The Dillard Airway Dilatation System is intended to dilate strictures of the airway tree.
Device Story
Dillard Airway Dilatation System is a balloon catheter used to dilate airway strictures. The device features a balloon at the distal end, a central lumen for guidewire use, and a proximal hub with a luer connector for inflation. It is operated by a clinician who uses a standard inflation device to apply hydraulic pressure to the balloon, expanding it to dilate the stricture. Radiopaque markers on the balloon assist with radiographic positioning. The device is supplied sterile for single use. By mechanically dilating the airway, the system aims to improve patient breathing and airway patency.
Clinical Evidence
Bench testing only. Performance testing included dimensional specifications, joint separation strength, accessory compatibility, balloon burst pressure, inflation/deflation times, balloon cycle fatigue, and balloon deflation challenge. Biocompatibility testing was conducted in compliance with ISO 10993-1 and FDA Bluebook Memorandum G-95-1, including cytotoxicity, sensitization, and mucosal irritation.
Technological Characteristics
Balloon catheter system; materials include Nylon (balloon), Nylon, Pebax/Plexar/HDPE tri-extrusion (catheter), and Platinum-Iridium (marker bands). Operates via hydraulic pressurization. Dimensions: 25cm/100cm working lengths, 5-12mm balloon diameters. Single-use; sterilized via 100% Ethylene Oxide gas. Compatible with 0.035" guidewires.
Indications for Use
Indicated for patients requiring dilation of strictures within the airway tree.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, ribbon-like appearance. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 27, 2016
Intuit Medical Products, LLC Mr. Jack Griffis VP, Research & Development 6018 Eagles Rest Trail Sugar Hill, GA 30518
Re: K153484
Trade/Device Name: Dillard Airway Dilatation System Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: KTI Dated: June 22, 2016 Received: June 29, 2016
Dear Mr. Griffis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K153484
Device Name Airway Dilatation System
Indications for Use (Describe) The Dillard Airway Dilatation System is intended to dilate strictures of the airway trees.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a profile view of a person's head and neck, with the internal structures of the nasal passages, mouth, and throat visible. These structures are highlighted in shades of blue and purple. To the right of the head is a blue rectangle containing the text "IMP" in large, bold letters, followed by "Intuit Medical Products, LLC" in smaller text below.
## 510(k) Summary
510(k) Number: K153484
July 25th, 2016 Date Revised:
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92.
- A. Submitter: Intuit Medical Products (IMP), LLC 6018 Eagles Rest Trail Sugar Hill, Georgia 30518
- B. Company Contact: Jack Griffis Vice President, Research & Development (404) 583-6889 (direct) jgriffis@intuitmedicalproducts.com
| C. | Device Information: | |
|----|---------------------|--------------------------------------------------------------------------|
| | Trade Name: | Dillard Airway Dilatation Catheter |
| | Common Name: | Airway Dilatation System |
| D. | Classification: | Bronchoscope (flexible or rigid) and accessories<br>KTI, 21 CFR 874.4680 |
- E. Predicate Device: Acclarent Airway Dilation Catheter (Inspira®), K090660
Class II
#### F. Physical Description:
The Dillard Airway Dilatation System includes a balloon catheter whose working length is 25cm or 100cm, and is available in balloon diameters from 5mm to 12mm and balloon lengths of 20mm and 40mm. The balloon component of the catheter has radiopaque markers to assist with radiographic positioning (as applicable). The proximal end of the device is a common catheter design consisting of a plastic hub and strain relief. The hub is used to inflate the balloon and the luer connector integrated into the hub is compatible with standard inflation devices. A second lumen within the catheter, intended for guidewire use, extends through the central lumen back to the proximal hub and through the distal tip. All catheters and accessories are supplied sterile and intended for single use.
NOTE: the proposed device is identical to that previously cleared under K143738, with the exception of additional balloon sizes and indication for use in airway dilatation.
{4}------------------------------------------------
- G. Indications for Use: The Dillard Airway Dilatation System is intended to dilate strictures of the airway tree.
- H. Comparison of Characteristics / Performance Testing / Substantial Equivalence: The Dillard Airway Dilatation System is substantially equivalent to the predicate device in intended use, indications for use, fundamental scientific technology, and important performance specifications. Refer to the Table below for a summary:
| Parameter | DAS Air Balloon Catheter<br>(Proposed) | Acclarent Inspira® AIR Balloon Catheter<br>(Predicate, K090660) |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Intended to dilate strictures of the airway tree. | |
| Principle of Operation | Operates on the principle of hydraulic pressurization applied through an inflatable<br>balloon attached to the distal end | |
| Guidewire Style | OTW | OTW and Integrated |
| Guidewire Compatibility | OTW: 0.035" | Integrated: Unknown<br>OTW: 0.035" |
| Catheter Lengths | 25cm and 140cm | 30cm |
| Balloon Lengths | 20 and 40mm | 24 and 40mm |
| Balloon Diameters | 5.0 thru 12.0mm | 5.0 thru 16.0mm |
| Nominal Inflation<br>Pressure | 6 ATM for all catheters | 6 ATM |
| MAX Recommended<br>Inflation Pressure | 12 ATM for balloon diameters 5.0<br>thru 10.0mm<br>10 ATM for balloon diameters<br>11.0 and 12.0mm | 16 ATM for balloon diameters 5.0 thru 7.0mm<br>12 ATM for balloon diameters 8.0 thru 11mm<br>10 ATM for balloon diameters 12.0 thru 14.0mm<br>8 ATM for balloon diameter 16.0mm |
| Single Use? | Yes | |
| Sterilization Method | 100% Ethylene Oxide gas | |
| Packaging | Tyvek-Mylar Pouch, Protective Tube, Stylet | |
| Balloon Material | Nylon | Nylon blend |
| Catheter Materials | Nylon, Pebax/Plexar/HDPE Tri-<br>Extrusion, Platinum-Iridium<br>Marker Bands | Polymer Extrusions (unknown), Platinum-Iridium<br>Marker Bands |
In addition, the device was subjected to the following performance tests to support the assertion of substantial equivalence:
- Dimensional specifications (including compliance)
- l Joint separation strength
- ' Compatibility with (standard) accessories
- l Balloon burst pressure
Image /page/4/Picture/9 description: The image shows a medical illustration of the human head and neck, with the nasal passages, sinuses, and throat highlighted in blue and purple. To the right of the illustration is the text "IMP Intuit Medical Products, LLC" in a blue box. The illustration is a side view of the head and neck, showing the internal structures of the respiratory system.
{5}------------------------------------------------
- Inflation and deflation times
- Balloon cycle fatigue
- Balloon deflation challenge testing
- Biocompatibility Testing in Compliance with the ISO 10993-1 and the FDA Bluebook Memorandum (G-95-1) as follows (and where applicable):
- O Cytotoxicity
- Sensitization (Guinea Pig Maximization) o
- Mucosal (oral) Irritation O
No new questions of safety or effectiveness were identified during device testing; therefore, the Dillard Airway Dilatation System is considered substantially equivalent to the predicate device.
Jack Griffis Vice President, Research & Development
Image /page/5/Picture/11 description: The image shows a logo for Intuit Medical Products, LLC. The logo features a side profile of a person's head and neck, with the inside of the head colored in shades of blue and purple to represent the respiratory system. To the right of the head is a light blue rectangle with the letters "IMP" in large, bold, white font. Below "IMP" are the words "Intuit Medical Products, LLC" in a smaller, white font.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.