The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0mm to 7.0 mm, using OCT-assisted orientation as an adjunct to fluoroscopy. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.
Device Story
The Pantheris System is a peripheral atherectomy catheter used by physicians in clinical settings to remove plaque from partially occluded peripheral vessels. The system integrates Optical Coherence Tomography (OCT) imaging to provide real-time, high-resolution visualization of the vessel wall, allowing the physician to orient the cutting element precisely toward the plaque under fluoroscopic guidance. The system consists of the Pantheris Catheter, a Lightbox Sled, and the Lightbox HS Imaging Console. The physician operates the catheter to excise plaque; the OCT imaging assists in distinguishing plaque from healthy vessel wall, potentially reducing the risk of vessel wall injury. The output is a visual display of the vessel interior, which informs the physician's decision-making during the atherectomy procedure. The device is designed for use in peripheral vasculature, excluding iliac, coronary, cerebral, renal, or carotid arteries.
Clinical Evidence
No human clinical data provided. Evidence consists of bench-top design verification (mechanical testing, life cycle, tensile strength, OCT image generation) and preclinical animal testing. A porcine study was conducted to assess acute and chronic histopathological outcomes, in vivo performance, OCT marker visibility, and thrombogenicity compared to the predicate device. All testing confirmed the device meets established specifications.
Technological Characteristics
Peripheral atherectomy catheter system utilizing OCT-assisted orientation. Components include a catheter with an apposition balloon, torque shaft, and handle assembly. Imaging is provided via the Lightbox HS Imaging Console. Connectivity involves a Sled interface for OCT data. Sterilization and packaging validation were performed. Technical specifications include 7Fr catheter option, 110cm working length, and integrated flush/venting systems.
Indications for Use
Indicated for patients requiring atherectomy to remove plaque from partially occluded peripheral vessels (3.0mm to 7.0mm diameter). Contraindicated for use in iliac, coronary, cerebral, renal, or carotid vasculature.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like figure with three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 1, 2016
Avinger, Inc. Ms. Patty Hevey Vice President, Clinical and Regulatory Affairs 400 Chesapeake Drive Redwood City, CA 94063
Re: K153460
Trade/Device Name: Pantheris System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: January 27, 2016 Received: January 28, 2016
Dear Ms. Hevey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153460
Device Name Pantheris System
Indications for Use (Describe)
The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0mm to 7.0 mm, using OCT-assisted orientation as an adjunct to fluoroscopy. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|
| <div> <span> <input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
510(k) Notification K_________________________________________________________________________________________________________________________________________________________
GENERAL INFORMATION
Applicant:
Avinger, Inc. 400 Chesapeake Drive Redwood City, CA 94063 U.S.A. Phone: 650-241-7900 Fax: 650-241-7901
#### Contact Person:
Patty Hevey VP, Clinical and Regulatory Affairs Avinger, Inc. Phone: 650-222-3666 Email: phevey@avinger.com
Date Prepared:
November 30th, 2015
#### DEVICE INFORMATION
Trade Name:
Pantheris System
Generic/Common Name:
Peripheral Atherectomy Catheter
Classification:
21 CFR§870.4875, Intraluminal Artery Stripper, Class II
Product Codes: MCW
{4}------------------------------------------------
#### PREDICATE DEVICES
The modified Pantheris Catheter is substantially equivalent to the original Pantheris Catheter cleared under K152275.
### DEVICE DESCRIPTION
The Pantheris System recently received clearance under K152275 to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature using OCTassisted orientation. The Pantheris Catheter reviewed as part of K152275 has since been modified to help improve manufacturability and usability of the catheter. The Pantheris System consists of the Pantheris Catheter, Lightbox Sled (referred to as Sled), the Lightbox HS Imaging Console and the Sterile Drape (accessory).
# INDICATIONS FOR USE
The Pantheris System is intended to remove plaque (atherectomy) from partially occluded vessels in the peripheral vasculature with a reference diameter of 3.0mm to 7.0 mm, using OCT-assisted orientation as an adjunct to fluoroscopy. The Pantheris System is NOT intended for use in the iliac, coronary, cerebral, renal or carotid vasculature.
# SUBSTANTIAL EQUIVALENCE
The modified Pantheris Catheter is essentially the same device as the original Pantheris Catheter cleared under K152275; both share the same basic design, principles of operations, mechanism of action and indications for use. Any differences between the modified Pantheris Catheter and the original Pantheris Catheter do not alter the intended use of the Pantheris System. The modifications made to the original Pantheris Catheter (K152275) to create the proposed device are provided in Table 1 below. The information included in this submission will establish the substantial equivalency of the modified Pantheris Catheter to the original Pantheris Catheter cleared under K152275.
{5}------------------------------------------------
| Catheter Section | Design Modifications |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Distal Assembly | Offering a 7Fr catheter option Modifying the shape of the Apposition Balloon Including additional viewing windows and increasing the length of the Nosecone Housing Replacing the flush channel with vent holes |
| Torque Shaft Assembly | Reducing the working length from 130cm to 110cm Relocating the Occlusion Balloon to the Torque Shaft (thereby obsoleting the optional accessory, Occlusion Sheath) Increasing the Torque Shaft stiffness |
| Handle Assembly | Improving the ergonomics of the handle design Extending the blue color coding across the entire Flush Lumen length |
| Packaged Accessories | Replacing 1mL Syringe with 6mL and 10mL Syringes |
# Table 1: Modified Pantheris Catheter Changes
# NON-CLINICAL TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION
To demonstrate the substantial equivalences of the Pantheris Catheter to the selected predicate device, the performance and technological characteristics were evaluated by completion of the following testing:
- o Design Verification
- Working Length O
- Catheter Flush O
- OCT Image Generation and Sled Interface O
- O Catheter Field of View
- Distal Tip Rotation O
- Guidewire compatibility O
- Distal Tip Max OD O
- Insertion Force O
- Force to Overcome User Slide Lock O
- Force to Remove Sled Bag o
- Force to Remove Sled Bag ring O
- Full 360° Image O
- O Inflation Cycles
- Cut/Pack Cycles O
- Packed Position Life Cycle O
- Active position Life Cycle O
- Balloon Burst O
{6}------------------------------------------------
- o Torque Shaft Torque Proof loading
- Drive Shaft Torque Proof loading o
- Inflation Luer Tensile Strength O
- Flush lumen Tensile Strength O
- Distal Catheter Joints Tensile Strength O
- Proximal Catheter Joints Tensile Strength O
- Simulated Use (Design Validation) 0
- Setup (Ease of Use & Sterility) O
- Functionality (Visualization, Flushing, Device Tracking, Target Sites, o Treatment, Markings, Safety)
- Insertion/Retraction force o
- Number of Insertions O
- Post-Retraction (Visualization, Device Tracking, Treatment, Markings, o Safety)
- Sterilization Validation ●
- Packaging Validation ●
- Product Shelf Life ●
- Biocompatibility ●
- O Cytotoxicity
- Sensitization O
- Intracutaneous Reactivity O
- o Systemic Toxicity
- Hemolysis extract O
- Hemolysis direct O
- Complement Activation o
- Material Mediated (Rabbit) Pyrogenicity о
- Preclinical Animal Testing ●
- O Porcine Study with acute and chronic histopathological assessments and in vivo performance comparison to a predicate device, with 8F Pantheris catheter
- Functionality o
- Radiopacity
- OCT marker visibility ■
- 트 Insertion/Retraction
- Advance to target location ■
- Ability to orient device to a landmark ■
- Minimum Effective Flow Rate (MEFR) for clear OCT imaging o
- Thrombogenicity O
- Safety Assessment O
{7}------------------------------------------------
The collective results of the non-clinical testing demonstrate that the modified Pantheris Catheter meets the established specifications necessary for consistent performance for its intended use.
#### CONCLUSION
The modified Pantheris Catheter has been carefully compared to the original Pantheris Catheter device with respect to intended use and fundamental technological characteristics. Performance testing was conducted to verify and validate the performance of the devices and ensure the modified Pantheris Catheter functions as intended and meet design specifications. The comparison and performance testing results demonstrate that the modified Pantheris Catheter is substantially equivalent to the original Catheter cleared under K152275 for the stated intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.