← Product Code [ORC](/productcode/ORC) · K153498

# Operio (K153498)

_Toul Meditech AB · ORC · Jul 20, 2016 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153498

## Device Facts

- **Applicant:** Toul Meditech AB
- **Product Code:** [ORC](/productcode/ORC.md)
- **Decision Date:** Jul 20, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5070
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K153498 · Jul 20, 2016 | Operio | Toul Meditech AB | Clinical environment measurements (Swedish University Hospital) | Clinical testing was performed in a hospital setting during ophthalmic surgeries to measure CFU levels and particulate counts at the surgical site and over instruments to demonstrate performance and substantial equivalence. | Ophthalmology; CFU reduction; Particulate counting; Clinical environment |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Clinical performance measurement in a hospital setting | Patients undergoing ophthalmic surgery; Sample Size: 23 patients (31 samples); Number of Sites: 1 (Swedish University Hospital) | Not applicable for this study | CFU levels and particulate counts at the surgical site and over instruments |

## Indications for Use

Operio is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m3 at the surgical site and instruments. The air flow is intended to be directed parallel to the surgical site and/or instruments within: 20" (50 cm) in width, 47" (120 cm) in length and 15" (40 cm) in height. Device effectiveness may not be reliably detectable at a distance of 47 inches (120 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

## Device Story

Operio is a mobile air-handling unit for surgical operating rooms; recirculates ambient air through HEPA filter to create directed, non-turbulent clean air zone. Device uses single-use sterile shield to direct airflow over surgical site and instruments; eliminates need for patient attachment. Operated by surgical staff; positioned via castors. Provides HEPA-filtered air (99.995% efficiency) to reduce particulate matter and microorganisms (< 5 CFU/m3) at surgical site. Output allows clinicians to maintain sterile field during ophthalmic procedures without physical contact with patient or obstruction of workspace. Benefits include reduced risk of surgical site contamination.

## Clinical Evidence

Clinical testing conducted at a Swedish University Hospital during ophthalmic surgery (n=23 patients, 31 samples). Primary endpoints were CFU levels and particulate counts at the surgical site/instruments. Results showed mean CFU of 0.4 CFU/m3 and mean particulate count of 1.1 particles of 0.5 µm/f3. Performance demonstrated statistically significant reduction in contaminants, meeting safety and effectiveness criteria compared to predicate.

## Technological Characteristics

Mobile unit; materials: aluminum, powder-coated steel, ABS plastic, stainless steel. HEPA filtration (Nelior membrane, 99.995% efficiency at 0.3 µm). Airflow: 235 CFM. Power: 120V AC, 60Hz. Sterile single-use shield (gamma-irradiated, 10^-6 SAL). Standards: IEC/EN 60601-1:2006, IEC/EN 60601-1-2:2007.

## Regulatory Identification

Air-handling apparatus for a surgical operating room is a device intended to produce a directed, nonturbulent flow of air that has been filtered to remove particulate matter and microorganisms to provide an area free of contaminants to reduce the possibility of infection in the patient.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an air handling bench apparatus, an air handling room apparatus, or an air handling enclosure apparatus, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

## Predicate Devices

- Air Barrier System ([K123006](/device/K123006.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 20, 2016

Toul Meditech AB Tomas Hansson CEO Tunbytorpsgatan 31 Västerås, 721 37 SE

Re: K153498

Trade/Device Name: Operio Regulation Number: 21 CFR 878.5070 Regulation Name: Air-Handling Apparatus For A Surgical Operating Room Regulatory Class: Class II Product Code: ORC Dated: January 7, 2016 Received: January 14, 2016

Dear Tomas Hansson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Michael J. Ryan -S

for Erin I. Keith, M.S.

Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K153498

Device Name Operio

#### Indications for Use (Describe)

Operio is a portable device for use in a surgical operating room that produces a directed, non-turbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m3 at the surgical site and instruments.

The air flow is intended to be directed parallel to the surgical site and/or instruments within: 20" (50 cm) in width, 47" (120 cm) in length and 15" (40 cm) in height. Device effectiveness may not be reliably detectable at a distance of 47 inches (120 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Toul Meditech. The word "toul" is written in blue, with the "o" inside of a light blue square. Below the word "toul" is the word "meditech" written in a smaller font, also in blue.

## 510(k) Summary

#### Submission information

| Submission date:  | 4th of December 2015                   |
|-------------------|----------------------------------------|
| 510(k) submitter: | Tomas Hansson                          |
|                   | Toul Meditech AB                       |
|                   | Tunbytorpsgatan 31                     |
|                   | 721 37 Västerås, Sweden                |
|                   | Phone: + 46 21 13 50 00                |
|                   | Fax: + 46 21 13 86 45                  |
|                   | Email: tomas.hansson@toulmeditech.com) |

#### Device information

| Trade name:          | Operio                                               |
|----------------------|------------------------------------------------------|
| Common name:         | Air zone unit                                        |
| Classification name: | Air-handling apparatus for a surgical operating room |
| Regulation number:   | 21 CFR 878.5070                                      |
| Product code         | ORC                                                  |
| Device class         | II                                                   |

### Legally marketed predicate device

Identification of the legally marketed predicate device to which Toul Meditech AB claims SE for.

| Code | Manufacturer | Device             | 510(k) # |
|------|--------------|--------------------|----------|
| ORC  | Nimbic, Inc  | Air Barrier System | K123006  |

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for "toul meditech". The word "toul" is in large, bold, blue letters, with the "o" in a light blue square. Below "toul" is the word "meditech" in smaller, blue letters.

### Device description

The air zone unit Operio is used in a surgical operating room for cleaning the air in a given area by re-circulating the ambient air and cleaning it from airborne particles. The air is filtered through a HEPA filter and the air is generated over the areas where the demands of clean air are especially high. The air zone unit Operio is equipped with a control panel for adjustments and for positioning of the unit.

The air zone unit Operio is a mobile unit that with the help of castors easily can be moved around the patient to an optimal position or, when needed, transported inside the hospital between wards. It is also equipped with an optional instrument tray.

By using a unique sterile shield as a protective barrier the air zone unit can be placed close to the OR table and deliver HEPA filtered air to the surgical site and instruments to reduce the presence of airborne particulate and microorganisms.

### Indications for use

Operio is a portable device for use in a surgical operating room that produces a directed, nonturbulent flow of air to the surgical site during ophthalmic surgery and to the sterile instruments used during surgery. The air flow from the device is HEPA-filtered to reduce the presence of particulate matter to reduce the microorganisms to a level of < 5 CFU per m3 at the surgical site and instruments.

The air flow is intended to be directed parallel to the surgical site and/or instruments within: 20" (50 cm) in width, 47" (120 cm) in length and 15" (40 cm) in height. Device effectiveness may not be reliably detectable at a distance of 47 inches (120 cm) from the air flow outlet, and effectiveness depreciates beyond this specified area.

Difference from predicate device:

The air zone unit Operio is suitable for ophthalmic surgery and sterile instruments used during surgery. It is possible to deliver the clean air without coming into contact with the patient or compromising the work space required by surgical staff. The clean air flow can also be used over instruments and the area of the air flow is bigger compared to the predicate device.

The difference in indications for use compared to the predicate device does not introduce any new hazards and test results can confirm that the subject device is as safe and as effective for use.

#### Summary of technological characteristics compared to predicate devices

The subject device Operio differs from the predicate device in that it doesn't use a sterile nozzle and device does not depend on attachment to the patient. The subject device requires a single use sterile shield which functions as an equipment cover as well as to direct the non-turbulent flow to the surgical site. The sterile shield is attached to the unit and is not in contact with the patient.

To eliminate contaminated particles entering the surgical site, HEPA filtration is used by both the predicate device and the subject device. The efficiency of HEPA filtration of the subject

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Toul Meditech. The logo is in blue and features the word "toul" in large, bold letters. The "o" in "toul" is a gradient blue circle. Below "toul" is the word "meditech" in smaller letters.

 device has a higher precision than the predicate device (subject device guarantees 99.995 % compared to 99.97 % for the predicate device).

| Comparison table: |  |
|-------------------|--|
|-------------------|--|

| Characteristic                                                        | Subject device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Predicate device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Review Panel                                                          | General Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | General Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Indications for use                                                   | Operio is a portable device for<br>use in a surgical operating room<br>that produces a directed, non-<br>turbulent flow of air to the<br>surgical site during ophthalmic<br>surgery and to the sterile<br>instruments used during surgery.<br>The air flow from the device is<br>HEPA-filtered to reduce the<br>presence of particulate matter to<br>reduce the microorganisms to a<br>level of < 5 CFU per m3 at the<br>surgical site and instruments.<br><br>The air flow is intended to be<br>directed parallel to the surgical<br>site and/or instruments within:<br>20" (50 cm) in width, 47" (120<br>cm) in length and 15" (40 cm) in<br>height. Device effectiveness may<br>not be reliably detectable at a<br>distance of 47 inches (120 cm)<br>from the air flow outlet, and<br>effectiveness depreciates beyond<br>this specified area. | The Air Barriers system is a<br>portable device for use in a<br>surgical operating room that<br>produces a directed, non-turbulent<br>flow of air to the surgical site. The<br>air flow from the device is HEPA-<br>filtered to reduce the presence of<br>particulate matter and<br>microorganisms at the surgical<br>site during hip arthroplasty and<br>posterior vertebral fusion and<br>laminoplasty.<br><br>The ABS Nozzle is intended to be<br>used only where: (1) it can be<br>placed on an anatomical surface<br>with no gap between the bottom<br>of the nozzle and the surface, (2)<br>the incision plane is parallel with<br>the direction of air flow, and (3)<br>the Incision dimensions are<br>within: 6' (15.2 cm) in width and<br>20b (50.8 cm) in length. Device<br>effectiveness may not be reliably<br>detectable at a distance of 20<br>inches from the Nozzle, and<br>effectiveness depreciates beyond<br>this specified area. |
| Physical<br>Dimension                                                 | 24 x 18 x 47-67 inches                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | 11 x 11 x 25 inches                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Weight                                                                | 92 lb                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | 48 lb                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Material                                                              | Aluminium, powder coated<br>steel, ABS plastic, stainless steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Stainless steel                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Electrical<br>Specifications                                          | 1. Mains power input: 120 V<br>(AC), 60 Hz<br>2. Power consumption: 290 VA                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | 1. Mains power input: 120 V (AC),<br>60 Hz<br>2. Unknown                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Type of device                                                        | Prescription use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Prescription use                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Movability                                                            | Operio is a portable unit.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | The Air Barrier System is a<br>portable unit.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Air filtration                                                        | HEPA filtered air with at least<br>99.995% efficiency against 0.3 µm particles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | HEPA filtered air with at least<br>99.97% efficiency against 0.3 µm particles                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| HEPA filter<br>media                                                  | Nelior membrane media.<br>Folded sheets of nelior.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Continuous pleated glass<br>microfiber                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Air inlet Pre-filter<br>Media                                         | Polyamide mesh                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Polyurethane Foam                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Regular<br>maintenance                                                | HEPA filter to be replaced after<br>2000 hours/yearly.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Annual                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Airflow capacity                                                      | Airflow is 235 CFM (400 m3/h)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Airflow 150 CFM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Efficiency                                                            | Bacteria carrying particles at the<br>surgical site is 0.4 CFU/m3.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Bacteria carrying particles in the<br>surgical incision is 1.57 CFU/m3                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Air volume<br>delivered to<br>incision in CFM                         | Airflow over surgical site is 235<br>CFM (400 m3/h)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | 41                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Measured<br>average<br>particulate<br>density at<br>surgical incision | 1.1 particulate of 0.5 µm/f³                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 68,122 particulate of 0.5 µm/f³                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Air flow position                                                     | Air flow is provided over the<br>surgical site and instruments. No<br>attachment to the patient is<br>necessary.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  | Air flow is provided to an<br>attachment, nozzle, attached on the<br>patient near the surgical site.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Sterilization                                                         | Single-use sterile shield.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Sterile single use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Sterility<br>Assurance (SAL)                                          | 10-6 SAL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | 10-6 SAL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Sterilization<br>Method                                               | Gamma irradiation                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | ETO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for "toul meditech". The word "toul" is in a bold, sans-serif font, with the "o" having a gradient fill. Below "toul" is the word "meditech" in a smaller, sans-serif font. The overall color scheme is blue.

### Summary of testing

The performance characteristics of the air zone unit Operio have been obtained through both a series of tests, clinical and non-clinical, and performance specifications. Particulate test and CFU test results clearly meet the criteria for achieving reduction of airborne contamination at the intended surgical site and over instruments. The efficiency of HEPA filtration has been tested and is found better than the predicate device.

In testing, the predicate demonstrated a level of particulate presence by 68,122 particles of 0.5 um/f³, while the subject device had a level of particulates by 1.1 particles of 0.5 µm/f3.

The predicate device demonstrated a reduction from 10.73 to 1.60 CFU's whereas the subject device demonstrated a reduction from 48.2 to 0.4 CFU in a clinical environment.

The following non-clinical tests has been conducted: Simulated use test, Particulate counting test, Colony Forming Units counting test, Air velocity test, Air leakage test, Smoke test and Turbulence tests to demonstrate air flow patterns.

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for TOUL Meditech. The logo is in blue and features the word "toul" in large letters, with the "o" in a square. Below the word "toul" is the word "meditech" in smaller letters. The logo is simple and modern.

The following standards have been used for performance testing:

- IEC/EN 60601-1:2006 Medical electrical equipment Part 1: General requirements ● for basic safety and essential performance
- IEC/EN 60601-1-2:2007 Medical electrical equipment - Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances - Requirements and tests

### Summary of clinical testing

Clinical tests have been carried out in surgical procedures such as Ophthalmology. Measurements where made near the surgical site and over the instruments where the results demonstrated a statistically significant reduction in CFU levels as well as particulates. The CFU value is based on 31 samples from 23 patients. The clinical tests have been carried out in a Swedish University Hospital. As a mean value, the subject device had a CFU counting of 0.4. For particulate counting, the subject device had a mean value of 1.1 particles of 0.5 um/f3.

Since both particulate counts and CFU level doesn't exceed the predicate device in a clinical environment it is concluded that substantial equivalence is achieved as clinical validation.

#### Conclusion

Based on a comparison of the performance characteristics, the tests performed, clinical and nonclinical, it is concluded that the subject device Operio is substantially equivalent to the predicate device and therefore is as safe and as effective.

---

**Source:** [https://fda-staging.innolitics.com/device/K153498](https://fda-staging.innolitics.com/device/K153498)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
