← Product Code [HRS](/productcode/HRS) · K153340

# Epic Extremity Plate System (K153340)

_Epic Extremity, LLC · HRS · Feb 9, 2016 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153340

## Device Facts

- **Applicant:** Epic Extremity, LLC
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Feb 9, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The Epic Extremity Plate System is intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes. The system may be used in both adult and pediatric patients. The plates/screws are intended for single use only.

## Device Story

Epic Extremity Plate System comprises metallic bone plates and 2.7mm/3.5mm locking and non-locking screws; includes associated surgical instruments. Used by orthopedic surgeons in clinical settings for internal fixation of small bone fractures, joint fusions, and reconstructions. Implants provide mechanical stabilization to bone segments to facilitate healing. System is single-use only.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing.

## Technological Characteristics

Metallic bone fixation plates and screws (2.7mm and 3.5mm). Materials are similar to predicate. Performance testing conducted per ASTM F-543 (metallic medical bone screws) and ASTM F-382 (metallic bone plates).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- DARCO Locking Bone Plate System ([K061808](/device/K061808.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 9, 2016

EPIC Extremity, LLC % Mr. Lee Strnad President Intrepid Orthopedics, LLC 3046 Brecksville Road. Suite 4 Richfield, Ohio 44286

Re: K153340 Trade/Device Name: Epic Extremity Plate System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: November 20, 2015 Received: November 23, 2015

Dear Mr. Strnad:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known) K153340

Device Name EPIC Extremity Plate System

Indications for Use (Describe)

The Epic Extremity Plate System is intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bone of the hand, fingers and toes. The system may be used in both adult and pediatric patients. The plates/screws are intended for single use only.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the text "K153340 Page 1 of 2" at the top. Below that is the word "EPIC" in large, bold, black letters. Underneath "EPIC" is the word "EXTREMITY" in smaller letters. The word "EXTREMITY" is in a gradient of blue and white.

# Epic Extremity Plate System 510(k) Summary

#### Submitter Information

| Applicant:          | EPIC Extremity, LLC<br>120 Marquerite Dr., Ste 301<br>Cranberry Twp., PA 16066                                                            |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:     | Lee A. Strnad<br>Management Representative<br>Intrepid Orthopedics<br>3046 Brecksville Rd, Ste 4<br>Richfield, OH 44286<br>(330) 659-0855 |
| Date Prepared:      | November 10, 2015                                                                                                                         |
| Name of Device:     | Epic Extremity Plate System                                                                                                               |
| Common Name:        | Bone Fixation Plate                                                                                                                       |
| Classification Name | Single/Multiple component metallic bone fixation appliances<br>and accessories (per 21 CFR 888.3030)                                      |
| Product Code/Panel: | HRS/Orthopedics                                                                                                                           |
| Predicate Devices:  | DARCO Locking Bone Plate System (K061808).                                                                                                |

#### Intended Use:

The Epic Extremity Plate System is intended for use in stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes. The system may be used in both adult and pediatric patients. The plates/screws are intended for single use only.

### Device Description

The Epic Extremity Plate System consists of multiple plate families of various anatomical sizes and shapes, 2.7mm and 3.5mm locking and nonlocking screws which mate into the plates, as well as various instruments to assist in implanting the system.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Epic Extremity. The logo is in black and white, with the word "EPIC" in large, bold letters. Below the word "EPIC" is the word "EXTREMITY" in smaller letters. At the top of the image, the text "K153340 Page 2 of 2" is visible.

# Epic Extremity Plate System 510(k) Summary

## Technological Characteristics

The EPIC Extremity Plate System have the same intended use as the predicate devices. The EPIC Extremity Plate System have similar indications for use as the predicate devices. The EPIC Extremity Plate System is manufactured from similar materials as the predicate devices. The range of sizes of the EPIC Extremity Plate System are similar to the predicate devices.

### Non-Clinical Performance Data Summary

1. ASTM F-543 2. ASTM F-382

### Clinical Performance Data Summary

No clinical testing was required.

### Non-Clinical and Clinical Performance Data Conclusions

Based on testing results and the comparisons provided, the EPIC Extremity Plate System are considered substantially equivalent to the Darco Locking Bone Plate System in material, construction, and performance characteristics.

---

**Source:** [https://fda-staging.innolitics.com/device/K153340](https://fda-staging.innolitics.com/device/K153340)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
