K153537 · Sol-Millennium Medical, Inc. · FMF · Jun 17, 2016 · General Hospital
Device Facts
Record ID
K153537
Device Name
Sol-M Insulin Syringe, Sol-M TB Syringe
Applicant
Sol-Millennium Medical, Inc.
Product Code
FMF · General Hospital
Decision Date
Jun 17, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Sol-M Insulin Syringe with needle, with the calibration unit of insulin for U-100, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin. The Sol-M TB Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.
Device Story
Sol-M Insulin and TB Syringes are sterile, single-use, 3-piece piston syringes with fixed needles. Components include a polypropylene barrel, plunger, and cap, with a Santoprene gasket. Operation involves manual aspiration of fluids into the barrel via plunger retraction and subsequent discharge via plunger depression. Used in clinical or home settings by healthcare providers or patients for insulin delivery or general fluid management. Output is the delivered fluid volume, visually verified by the user against barrel graduations. Benefits include precise, manual fluid administration.
Clinical Evidence
No clinical data submitted. Substantial equivalence is supported by bench testing verifying compliance with ISO 8537:2007 (insulin syringes), ISO 7886-1:1993 (hypodermic syringes), and ISO 9626:1991 (needle tubing).
Technological Characteristics
3-piece piston syringe; materials: polypropylene (barrel, plunger, cap), Santoprene (gasket); lubricant: Dow Corning Medical Silicone 360; fixed needle; sterilization: Ethylene Oxide gas; volumes: 0.3ml, 0.5ml, 1ml; standards: ISO 8537:2007, ISO 7886-1:1993, ISO 9626:1991, ISO 10993-1.
Indications for Use
Indicated for patients requiring manual aspiration and injection of insulin (Insulin Syringe) or general fluid injection/withdrawal (TB Syringe).
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of a bird or abstract human figures in profile, with three distinct head shapes layered on top of each other. The image is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2016
Sol-Millennium Medical, Inc. Mr. Jim Barley Director of Regulatory Affairs 1735 North Brown Road, Suite 100 Lawrenceville, Georgia 30043
Re: K153537
Trade/Device Name: Sol-M Insulin Syringe and Sol-M TB Syringe Regulation Number: 21 CFR 880.5860 Regulation Name: Piston syringe Regulatory Class: II Product Code: FMF Dated: May 2, 2016 Received: May 10, 2016
Dear Mr. Barley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Jim Barley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration |
| Indications for Use |
| Form Approved: OMB No. 0910-0120 |
|-----------------------------------|
| Expiration Date: January 31, 2017 |
| See PRA Statement below. |
| 510(k) Number (if known) | K153537 |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------|
| Device Name | Sol-M TB Syringe |
| Indications for Use (Describe) | The Sol-M TB Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body. |
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <div> <label><input checked="checked" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </div> <div> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------------------------------------|
| | PSC Publishing Services (301) 443-6740 EF |
{3}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES | |
|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Food and Drug Administration | |
| <b>Indications for Use</b> | |
| Form Approved: OMB No. 0910-0120 | |
| Expiration Date: January 31, 2017 | |
| See PRA Statement below. | |
| 510(k) Number ( <i>if known</i> ) | K153537 |
| Device Name | Sol-M Insulin Syringe |
| Indications for Use ( <i>Describe</i> ) | The Sol-M Insulin Syringe with needle, with the calibration unit of insulin for U-100, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin. |
| Type of Use ( <i>Select one or both, as applicable</i> ) | <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> <span> ☑ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14) Page 1 of 1
{4}------------------------------------------------
## 510(k) Summary for 510(k) # K153537
(As required by 21 CFR 807.92(a))
## 510(k) Summary for the Sol-M Insulin and TB Syringes
#### Date Prepared: June 16, 2016 A. Submitter Information
Sol-Millennium Medical, Inc. 1735 North Brown Road Suite 120 Lawrenceville, GA 30043 Contact: Phone Number:
Trade Names:
- B. Device Information Trade/Proprietary Name: Common name of device: Classification Name: Product Code: Regulatory Class: Classification Number: Reason for 510(k):
C. Predicate Device:
Predicate 510(k) #:
Predicate product code:
Reference Devices:
Jim Barley 404-973-2200
Sol-M Insulin Syringe Sol-M TB Syringe
Sol-M Insulin and TB Syringe Piston syringe Piston syringe 80 FMF II 880.5860 New device
KDL Syringes and Needles
K112057
FMF
Sol-Guard Insulin and Tuberculin Safety Syringe 510(k) # K132681
1 ml Fixed Needle, 3 ml Luer Lock, 5 ml Luer Lock, 10 ml Luer Lock, 20 ml LuerLock Safety Syringe 510(k) # K092430
Invirostripe Luer Lock Syringes (1, 3, 10, 20, 30 And 60 ml 510(k) #K101359
{5}------------------------------------------------
#### D. Device Description
The Sol-M Syringe is a sterile, single use, standard 3 piece piston syringe with a fixed needle. The Sol-M 0.3, 0.5 and 1 ml Insulin Syringe will be labeled with a Units Scale calibrated for U-100 Insulin. The 0.5 and 1 ml TB Syringe will be labeled with a mL Scale.
The piston type syringe is a plastic disposable syringe made of the following components:
- 1. Barrel – The barrel has a scale showing the capacity of the syringe as well as the company brand name with the Do Not Reuse symbol.
- 2. Plunger - The plunger is used to aspirate fluid into the barrel or push fluid out of the barrel.
- 3. Stopper - The Stopper maintains the fluid in the barrel between the base of the barrel and Plunger.
- Cap The Cap is used to cover and protect the tip of the barrel from being 4. damaged.
Following is a listing of model numbers and product descriptions included in this submission:
# Sol-M Insulin Syringes with Fixed Needle
| Sol-M 0.3 ml Insulin Syringe with Fixed Needle | |
|-------------------------------------------------|---------------------------------------|
| REF/Catalog # | Description |
| 16331518 | 0.3ml, 31G x 5/16"/10 Pack PE Bag |
| 16330518 | 0.3ml, 30G x 5/16"/10 Pack PE Bag |
| 16330128 | 0.3ml, 30G x 1/2"/10 Pack PE Bag |
| 16329128 | 0.3ml, 29G x 1/2"/10 Pack PE Bag |
| Sol-M 0.3 ml Insulin Syringe with Fixed Needle | |
| REF/Catalog # | Description |
| 16330516 | 0.3ml, 30G x 5/16"/1 ea. Blister Pack |
| 1633012 | 0.3ml, 30G x 1/2"/ 1 ea. Blister Pack |
| 1632912 | 0.3ml, 29G x 1/2"/ 1 ea. Blister Pack |
| Sol-M 0.5 ml Insulin Syringes with Fixed Needle | |
| REF/Catalog # | Description |
| 16531518 | 0.5ml, 31G x 5/16"/10 Pack PE Bag |
| 16530518 | 0.5ml, 30G x 5/16"/10 Pack PE Bag |
| 16530128 | 0.5ml, 30G x 1/2"/10 Pack PE Bag |
| 16529128 | 0.5ml, 29G x 1/2"/10 Pack PE Bag |
| 16130516 | 0.5ml, 30G x 5/16"/1 ea. Blister Pack |
| 1613012 | 0.5ml, 30G x 1/2"/ 1 ea. Blister Pack |
| 1612912 | 0.5ml, 29G x 1/2"/ 1 ea. Blister Pack |
{6}------------------------------------------------
| Sol-M 1 ml Insulin Syringes with Fixed Needle | |
|-----------------------------------------------|---------------------------------------|
| REF/Catalog # | Description |
| 16131518 | 1ml, 31G x 5/16"/10 Pack PE Bag |
| 16130518 | 1ml, 30G x 5/16"/10 Pack PE Bag |
| 16130128 | 1ml, 30G x 1/2"/10 Pack PE Bag |
| 16129128 | 1ml, 29G x 1/2"/10 Pack PE Bag |
| | |
| 16530516 | 1.0ml, 30G x 5/16"/1 ea. Blister Pack |
| 1653012 | 1.0ml, 30G x 1/2"/ 1 ea. Blister Pack |
| 1652912 | 1.0ml, 29G x 1/2"/ 1 ea. Blister Pack |
## Sol-M TB Syringes with Fixed Needle
| Sol-M 0.5 ml TB Syringes with Fixed Needle | |
|--------------------------------------------|-------------------------------------------|
| REF/Catalog # | Description |
| 180527 | 0.5ml, 27G x 1/2"/1 each Blister Pack |
| 180526IDB | 0.5ml, 26G x 3/8" IDB/1 each Blister Pack |
| 180527T | 0.5ml, 27G x 1/2"/25 each Tray Pack |
| 180527IDB | 0.5ml, 27G x 3/8"/25 Tray Pack |
| Sol-M 1 ml TB Syringes with Fixed Needle | |
| REF/Catalog # | Description |
| 181027 | 1ml, 27G x 1/2"/1 ea. Blister Pack |
| 181026IDB | 1ml, 26G x 3/8"IDB /1 ea. Blister Pack |
| 181026T | 1ml, 26G x 1/2"/25 each Tray Pack |
| 181126IDB | 1ml, 26G x 1/2"IDB/25 each Tray Pack |
| 181026IDB | 1ml, 26G x 3/8”IDB/25 each Tray Pack |
| 181027T | 1ml, 27G x 1/2"/25 each Tray Pack |
| 181127IDB | 1ml, 27G x 1/2"IDB/25 each Tray Pack |
| 181027IDB | 1ml, 27G x 3/8”IDB/25 each Tray Pack |
| 181023T | 1ml. 23G x 1/2"/25 ea. Blister Pack |
The Sol-M Syringes are sterilized by Ethylene Oxide Gas and supplied sterile in blister pack, tray pack or 10 pack.
{7}------------------------------------------------
- E. Statement of Indications for Use
## Sol-M Insulin Syringes
The Sol-M Insulin Syringe with needle, with the calibration unit of insulin for U-100, is a device intended for medical purposes for the manual aspiration of insulin, and for the injection of insulin into parts of the body below the surface skin.
## Sol-M TB Syringe
The Sol-M TB Syringe is intended to be used for medical purposes to inject fluids into or withdraw fluids from the body.
- F. Comparison of Required Technological Characteristics:
Information was submitted to demonstrate that there are no significant differences in technological characteristics between the Sol-M Insulin and KDL Insulin Syringes and the Sol-M TB and KDL Hypodermic Syringes. The following comparison chart shows that the subject device and the predicate device are substantially equivalent:
## Sol-M Insulin and TB Syringe versus the KDL Syringes and Needle
| ELEMENT OF<br>COMPARISON | Sol-M Insulin and<br>TB Syringe | KDL Syringes and Needle |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Syringe type | Piston | Piston |
| | Insulin | Insulin |
| Indications for<br>Use | The Sol-M Insulin Syringe<br>with needle, with the<br>calibration unit of insulin for<br>U-100, is a device intended<br>for medical purposes for the<br>manual aspiration of insulin,<br>and for the injection of insulin<br>into parts of the body below<br>the surface skin. | The sterile Insulin Syringe<br>for single use with needle is a<br>device intended for medical<br>purposes for the<br>manual aspiration of insulin,<br>and for the injection into parts<br>of the body below the surface<br>skin |
| | General Use | General Use |
| Indications for<br>Use | The Sol-M TB Syringe is<br>intended to be used for<br>medical purposes to inject<br>fluids into or withdraw fluids<br>from the body. | The Hypodermic Syringes<br>are intended to be used for<br>medical purposes to inject<br>fluids into or withdraw<br>fluids from the body. |
| Principle of<br>Operation | Three piece piston syringe.<br>Plunger is used to fill syringe<br>as well as discharge the fluid. | Three piece piston syringe.<br>Plunger is used to fill syringe<br>as well as discharge the<br>fluid. |
| Specific Drug<br>Use | Sol-M Insulin syringes -<br>Insulin<br><br>Sol-M TB syringes - General<br>Use | KDL Insulin syringes -<br>Insulin<br><br>KDL Hypodermic syringes -<br>General Use |
| Tip Type | Fixed Needle | Fixed Needle |
| Volume | Insulin Syringes - 0.3, 0.5<br>and 1ml<br><br>TB Syringes - 0.5 and 1 ml | Insulin Syringes - 0.3, 0.5<br>and 1ml<br><br>Hypodermic Syringes – 0.5<br>and 1 ml |
| Gradations<br>Legibility | Bold markings | Bold markings |
| Lubricant<br>Composition | Dow Corning Medical<br>Silicone 360 | Dow Corning Medical<br>Silicone 360 |
| Lubricant<br>amount/cm2 | 1ml-0.078mg/cm2 | 1 ml – 0.078 Mg/cm2 |
| Barrel<br>Transparency | Clear | Clear |
| Delivery<br>Accuracy | Insulin Syringes- Per ISO<br>8537:2007 | Insulin Syringes- Per ISO<br>8537:2007 |
| | Hypodermic Syringes - Per ISO 7886-1:1993 | Hypodermic Syringes - Per ISO 7886-1:1993 |
| Hub/Needle<br>Bond Strength | Per ISO 7886 | Per ISO 7886 |
| Biocompatibility | Per ISO 10993-1 | Per ISO 10993-1 |
| Materials | Barrel, Plunger and Cap –<br>Polypropylene<br><br>Gasket - Santoprene | Barrel, Plunger and Cap –<br>Polypropylene<br><br>Gasket - Santoprene |
| Labeling | Per 21 CFR 801 | Per 21 CFR 801 |
### SIDE BY SIDE COMPARISON TABLE
{8}------------------------------------------------
{9}------------------------------------------------
- G. Summary and Conclusion of Nonclinical and Clinical Tests:
Bench tests were conducted to verify that the proposed devices met all design specifications and were Substantially Equivalent (SE) to the predicate devices. The test results demonstrated that the proposed devices comply with the following standards:
Sol-M TB Syringes
- ISO 7886-1:1993 Sterile Hypodermic syringes for single use Part 1: Syringes for ● manual use;
- ISO 9626:1991 Stainless steel needle tubing for the manufacture of medical ● devices/Amendment:2001
Sol-M Insulin Syringes
- ISO 8537:2007 Sterile single use syringes, with and without needles, for insulin; ●
- ISO 9626:1991 Stainless steel needle tubing for the manufacture of medical ● devices/Amendment:2001
- H. Discussion of Clinical Tests:
None submitted
{10}------------------------------------------------
- Substantially Equivalent Conclusion: I.
The proposed device, Sol-M TB Syringes (various sizes), has been determined to be substantially equivalent to the predicate device, Sterile Hypodermic Syringe for single use, with needle, (various sizes), 510(k) # K112057.
The proposed device, Sol-M Insulin Syringes (various sizes), is determined to be substantially equivalent (SE) to the predicate device, Sterile Insulin Syringe for single use with needle, 510(k) # K112057.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.