFR 807.87(f)). | | | | |
|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------------|----------------------------------|--|
| Device<br>Modification<br>Summary | • Soft FSE, Soft FIR, Soft SE and Soft TOF, which are used for the Protocol Enhancements of Soft Sound, in pulse<br>sequences category are added to functional specification.<br>• Xeon 3.5GHz in CPU platform is added to product specification.<br>• ASL Perfusion (pCASL) in the protocol enhancement is added to product specification.<br>• VIVID image which enhances image quality in the 2D processing tasks is added to product specification.<br>• NCC function for improvement of MAC coil reconstruction image quality in the protocol enhancements is added to<br>product specification. | | | | |
| Significant<br>Changes | ☐ Manufacturing Process | ☐ Labeling | ☐ Technology | ☐ Performance | |
| | ☐ Engineering | ☐ Materials | ☐ Others | ☑ None (See rationale statement) | |
| HITACHI<br>Rationale<br>Statement | Modified functions do not constitute a new intended use. There are no significant changes in technological characteristics. For<br>safety, pulse sequences are controlled according to same safety limits as ECHELON Oval V5.0 (K151015). So safety and<br>effectivity of the device are equivalent to the ECHELON Oval V5.0 (K151015). | | | | |
#### Table 6 Functionality Comparison Analysis
{9}------------------------------------------------
# Substantial Equivalence
A summary decision was based on analysis of Table 7.
#### Table 7 Rationale Analysis: ECHELON Oval V5.1 MRI vs. Predicate
| ITEM | Overall Rationale Analysis |
|---------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Hardware | Different specifications do not constitute a new intended use. There are no significant changes in technological<br>characteristics, safety and effectiveness. |
| Coils | Modified functions don't constitute a new intended use. There are no significant changes in technological characteristics,<br>safety and effectiveness. |
| Functionality | Enhanced features do not constitute a new intended use. There are no significant changes in technological characteristics,<br>safety and effectiveness. The feature set of the device is generally equivalent to the Predicate. |
Therefore, based on a thorough analysis and comparison of the functions, scientific concepts, physical and performance characteristics, performance comparison and technological characteristics, the proposed ECHELON Oval MRI is considered substantially equivalent to the currently marketed predicate device (ECHELON Oval V5.0 MRI System (K151015)) in terms of design features, fundamental scientific technology, indications for use, and safety and effectiveness.
# Summary of Non-Clinical Testing
The revisions to the ECHELON Oval V5.1 MRI System software will have no effect on the standards tests which were conducted on the ECHELON Oval V5.0 MRI System (K151015) and included in the original submission.
Therefore, ECHELON Oval MRI System is in conformance with the applicable parts of the following standards:
- . NEMA MS 1-2008, Determination of Signal-to-noise Ratio (SNR) in Diagnostic Magnetic Resonance Images
- . NEMA MS 2-2008, Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images
- NEMA MS 3-2008, Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
- . NEMA MS 4-2010, Acoustic Noise Measurement Procedure for Diagnostic Magnetic Resonance Imaging Devices
- . NEMA MS 5-2010, Determination of Slice Thickness in Diagnostic Resonance Imaging
- . NEMA MS 8-2008, Characterization of the Specific Absorption Rate for Magnetic Resonance Imaging Systems
- . AAMI / ANSI ES60601-1:2005/(R) 2012 and A1:2012, c1:2009/(r) 2012 and a2:2010/(r ) 2012 (consolidated text) medical electrical equipment - part 1: general requirements for basic safety and essential performance (iec 60601-1:2005, mod).
- . IEC 60601-1-2 Edition 3:2007-03, medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility requirements and tests.
- IEC 60601-2-33 Edition 3.1 2013-04, medical electrical equipment - part 2-33: particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnostic.
- . IEC 62304 First edition 2006-05, medical device software - software life cycle processes.
In addition, tests were conducted on the new feature of the ECHELON Oval V5.1 MRI System. they include:
#### ● Soft Sound
Test results confirm the acoustic sound pressure levels of conventional sequences and soft sound sequences.
{10}------------------------------------------------
# Summary of Clinical Testing
There is no change from the previous predicate device that would affect clinical performance.
# Conclusions
It is the opinion of Hitachi Medical Systems America, Inc. the ECHELON Oval V5.1 MRI System is substantially equivalent with respect to hardware, base elements of the software, safety, effectiveness, and functionality to the ECHELON Oval V5.0 MRI System (K151015).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.