K153449 · Fresenius Medical Care Renal Therapies Group, LLC · KDI · Apr 26, 2016 · Gastroenterology, Urology
Device Facts
Record ID
K153449
Device Name
2008K2 Hemodialysis Machine
Applicant
Fresenius Medical Care Renal Therapies Group, LLC
Product Code
KDI · Gastroenterology, Urology
Decision Date
Apr 26, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5860
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Fresenius 2008K2 Hemodialysis Machine is intended for acute and chronic dialysis therapy.
Device Story
The 2008K2 Hemodialysis Machine performs hemodialysis by pumping patient blood through an extracorporeal circuit with a semi-permeable dialyzer membrane. It facilitates toxin removal via diffusion and excess water removal via ultrafiltration into a dialysate circuit. The machine mixes dialysate concentrates with purified water, heats, degasses, and delivers the fluid to the dialyzer. Balancing chambers regulate fluid flow. The system monitors venous/arterial blood pressures, air, and blood presence. Used in healthcare facilities by trained personnel. Output includes real-time monitoring of treatment parameters (blood/dialysate flow, fluid removal, heparin delivery). Healthcare providers use these outputs to manage dialysis therapy, ensuring patient safety and therapeutic efficacy.
Clinical Evidence
No clinical studies were performed. Evidence consists of bench testing, including software verification, mechanical testing (fluid removal/flow accuracy, heparin delivery), and electrical safety/EMC evaluation per IEC 60601-1-2:2007.
Technological Characteristics
Medical electrical equipment with programmable system. Materials: PAEK, PEI, PESU, PSU, PET, PUR, PP, PPO, PPS, PPSU, PTFE, PVDF, EPDM, Silicone, Stainless Steel, Titanium, Tantalum, Tungsten, Borosilicate Glass. Hydraulic system controls fluid flow, mixing, heating, and alarms. Connectivity: Standalone. Sterilization: Not specified. Software controls treatment functions.
Indications for Use
Indicated for acute and chronic hemodialysis therapy in a healthcare facility. Includes Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis for patients ≥ 20 kg. SLED is contraindicated for patients ≤ 40 kg. Not for plasma replacement, patients < 20 kg, or renal therapies using substitution fluid.
Regulatory Classification
Identification
A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).
Special Controls
*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 26, 2016
Fresenius Medical Care Renal Therapies Group, LLC Denise Oppermann Senior Director, Regulatory Affairs 920 Winter Street Waltham, MA 02451
Re: K153449
> Trade/Device Name: 2008K2 Hemodialysis Machines Regulation Number: 21 CFR§ 876.5860 Regulation Name: High Permeability Hemodialysis System Regulatory Class: II Product Code: KDI Dated: March 31, 2016 Received: April 1, 2016
Dear Denise Oppermann,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153449
Device Name 2008K2 Hemodialysis Machines
### Indications for Use (Describe)
The 2008K2 Hemodialysis Machine is indicated for acute and chronic hemodialysis therapy in a healtheare facility.
Additional therapy options for patients receiving hemodialysis include: Isolated Ultration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing ≥ 20 kg). This machine accommodates the use of both low flux and high flux dialyzers. The SLED therapy option is not to be used for patients weighing ≤ 40 kg. The 2008K2 Hemodialysis Machine is not to be used for plasma replacement therapies, for patients weighing less than 20 kg, or for renal therapies using substitution fluid.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| <div> <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Fresenius Medical Care. The logo consists of a blue geometric shape on the left and the text "FRESENIUS MEDICAL CARE" on the right. The geometric shape is made up of three downward-pointing chevrons stacked on top of each other. The text is in a bold, sans-serif font, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
### 510(K) SUMMARY 5.
This 510(k) Summary is in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content of this 510(k) summary is provided in conformance with 21 CFR § 807.92.
### 5.1. Submitter's Information
| Name: | Fresenius Medical Care Renal Therapies Group, LLC |
|-------------------|-------------------------------------------------------------------|
| Address: | 920 Winter Street<br>Waltham, MA<br>02451-1457 |
| Phone: | (781) 699-4479 |
| Fax: | (781) 699-9635 |
| Contact Person: | Denise Oppermann, Senior Director<br>Regulatory Affairs – Devices |
| Preparation Date: | 27 November 2015 |
### 5.2. Device Name
| Trade Name: | 2008K2 Hemodialysis Machine |
|------------------------------------------|-----------------------------------------------------------------|
| Common Name: | Dialyzer, High Permeability With Or<br>Without Sealed Dialysate |
| Classification Name : | High Permeability Hemodialysis System |
| Regulatory Class: | Class II per 21 CFR § 876.5860 |
| Product<br>Code/Classification<br>Panel: | KDI<br>KDI/Gastroenterology-Urology |
### 5.3. Legally Marketed Predicate Device
2008K Hemodialysis Machine, originally cleared under K994267. The 2008K Hemodialysis Machine was most recently cleared under K150708 for hardware, mechanical, software and labelling modifications required to address issues identified in the field.
### 5.4. Device Description
### 5.4.1. Device Identification:
The 2008K2 Hemodialysis Machine will be available in the configurations described in Table 1.
{4}------------------------------------------------
| 2008K2 Hemodialysis Machine |
|-----------------------------|
| Traditional 510(k) |
| Product<br>Line | Code | Part Number Description |
|-----------------|--------|------------------------------------------------------------------------|
| 2008K2 | 190610 | 2008K2 Hemodialysis System Online Clearance (OLC)/Diasafe<br>Plus (DP) |
| 2008K2 | 190628 | 2008K2 OLC only with Heparin Pump (HP) |
| 2008K2 | 190630 | 2008K2 Machine, Short Cabinet, OLC/DP, HP |
| 2008K2 | 190633 | 2008K2 Hemodialysis System, Optional Module* |
| Table 1: | 2008K2 Hemodialysis Machines |
|----------|------------------------------|
|----------|------------------------------|
*The optional module indicates a version where the heparin pump is placed in a lower cabin bay area.
#### 5.4.2. Device Characteristics
The machine described in this submission is medical electrical equipment containing a programmable electrical medical system. Software controls the functions of the machines during hemodialysis treatment, including fluid flow, mixing, heating, and alarms.
#### 5.4.3. Environment of Use
The 2008K2 Hemodialysis Machine is to be used in healthcare facilities.
### 5.4.4. Brief Written Description of the Device
The 2008K2 Hemodialysis machines is used for dialysis therapy. The machine pumps blood from the patient through an extracorporeal circuit containing a semi-permeable membrane (dialyzer). This membrane acts as an artificial kidney to transport toxins (diffusion) and excess water (ultrafiltration) from the blood into a separate fluid (dialysate) circuit. During this process, the extracorporeal blood circuit is monitored for venous and arterial blood pressures, and for the presence of air and blood. In the dialysate circuit, the dialysate concentrates are mixed with purified water, heated, degassed, and delivered to the dialyzer. Balancing chambers control the incoming flow and outgoing flow of the dialysate fluid during ultrafiltration.
### 5.4.5. Materials of Use
The 2008K2 Hemodialysis Machine's hydraulic system is composed of the following indirect, prolonged contact, externally communicating materials:
# Plastic/Rubber:
PAEK (Polyaryletherketone) PEI (Polyetherimide) PESU (Polyethersulfone) PSU (Polysulfone) PET (Polyethylene terephthalate) PUR (Polyurethane)
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Fresenius Medical Care. On the left is a blue icon that looks like three chevrons stacked on top of each other, pointing downwards. To the right of the icon, also in blue, is the text "FRESENIUS MEDICAL CARE" stacked on two lines.
PET (Polyethylene terephthalate) PP (Polypropylene) PPO (Polyphenylene oxide) PPS (Polyphenylene Sulfide) PPSU (Polyphenylsulfone) PTFE (Polytetrafluoroethylene) PVDF (Polyvinylidene fluoride) EPDM (Ethylene Propylene Diene Monomer Rubber) Silicone
Metals: Stainless Steel Titanium Tantalum Tungsten
Glass: Borosilicate Glass
The hydraulic lines of the machines are in contact with the dialysate circuit. The dialysate circuit has prolonged, indirect blood contact. No modifications to the dialysate circuit are proposed.
### 5.4.6. Key Performance Specifications/Characteristic
Key performance characteristics of the 2008K2 Hemodialysis Machine are:
- Blood flow rates
- Dialysis fluid flow rates ●
- Net fluid removal ●
- Dialysis time ●
- Dialysis fluid composition ●
- Dialysis fluid temperature ●
- Heparin delivery rate ●
### 5.5. Intended Use
The Fresenius 2008K2 Hemodialysis Machine is intended for acute and chronic dialysis therapy.
The intended use is identical to the predicate device's intended use.
### 5.6. Indications for Use
The 2008K2 Hemodialysis Machine is indicated for acute and chronic hemodialysis therapy in a healthcare facility.
{6}------------------------------------------------
Additional therapy options for patients receiving hemodialysis include: Isolated Ultrafiltration, Sustained Low Efficiency Dialysis (SLED), and low volume hemodialysis (patients weighing ≥ 20 kg and ≤ 40 kg). This machine accommodates the use of both low flux and high flux dialyzers. The SLED therapy option is not to be used for patients weighing < 40 kg. The 2008K2 Hemodialysis Machine is not to be used for plasma replacement therapies, for patients weighing less than 20 Kg, or for renal therapies using substitution fluid.
### 5.7. Comparison of Technological Characteristics with the Predicate Device
The technological characteristics of the 2008K2 Hemodialysis Machine are equivalent to the predicate device. Although modifications to functional board and the front panel board were necessary to accommodate additional kevs on the 2008K2 Machine, all other electrical components, mechanical components, hydraulic components, and safety and alarm monitoring systems are identical to the predicate device.
The following technical specifications of the 2008K2 machine remain the same as the predicate device:
Safety System System Performance Environmental Requirements Accessories Accuracy and Controls Protection against Mechanical Hazard Protection against Electrical Hazard Protection against excessive temperature or other hazards Transportation and Storage specifications Manufacturing location Manufacturing process (assembly, testing, shipping, installation, and service)
### 5.8. Performance Data
### Biocompatibility Testing 5.8.1.
Hydraulic components of the proposed device (i.e. valves, tubing, sensors and gauges, etc.) are in contact with the dialysate and are identical to the predicate device. Therefore, additional biocompatibility testing was not required.
### 5.8.2. Electrical Safety and Electromagnetic Compatibility (EMC)
The 2008K2 Hemodialysis Machine was evaluated to the electromagnetic compatibility (EMC) requirements of IEC 60601-1-2:2007 as well as to higher immunity test levels in supplemental testing to evaluate risk of additional healthcare facility EMI environmental considerations and risk of exposure to RF emitters.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left is a blue icon that looks like three chevrons stacked on top of each other, pointing downwards. To the right of the icon is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on top of "MEDICAL CARE".
K153449 Page 5 of 6
Proposed modifications to the machine are limited to software and labeling changes, with no modifications to the power supply, machine insulation, grounding, or critical components. Therefore, the existing electrical safety testing of the 2008K2 Hemodialysis Machine was not affected by the proposed modifications, and no additional electrical safety testing was performed.
### 5.8.3. Software Verification and Validation Testing
System level software verification testing was performed to demonstrate the effectiveness of the software modifications and confirm operation of the machine.
### 5.8.4. Mechanical and Acoustic Testing
Mechanical testing included net fluid removal accuracy evaluations. Other mechanical testing was limited to that which supported software verification. Examples of these evaluations include:
- Dialysate flow rate accuracy ●
- Regression tests for verifying impact of an additional dialysate flow rate (flow ● accuracy and temperature testing)
- Heparin syringe plunger rates and dispensing volumes ●
- . Simulated dialysis testing at the full range of dialysate and blood flow rates
- . General system regression testing (production)
No acoustic testing was necessary for the modifications to the 2008K2 Hemodialysis Machine, which were limited to software and labeling modifications.
### Other Bench Testing 5.8.5.
The following evaluations were performed, but are not classified under electrical, software, or mechanical categories of testing:
- Microbiological purity testing of dialysis fluid
- A risk-based analysis of software updates to determine the impact of software . modifications on usability.
### 5.8.6. Animal Studies
No animal studies were performed in support of the modifications.
### 5.8.7. Clinical Studies
No clinical studies were performed in support of the modifications.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Fresenius Medical Care. On the left is a blue icon that consists of three downward-pointing chevrons stacked on top of each other. To the right of the icon is the text "FRESENIUS MEDICAL CARE" in blue, with "FRESENIUS" on the top line and "MEDICAL CARE" on the bottom line.
K153449 Page 6 of 6
### 5.9. Conclusions
The 2008K2 Hemodialysis Machine has equivalent indications for use, intended use and technological characteristics as the predicate device. The information provided in this submission, including design verification and validation, risk management, and electrical safety demonstrate the device functions as intended and support the determination of substantial equivalence. Differences between the 2008K2 Hemodialysis Machine and the predicate device do not raise any new concerns with regard to safety or effectiveness. FMCRTG concludes that within the meaning of the Medical Device Amendments Act of 1976, the proposed modified 2008K2 Hemodialysis Machines is substantially equivalent to the predicate device, the 2008K Hemodialysis Machine (K994267, K150708).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.