Retrospective clinical performance evaluation; Study Period: 2013-2014 influenza season
Patients presenting with flu-like symptoms (archived specimens); Sample Size: 292 evaluable specimens; Number of Sites: 3
FDA-cleared multiplexed nucleic acid amplification test
Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) for Influenza A, Influenza B, and RSV
Indications for Use
The cobas® Influenza A/B & RSV Nucleic Acid Test for Use on the cobas® Liat System (cobas® Liat Influenza A/B & RSV) is an automated multiplex real-time RT-PCR assay for the rapid in vitro qualitative detection and discrimination of influenza A virus, influenza B virus and respiratory syncytial virus (RSV) RNA in nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors. The test is intended for use as an aid in the diagnosis and differentiation of influenza A, influenza B, and RSV in humans and is not intended to detect influenza C. Negative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. Performance characteristics for influenza A were established during the 2013-2014 and the 2014-2015 influenza seasons when influenza A/H3 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary. If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. The cobas® Influenza A/B & RSV Quality Control Kit contains External Controls for use with the cobas® Liat Influenza A/B & RSV assay. External Controls are run during the Add cobas® Liat Influenza A/B & RSV Tube Lot procedure. Additional External Controls should be tested in accordance with local, state, federal and/or accrediting organization requirements as applicable.
Device Story
The cobas Liat Influenza A/B & RSV assay is a rapid, automated, point-of-care diagnostic test. It accepts nasopharyngeal swab (NPS) specimens in viral transport media. The device uses a single-use, disposable cobas Liat Tube containing pre-packed reagents. The cobas Liat System integrates sample purification, nucleic acid extraction, and real-time RT-PCR amplification. The system uses mechanical actuators to compress tube segments, moving the sample through processing steps (lysis, binding, washing, elution, and amplification). An embedded microprocessor controls these processes. The system outputs qualitative results (Detected/Not Detected) for Influenza A, Influenza B, and RSV in approximately 20 minutes. It is intended for use by nurses and laboratory technologists in emergency rooms, urgent care, and outpatient clinics. The output is displayed on an LCD screen or printed, aiding clinicians in rapid diagnosis and patient management decisions, such as determining appropriate antiviral therapy or infection control measures.
Indicated for qualitative detection and discrimination of influenza A, influenza B, and RSV RNA in nasopharyngeal swab specimens from symptomatic patients. For prescription use only.
Regulatory Classification
Identification
A respiratory viral panel multiplex nucleic acid assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify multiple viral nucleic acids extracted from human respiratory specimens or viral culture. The detection and identification of a specific viral nucleic acid from individuals exhibiting signs and symptoms of respiratory infection aids in the diagnosis of respiratory viral infection when used in conjunction with other clinical and laboratory findings. The device is intended for detection and identification of a combination of the following viruses:(1) Influenza A and Influenza B; (2) Influenza A subtype H1 and Influenza A subtype H3; (3) Respiratory Syncytial Virus subtype A and Respiratory Syncytial Virus subtype B; (4) Parainfluenza 1, Parainfluenza 2, and Parainfluenza 3 virus; (5) Human Metapneumovirus; (6) Rhinovirus; and (7) Adenovirus.
Special Controls
*Classification.* Class II (special controls). The special controls are:(1) FDA's guidance document entitled “Class II Special Controls Guidance Document: Respiratory Viral Panel Multiplex Nucleic Acid Assay;”
(2) For a device that detects and identifies Human Metapneumovirus, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Human Metapneumovirus (hMPV) Using Nucleic Acid Assays;” and
(3) For a device that detects and differentiates Influenza A subtype H1 and subtype H3, FDA's guidance document entitled “Class II Special Controls Guidance Document: Testing for Detection and Differentiation of Influenza A Virus Subtypes Using Multiplex Nucleic Acid Assays.” See § 866.1(e) for the availability of these guidance documents.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K153544
B. Purpose for Submission:
New device 510k clearance for the cobas® Influenza A/B & RSV Nucleic Acid Test for Use on the cobas® Liat System (cobas® Liat Influenza A/B & RSV) and the cobas® Influenza A/B & RSV Quality Control Kit
C. Measurand:
Target RNA sequences of a well-conserved region of the matrix gene of influenza A (Inf A), the non-structural protein (NSP) gene of influenza B (Inf B), and the matrix gene of respiratory syncytial virus (RSV).
D. Type of Test:
Multiplexed nucleic acid assay for the qualitative detection of Influenza A, Influenza B and RSV RNA from Nasopharyngeal Swab (NPS) specimens, including nucleic acid isolation and multiplex real-time RT-PCR amplification using the cobas® Liat System.
E. Applicant:
IQuum, Inc. (a wholly-owned subsidiary of Roche Molecular Systems Inc.)
F. Proprietary and Established Names:
1) cobas® Influenza A/B & RSV Nucleic Acid Test for Use on the cobas® Liat System (cobas® Liat Influenza A/B & RSV)
2) cobas® Influenza A/B & RSV Quality Control Kit
G. Regulatory Information:
1. Regulation section:
21 CFR 866.3980, Respiratory Viral Panel Multiplex Nucleic Acid Assay
2. Classification:
Class II
3. Product code(s):
OCC - Respiratory virus panel nucleic acid assay system
OZE - Influenza A and influenza B multiplex nucleic acid assay
OOI - Real time nucleic acid amplification system
JJX - Single (specified) analyte controls (assayed and unassayed)
{1}
4. Panel:
Microbiology (83)
H. Intended Use:
1. Intended use(s):
1) The cobas® Influenza A/B & RSV Nucleic Acid Test for Use on the cobas® Liat System (cobas® Liat Influenza A/B & RSV) is an automated multiplex real-time RT-PCR assay for the rapid in vitro qualitative detection and discrimination of influenza A virus, influenza B virus and respiratory syncytial virus (RSV) RNA in nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors. The test is intended for use as an aid in the diagnosis and differentiation of influenza A, influenza B, and RSV in humans and is not intended to detect influenza C.
Negative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.
Performance characteristics for influenza A were established during the 2013-2014 and the 2014-2015 influenza seasons when influenza A/H3 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.
If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens.
2) The cobas® Influenza A/B & RSV Quality Control Kit contains External Controls for use with the cobas® Liat Influenza A/B & RSV assay. External Controls are run during the Add cobas® Liat Influenza A/B & RSV Tube Lot procedure. Additional External Controls should be tested in accordance with local, state, federal and/or accrediting organization requirements as applicable.
2. Indication(s) for use:
Same as Intended Use
3. Special conditions for use statement(s):
For prescription use only
4. Special instrument requirements: cobas® Liat System
{2}
I. Device Description:
Overview
The cobas® Liat Influenza A/B & RSV Nucleic Acid Test for Use on the cobas® Liat System (cobas® Liat Influenza A/B & RSV) is a rapid, automated in vitro diagnostic test for the qualitative detection of influenza A, influenza B, and RSV RNA in nasopharyngeal swab (NPS) specimens eluted in viral transport media.
The assay targets a well-conserved region of the matrix gene of influenza A (Inf A target), the non-structural protein gene of influenza B (Inf B target), and the matrix gene of RSV (RSV target). An Internal Process Control (IPC) is also included. The IPC is present to control for adequate processing of the target viruses and to monitor the presence of inhibitors in the sample preparation and RT-PCR.
The assay utilizes a single-use disposable cobas® Liat Tube that holds the sample purification and PCR reagents, and hosts the sample preparation and PCR processes. The cobas® Liat Tube uses a flexible tube as a sample vessel. It contains all required unit dose reagents pre-packed in tube segments, separated by peelable seals, in the order of reagent use.
The cobas® Liat System automates and integrates sample purification, nucleic acid amplification, and detection of the target sequence in biological samples. The cobas® Liat System performs all assay steps from clinical sample and reports assay result automatically. During the testing process, multiple sample processing actuators of the cobas® Liat System compress the cobas® Liat Tube to selectively release reagents from tube segments, move the sample from one segment to another, and control reaction volume, temperature, and time to conduct sample preparation, nucleic acid extraction, target enrichment, inhibitor removal, nucleic acid elution, and real-time PCR. An embedded microprocessor controls and coordinates the actions of these sample processors to perform all required assay steps within the closed cobas® Liat Tube.
3
{3}
cobas® Liat Influenza A/B & RSV Nucleic Acid Test for Use on the cobas® Liat System (cobas® Liat Influenza A/B & RSV) Component
# Materials Provided
The cobas® Liat Influenza A/B & RSV Assay Pack (Cat # 07341903190) contains sufficient reagents to process 20 specimens or quality control samples. The pack contains 20 sets of a cobas® Liat Influenza A/B & RSV Tube and a transfer pipette. A Package Insert Barcode Card (Cat # 34-04895) with lot-specific barcode is also included.
# Equipment
cobas® Liat System (Cat # 07341920190) (Analyzer)
# Materials Required but Not Provided
- Acceptable specimen collection kits include Universal Transport Medium (UTM) Swab Collection Kits (BD Cat # 220531 or Copan Cat # 305C), each kit containing a Collection Swab and a tube containing 3 mL of UTM, or equivalent.
- cobas® Liat Influenza A/B & RSV Quality Control Kit, Cat# 07402686190, containing:
- cobas® Liat Influenza A/B & RSV Positive Control (Positive Control), Cat # 20-04515
- cobas® Liat Influenza A/B & RSV Dilution UTM (Dilution UTM), Cat # 20-04514
- Transfer Pipette, Cat # 20-03795
- Control Kit Barcode Card, Cat # 34-04894
- Negative Control Barcode Label, Cat # 34-04535
- Positive Control Barcode Label, Cat # 34-04534
# Quality Control
The cobas® Liat Influenza A/B & RSV assay has three controls: (1) internal process control, (2) positive control, and (3) negative control.
# Internal Process Control
The internal process control (IPC) is comprised of a bacteriophage MS2 that is pre-packed in each cobas® Liat tube. The IPC is designed to be detected at a cycle threshold (Ct) of ~31 on the cobas® Liat Influenza A/B & RSV assay, similar to the Ct of approximately 4x Limit of Detection (LOD) concentration of target viruses. When conducting an assay, IPC is first mixed with a sample and then it goes through all the test processes to monitor both the sample preparation and RT-PCR performance. The IPC RNA is detected in a separate channel by MS2 specific primers and probe. If IPC Ct does not meet the acceptance criteria and Inf A, Inf B, and RSV are not detected, the assay run is called "Invalid" to avoid false negative results due to excessive sample inhibition or system operation outside the normal range.
# Positive Control
Positive control is provided in the cobas® Liat Influenza A/B & RSV assay Quality Control Kit. The positive control is comprised of a pooled sample of inactivated Influenza A, Influenza B, and RSV. The individual inactivated viruses are pooled to make the positive control solution, and then dispensed and "freeze-dried" in unit-dose quantities in individual Positive Control tubes. To use the positive control, an operator transfers a unit dose of Universal Transport Media (UTM) from the Dilution UTM tube into the Positive Control tube using an included transfer pipette to dissolve and mix the dried positive control. The
{4}
entire mixture is then transfer into the cobas® Liat tube, and the cobas® Liat tube is run on a cobas® Liat Analyzer according to the Instructions for Use.
The positive control is required to be run during the “Add Liat Tube Lot” process, in which the cobas® Liat tube lot and end user site procedures are checked. Additional positive control runs may be performed by the end user to confirm the performance of a cobas® Liat Analyzer and a cobas® Liat tube lot through detection of Influenza A, Influenza B and RSV targets, or as required by the end user’s quality control standards.
## Negative Control
Negative control is provided in the cobas® Liat Influenza A/B & RSV assay Quality Control Kit. The negative control is comprised of UTM. The solution is provided in unit dose quantity and labeled as Dilution UTM. To use the negative control, an operator transfers the entire contents of the Dilution UTM tube into the cobas® Liat tube using a transfer pipette and runs the assay following the Instructions for Use.
The negative control is required to be run during the “Add Liat Tube Lot” process, in which potential contamination and end user site procedures are checked. Additional negative control runs may be performed by the end user to check if there is contamination resulting in a false positive result, or as required by the end user’s quality control standards.
## Workflow
To perform the cobas® Liat Influenza A/B & RSV assay, an operator first collects a nasopharyngeal swab and places the swab into UTM. The operator transfers the sample into cobas® Liat Influenza A/B & RSV assay tube using a transfer pipette, and scans the tube barcode to identify the test and the sample barcode to code the sample ID with the assay run on the cobas® Liat System. The cobas® Liat Tube is then inserted into the cobas® Liat System. The system performs all the test steps and outputs interpreted results (e.g., Influenza A Detected, Influenza B Not Detected, RSV Not Detected) in ~20 minutes. A report of the interpreted results can be viewed on the cobas® Liat System’s LCD screen, and printed directly through a USB or network connected printer. No reagent preparation or additional steps are required other than adding the sample to the cobas® Liat Tube. Because all the reagents are contained within the cobas® Liat Tube and no sample or reagent needs to be removed from the tube, cross-contamination between samples is minimized.
## Results Interpretation
The results are interpreted by the cobas® Liat System software from measured fluorescent signals and real time curve recognition algorithm. No manual data interpretation or reviewing PCR curves is required. The cobas® Liat Analyzer automatically interprets the results from measured fluorescent signals. Embedded algorithms determine the PCR cycle threshold (Ct), and evaluate the Ct against a valid range (i.e., minimum value to assay cutoff) to generate a positive or negative PCR result. Additionally, pattern recognition algorithms inspect the baseline, logarithmic, and plateau phases of the PCR curve to determine if the curve is typical/normal or abnormal. If a PCR curve is determined to be abnormal, the result may be called “Invalid” or “Indeterminate.” This pattern recognition provides additional error checking capabilities to ensure the quality of the results. In cases where one of Inf A, Inf B, and RSV is detected, but the PCR curve pattern is abnormal, the result is reported as “Invalid” to avoid false positive results. When at least two of Inf A, Inf B and RSV are detected but the PCR curve pattern for one target is abnormal, the target(s) with the normal curve pattern is called “Detected” while the target with the abnormal curve is called “Indeterminate”.
All possible final test results and the respective follow-up actions are described in Table 1 below:
5
{5}
Table 1: Interpretation of Results
| Result Report | | Interpretation |
| --- | --- | --- |
| Influenza A | Influenza A Not Detected | Negative test for Influenza A (no Influenza A RNA detected) |
| | Influenza A Detected | Positive test for Influenza A (Influenza A RNA present) |
| | Influenza A Indeterminate Repeat Assay | Presence or absence of Influenza A cannot be determined. Repeat assay with same sample or, if possible, new sample. |
| Influenza B | Influenza B Not Detected | Negative test for Influenza B (no Influenza B RNA detected) |
| | Influenza B Detected | Positive test for Influenza B (Influenza B RNA present) |
| | Influenza B Indeterminate Repeat Assay | Presence or absence of Influenza B cannot be determined. Repeat assay with same sample or, if possible, new sample. |
| RSV | RSV Not Detected | Negative test for RSV (no RSV RNA detected) |
| | RSV Detected | Positive test for RSV (RSV RNA present) |
| | RSV Indeterminate Repeat Assay | Presence or absence of RSV cannot be determined. Repeat assay with same sample or, if possible, new sample. |
| Assay Invalid. Repeat Assay | | Presence or absence of Influenza A, Influenza B, and RSV cannot be determined. Repeat assay with same sample or, if possible, new sample. |
| [Error]. Assay Aborted | | |
If the test result is "Indeterminate" or "Invalid", the assay should be repeated with the same patient specimen, or if possible, with a newly collected specimen. Specimens that have repeat "Indeterminate" or "Invalid" results should be sent to a laboratory for confirmatory testing.
If an assay is aborted due to run error, or if an assay is aborted by user, the test should be repeated with the same sample. Or, if possible, the test should be repeated with a new sample. Manufacturer Service Representative should be contacted if repeat "Errors" are reported.
Dual infections of Influenza A and Influenza B are rare. If the test result is "Influenza A Detected" and "Influenza B Detected", the assay should be repeated with the same patient specimen, or if possible, with a newly collected specimen. Specimens that have repeat "Influenza A Detected" and "Influenza B Detected" results should be sent to a laboratory for confirmatory testing.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Hologic Prodesse ProFlu+TM
2. Predicate 510(k) number:
K132129
{6}
3. Comparison with predicate:
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | cobas® Influenza A/B & RSV Nucleic Acid Test for Use on the cobas® Liat System (K153544) | Hologic Prodesse ProFlu+TM (K132129) |
| Assay Analyte | Influenza A, Influenza B, and RSV | Same |
| Intended Use | The cobas® Influenza A/B & RSV Nucleic Acid Test for Use on the cobas® Liat System (cobas® Liat Influenza A/B & RSV) is an automated multiplex real-time RT-PCR assay for the rapid in vitro qualitative detection and discrimination of influenza A virus, influenza B virus and respiratory syncytial virus (RSV) RNA in nasopharyngeal swab specimens from patients with signs and symptoms of respiratory infection in conjunction with clinical and epidemiological risk factors. The test is intended for use as an aid in the diagnosis and differentiation of influenza A, influenza B, and RSV in humans and is not intended to detect influenza C.Negative results do not preclude influenza virus or RSV infection and should not be used as the sole basis for treatment or other patient management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease.Performance characteristics for influenza A were established during the 2013-2014 and the 2014-2015 influenza seasons when influenza A/H3 and A/H1N1 pandemic were the predominant influenza A viruses in circulation. When other influenza A viruses are emerging, performance characteristics may vary.If infection with a novel influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. | The ProFlu+TM Assay is a multiplex Real-Time PCR (RT-PCR) in vitro diagnostic test for the rapid and qualitative detection and discrimination of Influenza A Virus, Influenza B Virus, and Respiratory Syncytial Virus (RSV) nucleic acids isolated and purified from nasopharyngeal (NP) swab specimens obtained from symptomatic patients. This test is intended for use to aid in the differential diagnosis of Influenza A, Influenza B and RSV viral infections in humans and is not intended to detect Influenza C.Negative results do not preclude influenza or RSV virus infection and should not be used as the sole basis for treatment or other management decisions. Conversely, positive results do not rule-out bacterial infection or co-infection with other viruses. The agent detected may not be the definite cause of disease. The use of additional laboratory testing and clinical presentation must be considered in order to obtain the final diagnosis of respiratory viral infection.Performance characteristics for Influenza A Virus were established when Influenza A/H3 and A/H1 were the predominant Influenza A viruses in circulation (2006 – 2007 respiratory season). Performance characteristics for Influenza A were confirmed when Influenza A/H1, Influenza A/H3, and Influenza A/2009 H1N1 were the predominant Influenza A viruses in circulation (2008 and 2009). When other Influenza A viruses are emerging, performance characteristics may vary.If infection with a novel Influenza A virus is suspected based on current clinical and epidemiological screening criteria recommended by public health authorities, specimens should be collected with appropriate infection control precautions for novel virulent Influenza viruses and sent to state or local health department for testing. Viral culture should not be attempted in these cases unless a BSL 3+ facility is available to receive and culture specimens. |
| Sample Type | Nasopharyngeal swab eluted in VTM | Same |
| Internal Control | Yes for sample preparation/nucleic acid extraction and RT-PCR performance using | Yes for RT-PCR performance only using in vitro transcribed RNA. Extraction control is |
| | in situ. The sample preparation was then performed using the same method as in the previous study. The sample preparation was then performed using the same method as in the previous study. The sample preparation was then performed using the same method as in the previous study. | in situ. The sample preparation was then performed using the same method as in the previous study. The sample preparation was then performed using the same method as in the previous study. |
{7}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | cobas® Influenza A/B & RSV Nucleic Acid Test for Use on the cobas® Liat System (K153544) | Hologic Prodesse ProFlu+™ (K132129) |
| | encapsulated RNA. | recommended but not provided. |
| Influenza A Viral Target | Well conserved region of the matrix gene | Same |
| Influenza B Viral Target | Well conserved region of the non-structure protein (NSP) gene | Same |
| RSV Viral Target | Well conserved region of the matrix gene | Polymerase gene |
| Extraction Method | Integrated silica-magnetic bead-based nucleic acid extraction | Silica-magnetic bead-based nucleic acid extraction using Roche MagNA Pure LC System, or bioMérieux NucliSENS easyMag |
| Assay Method | RT-PCR | Same |
| Detection | Multiplex assay using different reporter dyes for each target | Same |
| Assay Result | Qualitative | Same |
| Assay Instrument | cobas® Liat System | Roche MagNA Pure LC System, or bioMérieux NucliSENS easyMag
Cepheid SmartCycler II Real Time Instrument |
| Self-contained System | Yes | No |
| All Assay Reagents Contained in Disposable | Yes | No |
| Sample Volume Detection | Yes | No |
| Automated Assay | Yes | No |
| Error Diagnostic System | Yes | No |
| PCR Curve Pattern Recognition | Yes | No |
| Automated Result Interpretation | Yes | No |
| User | Hospital nurse and CLIA moderate complexity laboratory technologist.
Users in various CLIA waived settings, such as nurses and medical assistants at emergency rooms, urgent care clinics, and outpatient clinics, including physician’s offices. | CLIA high complexity laboratory technologist. |
{8}
| Similarities and Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Test Availability | cobas® Influenza A/B & RSV Nucleic Acid
Test for Use on the cobas® Liat System
(K153544) | Hologic Prodesse ProFlu+™ (K132129) |
| | Random access, on-demand test | Batch processing |
| Time to Result | ~20 minutes | ≥4 hours |
## K. Standard/Guidance Document Referenced (if applicable):
Guidance for Industry and FDA Staff – Establishing the Performance Characteristics of *In Vitro Diagnostic Devices for the Detection or Detection and Differentiation of Influenza Viruses*. Document issued on July 15, 2011. Docket number FDA-2008-D-0095.
## L. Test Principle:
The cobas® Liat Influenza A/B & RSV assay uses well established nucleic acid test chemistry and assay protocol for viral RNA detection. The sample preparation methodology is based on chaotropic agent-based lysis and magnetic particle nucleic acid purification. First, the NPS sample is diluted and mixed with an internal process control (IPC) (encapsulated RNA). Chaotropic and proteolytic reagent then disrupts the three dimensional structure in macromolecules such as proteins and nucleic acids in the sample, and denatures them. Second, nucleic acids are isolated from lysates through binding to the surface of silica magnetic beads in the presence of a chaotropic salt, which removes water from hydrated molecules in solution. Third, the beads are separated from the lysates using a magnetic field, and the lysate removed. Fourth, the beads with captured nucleic acids are washed to remove possible inhibitors in the sample. Finally, the captured nucleic acids are eluted under low-salt conditions into a small volume of elution buffer.
Eluted viral RNA is first transcribed into cDNA using reverse transcriptase. This cDNA then undergoes a polymerase chain reaction (PCR) where the reaction mixture is repeatedly heated to denature the nucleic acid, and cooled to allow annealing of primers and extension of annealed primers by DNA polymerase to logarithmically amplify the specific region of the cDNA. Dual-labeled fluorogenic hydrolysis (TaqMan) probes anneal to the specific target sequences located between the binding regions of forward and reverse primers. During the extension phase of the PCR cycle, the 5′ nuclease activity of the polymerase degrades the probes, causing the reporter dyes to separate from the quenchers, thus generating fluorescent signals. Fluorescence intensities are monitored at each PCR cycle. When the fluorescence intensities exceed predetermined thresholds, cycle threshold (Ct) values are generated for the specific analyte corresponding to the fluorescence channel.
All these sample preparation/nucleic acids extractions and real-time PCR amplification processes are conducted in a closed cobas® Liat Tube in ~20 minutes.
## M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
{9}
10
# Reproducibility Study
A reproducibility study was conducted assessing the total variability of the cobas® Liat Influenza A/B & RSV assay across operators, study sites, testing days, cobas® Liat Analyzers, and cobas® Liat assay tube lots. The cobas® Liat Influenza A/B & RSV assay was evaluated at three CLIA waived sites. Two (2) operators at each of the three sites tested a 10-member reproducibility panel in triplicate on five different days, for a total of ~900 runs (10 panel members × 3 replicates × 2 operators × 5 days × 3 sites). The reproducibility panel contained a true negative; and a high negative, a low positive and a moderate positive of each of Influenza A, Influenza B and RSV.
The reproducibility panel samples were constructed by spiking viruses of known titer into negative NPS in UTM sample matrix. A/Brisbane/59/07, B/Malaysia/2506/04, and RSV-A 2006 Isolate with known $\mathrm{TCID}_{50} / \mathrm{mL}$ titers were obtained from a vendor. The moderate positive, low positive, and high negative concentrations used for each of the strains corresponded to $C_{100}$ (4xLOD), $C_{95}$ (1xLOD), and $C_5$ (as determined by Probit analysis using LOD data), respectively. The true negative sample comprised negative NPS sample matrix in UTM. For a given virus, the expected result for the true negative and the high negative panel member is "Not Detected", while the expected result for the low positive and moderate positive panel member is "Detected".
Three (3) CLIA waived sites and six operators (two operators per site) participated in this reproducibility study. The six operators consisted of two nurses, three medical assistants, and one administrative assistant with no formal medical laboratory training. All operators had limited or no training or hands-on experience in conducting laboratory testing.
The six operators at the three sites tested the members of the reproducibility panel in triplicate on five non-consecutive days. Three (3) cobas® Liat Analyzers were used at each site for a total of nine cobas® Liat Analyzers. Each site also used a different lot of cobas® Liat Influenza A/B & RSV assay tubes for a total of three lots.
Percent agreement with expected result, mean Ct, and Ct %CV for each site are show in Table 2 to Table 4 below:
{10}
Table 2: Influenza A Reproducibility Study Results
| | Site 1 | | | Site 2 | | | Site 3 | | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Agreement | Ave. Ct | Ct % CV | Agreement | Ave. Ct | Ct % CV | Agreement | Ave. Ct | Ct % CV | Agreement | 95% CI |
| Negative | 30 / 30 | - | - | 31 / 31 | - | - | 30 / 30 | - | - | 91 / 91 (100.0%) | 96.0% - 100.0% |
| Flu A High Negative (C5) | 29 / 30 | 37.0 | - | 30 / 30 | - | - | 29 / 30 | 35.7 | - | 88 / 90 (97.8%) | 92.3% - 99.4% |
| Flu A Low Positive (C95) | 30 / 30 | 32.7 | 2.9% | 30 / 30 | 32.1 | 1.6% | 30 / 30 | 32.3 | 1.6% | 90 / 90 (100.0%) | 95.9% - 100.0% |
| Flu A Moderate Positive (C100) | 30 / 30 | 30.4 | 1.0% | 30 / 30 | 30.0 | 1.2% | 30 / 30 | 30.1 | 0.9% | 90 / 90 (100.0%) | 95.9% - 100.0% |
| Flu B High Negative (C5) | 30 / 30 | - | - | 31 / 31 | - | - | 30 / 30 | - | - | 91 / 91 (100.0%) | 96.0% - 100.0% |
| Flu B Low Positive (C95) | 30 / 30 | - | - | 30 / 30 | - | - | 29 / 291 | - | - | 89 / 89 (100.0%) | 95.9% - 100.0% |
| Flu B Moderate Positive (C100) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| RSV High Negative (C5) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| RSV Low Positive (C95) | 30 / 30 | - | - | 31 / 31 | - | - | 30 / 30 | - | - | 91 / 91 (100.0%) | 96.0% - 100.0% |
| RSV Moderate Positive (C100) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| Total Agreement | 299 / 300 (99.7%) | | | 303 / 303 (100.0%) | | | 298 / 299 (99.7%) | | | 900 / 902 (99.8%) | 99.2% - 99.9% |
30 Flu B Low Positive $(\mathrm{C}_{95})$ replicates yielded an "Assay Invalid. Repeat Assay." result, and was not repeated.
Table 3: Influenza B Reproducibility Study Results
| | Site 1 | | | Site 2 | | | Site 3 | | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Agreement | Ave. Ct | Ct % CV | Agreement | Ave. Ct | Ct % CV | Agreement | Ave. Ct | Ct % CV | Agreement | 95% CI |
| Negative | 30 / 30 | - | - | 31 / 31 | - | - | 30 / 30 | - | - | 91 / 91 (100.0%) | 96.0% - 100.0% |
| Flu A High Negative (C5) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| Flu A Low Positive (C95) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| Flu A Moderate Positive (C100) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| Flu B High Negative (C5) | 29 / 30 | 35.1 | - | 31 / 31 | - | - | 30 / 30 | - | - | 90 / 91 (98.9%) | 94.0% - 99.8% |
| Flu B Low Positive (C95) | 30 / 30 | 31.9 | 1.8% | 30 / 30 | 31.6 | 1.4% | 29 / 291 | 31.6 | 1.5% | 89 / 89 (100.0%) | 95.9% - 100.0% |
| Flu B Moderate Positive (C100) | 30 / 30 | 30.8 | 1.3% | 30 / 30 | 30.4 | 1.4% | 30 / 30 | 30.5 | 1.3% | 90 / 90 (100.0%) | 95.9% - 100.0% |
| RSV High Negative (C5) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| RSV Low Positive (C95) | 30 / 30 | - | - | 31 / 31 | - | - | 30 / 30 | - | - | 91 / 91 (100.0%) | 96.0% - 100.0% |
| RSV Moderate Positive (C100) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| Total Agreement | 299 / 300 (99.7%) | | | 303 / 303 (100.0%) | | | 299 / 299 (100.0%) | | | 901 / 902 (99.9%) | 99.4% - 100.0% |
30 Flu B Low Positive $(\mathrm{C}_{95})$ replicates yielded an "Assay Invalid. Repeat Assay." result, and was not repeated.
{11}
Table 4: RSV Reproducibility Study Results
| | Site 1 | | | Site 2 | | | Site 3 | | | Total | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Sample | Agreement | Ave. Ct | Ct % CV | Agreement | Ave. Ct | Ct % CV | Agreement | Ave. Ct | Ct % CV | Agreement | 95% CI |
| Negative | 30 / 30 | - | - | 31 / 31 | - | - | 30 / 30 | - | - | 91 / 91 (100.0%) | 96.0% - 100.0% |
| Flu A High Negative (C5) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| Flu A Low Positive (C95) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| Flu A Moderate Positive (C100) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| Flu B High Negative (C5) | 30 / 30 | - | - | 31 / 31 | - | - | 30 / 30 | - | - | 91 / 91 (100.0%) | 96.0% - 100.0% |
| Flu B Low Positive (C95) | 30 / 30 | - | - | 30 / 30 | - | - | 29 / 291 | - | - | 89 / 89 (100.0%) | 95.9% - 100.0% |
| Flu B Moderate Positive (C100) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| RSV High Negative (C5) | 30 / 30 | - | - | 30 / 30 | - | - | 30 / 30 | - | - | 90 / 90 (100.0%) | 95.9% - 100.0% |
| RSV Low Positive (C95) | 29 / 30 | 33.0 | 3.7% | 31 / 31 | 32.8 | 3.4% | 30 / 30 | 32.8 | 2.7% | 90 / 91 (98.9%) | 94.0% - 99.8% |
| RSV Moderate Positive (C100) | 30 / 30 | 30.6 | 2.9% | 30 / 30 | 30.9 | 1.6% | 30 / 30 | 30.5 | 2.5% | 90 / 90 (100.0%) | 95.9% - 100.0% |
| Total Agreement | 299 / 300 (99.7%) | | | 303 / 303 (100.0%) | | | 299 / 299 (100.0%) | | | 901 / 902 (99.9%) | 99.4% - 100.0% |
30 Flu B Low Positive (C95) replicates yielded an "Assay Invalid. Repeat Assay." result, and was not repeated.
b. Linearity/assay reportable range:
Not applicable
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
## Controls
External Positive and Negative Controls
Positive and negative controls are provided in the cobas® Liat Influenza A/B & RSV assay Quality Control Kit. Refer to the "Device Description" section of this summary for details regarding these controls.
External positive and negative controls are required to be run during the Add cobas® Liat Influenza A/B & RSV Tube Lot procedure. Additional External positive and negative controls may be tested in accordance with local, state, federal, and/or accrediting organization requirements as applicable.
During the clinical study testing both prospective and retrospective specimens (during the 2014 to 2015 influenza season), quality control testing using the external positive and negative controls was performed per the product Instructions for Use at all three study sites. A total of 173 external positive controls and 173 external negative controls were tested during the clinical study. Two (2) external positive controls (2/173, 1.2%, 95% CI, 0.3% - 4.1%) and three external negative controls (3/173, 1.7%, 95% CI, 0.6% - 5.0%) failed initially, but succeeded on repeat testing per the product Instructions for Use.
## Specimen Stability Study
Specimen stability was evaluated at temperatures specified in the cobas® Liat Influenza A/B & RSV
{12}
Package Insert. Influenza A, Influenza B and RSV strains at 3x LOD were spiked into negative nasopharyngeal swab in UTM matrix, stored at $-80^{\circ}\mathrm{C}$ and $4^{\circ}\mathrm{C}$ , and tested at different time points.
General recommendations for the storage and transport of nasopharyngeal swab in UTM specimens with regard to nucleic acid-based influenza and RSV testing are to store specimens refrigerated (2- $8^{\circ}\mathrm{C}$ ) for up to 72 hours prior to testing. For long duration transport or storage, specimens should be frozen at $-70^{\circ}\mathrm{C}$ or colder and transported on dry ice (i.e. $-78.5^{\circ}\mathrm{C}$ ) (CDC Influenza Specimen Collection, CS246972; Copan Universal Transport Media System Package Insert, 44A-Rev.07061). The cobas® Liat Influenza A/B & RSV assay Package Insert specifies the same storage and transport conditions. This specimen stability study evaluated the effect of these specimen storage temperatures on influenza and RSV detection by the cobas® Liat Influenza A/B & RSV assay.
Each temperature condition was tested using a sample panel containing a positive sample (3x LOD) and a negative sample. The positive sample contained A/Brisbane/59/2007 at $6.0 \times 10^{-3} \mathrm{TCID}_{50} / \mathrm{mL}$ , B/Malaysia/2506/04 at $1.2 \times 10^{-2} \mathrm{TCID}_{50} / \mathrm{mL}$ , and RSV-A 2006 Isolate at $1.2 \times 10^{0} \mathrm{TCID}_{50} / \mathrm{mL}$ ) spiked into negative NPS in UTM sample matrix at 3x LOD concentration. Influenza and RSV strains used were from the same source as identified in the "Detection Limit" section of this summary. The negative sample contained negative NPS in UTM sample matrix.
Five (5) positive and five negative samples were tested using the cobas® Liat Influenza A/B & RSV assay immediately after spiking. Other samples were stored at $-80^{\circ}\mathrm{C}$ for 30 days; and at $4^{\circ}\mathrm{C}$ for 2, 3, and 7 days. At each time point, another five positive and five negative samples were tested using the cobas® Liat Influenza A/B & RSV assay. Samples stored at $-80^{\circ}\mathrm{C}$ were thawed prior to testing.
A total of 60 runs were performed in this study. Runs were performed on 14 cobas® Liat Analyzers and used 4 lots of cobas® Liat tubes.
Table 5 below shows the results from this study. The cobas® Liat Influenza A/B & RSV assay yielded the expected result for positive ("Influenza A Detected, Influenza B Detected, RSV Detected") and negative ("Influenza A Not Detected, Influenza B Not Detected, RSV Not Detected") samples in all replicates tested at each time point.
Table 5: Specimen Stability Study Results
| Temp. | Time (Days) | Sample | Inf A Ct Ave. | Inf B Ct Ave. | RSV Ct Ave. | IPC Ct Ave. | Inf A Detected | Inf A Not Detected | Inf B Detected | Inf B Not Detected | RSV Detected | RSV Not Detected |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| -80°C | 0 | Positive | 31.7 | 29.2 | 31.2 | 30.4 | 5/5 | - | 5/5 | - | 5/5 | - |
| | | Negative | - | - | - | 30.7 | - | 5/5 | - | 5/5 | - | 5/5 |
| | 30 | Positive | 31.9 | 29.6 | 30.9 | 30.7 | 5/5 | - | 5/5 | - | 5/5 | - |
| | | Negative | - | - | - | 30.9 | - | 5/5 | - | 5/5 | - | 5/5 |
| 4°C | 0 | Positive | 32.2 | 30.7 | 31.0 | 31.9 | 5/5 | - | 5/5 | - | 5/5 | - |
| | | Negative | - | - | - | 32.4 | - | 5/5 | - | 5/5 | - | 5/5 |
| | 2 | Positive | 32.2 | 30.9 | 30.7 | 31.9 | 5/5 | - | 5/5 | - | 5/5 | - |
| | | Negative | - | - | - | 32.5 | - | 5/5 | - | 5/5 | - | 5/5 |
| | 3 | Positive | 33.1 | 31.0 | 31.6 | 32.6 | 5/5 | - | 5/5 | - | 5/5 | - |
| | | Negative | - | - | - | 32.1 | - | 5/5 | - | 5/5 | - | 5/5 |
| | 7 | Positive | 32.8 | 30.8 | 31.4 | 32.0 | 5/5 | - | 5/5 | - | 5/5 | - |
| | | Negative | - | - | - | 32.1 | - | 5/5 | - | 5/5 | - | 5/5 |
Overall, the study results show that nasopharyngeal swab samples may be stored at $-80^{\circ}\mathrm{C}$ for up to 30 days and at $4^{\circ}\mathrm{C}$ for up to 7 days prior to testing. The results support the package insert instructions that specimens may be refrigerated $(2 - 8^{\circ}\mathrm{C})$ for up to 72 hours (3 days); and that, for long duration transport or storage, specimens should be frozen at $-70^{\circ}\mathrm{C}$ or colder and transported on dry ice.
{13}
14
# Performance using Fresh vs. Frozen Specimen
The cobas® Liat Influenza A/B & RSV assay was tested by comparing its performance using fresh and frozen specimens. One (1) influenza A (A/Brisbane/59/07), one influenza B (B/Malaysia/2506/04), and one RSV strain (RSV-A 2006 Isolate) was individually spiked into nasopharyngeal swab in UTM matrix at different viral loads, including levels near the LOD and levels reflecting the clinical sample range. Influenza and RSV strains used were from the same source as identified in the "Detection Limit" section of this summary. For each strain, at least 60 samples were tested immediately while at least another 60 samples were frozen at -80°C for 7 days, thawed and then tested. Ten (10) negative samples were also tested fresh and after being frozen at -80°C for 7 days. The negative sample consisted of negative NPS in UTM sample matrix.
A total of 381 runs were performed in this study. Runs were performed on 26 cobas® Liat Analyzers using one lot of cobas® Liat tubes.
Table 6 below shows the results from the study:
{14}
Table 6: Specimen Fresh vs. Frozen Study Results
| | x LOD | Conc. (TCID50/mL) | Fresh | | | | | Frozen | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | | n | Positive Call Rate | Ct Avg. | Ct SD | Ct %CV | n | Positive Call Rate | Ct Avg. | Ct SD | Ct %CV |
| Influenza A/Brisbane/59/07 | 3 | 6.0 x 10-3 | 30 | 30/30 | 32.7 | 1.3 | 3.9% | 30 | 30/30 | 32.5 | 0.9 | 2.6% |
| | 20 | 4.0 x 10-2 | 10 | 10/10 | 30.2 | 0.6 | 2.0% | 10 | 10/10 | 30.5 | 0.6 | 2.0% |
| | 200 | 4.0 x 10-1 | 10 | 10/10 | 26.2 | 1.3 | 4.9% | 10 | 10/10 | 26.8 | 0.4 | 1.6% |
| | 2000 | 4.0 x 100 | 5 | 5/5 | 23.0 | 0.4 | 1.8% | 5 | 5/5 | 23.1 | 0.4 | 1.9% |
| | 20000 | 4.0 x 101 | 5 | 5/5 | 19.4 | 0.1 | 0.7% | 5 | 5/5 | 19.5 | 0.2 | 1.0% |
| Influenza B/Malaysia/2506/04 | 3 | 1.2 x 10-2 | 30 | 30/30 | 29.9 | 0.4 | 1.4% | 30 | 30/30 | 30.0 | 0.3 | 1.0% |
| | 20 | 8.0 x 10-2 | 10 | 10/10 | 26.7 | 0.3 | 1.2% | 10 | 10/10 | 26.6 | 0.3 | 1.3% |
| | 200 | 8.0 x 10-1 | 10 | 10/10 | 23.5 | 0.3 | 1.3% | 10 | 10/10 | 23.4 | 0.1 | 0.6% |
| | 2000 | 8.0 x 100 | 5 | 5/5 | 20.5 | 0.1 | 0.4% | 5 | 5/5 | 20.4 | 0.4 | 1.9% |
| | 20000 | 8.0 x 101 | 5 | 5/5 | 17.1 | 0.3 | 1.7% | 5 | 5/5 | 17.2 | 0.0 | 0.3% |
| RSV A 2006 Isolate | 3 | 1.2 x 100 | 30 | 30/30 | 31.8 | 0.6 | 2.0% | 31 | 31/31 | 31.9 | 0.6 | 2.0% |
| | 20 | 8.0 x 100 | 10 | 10/10 | 29.1 | 0.4 | 1.4% | 10 | 10/10 | 29.1 | 0.5 | 1.6% |
| | 200 | 8.0 x 101 | 10 | 10/10 | 26.0 | 0.2 | 0.7% | 10 | 10/10 | 25.9 | 0.1 | 0.6% |
| | 2000 | 8.0 x 102 | 5 | 5/5 | 22.4 | 0.1 | 0.3% | 5 | 5/5 | 22.5 | 0.2 | 1.0% |
| | 20000 | 8.0 x 103 | 5 | 5/5 | 19.2 | 0.1 | 0.4% | 5 | 5/5 | 19.2 | 0.2 | 1.2% |
| Negative | - | - | 10 | 0/10 | 30.4 | 0.5 | 1.8% | 10 | 0/10 | 30.8 | 0.9 | 2.9% |
Note: Ct Average (Ave.) and Ct Standard Deviation (SD) for Negative sample is that of the IPC.
Fresh and frozen positive samples demonstrated $100\%$ detection across all levels of viral load, including those near LOD. Ct values were also very similar between fresh and frozen samples. Fresh and frozen negative samples all demonstrated $0\%$ detection. The results showed that freezing and thawing samples had no effect on the detection of Influenza A, Influenza B, and RSV by the cobas® Liat Influenza A/B & RSV assay.
# d. Detection Limit:
The Limit of Detection (LOD) of the cobas® Liat Influenza A/B & RSV assay was tested using three strains of Influenza A (A/Brisbane/10/2007, A/Brisbane/59/2007, and A/NY/01/2009), two strains of Influenza B (B/Malaysia/2506/04 and B/Florida/04/06), and two strains of RSV (RSV-A 2006 isolate and RSV-B Ch93(18)-18). The LOD was determined by limiting dilution studies using these titered viruses. The viruses were spiked into nasopharyngeal swab (NPS) in UTM sample matrix, and then tested using the cobas® Liat Influenza A/B & RSV assay. The LOD was determined as the lowest virus concentration that was detected $\geq 95\%$ of the time (i.e., concentration at which at least 19 out of 20 replicates tested positive).
Virus strains (Table 7 below) were cultured and re-titered using $\mathrm{TCID}_{50}$ procedure by a third party laboratory before their use in LOD determination. Each characterized strain was first adjusted to the highest $\log_{10} \mathrm{TCID}_{50} / \mathrm{mL}$ in UTM as diluent.
{15}
Table 7: Influenza and RSV Strains and Viral Titers Used in the LOD Study
| Viral Strain | Virus Type | TCID50/mL |
| --- | --- | --- |
| A/Brisbane/10/2007 | Influenza A/H3N2 | 1.4 x 103 |
| A/Brisbane/59/2007 | Influenza A/H1N1 | 1.4 x 103 |
| A/NY/01/2009 | Influenza A/H1N1 pdm | 7.2 x 103 |
| B/Florida/04/06 | Influenza B Yamagata Lineage | 1.4 x 103 |
| B/Malaysia/2506/04 | Influenza B Victoria Lineage | 2.5 x 103 |
| RSV-A 2006 isolate | RSV Type A | 6.6 x 106 |
| RSV-B Ch93(18)-18 | RSV Type B | 1.3 x 106 |
To prepare the testing sample, the virus at different concentrations was spiked into a pooled NPS in UTM sample matrix. NPS samples were collected throughout the study period from healthy donors that did not show any flu-like symptoms. The swabs were eluted in UTM, pooled, and then tested to confirm that the matrix pool was negative for Influenza A, Influenza B and RSV. Virus was spiked into the NPS in UTM sample matrix at no more than $5\%$ the final volume to prepare the test sample.
For each run, the spiked sample was tested using the cobas® Liat Influenza A/B & RSV assay following the Instructions for Use. Dilutions of each virus strain were first tested in triplicates. The lowest concentration at which all three replicates tested positive was treated as the tentative LOD for each strain. The LOD was then confirmed by testing at least 20 replicates with concentrations at the tentative LOD. The confirmed LOD of each strain was determined as the lowest concentration resulting in positive detection of at least 19 out of 20 replicates. If more than one replicate is detected as negative, a two-fold or higher concentration of virus was tested with 20 additional replicates. The process was iterated until the concentration at which the virus was detected in at least 19 of 20 replicates.
A total of 263 runs were performed in this study. Runs were performed on 22 cobas® Liat Analyzers using three lots of cobas® Liat assay tubes.
The LOD Study results are presented in Table 8 below:
Table 8: LOD Study Results
| Virus Strain | Analyte | Stock Virus Concentration (TCID50/mL) | Test Virus LOD Concentration (TCID50/mL) | Positive Call Rate | Ct Ave. | Ct %CV |
| --- | --- | --- | --- | --- | --- | --- |
| A/Brisbane/10/2007 | Inf A | 1.4 x 105 | 2.0 x 10-2 | 23/23 | 33.6 | 1.2% |
| A/Brisbane/59/2007 | Inf A | 1.4 x 105 | 2.0 x 10-3 | 23/23 | 34.1 | 2.8% |
| A/NY/01/2009 | Inf A | 7.2 x 105 | 2.0 x 10-2 | 23/23 | 32.8 | 1.4% |
| B/Florida/04/06 | Inf B | 1.4 x 105 | 2.0 x 10-3 | 23/23 | 33.7 | 2.1% |
| B/Malaysia/2506/04 | Inf B | 2.5 x 105 | 4.0 x 10-3 | 23/23 | 32.8 | 1.9% |
| RSV-A 2006 isolate | RSV | 6.6 x 106 | 4.0 x 10-1 | 23/23 | 32.4 | 2.6% |
| RSV-B Ch93(18)-18 | RSV | 1.3 x 106 | 4.0 x 10-1 | 23/23 | 33.0 | 2.7% |
# e. Analytical Reactivity:
An analytical reactivity study was conducted evaluating the ability of the cobas® Liat Influenza A/B & RSV assay to detect influenza and RSV strains representing temporal and geographical diversity. The cobas® Liat Influenza A/B & RSV assay was evaluated for reactivity with 28 Influenza A strains, 15 Influenza B strains, and 7 RSV strains. Influenza A strains included 14 Influenza A/H1
{16}
strains (including 3 H1N1 pdm09 strains), 12 Influenza A/H3 strains (including one H3N2v strain), one Influenza A/H7N9 strain, and one Influenza A/H5N1 reassortant strain. Influenza B strains included strains from both the Victoria lineage and Yamagata lineage. RSV strains included both RSV Type A and Type B strains.
Cultured viral strains were obtained from vendors. All strains were frozen and were accompanied by Certificate of Analysis including $\mathrm{TCID}_{50} / \mathrm{mL}$ or $\mathrm{CEID}_{50} / \mathrm{mL}$ titer information. Each viral strain under test was serially diluted in 10 fold increments in UTM using the same procedure as that for the LOD study. The virus was then spiked in negative NPS in UTM sample matrix, and the spiked sample was tested using the cobas® Liat Influenza A/B & RSV assay following the Instructions for Use. Each viral strain was tested at the LOD for strains used in LOD studies, and at "near LOD" level for other strains. To estimate "near LOD" level, each strain was first tested at $1.0 \times 10^{2} \mathrm{TCID}_{50} / \mathrm{mL}$ or $1.0 \times 10^{2} \mathrm{CEID}_{50} / \mathrm{mL}$ , and then higher or lower concentrations (as appropriate) until all triplicate runs at a concentration were detected, and gave a Ct between 31 and 34. Given that the Ct at LOD was 31-34 based on LOD study data, a Ct value in this range was used to estimate the "near LOD" level.
A total of 243 runs were performed on 20 cobas® Liat Analyzers using three lots of cobas® Liat tubes.
Table 9 below shows the strains tested for reactivity and the "near LOD" concentration of each strain that yielded $3 / 3$ positive results.
Table 9: Analytical Reactivity Study Results
| Virus Strain | Type / Subtype | Test Concentration | Inf A Result | Inf B Result | RSV Result |
| --- | --- | --- | --- | --- | --- |
| A/Aichi/2/68 | Influenza A/H3N2 | 1.0×102CEID50/mL | + | - | - |
| A/Alice | Influenza A/H3N2 | 5.0×101CEID50/mL | + | - | - |
| A/Anhui/1/2013 | Influenza A/H7N9 (Eurasian lineage) | 1.0×103TCID50/mL | + | - | - |
| A/Brisbane/10/07 | Influenza A/H3N2 | 2.0×10-2TCID50/mL | + | - | - |
| A/Brisbane/59/07 | Influenza A/H1N1 | 2.0×10-3TCID50/mL | + | - | - |
| A/Cambodia/X0810301/2013 (H5N1)-PR8-IDCDC-RG34B | Influenza A/H5N1 reassortant | 2.5×101CEID50/mL | + | - | - |
| A/Denver/1/57 | Influenza A/H1N1 | 1.0×102CEID50/mL | + | - | - |
| A/FM/1/47 | Influenza A/H1N1 | 1.0×102CEID50/mL | + | - | - |
| A/H3/Perth/16/09 | Influenza A/H3N2 | 2.5×10-1TCID50/mL | + | - | - |
| A/Hong Kong/8/68 | Influenza A/H3N2 | 1.0×102TCID50/mL | + | - | - |
| A/Indiana/8/2011 | Influenza A/H3N2v | 5.0×10-1TCID50/mL | + | - | - |
| A/Mal/302/54 | Influenza A/H1N1 | 4.0×102CEID50/mL | + | - | - |
| A/MRC2 | Influenza A/H3 | 1.0×102CEID50/mL | + | - | - |
| A/New Caledonia/20/99 | Influenza A/H1N1 | 1.0×102TCID50/mL | + | - | - |
| A/New Jersey/8/76 | Influenza A/H1N1 | 1.0×101CEID50/mL | + | - | - |
| A/NY/01/2009 | Influenza A/H1N1 pdm09 | 2.0×10-2TCID50/mL | + | - | - |
{17}
| Virus Strain | Type / Subtype | Test Concentration | Inf A Result | Inf B Result | RSV Result |
| --- | --- | --- | --- | --- | --- |
| A/NY/02/2009 | Influenza A/H1N1 pdm09 | 2.5×10-2TCID50/mL | + | - | - |
| A/NY/03/2009 | Influenza A/H1N1 pdm09 | 2.0×10-1TCID50/mL | + | - | - |
| A/Port Chalmers/1/73 | Influenza A/H3N2 | 1.0×102CEID50/mL | + | - | - |
| A/PR/8/34 | Influenza A/H1N1 | 5.0×100TCID50/mL | + | - | - |
| A/Solomon Island/3/2006 | Influenza A/H1N1 | 5.0×10-2TCID50/mL | + | - | - |
| A/Swine/1976/31 | Influenza A/H1N1 | 1.0×101CEID50/mL | + | - | - |
| A/Swine/Iowa/15/30 | Influenza A/H1N1 | 1.0×102CEID50/mL | + | - | - |
| A/Texas/50/2012 | Influenza A/H3N2 | 1.0×10-1TCID50/mL | + | - | - |
| A/Victoria/3/75 | Influenza A/H3N2 | 1.0×102CEID50/mL | + | - | - |
| A/Victoria/361/2011 | Influenza A/H3N2 | 2.0×10-2TCID50/mL | + | - | - |
| A/Weiss/43 | Influenza A/H1N1 | 1.0×103TCID50/mL | + | - | - |
| A/Wisconsin/67/05 | Influenza A/H3N2 | 5.0×10-1TCID50/mL | + | - | - |
| B/Allen/45 | Influenza B | 5.0×10-1TCID50/mL | - | + | - |
| B/Brisbane/60/2008 | Influenza B (Victoria lineage) | 1.0×10-2TCID50/mL | - | + | - |
| B/Florida/04/06 | Influenza B (Yamagata lineage) | 2.0×10-3TCID50/mL | - | + | - |
| B/Florida/07/04 | Influenza B (Yamagata lineage) | 5.0×10-2TCID50/mL | - | + | - |
| B/GL/1739/54 | Influenza B | 2.0×100TCID50/mL | - | + | - |
| B/HongKong/5/72 | Influenza B | 2.5×10-1TCID50/mL | - | + | - |
| B/Lee/40 | Influenza B | 2.5×10-1TCID50/mL | - | + | - |
| B/Malaysia/2506/04 | Influenza B (Victoria lineage) | 4.0×10-3TCID50/mL | - | + | - |
| B/Maryland/1/59 | Influenza B | 5.0×10-2TCID50/mL | - | + | - |
| B/Mass/3/66 | Influenza B | 1.0×101TCID50/mL | - | + | - |
| B/Massachusetts/2/2012 | Influenza B (Yamagata lineage) | 5.0×10-3TCID50/mL | - | + | - |
| B/Nevada/03/2011 | Influenza B (Victoria lineage) | 2.5×10-1CEID50/mL | - | + | - |
| B/Taiwan/2/62 | Influenza B | 1.0×100TCID50/mL | - | + | - |
| B/Texas/6/2011 | Influenza B (Yamagata lineage) | 1.0×10-1TCID50/mL | - | + | - |
| B/Wisconsin/1/2010 | Influenza B (Yamagata lineage) | 5.0×10-1TCID50/mL | - | + | - |
| RSV A 2006 isolate | RSV A | 4.0×10-1TCID50/mL | - | - | + |
| RSV A Long | RSV A | 1.0×102TCID50/mL | - | - | + |
| RSV A2 | RSV A | 1.0×100TCID50/mL | - | - | + |
| RSV B 9320 | RSV B | 1.0×100TCID50/mL | - | - | + |
| RSV B Ch93(18)-18 | RSV B | 4.0×10-1TCID50/mL | - | - | + |
| RSV B Ch10 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch11 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch12 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch13 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch14 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch15 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch16 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch17 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch18 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch19 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch20 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch21 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch22 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch23 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch24 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch25 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch26 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch27 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
| RSV B Ch28 | RSV B | 1.0×10-1TCID50/mL | - | - | + |
{18}
| Virus Strain | Type / Subtype | Test Concentration | Inf A Result | Inf B Result | RSV Result |
| --- | --- | --- | --- | --- | --- |
| RSV B Wash/18537 | RSV B | 1.0×10^0 TCID50/mL | - | - | + |
| RSV B WV/14617/85 | RSV B | 1.0×10^-1 TCID50/mL | - | - | + |
Data demonstrates that the cobas® Liat Influenza A/B & RSV assay is reactive against all strains at the specified concentrations tested.
The cobas® Liat Influenza A/B & RSV assay was shown to be reactive against viral strains with titers ranging from $10^{-2}$ to $10^{3}$ TCID $_{50}$ /mL or CEID $_{50}$ /mL. The observed difference in titers is most likely due to the fact that TCID $_{50}$ /mL or CEID $_{50}$ /mL determination and RT-PCR testing measure different targets. TCID $_{50}$ /mL or CEID $_{50}$ /mL is the measure of infectious virus titer, or the virus required to produce a cytopathic effect in $50\%$ of inoculated tissue culture cells. This is a measure of only live, intact virus, and takes into account the infection and cytopathic efficiency of the virus for the host cells. On the other hand, PCR-based assay detect nucleic acids, including that from intact and non-intact virions, and free nucleic acids in solution. Because of this measurement difference, TCID $_{50}$ /mL or CEID $_{50}$ /mL can vary widely for a substantially equivalent quantity of RNA, and vice versa, resulting in the difference observed. Also, titering methods have been reported to have an inherent variability of up to one order of magnitude, which also may contribute to the observed difference (Propagation and purification of influenza virus, GE Healthcare Life Sciences, Application note 28-9843-41 AB, https://www.gelifesciences.com/gehcls_images/GELS/Related%20Content/Files/1354807438862/lit doc28984341_20130610001258.pdf).
{19}
In addition, based on in silico analysis, there is no evidence to suggest that the observed difference in analytical reactivity is principally caused by sequence mismatch between the primer/probes of the cobas® Liat Influenza A/B & RSV assay to the strains tested.
# f. Analytical Specificity:
A cross reactivity study was carried out to evaluate the cobas® Liat Influenza A/B & RSV assay's potential cross-reactivity with microorganisms that may be present in nasopharyngeal swab samples. The cobas® Liat Influenza A/B & RSV assay was evaluated against a panel containing human genomic DNA and 35 microorganisms. Bacteria and Candida albicans were tested at $\geq 10^{6}$ CFU/mL, and viruses were tested at $\geq 10^{5}$ TCID $_{50}$ /mL, or the highest available concentration. For Mycobacterium tuberculosis and Mycoplasma pneumoniae, testing was performed with genomic DNA due to difficulties in propagation of these bacteria. Each microorganism was spiked into negative NPS in UTM matrix at the concentration indicated in Table 10 below. The spiked volume did not exceed $5\%$ of the sample volume. The spiked samples were tested using the cobas® Liat Influenza A/B & RSV assay in triplicate. A control of negative NPS in UTM matrix and no microorganisms was also tested.
A total of 114 runs were performed on 13 cobas® Liat Analyzers using 6 lots of cobas® Liat assay tubes.
Table 10 below shows the cross reactivity study results:
Table 10: Cross Reactivity Study Results
| Microorganism | Test Concentration | Inf A Result | Inf B Result | RSV Result |
| --- | --- | --- | --- | --- |
| Adenovirus Type 1 | 9.0×105TCID50/mL | - | - | - |
| Adenovirus Type 7 | 1.4×105TCID50/mL | - | - | - |
| Cytomegalovirus | 4.5×104TCID50/mL | - | - | - |
| Epstein Barr Virus | 2.5×105TCID50/mL | - | - | - |
| Herpes Simplex Virus | 1.4×105TCID50/mL | - | - | - |
| Human Coronavirus 229E | 8.0×103TCID50/mL | - | - | - |
| Human Coronavirus OC43 | 8.0×104TCID50/mL | - | - | - |
| Human Enterovirus 68 | 1.0×105TCID50/mL | - | - | - |
| Human Metapneumovirus | 7.0×103TCID50/mL | - | - | - |
| Human Parainfluenza Type 1 | 3.7×105TCID50/mL | - | - | - |
| Human Parainfluenza Type 2 | 7.5×105TCID50/mL | - | - | - |
| Human Parainfluenza Type 3 | 4.5×105TCID50/mL | - | - | - |
| Human Rhinovirus Type 1A | 8.0×105TCID50/mL | - | - | - |
| Measles | 8.0×104TCID50/mL | - | - | - |
| Mumps Virus | 8.0×104TCID50/mL | - | - | - |
| Varicella-Zoster Virus | 4.4×103TCID50/mL | - | - | - |
| Bordetella pertussis | 2.2×106CFU/mL | - | - | - |
| Candida albicans | 4.2×106CFU/mL | - | - | - |
| Chlamydia pneumoniae | 8.0×104TCID50/mL | - | - | - |
{20}
| Microorganism | Test Concentration | Inf A Result | Inf B Result | RSV Result |
| --- | --- | --- | --- | --- |
| Corynebacterium sp. | 3.6×106CFU/mL | - | - | - |
| Escherichia coli | 1.9×106CFU/mL | - | - | - |
| Haemophilus influenzae | 2.3×106CFU/mL | - | - | - |
| Lactobacillus sp. | 1.9×106CFU/mL | - | - | - |
| Legionella pneumophila | 6.7×106CFU/mL | - | - | - |
| Moraxella catarrhalis | 2.5×106CFU/mL | - | - | - |
| Mycobacterium tuberculosis | 2.8×106copies/mL1 | - | - | - |
| Mycoplasma pneumoniae | 2.9×106copies/mL1 | - | - | - |
| Neisseria elongata | 2.0×106CFU/mL | - | - | - |
| Neisseria meningitidis | 2.2×106CFU/mL | - | - | - |
| Pseudomonas aeruginosa | 2.3×106CFU/mL | - | - | - |
| Staphylococcus aureus | 2.4×106CFU/mL | - | - | - |
| Staphylococcus epidermidis | 1.9×106CFU/mL | - | - | - |
| Streptococcus pneumoniae | 1.8×106CFU/mL | - | - | - |
| Streptococcus pyogenes | 2.5×106CFU/mL | - | - | - |
| Streptococcus salivarius | 4.3×106CFU/mL | - | - | - |
| Human genomic DNA | 1.0×104copies/mL | - | - | - |
1 Testing was performed with genomic DNA due to difficulties in propagation of these bacteria.
The cobas® Liat Influenza A/B & RSV assay showed no cross reactivity for the human genomic DNA or the microorganisms at the concentrations tested.
# g. Potentially Interfering Substances:
The cobas® Liat Influenza A/B & RSV assay was evaluated with potentially interfering substances that may be encountered in nasopharyngeal swab specimens. Medically and/or physiologically relevant concentrations of potential interferents (Table 11 below) were tested with two influenza A strains, two influenza B strains, and two RSV strains at $3\mathrm{x}$ LOD $(10^{0} - 10^{-2}\mathrm{TCID}_{50} / \mathrm{mL})$
{21}
Table 11: Interfering Substances Panel
| Potentially Interfering Substances | Active Ingredient | Concentration |
| --- | --- | --- |
| Mucin: bovine submaxillary gland, type I-S | Purified mucin protein | 5 mg/ml |
| Blood | - | 5% (v/v) |
| Nasal spray – Afrin | Oxymetazoline | 5% (v/v) |
| Nasal corticosteroids – Veramyst | Fluticasone | 5% (v/v) |
| Nasal gel – Zicam | Galphimia glauca, Histaminum hydrochloricum, Luffa operculata, Sulphur | 5% (v/v) |
| Throat lozenges, oral anesthetic and analgesic – Cepacol | Benzocaine, Menthol | 5 mg/ml |
| Antibiotic, nasal ointment – Bactroban | Mupirocin | 5 mg/ml |
| Antiviral drug – Relenza | Zanamivir | 5 mg/ml |
| Antiviral drug – Tamiflu | Oseltamivir | 7.5 mg/ml |
| Antimicrobial, systemic | Tobramycin | 4 μg/ml |
The potentially interfering substances were evaluated at a concentration equivalent to: (1) $\geq 3$ times the average concentration found in serum (or nasal secretion for mucin), (2) $5\%$ v/v for liquids or 5 mg/mL for solids, or (3) the highest soluble concentration in UTM.
Interfering substance testing was performed on a sample panel containing a negative sample and six positive samples (Table 12). The negative sample contained negative NPS in UTM sample matrix. The positive samples contained two influenza A strains, two influenza B strains, and two RSV strains, each spiked into negative NPS in UTM sample matrix at its 3x LOD concentration.
Table 12: Concentration of Influenza and RSV Strains Used in Potentially Interfering Substances Study
| Sample | Concentration (TCID50/mL) |
| --- | --- |
| A/Brisbane/10/07 | 6.0×10-2 |
| A/Brisbane/59/07 | 6.0×10-3 |
| B/Florida/04/06 | 6.0×10-3 |
| B/Malaysia/2506/04 | 1.2×10-2 |
| RSV-A 2006 Isolate | 1.2×100 |
| RSV-B Ch93(18)-18 | 1.2×100 |
| Negative | - |
Each interfering substance was spiked into each sample and tested using the cobas® Liat Influenza A/B & RSV assay in triplicate. Controls containing the samples and no interfering substance were also tested.
A total of 231 runs were performed on 18 cobas® Liat Analyzers using one lot of cobas® Liat tubes.
Table 13 below shows the potentially interfering substances study results:
{22}
Table 13: Potentially Interfering Substances Study Results
| Potentially Interfering Substances | Conc. | A/Brisbane/10 /7 | | A/Brisbane/59/ 07 | | B/Florida/04/0 6 | | B/Malaysia/2 506 /04 | | RSV A 2006 Isolate | | RSV B Ch93(18)-18 | | Negative | |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| | | Inf A Detected | Inf A Ct. Ave. | Inf A Detected | Inf A Ct. Ave. | Inf B Detected | Inf B Ct. Ave. | Inf B Detected | Inf B Ct. Ave. | RSV Detected | RSV Ct. Ave. | RSV Detected | RSV Ct. Ave. | Inf A ND Inf B ND RSV ND | IPC Ct. Ave. |
| Control | - | 3/3 | 31.9 | 3/3 | 32.2 | 3/3 | 31.8 | 3/3 | 31.2 | 3/3 | 31.0 | 3/3 | 31.5 | 3/3 | 29.5 |
| Afrin | 5% (v/v) | 3/3 | 32.1 | 3/3 | 32.5 | 3/3 | 30.8 | 3/3 | 31.1 | 3/3 | 30.8 | 3/3 | 31.6 | 3/3 | 29.5 |
| Bactroban | 5 mg/mL | 3/3 | 31.7 | 3/3 | 32.9 | 3/3 | 31.4 | 3/3 | 31.2 | 3/3 | 31.4 | 3/3 | 31.1 | 3/3 | 29.8 |
| Blood | 5% (v/v) | 3/3 | 33.0 | 3/3 | 33.4 | 3/3 | 32.1 | 3/3 | 31.4 | 3/3 | 31.4 | 3/3 | 32.0 | 3/3 | 31.6 |
| Cepacol | 5 mg/mL | 3/3 | 31.8 | 3/3 | 32.4 | 3/3 | 31.2 | 3/3 | 31.3 | 3/3 | 31.2 | 3/3 | 31.3 | 3/3 | 29.5 |
| Mucin | 5 mg/mL | 3/3 | 31.0 | 3/3 | 32.0 | 3/3 | 31.9 | 3/3 | 30.2 | 3/3 | 30.6 | 3/3 | 31.0 | 3/3 | 30.4 |
| Relenza | 5 mg/mL | 3/3 | 31.8 | 3/3 | 32.5 | 3/3 | 31.6 | 3/3 | 31.3 | 3/3 | 31.5 | 3/3 | 31.0 | 3/3 | 29.3 |
| Tamiflu | 7.5 mg/mL | 3/3 | 33.2 | 3/3 | 32.7 | 3/3 | 32.0 | 3/3 | 31.7 | 3/3 | 32.5 | 3/3 | 32.7 | 3/3 | 30.8 |
| Tobramycin | 4 ug/mL | 3/3 | 31.9 | 3/3 | 32.3 | 3/3 | 31.8 | 3/3 | 31.7 | 3/3 | 31.2 | 3/3 | 31.5 | 3/3 | 29.5 |
| Veramyst | 5% (v/v) | 3/3 | 31.9 | 3/3 | 32.4 | 3/3 | 32.0 | 3/3 | 31.3 | 3/3 | 32.0 | 3/3 | 31.8 | 3/3 | 29.6 |
| Zicam | 5% (v/v) | 3/3 | 31.6 | 3/3 | 32.3 | 3/3 | 32.0 | 3/3 | 31.3 | 3/3 | 32.0 | 3/3 | 31.0 | 3/3 | 29.2 |
Results show that substances at the concentrations tested did not interfere in the detection of influenza A, influenza B and RSV strains.
# h. Potential Microbial Interference Study:
A potentially interfering microorganism study was conducted to evaluate whether microorganisms that may be present in nasopharyngeal swab samples can interfere in the detection of Influenza A, Influenza B or RSV by the cobas® Liat Influenza A/B & RSV assay. The panel of human genomic DNA and 35 microorganisms tested in the cross-reactivity study was tested for potential interference. Bacteria and Candida albicans were tested at $\geq 10^{6}$ CFU/mL and viruses were tested at $\geq 10^{5}$ TCID $_{50}$ /mL or the highest available concentration, in the presence of one Influenza A strain, one Influenza B strain and one RSV strain at 3x LOD concentrations in negative nasopharyngeal swab in UTM matrix. For Mycobacterium tuberculosis and Mycoplasma pneumoniae, testing was performed with genomic DNA due to difficulties in propagation of these bacteria.
Interfering microorganism testing was performing using a positive sample comprised of A/Brisbane/59/2007 at $6.0 \times 10^{-3}$ TCID $_{50}$ /mL, B/Malaysia/2506/04 at $1.2 \times 10^{-2}$ TCID $_{50}$ /mL), and RSV-A 2006 Isolate at $1.2 \times 10^{0}$ TCID $_{50}$ /mL, spiked into negative NPS in UTM sample matrix (3x LOD concentration). Each potentially interfering microorganism was spiked into each sample at the concentration indicated in Table 14. The spiked volume did not exceed $5\%$ of the sample volume. The spiked samples were tested using the cobas® Liat Influenza A/B & RSV assay in triplicate. A control comprised of the positive sample and no interfering microorganism was also tested.
A total of 111 runs were performed on 12 cobas® Liat Analyzers using three lots of cobas® Liat assay tubes
{23}
Table 14: Potentially Interfering Microorganisms Study Results
| Microorganism | Test Concentration | One Flu A, One Flu B & One RSV strain at 3x LOD | | |
| --- | --- | --- | --- | --- |
| | | Inf A Result | Inf B Result | RSV Result |
| Adenovirus Type 1 | 9.0×105TCID50/mL | + | + | + |
| Adenovirus Type 7 | 1.4×105TCID50/mL | + | + | + |
| Cytomegalovirus | 4.5×104TCID50/mL | + | + | + |
| Epstein Barr Virus | 2.5×105TCID50/mL | + | + | + |
| Herpes Simplex Virus | 1.4×105TCID50/mL | + | + | + |
| Human Coronavirus 229E | 8.0×103TCID50/mL | + | + | + |
| Human Coronavirus OC43 | 8.0×104TCID50/mL | + | + | + |
| Human Enterovirus 68 | 1.0×105TCID50/mL | + | + | + |
| Human Metapneumovirus | 7.0×103TCID50/mL | + | + | + |
| Human Parainfluenza Type 1 | 3.7×105TCID50/mL | + | + | + |
| Human Parainfluenza Type 2 | 7.5×105TCID50/mL | + | + | + |
| Human Parainfluenza Type 3 | 4.5×105TCID50/mL | + | + | + |
| Human Rhinovirus Type 1A | 8.0×105TCID50/mL | + | + | + |
| Measles | 8.0×104TCID50/mL | + | + | + |
| Mumps Virus | 8.0×104TCID50/mL | + | + | + |
| Varicella-Zoster Virus | 4.4×103TCID50/mL | + | + | + |
| Bordetella pertussis | 2.2×106CFU/mL | + | + | + |
| Candida albicans | 4.2×106CFU/mL | + | + | + |
| Chlamydia pneumoniae | 8.0×104TCID50/mL | + | + | + |
| Corynebacterium sp | 3.6×106CFU/mL | + | + | + |
| Escherichia coli | 1.9×106CFU/mL | + | + | + |
| Haemophilus influenzae | 2.3×106CFU/mL | + | + | + |
| Lactobacillus sp | 1.9×106CFU/mL | + | + | + |
| Legionella pneumophila | 6.7×106CFU/mL | + | + | + |
| Moraxella catarrhalis | 2.5×106CFU/mL | + | + | + |
| Mycobacterium tuberculosis | 2.8×106copies/mL1 | + | + | + |
| Mycoplasma pneumoniae | 2.9×106copies/mL1 | + | + | + |
| Neisseria elongata | 2.0×106CFU/mL | + | + | + |
| Neisseria meningitidis | 2.2×106CFU/mL | + | + | + |
| Pseudomonas aeruginosa | 2.3×106CFU/mL | + | + | + |
| Staphylococcus aureus | 2.4×106CFU/mL | + | + | + |
| Staphylococcus epidermidis | 1.9×106CFU/mL | + | + | + |
| Streptococcus pneumoniae | 1.8×106CFU/mL | + | + | + |
| Streptococcus pyogenes | 2.5×106CFU/mL | + | + | + |
| Streptococcus salivarius | 4.3×106CFU/mL | + | + | + |
| Human Genomic DNA | 1.0×104copies/mL | + | + | + |
1 Testing was performed with genomic DNA due to difficulties in propagation of these bacteria.
Results show that the presence of human genomic DNA or the microorganisms at the concentrations tested did not interfere with the detection of Influenza A, Influenza B, or RSV.
{24}
i. Carry-Over Study:
A study was conducted to demonstrate that the single-use, self-contained cobas® Liat assay tube and the cobas® Liat System are not susceptible to carry-over contamination when alternating high positive and negative samples are tested. High positive samples consisted of an influenza A virus spiked into negative nasopharyngeal swab in UTM matrix at a concentration greater than that expected in 95% or more of specimens of diseased patients in the intended use population. In clinical studies, the lowest Ct values of Influenza A detected specimens, which corresponds to the highest concentration specimens, was approximately 12 to 13. The highest titer available for Influenza A/Brisbane/59/07 allowed for a test concentration of 7.1×10³ TCID50/mL. This test concentration gave a Ct of approximately 12.7. To test the worst case scenario, this highest available concentration of Influenza A was used as the high positive sample concentration. Negative samples contained negative nasopharyngeal swab in UTM matrix.
Twenty (20) high positive and 20 negative samples were tested on each of two cobas® Liat Systems (i.e. 40 high positive and 40 negative samples in total). High positive and negative samples were run alternating analyzer-to-analyzer and run-to-run. All 40 high positive samples tested were correctly reported as "Influenza A Detected". All 40 negative samples tested were correctly reported as "Influenza A Not Detected". There was no carry-over or cross contamination observed during this study.
j. Assay cut-off:
The assay cutoff for each analyte in terms of cycle threshold (Ct) was determined based on two considerations: (1) the Ct cutoff should be greater than the latest Ct values of positive runs at the LOD of the virus strains tested to ensure analytical sensitivity; and (2) the Ct cutoff should be early enough to minimize the risk of false positive detection to ensure the specificity for target detection.
For the Influenza A, A/Brisbane/59/07 had the latest Ct in the LOD studies, generating an average LOD Ct of 34.1 with a standard deviation of 0.97. As such, the latest expected Ct value of positive runs at the LOD was calculated to be 36.7. The earliest Ct for negative samples was 37.2. Therefore, the Ct cutoff for Influenza A was set at 37.0.
Similarly, for Influenza B, B/Florida/04/06 had the latest Ct in LOD studies, generating an average LOD Ct of 33.7 with a standard deviation of 0.72. As such, the latest expected Ct value of positive runs at the LOD was calculated to be 35.6. The earliest Ct for negative samples was 42.0. Therefore, the Ct cutoff for Influenza B was set at 38.5.
For RSV, RSV-B Ch 93(18)-18 had the latest Ct in LOD studies, generating an average LOD Ct of 33.0 with a standard deviation of 0.90. As such, the latest expected Ct value of positive runs at the LOD was calculated to be 35.4. The earliest Ct for negative samples was 37.7. Therefore, the Ct cutoff for RSV was set at 37.3.
These Ct cut-offs were further verified in clinical studies.
2. Comparison studies:
a. Method comparison with predicate device:
{25}
Not applicable. Performance of the cobas® Liat Influenza A/B & RSV assay was evaluated against the comparator method of an FDA-cleared nucleic acid test in a prospective clinical study.
# b. Matrix comparison:
Not applicable.
# 3. Clinical studies:
# a. Prospective Study:
Clinical performance characteristics of the cobas® Liat Influenza A/B & RSV assay were evaluated in multi-site prospective studies during the 2013-2014 and the 2014-2015 influenza seasons in the U.S. A total of 12 investigational sites (CLIA waived sites) throughout the U.S. participated in the study. To be enrolled in the study, patients had to be presenting at the participating study sites with flu-like symptoms, and provided informed consents.
Nasopharyngeal swab (NPS) samples were prospectively collected and tested at 6 of the 12 sites in the 2013-2014 influenza season from January 2014 to May 2014, and at 9 of the 12 sites in the 2014-2015 influenza season from October 2014 to April 2014. Three (3) of the 12 sites participated in the study during both time periods.
A FDA-cleared multiplexed nucleic acid amplification test for Influenza A, B, and RSV was performed by a central reference laboratory as the comparator method for this study.
One NPS sample, placed into UTM after specimen collection, was used for the cobas® Liat Influenza A/B & RSV assay testing according to the product instructions for use. The remaining sample in UTM was labeled and stored frozen at $-80^{\circ}\mathrm{C}$ for later shipment to the central reference laboratory for comparator testing.
PCR and bi-directional sequencing were used to investigate discordant results between the cobas® Liat Influenza A/B & RSV assay and the comparator test. Nested PCR was used to increase the sensitivity of the PCR/sequencing method which uses primers that are distinguished from those used in the cobas® Liat Influenza A/B & RSV assay.
A total of 1421 nasopharyngeal swab (NPS) specimens eluted in VTM were enrolled in this study. Of those, 41 specimens did not meet eligibility criteria. Additionally, 17 and 13 specimens were excluded due to invalid results from the cobas® Liat Influenza A/B & RSV assay and the comparator assay, respectively. A total of 1350 prospective NPS specimens were included in the performance analysis. Patient age and gender distribution for all the 1421 prospective specimens is presented in Table 15 below.
Table 15: Patient Demographics - Prospective Study
| Age Group | Number | Percent of Patients |
| --- | --- | --- |
| ≤ 5 years | 346 | 24.3% |
| 6 to 21 years | 428 | 30.1% |
| 22 to 59 years | 514 | 36.2% |
| ≥ 60 years | 133 | 9.4% |
| Total | 1421 | 100% |
| Sex | Number | Percent of Patients |
| Male | 692 | 48.7% |
| Female | 729 | 51.3% |
| Total | 636 | 100% |
{26}
Compared to the comparator test, the performance of the cobas® Liat Influenza A/B & RSV assay for influenza A, influenza B and RSV are presented in Table 16, Table 17, and Table 18 below, respectively:
Table 16: Influenza A Performance – Prospective NPS Specimens
| Flu A (Prospective) | Comparator Positive | Comparator Negative | Total |
| --- | --- | --- | --- |
| Liat Positive | 172 | 47^{a} | 219 |
| Liat Negative | 3 | 1128 | 1131 |
| Total | 175 | 1175 | 1350 |
| Positive Percent Agreement (PPA) = 98.3% (172/175), 95%CI: 95.1% to 99.4% | | | |
| Negative Percent Agreement (NPA) = 96.0% (1128/1175), 95%CI: 94.7% to 97.0% | | | |
a 41 Liat positive, comparator negative specimens were tested by PCR/sequencing. Of these, 18 were positive and 23 were negative by PCR/sequencing.
Table 17: Influenza B Performance – Prospective NPS Specimens
| Flu B (Prospective) | Comparator Positive | Comparator Negative | Total |
| --- | --- | --- | --- |
| Liat Positive | 40 | 8^{a} | 48 |
| Liat Negative | 2 | 1300 | 1302 |
| Total | 42 | 1308 | 1350 |
| Positive Percent Agreement (PPA) = 95.2% (40/42), 95%CI: 84.2% to 98.7% | | | |
| Negative Percent Agreement (NPA) = 99.4% (1300/1308), 95%CI: 98.8% - 99.7% | | | |
a 6 Liat positive, comparator negative specimens were tested by PCR/sequencing. Of these, 5 were positive and 1 was negative by PCR/sequencing.
Table 18: RSV Performance – Prospective NPS Specimens
| RSV (Prospective) | Comparator Positive | Comparator Negative | Total |
| --- | --- | --- | --- |
| Liat Positive | 96 | 16^{a} | 112 |
| Liat Negative | 3 | 1235 | 1238 |
| Total | 99 | 1251 | 1350 |
| Positive Percent Agreement (PPA) = 97.0% (96/99), 95%CI: 91.5% to 99.0% | | | |
| Negative Percent Agreement (NPA) = 98.7% (1235/1251), 95%CI: 97.9% to 99.2% | | | |
a 15 Liat positive, comparator negative specimens were tested by PCR/sequencing. Of these, 3 were positive and 12 were negative by PCR/sequencing.
Compared to the comparator test, the performance of the cobas® Liat Influenza A/B & RSV assay for influenza A, influenza B and RSV in the prospective study was also analyzed by study site and by patient age group. The cobas® Liat Influenza A/B & RSV assay demonstrated similar performance for influenza A, influenza B and RSV across study sites and patient age groups.
The Liat Influenza A/B & RSV assay detected six mixed influenza A, B, and RSV infections in this prospective clinical evaluation. One (1) sample tested positive for influenza A and influenza B by the Liat Influenza A/B & RSV assay, while the comparator assay and/or PCR-sequencing results were negative for influenza A and influenza B. Four (4) samples tested positive for influenza A and RSV, while the comparator assay was positive for influenza A and RSV for two of these samples, positive for influenza A and negative for RSV for one sample, and negative for influenza A and positive for RSV for one sample; PCR-sequencing of the two discrepant samples yielded negative results. One (1) sample tested positive for influenza B and RSV, and the comparator assay result was also positive for influenza B and RSV. No samples tested positive for influenza A, B, and RSV.
The following statement is included in the "Reasons to Repeat the Assay" section of the Product
{27}
Insert: "Dual infections of Influenza A and Influenza B are rare. If the test result is "Influenza A Detected" and "Influenza B Detected", the assay should be repeated with the same patient specimen, or if possible, with a newly collected specimen. Specimens that have repeat "Influenza A Detected" and "Influenza B Detected" results should be sent to a laboratory for confirmatory testing."
# b. Retrospective Study:
In addition to the prospective clinical study, retrospective NPS samples were obtained from two reference laboratories. Specimens obtained were collected in the 2013 - 2014 influenza season. All samples were archived at $-80^{\circ}\mathrm{C}$ upon receipt at IQuum. A total of 300 selected retrospective samples were distributed to and tested at 3 of the 12 sites. The retrospective samples were worked into the daily workload of those sites for testing.
The same FDA-cleared multiplexed nucleic acid amplification test for Influenza A, B and RSV that was used in the prospective study was also the comparator test for this study.
The same PCR and bi-directional sequencing method that was used in the prospective study for investigating discordances between the cobas® Liat Influenza A/B & RSV assay and the comparator method was also utilized in this study.
Of the 300 enrolled NPS specimens, no cobas® Liat Influenza A/B & RSV assay result was obtained for five specimens due to invalid results. These specimens were excluded from the data analysis, but are reported as a part of the invalid rate. Three (3) specimens had an invalid comparator result (i.e. target and internal control not detected), and were removed from the performance analyses.
As such, 292 retrospective NPS specimens were available for performance analyses.
Compared to the comparator test, the performance of the cobas® Liat Influenza A/B & RSV assay for influenza A, influenza B and RSV in this retrospective study are presented in Table 19, Table 20, and Table 21 below, respectively:
Table 19: Influenza A Performance - Retrospective NPS Specimens
| Flu A (Retrospective) | Comparator Positive | Comparator Negative | Total |
| --- | --- | --- | --- |
| Liat Positive | 76 | 2a | 78 |
| Liat Negative | 1 | 213 | 214 |
| Total | 77 | 215 | 292 |
| Positive Percent Agreement (PPA) = 98.7% (76/77), 95%CI: 93.0% to 99.8% | | | |
| Negative Percent Agreement (NPA) = 99.1% (213/215), 95%CI: 96.7% to 99.7% | | | |
a 1 Liat positive, comparator negative specimen was tested by PCR/sequencing. This sample was negative by PCR/sequencing.
{28}
Table 20: Influenza B Performance – Retrospective NPS Specimens
| Flu B (Retrospective) | Comparator Positive | Comparator Negative | Total |
| --- | --- | --- | --- |
| Liat Positive | 97 | 1 | 98 |
| Liat Negative | 1 | 193 | 194 |
| Total | 98 | 194 | 292 |
| Positive Percent Agreement (PPA) = 99.0% (97/98), 95%CI: 94.4% to 99.8% | | | |
| Negative Percent Agreement (NPA) = 99.5% (193/194), 95%CI: 97.1% to 99.9% | | | |
Table 21: RSV Performance – Retrospective NPS Specimens
| RSV (Retrospective) | Comparator Positive | Comparator N…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.