The ClariVein IC is indicated for infusion of physician specified agents in the peripheral vasculature.
Device Story
ClariVein IC is a specialty infusion catheter featuring a 360° rotatable fluid dispersion wire connected to a proximal, battery-powered motor drive unit (MDU). The MDU includes a speed selector, handle grip, and syringe support. The device is introduced via a microintroducer and navigated to the treatment site using vascular imaging. Fluid is delivered through an infusion port, surrounding the dispersion wire, and exiting at the distal tip. The rotating wire enhances fluid delivery into the vessel wall. The physician controls the rotation speed (four settings: 2,000–3,500 RPM) and infusion via the MDU trigger. Used in clinical settings by physicians, the device is single-use, sterile, and disposable, with no user-serviceable parts. It benefits patients by facilitating controlled, localized infusion of therapeutic agents.
Clinical Evidence
No clinical or animal testing was performed. Substantial equivalence was supported by bench testing, including biocompatibility (ISO 10993), electrical safety (IEC 60601-1), EMC (IEC 60601-2), and mechanical performance testing (tensile, fatigue, leak, and rotation speed verification).
Technological Characteristics
Sterile, single-use infusion catheter. Features a 360° rotatable dispersion wire driven by an integral 9V battery-powered MDU. Four selectable rotation speeds (2,000–3,500 RPM). Materials are biocompatible, phthalate-free, and latex-free. Complies with ISO 10555-1, ISO 11607-1, ISO 11135-1, and IEC 60601 standards. Sterilized via ethylene oxide (EO).
Indications for Use
Indicated for infusion of physician-specified agents into the peripheral vasculature.
Regulatory Classification
Identification
A continuous flush catheter is an attachment to a catheter-transducer system that permits continuous intravascular flushing at a slow infusion rate for the purpose of eliminating clotting, back-leakage, and waveform damping.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a logo for the U.S. Department of Health & Human Services. The logo features the department's symbol, which consists of a stylized human figure with three profiles facing right, representing the department's focus on people and health. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 6, 2016
Vascular Insight, LLC c/o Lorraine Hanlev Vice President, Regulatory Affairs 1 Pine Hill Drive Two Batterymarch Park Suite 100 Quincy, MA 02169
Re: K153502
Trade/Device Name: ClariVein IC Regulation Number: 21 CFR 870.1210 Regulation Name: Continuous Flush Catheter Regulatory Class: Class II Product Code: KRA Dated: December 1, 2015 Received: December 7, 2015
Dear Ms. Hanley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153502
Device Name ClariVein IC
Indications for Use (Describe)
The Clari Vein IC is indicated for infusion of physician specified agents in the peripheral vasculature.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Vascular Insights. The logo is white text on a gray background. The word "VASCULAR" is on the top line, and the word "INSIGHTS" is on the bottom line. A curved line with a dot in the middle connects the two words.
## 510(k) SUMMARY (Page 1 of 3)
#### Device Name: ClariVein® IC
Date Prepared: 6 January 2016
#### A. Submitter/Sponsor
Vascular Insights. LLC. 1 Pine Hill Drive, Two Batterymarch Park Suite 100 Quincy, MA 02169 Telephone Number/Fax: 617-519-1109/203-350-0311 Contact Person: Lorraine M. Hanley, Vice President Regulatory Affairs
#### B. Device Name
| Trade Name: | ClariVein® IC |
|------------------------|----------------------------------------------------|
| Common/Usual names: | Intravascular Catheter, Infusion Catheter, Cannula |
| Classification Names: | Continuous Flush Catheter |
| Regulation Number | 21 CFR§870.1210 |
| Classification: | Class II |
| ProCode | KRA |
| C. Predicate Device(s) | ClariVein® IC |
| 510(k) | K071468 |
| | |
| Common/Usual names: | Intravascular Catheter, Infusion Catheter, Cannula |
| Classification Names: | Continuous Flush Catheter |
| Regulation Number | 21 CFR§870.1210 |
| Classification: | Class II |
| ProCode | KRA |
#### D. Device Description
The ClariVein IC is a specialty infusion catheter with 360°rotatable fluid dispersion wire connected to a proximally located integral battery powered motor drive unit (MDU). The MDU includes the speed selector, handle grip and syringe support features which facilitate physician-controlled infusion of the selected agent. The ClariVein IC is introduced through a microintroducer. Utilizing vascular imaging, the coaxial catheter sheath with dispersion wire is navigated through the vasculature to the treatment site. Fluid delivered through the catheter assembly's infusion port, surrounds the dispersion wire and exits via an opening at the distal end of the catheter. The Clari Vein IC has no user serviceable parts or capital equipment. It is provided to the user sterile, for single patient use and is fully disposable.
{4}------------------------------------------------
## 510(k) SUMMARY (Page 2 of 3)
## E. Indication For Use
The ClariVein® IC is indicated for infusion of physician specified agents in peripheral vasculature.
## F. Technology Characteristics
The ClariVein IC is a sterile, single use, low profile infusion catheter available in multiple lengths and configurations for physician selection. Its distally located catheter assembly, including the catheter shaft and a 360° rotatable dispersion wire, is connected to the proximally located motor drive unit (MDU) powered by an integral, self-contained 9V battery and circuitry. Infusion is through an opening at the distal end of the catheter and the fluid delivery is enhanced by use of a rotating dispersion wire to deliver the infused fluid into the vessel wall. The rate of rotation of the dispersion wire is physician determined and controlled via the MDU's speed selector. The device provides the user with four optional dispersion wire rotation speeds (RPM); i.e. 2,000 (Low), 2,500 (Medium), 3,000 (Medium-High) and 3,500 (High). Wire rotation is physiciancontrolled by selecting the desired rotation speed and depressing the trigger. The ClariVein IC is provided with labeling which presents instruction for the safe and effective use of the device.
The Clari Vein IC is manufactured of materials demonstrated to be biocompatible for its intended use. The ClariVein IC is not made with phthalate material or natural rubber latex. The ClariVein product does not include medicants. There are no user serviceable parts or capital equipment. The device is entirely disposable upon completion of the procedure.
## G. Performance Data
The ClariVein IC has been assessed in accordance with relevant industry standards. FDA recognized standards, and FDA Guidance documents. The product has been determined to meet all relevant parts of the following industry standards and guidance documents:
| ISO 10555-1 Singular Use Intravascular Catheter | ISO 11607-1 Packaging |
|---------------------------------------------------------------------------------|--------------------------------|
| ISO 10993-1, -4, -5, -7, -10, -11 Biocompatibility | ISTA 2A - Transportation |
| IEC 60601-1 3rd Ed. - Electrical Safety | ISO 11135-1 - Sterilization |
| IEC 60601-2 – Electromagnetic Compatibility (EMC) | ISO 15223-1, EN 1041, Labeling |
| ISO 11070 Sterile single use intravascular introducers, dilators and guidewires | |
## Summary of Nonclinical Tests
Nonclinical testing was performed on the Clari Vein IC to support a determination of substantial equivalence. The results of these tests provide reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements for its intended use. No new safety or performance issues were raised during testing. Testing included the follow:
#### Biocompatibility
| Hemocompatibility - Hemolysis | Hemocompatibility-Coagulation | Cytotoxicity |
|-----------------------------------|-----------------------------------------|-----------------------------------|
| Direct/Indirect Contact | Unactivated Partial Thromboplastin Time | L929 Natural Red Uptake |
| Hemocompatibility | Irritation/Sensitization | Irritation/Sensitization |
| Thrombogenicity | Kligman Maximization | Intracutaneous Reactivity |
| Systemic Toxicity (acute) | Systemic Toxicity - Acute | Systemic Toxicity (acute) |
| Materials Mediated Rabbit Pyrogen | | Materials Mediated Rabbit Pyrogen |
{5}------------------------------------------------
## 510(k) SUMMARY (Page 3 of 3)
#### Performance Testing
| Visual Quality | Corrosion | Labeling Nominals |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| Test demonstrates that there are no | Test demonstrates that there are no | Test demonstrates that labeling |
| extraneous materials. | signs of corrosion. | meets unit expressions. |
| Peak Tensile Tests:<br>Distal Tip to Sheath<br>Hub to sheath at strain relief<br>Hub to Guide Wing<br>Female coupling to male coupling<br>Male coupling to motor drive shaft<br>Wire to female coupling<br>Test demonstrates that all joints meet<br>requirements. | Dispersion Wire Tests:<br>Tensile<br>Fatigue<br>Test demonstrates that all joints<br>meet requirements. | Simulated Use<br>Test demonstrates functionality<br>in a simulated fixture. |
| Liquid leak under pressure<br>Test demonstrates no leakage | Wire rotation Speed<br>Test demonstrates required rotation<br>speed. | Torque Force to drive train<br>Test demonstrates torque<br>requirements. |
| Syringe and check valve functionality<br>Test demonstrates compatible functionality. | | |
#### H. Clinical Testing
Not applicable.
## I. Animal Testing
Not applicable.
## J. Conclusion
Responses to questions posed in the FDA's Substantial Equivalence Decision Flow Chart lead to a determination of "substantial equivalence" for the subject ClariVein IC.
The ClariVein IC is substantially equivalent to the predicate in the following aspects:
- . Design, Function and Manufacture - the subject and predicate devices are substantially equivalent to the technological characteristics of both designs. The predicate ClariVein IC is available in 45 and 65 cm lengths; the proposed ClariVein IC introduces an 85 cm length device.
- Materials - There are no new materials used between the predicate and proposed devices.
- Indications Both the predicate and proposed devices have identical indications for use. ●
- Packaging - The subject and predicate devices utilize similar packaging configurations using industry accepted materials.
- . Sterilization - The subject and predicate devices are both sterilized utilizing a validated EO sterilization process in accordance to ISO 11135.
- . Labeling - Both the subject and predicate device have similar labeling.
Testing was conducted to show that no new risks were identified and that the performance profile is similar to the well-established predicate device cleared for market.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.