← Product Code [DTZ](/productcode/DTZ) · K153295

# TandemLung Oxygenator (K153295)

_Cardiacassist, Inc. · DTZ · Feb 26, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153295

## Device Facts

- **Applicant:** Cardiacassist, Inc.
- **Product Code:** [DTZ](/productcode/DTZ.md)
- **Decision Date:** Feb 26, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.4350
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The TandemLung Oxygenator (TLO) is intended to be used for adult patients for extracorporeal circulation during cardiopulmonary bypass in the field of open-heart surgery. Within the indicated flow rates, blood is oxygenated and carbon dioxide is removed. The utilization period of this device is restricted to six hours.

## Device Story

TandemLung Oxygenator (TLO) facilitates extracorporeal gas exchange during open-heart surgery; utilizes injection-molded cylindrical shell containing Polymethylpentene (PMP) hollow fiber membrane bundle; sweep gas (oxygen) supplied to fiber lumens; oxygen/CO2 exchange occurs via simple diffusion across fiber walls; blood flows radially across membranes via internal flow separator; arterialized blood exits to patient; requires external pump, cannulae, and tubing; operated by clinical staff in surgical settings; provides temporary oxygenation and decarbonation; benefits patient by maintaining gas exchange during cardiopulmonary bypass.

## Clinical Evidence

Bench testing only. In vitro testing performed per FDA guidance for cardiopulmonary bypass oxygenators, including hemolysis, gas exchange, pressure drop, blood/gas pathway integrity, and physical properties.

## Technological Characteristics

Polymethylpentene (PMP) hollow fiber membranes; polycarbonate shell; two-part polyurethane potting compound; cylindrical form factor with internal flow separator; passive gas exchange via diffusion; standalone device requiring external pump/tubing.

## Regulatory Identification

A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”

## Predicate Devices

- Novalung sLA Membrane Lung ([K072362](/device/K072362.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

DEPARTMENT OF HEALTH & HUMAN SERVICES - USA

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 26, 2016

Cardiac Assist. Inc. Greg Johnson, Ph.D. Director of Regulatory Affairs 240 Alpha Dr. Pittsburgh, PA 15238

Re: K153295

Trade/Device Name: TandemLung Oxygenator Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: January 26, 2016 Received: January 27, 2016

Dear Greg Johnson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

M.A. Hillebrand

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration                                                                                                                                                                                                                                                                                                                                                             | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| 510(k) Number (if known)<br>K153295                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                   |
| Device Name<br>TandemLung Oxygenator                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                   |
| Indications for Use (Describe)<br>The TandemLung Oxygenator (TLO) is intended to be used for adult patients for extracorporeal circulation during<br>cardiopulmonary bypass in the field of open-heart surgery. Within the indicated flow rates, blood is oxygenated and<br>carbon dioxide is removed. The utilization period of this device is restricted to six hours.                                                            |                                                                                                   |
| Type of Use (Select one or both, as applicable)                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                   |
| <span style="text-decoration: underline;"><b>Prescription Use</b></span> (Part 21 CFR 801 Subpart D)                                                                                                                                                                                                                                                                                                                                | <span style="text-decoration: underline;">Over-The-Counter Use</span> (21 CFR 801 Subpart C)      |
| CONTINUE ON A SEPARATE PAGE IF NEEDED.                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                   |
| This section applies only to requirements of the Paperwork Reduction Act of 1995.                                                                                                                                                                                                                                                                                                                                                   |                                                                                                   |
| *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*                                                                                                                                                                                                                                                                                                                                                             |                                                                                                   |
| The burden time for this collection of information is estimated to average 79 hours per response, including the<br>time to review instructions, search existing data sources, gather and maintain the data needed and complete<br>and review the collection of information. Send comments regarding this burden estimate or any other aspect<br>of this information collection, including suggestions for reducing this burden, to: |                                                                                                   |
| Department of Health and Human Services<br>Food and Drug Administration<br>Office of Chief Information Officer<br>Paperwork Reduction Act (PRA) Staff<br>PRAStaff@fda.hhs.gov                                                                                                                                                                                                                                                       |                                                                                                   |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of<br>information unless it displays a currently valid OMB number."                                                                                                                                                                                                                                                                 |                                                                                                   |

PSC Publishing Services (301) 443-6740

Page 1 of 1

FORM FDA 3881 (8/14)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for CardiacAssist, Inc. The logo consists of a stylized symbol to the left of the company name. The symbol appears to be a circular design with arrows pointing in different directions. The company name, "CardiacAssist, Inc.", is written in a bold, sans-serif font.

Section 5 510(k) Summary 510(k) Traditional

Date: 11/11/2015

# Applicant

CardiacAssist, Inc. 240 Alpha Drive Pittsburgh, PA 15238 Telephone: 412-963-7770 Fax: 412-963-0800

## Contact person

Greg Johnson, PhD Title: Director of Regulatory Affairs Phone: 412-963-7770 x266 e-mail: gjohnson@tandemheart.com

## Device

Trade/Proprietary Name: Common Name: Classification Name:

TandemLung Oxygenator Cardiopulmonary Bypass Oxygenator Oxygenator, Cardiopulmonary Bypass (21 CFR 870.4350, Product Code DTZ)

## Predicate Device

Novalung sLA Membrane Lung (K072362).

## Device Description

The TandemLung Oxygenator TLO consists of an injection molded exterior shell containing a bundle of Polymethylpentene (PMP) hollow fiber membranes in which blood is oxygenated and decarbonated. Sweep gas, primarily oxygen, is supplied to interior lumens of the fibers and diffuses from the fiber wall to the blood plasma via simple diffusion. Similarly, carbon dioxide diffuses from the blood into the inner lumen of the fibers and is exhausted. Arterialized blood exits out of the blood outlet port and is delivered to the patient.

A two part polyurethane "potting" compound is used to both mechanically fix the end of the hollow fiber membranes in place and to create a barrier between the blood and gas pathways within the TLO.

Use of the TLO requires use of a pump and cannula(s), as well as tubing and connectors/fittings.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for CardiacAssist, Inc. The logo consists of a circular graphic with arrows pointing in different directions, followed by the company name in a bold, blue font. The text "CardiacAssist, Inc." is written to the right of the graphic.

Section 5 510(k) Summary 510(k) Traditional

## Intended Use

The TandemLung Oxygenator (TLO) is intended to be used for adult patients for extracorporeal circulation during cardiopulmonary bypass in the field of open-heart surgery. Within the indicated flow rates, blood is oxygenated and carbon dioxide is removed. The utilization period of this device is restricted to six hours.

#### Comparison of Technological Characteristics

The TandemLung Oxygenator is functionally and technologically similar to the predicate Novalung sLA Membrane Lung (K072362). Both products consist of polymethylpentene hollow fiber membrane material encased in polycarbonate shell material. Technological differences between the TandemLung Oxygenator and the predicate Novalung sLA Membrane Lung include: 1) The geometry and the blood flow paths of the two devices differ. The Novalung has a square overall shape, whereas the TandemLung is cylindrical with flow moving from the top to bottom of the cylinder, 2) The TandemLung incorporates a flow separator designed to optimize flow and move blood in a radial (i.e. transverse) direction across the hollow fiber membranes. The Novalung does not have a flow separator, and 3) The TandemLung has a smaller priming volume than the predicate Novalung.

#### Performance Data

In vitro bench tests were carried out to demonstrate performance and substantial equivalence to the predicate Novalung, according to the requirements of FDAs document "Guidance for Cardiopulmunary Bypass Oxygenators 510(k) Submission", issued on November 13, 2000. Specifically hemolysis, gas exchange, pressure drop, blood pathway integrity, gas pathway integrity, and physical properties were tested. This testing confirmed that the TandemLung Oxygenator does not raise any new issues of safety or effectiveness and that it is substantially equivalent to the predicate.

{End of Section}

---

**Source:** [https://fda-staging.innolitics.com/device/K153295](https://fda-staging.innolitics.com/device/K153295)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
