Diazyme Human Kappa Free Light Chain Assay, Diazyme Human Lambda Free Light Chain Assay, Diazyme Human Kappa/Lamda Free Light Chain Calibrator, Diazyme Human Kappa/Lamda Free Light Chain Calibrator
Device Facts
| Record ID | K153394 |
|---|---|
| Device Name | Diazyme Human Kappa Free Light Chain Assay, Diazyme Human Lambda Free Light Chain Assay, Diazyme Human Kappa/Lamda Free Light Chain Calibrator, Diazyme Human Kappa/Lamda Free Light Chain Calibrator |
| Applicant | Diazyme Laboratories |
| Product Code | DFH · Immunology |
| Decision Date | Nov 18, 2016 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.5550 |
| Device Class | Class 2 |
| Attributes | Real-World Evidence |
Real-World Evidence
| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
|---|---|---|---|---|---|
| K153394 · Nov 18, 2016 | Diazyme Human Kappa Free Light Chain Assay, Diazyme Human Lambda Free Light Chain Assay, Diazyme Human Kappa/Lamda Free Light Chain Calibrator, Diazyme Human Kappa/Lamda Free Light Chain Calibrator | Diazyme Laboratories | Clinical patient serum samples | Retrospective clinical patient samples were used to perform a method comparison study against the predicate device to demonstrate substantial equivalence and clinical performance. | Method comparison; Clinical patient samples; Multiple Myeloma; MGUS |
Clinical Evidence
| Study Design | Population | Comparator | Key Endpoints |
|---|---|---|---|
| Method comparison for Diazyme Human Kappa (κ) and Lambda (λ) FLC Assays; Method comparison study | 126 serum samples including patients with Multiple Myeloma (MM), Monoclonal Gammopathy of Undetermined Significance (MGUS), other abnormal disease states, and normal (no disease) patients.; Sample Size: 126 | Freelite® Human Kappa/Lambda Free Kit (Predicate) | Concordance, linear regression, Deming regression, positive percent agreement (PPA), negative percent agreement (NPA), and overall percent agreement (TPA). |
Indications for Use
The Diazyme Human Kappa Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Kappa Free Light Chain (FLC) concentration in serum on Hitachi 917 analyzers. The measurement of Kappa FLC in conjunction with Lambda FLC aids in the diagnosis of multiple myeloma in conjunction with other laboratory findings. For in-vitro diagnostic use only. Diazyme Human Kappa/Lambda FLC Calibrator (DZ169A-CAL) is intended for use in establishing the calibration curve for the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only. Diazyme Human Kappa/Lambda FLC Control (DZ169A-CON) is intended for use in monitoring the quality control of results obtained with the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only. The Diazyme Human Lambda Free Light Chain Assay is intended as a latex particle enhanced immunoturbidimetric assay for the quantitative determination of Lambda Free Light Chain (FLC) concentration in serum on Hitachi 917 analyzers. The measurement of Lambda FLC in conjunction with Kappa FLC aids in the diagnosis of multiple myeloma in conjunction with other laboratory findings. For in-vitro diagnostic use only. Diazyme Human Kappa/Lambda FLC Calibrator (DZ169A-CAL) is intended for use in establishing the calibration curve for the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only. Diazyme Human Kappa/Lambda FLC Control (DZ169A-CON) is intended for use in monitoring the quality control of results obtained with the Diazyme Human Kappa/Lambda Free Light Chain Assay. For in-vitro diagnostic use only.
Device Story
Device is a latex particle-enhanced immunoturbidimetric assay for quantitative measurement of Kappa and Lambda FLC in human serum. Input: patient serum sample. Principle: FLC in sample binds to specific anti-FLC antibodies coated on polystyrene latex particles, causing agglutination; turbidity measured optically at 570 nm. Output: FLC concentration calculated via interpolation of a 6-point calibration curve. Used in clinical laboratories; operated by trained professionals on Roche Hitachi 917 analyzers. Provides quantitative data to assist physicians in diagnosing multiple myeloma and related plasma cell disorders. Benefits patient by enabling standardized, automated measurement of FLC levels.
Clinical Evidence
Bench testing only. Precision evaluated per CLSI EP5-A (n=240, multiple lots/analyzers); Kappa CVs 1.2-11.7%, Lambda CVs 1.1-11.4%. Linearity per CLSI EP6-A (Kappa 4.5-150 mg/L standard, 90-3000 mg/L extended; Lambda 6.1-200 mg/L standard, 122-4000 mg/L extended). Method comparison (n=126) against predicate showed high concordance (Kappa 98%, Lambda 95%) and strong correlation (Kappa R²=0.977, Lambda R²=0.972). Interference testing showed <10% deviation for common blood substances.
Technological Characteristics
Latex particle-enhanced immunoturbidimetric assay. Reagents: Tris buffer, sodium azide, polystyrene beads coated with polyclonal rabbit/goat anti-human FLC antibodies. Energy source: Optical measurement (570 nm) on clinical chemistry analyzer. Connectivity: Standalone application on Roche Hitachi 917. Form factor: Liquid stable reagents, calibrators, and controls.
Indications for Use
Indicated for the quantitative determination of Kappa and Lambda free light chains in serum to aid in the diagnosis of multiple myeloma in conjunction with other laboratory findings. For prescription use only.
Regulatory Classification
Identification
An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.
Predicate Devices
- Freelite® Human Kappa Free Kit for use on the Hitachi 911/912/917/Modular P analyzers (K023009 and K040009)
- Freelite® Human Lambda Free Kit for use on the Hitachi 911/912/917/Modular P analyzers (K023009 and K023131)