Repagyn is a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.
Device Story
Repagyn Vaginal Suppositories are semi-solid, non-sterile preparations containing hyaluronic acid and semi-synthetic glycerides; packaged in PVC/PE blisters. Designed for vaginal application; inserted manually into vaginal cavity. Upon contact with vaginal mucosa, body temperature causes suppository to melt into viscous mass; provides lubrication and moisture. Used by patients for personal care; intended to enhance comfort during intimate sexual activity. Benefits include supplementation of natural lubrication. Not compatible with natural rubber latex, polyurethane, or polyisoprene condoms.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993 standards: acute systemic toxicity (ISO 10993-11:2006), cytotoxicity (ISO 10993-5:2009), vaginal irritation (ISO 10993-10:2010), and guinea pig maximization test (ISO 10993-10:2010). Endotoxin levels verified via USP <85>. Shelf life of 24 months established via real-time aging. Condom compatibility tested per ASTM D7661-10.
Technological Characteristics
Hyaluronic acid and semi-synthetic glycerides formulation. 2g weight per suppository. PVC/PE blister packaging. Melting point corresponds to body temperature. Biocompatibility per ISO 10993. Condom compatibility tested per ASTM D7661-10.
Indications for Use
Indicated for individuals requiring vaginal moisturization and lubrication to enhance comfort during intimate sexual activity and supplement natural lubrication.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an image of three stylized human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the largest and the bottom profile being the smallest.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 30, 2016
Farma-Derma s.r.1. Mara Calzolari Business Development Manager Via Dei Bersaglieri, 10 40010 Sala Bolognese Bologna, Italy
Re: K153372
> Trade/Device Name: Repagyn Vaginal Suppositories Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: Class II Product Code: NUC Dated: Mav 31, 2016 Received: June 2, 2016
Dear Mara Calzolari,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
K153372
Device Name
Repagyn vaginal suppositories
Indications for Use (Describe)
Reparyn is a personal lubricant, for vaginal application, intended to moisturize, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary for Repagyn Vaginal Suppositories
This 510(k) Summary is being submitted in accordance with the requirements of 21 C.F.R. 807.92.
#### General Information 1.
| Submitter: | FARMA-DERMA S.r.l.<br>Via dei Bersaglieri, 10<br>40010 Sala Bolognese (BO)<br>ITALY |
|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person: | Ms. Mara Calzolari<br>FARMA-DERMA srl<br>Via dei Bersaglieri 10<br>40010 Sala Bolognese (BO)<br>Italy<br>Mobile: +39 3486914895<br>Office: +39 051221606<br>Fax: +39 0516814833<br>Email: mara.calzolari@farmaderma.it |
| Additional contact person: | Ms. Giulia Canzano<br>FARMA-DERMA srl<br>Via dei Bersaglieri 10<br>40010 Sala Bolognese (BO)<br>Italy<br>Mobile: +39 3486914892<br>Office: +39 0516814181<br>Fax: +39 0516814833<br>Email: giulia.canzano@farmaderma.it |
| Summary Preparation Date: | November 19, 2015 |
| 2. Names | |
| Device Name: | Repagyn Vaginal Suppositories |
| Regulation number: | 21 C.F.R. 884.5300 |
| Classification: | II |
| Product Code: | NUC |
| Common Name: | Lubricant Personal |
{4}------------------------------------------------
#### 3. Predicate Devices
Repagyn Vaginal Suppositories are substantially equivalent to Hyalo Gyn, K094039, registered by Fidia Pharmaceuticals.
## 4. Device Description
The Farma-Derma Repagyn Vaginal Suppositories are a non-sterile hyaluronic acid and glycerides based semi-solid preparation presented in the form of vaginal suppositories of 2 g in weight individually packed in a polyvinyl chloride (PVC)/polyethylene (PE) blister. The device is available for sale in a pack containing 10 vaginal suppositories (two blisters of 5 vaginal suppositories each) while the sample pack contains 3 vaginal suppositories (one blister).
Due to its specific shape the device can be introduced into the vaginal cavity using a finger.
Repagyn Vaginal Suppositories are a Hyaluronic acid (HA) based product; the HA is mixed in with semi-synthetic glycerides that represent the major quantity of material used in the formulation.
When the vaginal suppository is in contact with the vaginal mucosa it melts due to body temperature and becomes a viscous mass which remains in contact with the vaginal mucosa.
It acts similarly to other vaginal creams or gels for topical application with Hyaluronic acid based formulations.
The specifications for Repagyn Vaginal Suppositories are described in Table 1.
| Specifications | |
|-----------------------------------------------------------------------|-------------------------------------------|
| Appearance | Off-white opaque vaginal<br>suppositories |
| Disintegration time | According to Eu. Ph. |
| Average weight | 2 g |
| Uniformity of weight | According to Eu. Ph. |
| Content of HA | 5 mg |
| TAMC | ≤100 UFC/g |
| TYMC | ≤10 UFC/g |
| Pseudomonas Aeruginosa,<br>Staphilococcus Aureus, Candida<br>Albicans | Absent/g |
#### Table 1: Repagyn Vaginal Suppositories specifications
Furthermore the endotoxins level will be checked according to the kinetic test (LAL test) on each lot of finished product.
{5}------------------------------------------------
#### 5. Indications for Use
The intended use for Repagyn vaginal suppositories is:
"Repagyn is a personal lubricant, for vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyurethane, and polyisoprene condoms."
#### 6. Technological Characteristics
comparison between the technological characteristics of Repagyn vaginal র suppositories and the predicate device Hyalo Gyn is provided in Table 2.
| | Repagyn Vaginal<br>Suppositories | Hyalo Gyn<br>(Predicate Device<br>K094039) |
|----------------------|----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------|
| Base type | Glycerides | Water |
| Primary ingredient | Glycerides | Propylene glycol and water |
| Condom compatibility | Not compatible with natural<br>rubber latex, polyisoprene,<br>and polyurethane condoms | Compatible with<br>lubricated/non-lubricated<br>latex, lubricated<br>polyurethane, lubricated<br>natural skin condoms |
| Packaging | PVC/PE blister of 3 or 5<br>vaginal suppositories | Tube |
Table 2: Technological characteristics - Comparison between Repagyn vaginal suppositories and Hyalo Gyn
Repagyn Vaginal Suppositories and Hyalo Gyn differ in their shape, Repagyn is in the form of vaginal suppositories and Hyalo Gyn is an aqueous gel. However, when the suppositories are introduced in the vaginal cavity, they melt because of body temperature and become a viscous mass that act in the same way as the predicate device.
The components used for the predicate device are similar to the ones used for Repagyn vaginal suppositories.
The area of application for both medical devices is the vaginal mucosa. The only difference is that the predicate device Hyalo Gyn (K094039) can be applied also on the penis during sexual intercourses while Repagyn is designed for vaginal application only.
{6}------------------------------------------------
#### 7. Performance Data
#### Biocompatibility:
Biocompatibility studies including acute system toxicity, vaginal irritation, cytotoxicity, and guinea maximization sensitization were performed according to the ISO standards as described below:
- Acute Systemic Toxicity: ISO 10993-11:2006 ●
- Cytotoxicity: ISO 10993-5:2009
- Vaginal Irritation Testing: ISO 10993-10:2010 ●
- Guinea Pig Maximization Test (GPMT): ISO 10993-10:2010. ●
- Endotoxin levels determination: USP <85> ●
#### Shelf Life:
The shelf life of Repagyn vaginal suppositories is 24 months. This is based on the results of real time aging studies that demonstrated that the device maintains its specifications over the duration of its shelf life.
#### Condom Compatibility:
Condom compatibility was performed using the FDA recognized consensus standard ASTM D7661-10 Standard Test Method for Determining Compatibility of Personal Lubricants with Natural Rubber Latex Condoms. The Repagyn Vaginal Suppositories were found to be not compatible with natural rubber latex, polyisoprene and polyurethane condoms.
#### 8. Conclusion
Repagyn Vaginal Suppositories are substantially equivalent to the predicate device, Hyalo Gyn.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.