← Product Code [EBF](/productcode/EBF) · K153325

# Gradia Plus (K153325)

_GC America, Inc. · EBF · Jul 12, 2016 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153325

## Device Facts

- **Applicant:** GC America, Inc.
- **Product Code:** [EBF](/productcode/EBF.md)
- **Decision Date:** Jul 12, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3690
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

1. Crown and Bridgework (with metal backing) 2. Metal free restorations: Jacket Crowns, Inlays, Onlays and Laminated Veneers 3. Implant superstructures 4. Reproduction of gum tissue for crown restorations

## Device Story

GRADIA PLUS is a light-cured, nano-filled, radiopaque dental composite resin supplied in syringes. It consists of cross-linked polymeric material reinforced with filler particles. The device is used by dental professionals in laboratory settings to fabricate dental restorations, including crown and bridgework, jacket crowns, inlays, onlays, veneers, and implant superstructures, as well as to reproduce gingival tissue. The material is applied in layers and cured using a photo-initiator system. The process involves preparation, build-up, light-curing, and a final surface finish achieved through coating or polishing. The device provides wear resistance, gloss retention, and shade matching to restore dental function and aesthetics. It benefits patients by providing durable, natural-looking prosthetic restorations.

## Clinical Evidence

Bench testing only. The device was evaluated for sensitivity to ambient light, depth of cure, surface finish, flexural strength, water sorption, solubility, shade consistency, and color stability. Results confirmed conformity to ISO 10477:2004 and internal company specifications.

## Technological Characteristics

Light-cured nano-filled radiopaque composite resin. Contains ultra-fine barium glass and pre-polymerized filler in a methacrylate ester monomer matrix. Curing via photo-initiator system. Available in 73 shades across 5 paste types (Opaque, HB, LB, One Body, Gum Shades, Lustre Paint). Complies with ISO 10477:2004.

## Regulatory Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

## Predicate Devices

- GRADIA ([K001518](/device/K001518.md))

## Reference Devices

- GRADIA gum shades ([K033808](/device/K033808.md))
- G-aenial Sculpt (MFP-051) ([K123631](/device/K123631.md))
- OPTIGLAZE color ([K133836](/device/K133836.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name in a circular arrangement around an emblem. The emblem depicts a stylized image of three faces in profile, layered on top of each other, suggesting a sense of community and care. The seal is black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 12, 2016

GC Americica Inc. Mark Heiss, D.D.S. Director, Regulatory and Academic Affairs 3737 W 127th Street Alsip, Illinois 60803

Re: K153325

Trade/Device Name: GRADIA PLUS Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: II Product Code: EBF Dated: June 2, 2016 Received: June 3, 2016

Dear Dr. Heiss:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Mark Heiss, D.D.S.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Section 4 - Indications for Use Statement

510(k) Number (if known): K153325

Device Name: GRADIA PLUS

Indications for Use:

- 1. Crown and Bridgework (with metal backing)
- 2. Metal free restorations: Jacket Crowns, Inlays, Onlays and Laminated Veneers
- 3. Implant superstructures
- 4. Reproduction of gum tissue for crown restorations

Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the letters 'CCI' in a stylized, sans-serif font. The letters are a vibrant green color, and they appear to be slightly italicized or slanted to the right. The letters are enclosed in single quotation marks.

GC AMERICA INC.
3737 WEST 127TH STREET
ALSIP, ILLINOIS 60003
TEL (708) 597-0900
FAX (708) 371-5103

- 1. Submitter Information:
GC AMERICA INC. 3737 W. 127th Street Alsip, IL 60803

| Contact Person:    | Mark Heiss, D.D.S. |
|--------------------|--------------------|
| Phone:             | (708) 926-3090     |
| Alternate Contact: | Lori Rietman       |
|                    | (708) 926-3092     |
| Fax:               | (708) 926-9100     |

Date Prepared: June 1, 2016

- 2. Device Name:

| Proprietary Name:      | GRADIA PLUS                |
|------------------------|----------------------------|
| Classification Name:   | Tooth shade resin material |
| Device Classification: | Class II, 872.3690         |
| Product Code:          | EBF                        |

### 3. Predicate Devices:

| Company         | Device                    | 510(k) No. | Date Cleared | Predicate |
|-----------------|---------------------------|------------|--------------|-----------|
| GC America Inc. | GRADIA                    | K001518    | 07/05/2000   | Primary   |
| GC America Inc. | GRADIA gum shades         | K033808    | 02/02/2004   | Reference |
| GC America Inc. | G-aenial Sculpt (MFP-051) | K123631    | 07/23/2013   | Reference |
| GC America Inc. | OPTIGLAZE color           | K133836    | 03/06/2014   | Reference |

### 4. Description of Device:

GRADIA PLUS is a dental composite consisting of cross-linked polymeric material reinforced by a dispersion of filler particles bonded to the resin matrix. The product contains a photo initiator to allow for curing/conversion. The end result is a conversion from a paste to a hard solid material.

GRADIA PLUS is composed of 4 types of pastes, such as Opaque paste, Gum shade paste and Lustre paint.

Opaque paste is used for masking the metal color.

Body paste has two different consistencies. Paste HB and Paste LB. Both pastes can be used for creating several types of anterior or posterior restorations. There are several shades for dentin shade, Enamel shade and effect shade. The dentin layers have an enhanced brightness and masking ability that adjusts the opalescence of the composites. This gives a bright color tone in the oral cavity and eliminates unnatural translucency at the incisal edge.

{4}------------------------------------------------

In addition, it has several red colors as a Gum shade paste with different transparency and color tone for reproducing of gingival tissue over crown.

Lustre Paint is used for staining, characterizing and coating.

The device is used for crown and bridgework with metal backing, jacket crowns, inlays, veneers, implant super structures and reproduction of gum tissue for crown restorations. In particular, Gradia Plus is intended to fabricate the tertiary element/exostructure of the implant superstructure, such as crown and bridges.

GRADIA PLUS is filled in a syringe and is available in 73shades: HB-DA1, HB-DA3, HB-DA3.5, HB-DB1, HB-DB3, HB-DC3, HB-DD2, HB-DW, HB-ODA, HB-ODB, HB-ODC, HB-ODD, HB-ODW, HB-EL, HB-ED, HB-CLF, HB-PE, LB-BaseD, LB-BaseCLF, LB-BaseCLF, LB-BaseOD, LB-BaseOpal, LB-DW, LB-Orange, LB-Red, LB-Yellow, LB-Blue, LB-Grey, LB-Milky, LB-Inlay E, LB-Inlay TD, LB-A, LB-B, LB-C, LB-D, LB-W, GLB-1, GLB-2, GLB-3, GLB-CL, GHB-1, GHB-2, GHB-3, GHB-CL, O-BASE, OA, OB, OC, OD, OA1, OA2, OA3, OA3.5, OB1, OB3, OC3, OD2, GO-1, GO-2, LP-A, LP-B, LP-C, LP-D, LP-CLF, LP-Blue, LP-Cream, LP-Grey, LP-Lavender, GLP-Bright red, GLP-Violet, LP-CL

### ર Performance Bench Tests

It is confirmed that the device conforms to the required specifications of ISO 10477:2004 and company standard are suitable for its intended use. Performance testing includes:

- Sensitivity to ambient light ●
- Depth of cure ●
- Surface finish ●
- Flexural strength ●
- Water sorption
- Solubility ●
- Shade consistency ●
- Color stability ●

## 6. Packaging

- 1) OPAQUE, 2.0mL syringe (in 13 shades)
	- O-BASE, OA, OB, OC, OD, OA1, OA2, OA3, OA3.5, OB1, OB3, OC3, OD2
- 2) PASTE HB, 3.3mL syringe (in 18 shades) HB-DA1, HB-DA2, HB-DA3, HB-DA3.5, HB-DB1, HB-DB3, HB-DC3, HB-DD2, HB-DW, HB-
- ODA, HB-ODB, HB-ODC, HB-ODD, HB-ODW, HB-EL, HB-ED, HB-CLF, HB-PE 3) PASTE LB, 2.0mL syringe (in 15 shades) LB-BaseD, LB-BaseE, LB-BaseCLF, LB-BaseOD, LB-BaseOpal, LB-DW, LB-ODW, LB-Orange,
- LB-Red, LB-Yellow, LB-Blue, LB-Grey, LB-Milky, LB-Inlay E, LB-Inlay TD
- 4) ONE BODY, 2.0mL syringe (in 5 shades)
- LB-A, LB-B, LB-C, LB-D, LB-W
- 5) GUM SHADES, 2.0mL syringe (in 10 shades)

G0-1, GO-2, GLB-1, GLB-2, GLB-3, GLB-CL, GHB-1, GHB-2, GHB-3, GHB-CL

- 6) LUSTRE PAINT, 0.8mL syringe (in 12 shades)
LP-A, LP-B, LP-C, LP-D, LP-CLF, LP-Blue, LP-Cream, LP-Grey, LP-Lavender, GLP-Bright red, GLP-Violet, LP-CL

{5}------------------------------------------------

Accessories 1) MODELING LIQUID 3mL Bottle 2) LP DILUTING LIQUID 3mL Bottle 3) AIR BARRIER 10mL Bottle 4) SEPARATOR 5mL Bottle 5) DIE-HARDNER 5mL Bottle

- 7. Shelf Life Evaluation and Storage Conditions:
	- Shelf Life 3 years
	- Recommended for optimal performance, store in a cool place (4-25℃ / 39.2-77.0 °F) away from high temperatures of direct sunlight.
- 8. Indications for Use:
	- 1. Crown and Bridgework (with metal backing)
	- 2. Metal free restorations: Jacket Crowns, Inlays, Onlays and Laminated Veneers
	- 3. Implant superstructures
	- 4. Reproduction of gum tissue for crown restorations
- 9. Technological characteristics:

All the components of the applicant device, GRADIA PLUS, have already been used in the predicate devices. GRADIA PLUS contains ultra-fine barium glass and pre-polymerized filler which exhibit wear resistance and gloss retention. The applicant device also shows equivalence in flexural strength, water sorption and solubility, which indicate the stability of materials in oral environment. The curing mechanism of the predicates and GRADIA PLUS is polymerization of uncured methacrylate ester monomers. This reaction is caused by photo initiator system.

#### 10. Substantial equivalence

The curing mechanism of the new and predicate devices is substantially equivalent in principle. Therefore, the applicant and primary predicate device are the same in function, and similar in composition and intended use. This supports that the compatibility of the applicant device is substantially equivalent to the primary predicate device.

### 11. Differences

- The following differences may be noted between the predicate device and GRADIA PLUS:
- GRADIA PLUS is available in 73 shades ●
- GRADIA PLUS is available in 5 types of paste: Paste HB, Paste LB, Gum Shade LB, Opaque, ● Lustre Paint (Stain)
- GRADIA PLUS is used for crown and bridgework with metal backing, jacket crowns, inlays, ● onlays, veneers, implant super structures and reproduction of gum tissue for crown restorations. The difference is that Gradia is not indicated for shade reproduction of gum tissue. However, this indication is found in Gradia Gum, K033808.
- . Applicant device requires coating or polishing as a surface finish. Gradia requires only polishing and Optiglaze Color requires only Coating.
- Gradia uses light-cured hybrid fillers, while Gradia Plus uses light-cured nano fillers. .
- . Gradia Plus has improved surface gloss which utilizes coating and polishing versus primary predicate only requiring polishing.

{6}------------------------------------------------

|                         | Applicant device                                                                                                                                                                                                                                                                                   | Primary Predicate device                                                                                                                                                                                                                                                                                     | Difference | Impact on Effectiveness and                                                                                                                                                                                                                                                                   |
|-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name              | GRADIA PLUS                                                                                                                                                                                                                                                                                        | GRADIA                                                                                                                                                                                                                                                                                                       | Y/N        | rationale                                                                                                                                                                                                                                                                                     |
| Product category        | Light-cured composite for crown and bridge                                                                                                                                                                                                                                                         | Light-cured composite for crown and<br>bridge                                                                                                                                                                                                                                                                |            |                                                                                                                                                                                                                                                                                               |
| Company                 | GC Corporation                                                                                                                                                                                                                                                                                     | GC Corporation                                                                                                                                                                                                                                                                                               |            |                                                                                                                                                                                                                                                                                               |
| 510(k) No.              | K153325                                                                                                                                                                                                                                                                                            | K001518                                                                                                                                                                                                                                                                                                      |            |                                                                                                                                                                                                                                                                                               |
| Number of shades        | 73                                                                                                                                                                                                                                                                                                 | 137                                                                                                                                                                                                                                                                                                          | Y          | Shade number does not impact<br>on safety or efficacy.                                                                                                                                                                                                                                        |
| Indications for use     | 1. Crown and Bridgework (with metal<br>backing)<br>2. Metal free restorations: Jacket Crowns,<br>Inlays, Onlays and Laminated Veneers<br>3. Implant superstructures<br>4. Reproduction of gum tissue for crown<br>restorations                                                                     | 1. Crown and Bridgework (with metal<br>backing)<br>2. Metal free restorations: Jacket<br>Crowns, Inlays, Onlays and Laminated<br>Veneers<br>3. Implant superstructures                                                                                                                                       | Y          | Applicant device includes for<br>reproduction of gum tissue.<br>Comparative device does not<br>include for reproduction of gum<br>tissue.<br>Product performs within<br>requirements.                                                                                                         |
| Product<br>description  | GRADIA PLUS is a light-cured nano-filled<br>radiopaque composite resin filled in a<br>syringe. The device is used for crown and<br>bridgework with metal backing, jacket<br>crowns, inlays, onlays, veneers, implant<br>super structures and reproduction of gum<br>tissue for crown restorations. | GRADIA is a light-cured hybrid MFR<br>(micro-filled-resin) hard resin for<br>laboratory fabrication of crown & bridge,<br>jacket crowns, inlays, onlays and<br>laminated veneer restorations. Gradia<br>can be used for both metal free and metal<br>based restorations                                      | Y          | Applicant device includes for<br>reproduction of gum tissue.<br>Comparative device does not<br>include for reproduction of gum<br>tissue.<br>Product performs within<br>requirements.                                                                                                         |
| Instructions for<br>use | 1.Preparation<br>2.Treatment<br>3.Build up<br>4.Light curing<br>5.Glazing (Coating and Polishing)                                                                                                                                                                                                  | 1.Preparation<br>2.Treatment<br>3.Build up<br>4.Light curing<br>5. Polishing                                                                                                                                                                                                                                 | Y          | Applicant device requires coating<br>or polishing as a surface finish.<br>Predicate device requires only<br>polishing.<br>Product performs within<br>requirements.                                                                                                                            |
|                         | Applicant device                                                                                                                                                                                                                                                                                   | Reference Predicate device                                                                                                                                                                                                                                                                                   | Difference | Impact on Effectiveness                                                                                                                                                                                                                                                                       |
| Trade name              | GRADIA PLUS                                                                                                                                                                                                                                                                                        | GRADIA gum shades                                                                                                                                                                                                                                                                                            | Y/N        | and rationale                                                                                                                                                                                                                                                                                 |
| Product category        | Light-cured composite for crown and bridge                                                                                                                                                                                                                                                         | Light-cured gum shades composite<br>system                                                                                                                                                                                                                                                                   |            |                                                                                                                                                                                                                                                                                               |
| Company                 | GC Corporation                                                                                                                                                                                                                                                                                     | GC Corporation                                                                                                                                                                                                                                                                                               |            |                                                                                                                                                                                                                                                                                               |
| 510(k) No.              | K153325                                                                                                                                                                                                                                                                                            | K033808                                                                                                                                                                                                                                                                                                      |            |                                                                                                                                                                                                                                                                                               |
| Shade number            | 73                                                                                                                                                                                                                                                                                                 | 17                                                                                                                                                                                                                                                                                                           | Y          | Shade number does not impact<br>on safety or efficacy.                                                                                                                                                                                                                                        |
| Indications for use     | 1. Crown and Bridgework (with metal<br>backing)<br>2. Metal free restorations: Jacket Crowns,<br>Inlays, Onlays and Laminated Veneers<br>3. Implant superstructures<br>4. Reproduction of gum tissue for crown<br>restorations                                                                     | GRADIA GUM is a light-cured composite<br>resin specifically used for the<br>reproduction of the gum on crown and<br>bridge prostheses.                                                                                                                                                                       | Y          | Applicant device is for crown ,<br>inlays, onlays, veneers, Implant<br>superstructure and shade<br>reproduction of gum tissue.<br>Comparative device is only for<br>gum tissue, such as implant<br>superstructures and removable<br>metal frames.<br>Product performs within<br>requirements. |
| Product<br>description  | GRADIA PLUS is a light-cured nano-filled<br>radiopaque composite resin filled in a<br>syringe. The device is used for crown and<br>bridgework with metal backing, jacket<br>crowns, inlays, onlays, veneers, implant<br>super structures and reproduction of gum<br>tissue for crown restorations. | GRADIA gum shades is a light-cured<br>MFR (micro-filled-resin) composite for<br>reproduction of gum tissue for crown<br>restorations including implant<br>superstructures and removal of metal<br>frames. Five gum shades are available<br>along with the Opaque, Translucent,<br>Modifiers and Accessories. | Y          | Applicant device is for crown,<br>inlays, onlays, veneers, Implant<br>superstructure and reproduction<br>of gum tissue. Comparative<br>device is only for gum tissue,<br>such as implant superstructures<br>and removable metal frames.<br>Product performs within<br>requirements.           |
| Instructions for<br>use | 1.Preparation<br>2. Treatment<br>3.Build up<br>4.Light curing<br>5.Glazing (Coating and Polishing)                                                                                                                                                                                                 | 1.Preparation<br>2. Treatment<br>3.Build up<br>4.Light curing<br>5. Polishing                                                                                                                                                                                                                                | Y          | Applicant device require coating<br>or polishing as a surface finish.<br>Predicate device require only<br>polishing.<br>Product performs within<br>requirements.                                                                                                                              |

# Table 5.1 Substantial equivalence to the primary predicate device (GRADIA)

{7}------------------------------------------------

|  | Table 5.2 Substantial equivalence to the reference predicate device (GRADIA GUM) |
|--|----------------------------------------------------------------------------------|
|  |                                                                                  |

{8}------------------------------------------------

|                         | Applicant device                                                                                                                                                                                                                                                                                                | Reference devices                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                               | Difference<br>Y/N | Impact on Effectiveness and<br>rationale                                                                                                                                             |
|-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name              | GRADIA PLUS                                                                                                                                                                                                                                                                                                     | G-aenial Sculpt (MFP-051)                                                                                                                                                                                        | OPTIGLAZE color (Nanocoat color)                                                                                                                                                                                                                                                                                              |                   |                                                                                                                                                                                      |
| Product<br>category     | Light-cured composite for<br>crown and bridge                                                                                                                                                                                                                                                                   | Light-cured composite<br>restorative                                                                                                                                                                             | Light-cured<br>characterization material                                                                                                                                                                                                                                                                                      |                   |                                                                                                                                                                                      |
| Manufacturer            | GC Corporation                                                                                                                                                                                                                                                                                                  | GC Corporation                                                                                                                                                                                                   | GC Corporation                                                                                                                                                                                                                                                                                                                |                   |                                                                                                                                                                                      |
| 510(K) No.              | K153325                                                                                                                                                                                                                                                                                                         | K123631                                                                                                                                                                                                          | K133836                                                                                                                                                                                                                                                                                                                       |                   |                                                                                                                                                                                      |
| Number of<br>Shades     | 73                                                                                                                                                                                                                                                                                                              | 17                                                                                                                                                                                                               | 17                                                                                                                                                                                                                                                                                                                            | Y                 | Shade number does not<br>impact on safety or efficacy.                                                                                                                               |
| Indications<br>for use  | 1. Crown and Bridgework<br>(with metal backing)<br>2. Metal free restorations:<br>Jacket Crowns, Inlays, Onlays<br>and Laminated Veneers<br>3. Implant superstructures<br>4. Reproduction of gum tissue<br>for crown restorations                                                                               | 1. Direct restorative for class I,<br>II, III, IV, V cavities.<br>2. Direct restorative for wedge-<br>shaped defects and root<br>surface cavities.<br>3. Direct restorative for veneers<br>and diastema closure. | 1. Characterization of direct &<br>indirect composite restorations,<br>acrylic denture base and artificial<br>acrylic teeth.<br>2. For obtaining surface<br>smoothness and wear resistance<br>of restorations made of<br>composite resin, acrylic denture<br>base and artificial acrylic teeth.                               | Y                 | Applicant device includes for<br>reproduction of gum tissue.<br>Comparative devices do not<br>include for reproduction of<br>gum tissue.<br>Product performs within<br>requirements. |
| Product<br>description  | GRADIA PLUS is a light-<br>cured nano-filled radiopaque<br>composite resin filled in a<br>syringe. The device is used<br>for crown and bridgework<br>with metal backing, jacket<br>crowns, inlays, onlays,<br>veneers, implant super<br>structures and reproduction<br>of gum tissue for crown<br>restorations. | G-aenial Sculpt (MFP-051) is a<br>light cured nano-filled<br>radiopaque composite resin<br>filled in a syringe and unitip.<br>The device is used for the<br>restorations of both anterior<br>and posterior teeth | OPTIGLAZE Color is a light-<br>cured glazing and characterizing<br>material for such restorations as<br>direct and indirect composites. It<br>is also used for obtaining<br>surface smoothness and wear<br>resistance of restorations made<br>of composite resins and acrylics.<br>The material is available in 17<br>shades. | Y                 | Applicant device requires<br>coating or polishing as a<br>surface finish.<br>Product performs within<br>requirements.                                                                |
| Instructions<br>for Use | 1. Preparation<br>2. Treatment<br>3. Build up<br>4. Light curing<br>5. Glazing (Coating and<br>Polishing)                                                                                                                                                                                                       | 1. Cavity Preparation<br>2. Bonding Treatment<br>3. Placement<br>4. Light Curing<br>5. Finishing and polishing                                                                                                   | 1. Preparation<br>2. Clean and dry<br>3. Apply<br>4. Light curing (Coating)                                                                                                                                                                                                                                                   |                   |                                                                                                                                                                                      |

{9}------------------------------------------------

# 12. Conclusion

Based on formulation, testing and meeting and/or exceeding ISO standards related to performance, we find the applicant device substantial equivalent to the noted predicates.

---

**Source:** [https://fda-staging.innolitics.com/device/K153325](https://fda-staging.innolitics.com/device/K153325)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
