K153274 · Snyder Industries, Inc. · MMK · Aug 4, 2016 · General Hospital
Device Facts
Record ID
K153274
Device Name
Med-Tainer Single Deposit Container
Applicant
Snyder Industries, Inc.
Product Code
MMK · General Hospital
Decision Date
Aug 4, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
Med-Tainer Single Deposit Containers (SDC) are intended for the safe collection and transport of medical sharps. The reusable 10 gallon and 17 gallon containers are to be used in supervised areas of healthcare institutions such as hospitals, clinics, veterinary offices and laboratories.
Device Story
Med-Tainer Single Deposit Container (SDC) is a reusable, rotationally molded, low-density polyethylene container for medical sharps collection. Available in 10 and 17-gallon sizes with various lid configurations (locking transport, lab, restrictive). Used in supervised healthcare areas; operated by healthcare staff. Containers are supplied to waste management companies for transport, decanting, decontamination, and destruction. Reusable up to 780 cycles. Provides safe containment of sharps to prevent injury during disposal and transport.
Clinical Evidence
Bench testing only. Performance validated through 780-cycle life span simulation, transport vibration, leak resistance, container stability, handle strength, drop testing (1m), puncture resistance (ASTM F2132-01), stack testing, and cleaning validation per FDA guidance.
Technological Characteristics
Rotationally molded linear low-density polyethylene. 10 and 17-gallon capacities. Reusable (up to 780 cycles). Various lid types (locking, lab, restrictive). Tested per ISO 23907, ASTM F2132-01, and 49 CFR 178.603/606/608.
Indications for Use
Indicated for safe collection and transport of medical sharps in supervised healthcare settings (hospitals, clinics, veterinary offices, laboratories).
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three human profiles incorporated into its design. The eagle's head is formed by the first profile, and its body is formed by the other two profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 4, 2016
Snyder Industries, Inc. % Claudia Fell Regulatory Affairs Paxmed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K153274
Trade/Device Name: Med-tainer Single Deposit Container (SDC) Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: MMK Dated: July 6, 2016 Received: July 7, 2016
Dear Claudia Fell:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Tina Kiang Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K153274 Device Name
Med-Tainer Single Deposit Container (SDC)
#### Indications for Use (Describe)
Med-Tainer Single Deposit Containers (SDC) are intended for the safe collection and transport of medical sharps. The reusable 10 gallon and 17 gallon containers are to be used in supervised areas of healthcare institutions such as hospitals, clinics, veterinary offices and laboratories.
| Med-Tainer Single Deposit Containers | | | |
|-----------------------------------------------------|-----------------------------------|---------------------------|--------------------------------------------------------------------------|
| Med-Tainer Single Deposit Container (SDC) 10 Gallon | | | |
| Model Number | Description | Approximate Size (inches) | |
| 780300E74100 | Med-Tainer SDC, 10 gallon, Red | 18 L x 13 W x 16 H | |
| 780300E74200 | Med-Tainer SDC, 10 gallon, Teal | 18 L x 13 W x 16 H | |
| 780300E74300 | Med-Tainer SDC, 10 gallon, Yellow | 18 L x 13 W x 16 H | |
| 780300E74400 | Med-Tainer SDC, 10 gallon, Blue | 18 L x 13 W x 16 H | |
| 780300E74500 | Med-Tainer SDC, 10 gallon, Green | 18 L x 13 W x 16 H | |
| Med-Tainer Single Deposit Container (SDC) 17 Gallon | | | |
| Model Number | Description | Approximate Size (inches) | |
| 780310E74100 | Med-Tainer SDC, 17 gallon, Red | 18 L x 13 W x 25 H | |
| 780310E74200 | Med-Tainer SDC, 17 gallon, Teal | 18 L x 13 W x 25 H | |
| 780310E74300 | Med-Tainer SDC, 17 gallon, Yellow | 18 L x 13 W x 25 H | |
| 780310E74400 | Med-Tainer SDC, 17 gallon, Blue | 18 L x 13 W x 25 H | |
| 780310E74500 | Med-Tainer SDC, 17 gallon, Green | 18 L x 13 W x 25 H | |
| Med-Tainer Single Deposit Container Lids | | | |
| Model Number | Description | Approximate Size (Inches) | Aperture (inches) |
| 34300474 | Locking Transport Lid, Black | 17 L x 11 W | No opening - use to seal container for transport |
| 34300476 | Large Lab Lid, Black | 17 L x 11 W | 5.75 x 9.75 - use only in areas that have no unsupervised patient access |
| 34300477 | Small Lab Lid, Black | 13 L x 9 W | No opening - use to cover large lab lid when not in use |
| 78033300E74101 | Restrictive Lid, Red | 18.75 L x 13.5 W x 11.6 H | 3.8 H x 13.8 W |
| 78033300E74201 | Restrictive Lid, Teal | 18.75 L x 13.5 W x 11.6 H | 3.8 H x 13.8 W |
| 78033300E74301 | Restrictive Lid, Yellow | 18.75 L x 13.5 W x 11.6 H | 3.8 H x 13.8 W |
| 78033300E74401 | Restrictive Lid, Blue | 18.75 L x 13.5 W x 11.6 H | 3.8 H x 13.8 W |
| 78033300E74501 | Restrictive Lid, Green | 18.75 L x 13.5 W x 11.6 H | 3.8 H x 13.8 W |
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Snyder Industries, Inc.
# Med-Tainer Single Deposit Container (SDC)
August 3, 2016
| ADMINISTRATIVE INFORMATION | |
|----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer Name | Snyder Industries, Inc.<br>6940 O Street, Suite 100<br>Lincoln, NE 68510<br>Telephone: +1 (402) 467-5221<br>Fax: +1 (402) 465-1220 |
| Official Contact | Herbert H. Spann<br>Engineering Manager |
| Representative/Consultant | Claudia H. Fell<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone: +1 (858) 792-1235<br>Fax: +1 (858) 792-1236<br>Email: cfell@paxmed.com |
# DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Med-Tainer Single Deposit Container (SDC) |
|----------------------------|-------------------------------------------|
| Common Name | Container, Sharps |
| Classification Name | Hypodermic single lumen needle |
| Classification Regulations | 21 CFR 880.5570, Class II |
| Product Code | MMK |
| Classification Panel | General Hospital |
| Reviewing Branch | Infection Control Devices Branch |
{4}------------------------------------------------
#### PREDICATE DEVICE INFORMATION
### Primary predicate device: K111085, Rehrig Pacific Company, Sharps Tank Reusable Sharps Container
#### INDICATIONS FOR USE
Med-Tainer Single Deposit Containers (SDC) are intended for the safe collection and transport of medical sharps. The reusable 10 gallon and 17 gallon containers are to be used in supervised areas of healthcare institutions such as hospitals, clinics, veterinary offices and laboratories.
| Med-Tainer Single Deposit Containers | | | |
|-----------------------------------------------------|-----------------------------------|---------------------------|--------------------------------------------------------------------------|
| Med-Tainer Single Deposit Container (SDC) 10 Gallon | | | |
| Model Number | Description | Approximate Size (inches) | |
| 780300E74100 | Med-Tainer SDC, 10 gallon, Red | 18 L x 13 W x 16 H | |
| 780300E74200 | Med-Tainer SDC, 10 gallon, Teal | 18 L x 13 W x 16 H | |
| 780300E74300 | Med-Tainer SDC, 10 gallon, Yellow | 18 L x 13 W x 16 H | |
| 780300E74400 | Med-Tainer SDC, 10 gallon, Blue | 18 L x 13 W x 16 H | |
| 780300E74500 | Med-Tainer SDC, 10 gallon, Green | 18 L x 13 W x 16 H | |
| Med-Tainer Single Deposit Container (SDC) 17 Gallon | | | |
| Model Number | Description | Approximate Size (inches) | |
| 780310E74100 | Med-Tainer SDC, 17 gallon, Red | 18 L x 13 W x 25 H | |
| 780310E74200 | Med-Tainer SDC, 17 gallon, Teal | 18 L x 13 W x 25 H | |
| 780310E74300 | Med-Tainer SDC, 17 gallon, Yellow | 18 L x 13 W x 25 H | |
| 780310E74400 | Med-Tainer SDC, 17 gallon, Blue | 18 L x 13 W x 25 H | |
| 780310E74500 | Med-Tainer SDC, 17 gallon, Green | 18 L x 13 W x 25 H | |
| Med-Tainer Single Deposit Container Lids | | | |
| Model Number | Description | Approximate Size (Inches) | Aperture (inches) |
| 34300474 | Locking Transport Lid, Black | 17 L x 11 W | No opening - use to seal container for transport |
| 34300476 | Large Lab Lid, Black | 17 L x 11 W | 5.75 x 9.75 - use only in areas that have no unsupervised patient access |
| 34300477 | Small Lab Lid, Black | 13 L x 9 W | No opening - use to cover large lab lid when not in use |
| 78033300E74101 | Restrictive Lid, Red | 18.75 L x 13.5 W x 11.6 H | 3.8 H x 13.8 W |
| 78033300E74201 | Restrictive Lid, Teal | 18.75 L x 13.5 W x 11.6 H | 3.8 H x 13.8 W |
| 78033300E74301 | Restrictive Lid, Yellow | 18.75 L x 13.5 W x 11.6 H | 3.8 H x 13.8 W |
| 78033300E74401 | Restrictive Lid, Blue | 18.75 L x 13.5 W x 11.6 H | 3.8 H x 13.8 W |
| 78033300E74501 | Restrictive Lid. Green | 18.75 L x 13.5 W x 11.6 H | 3.8 H x 13.8 W |
### DEVICE DESCRIPTION
Med-Tainer Single Deposit Container (SDC) is used to collect medical sharps in supervised areas of healthcare facilities. The containers are rotationally molded and made of low density polyethylene. They are available in 17 gallon and 10 gallon sizes and in five colors: red, teal, yellow, blue and green. Med-Tainer SDC is a reusable container intended to be reprocessed up to 780 times. The products are supplied to waste management companies that are responsible for transporting, decanting, decontaminating, and destroying the containers. The accompanying lids vary by purpose. The black transport lid is used during transport of the filled container. The combination of black large lab lid, with an opening to discard sharps, and the black small lab lid, used to slide across the large lab lid when not in use, allows easy disposal of sharps. The color coordinating restrictive lid with a mail box type opening is intended to prevent hand or finger access.
{5}------------------------------------------------
# PERFORMANCE DATA
#### Life Span Processing Simulation per CAN/CSA Z316.6-14 BA.8.3.1 Sharps injury protection -Requirements and test methods – Sharps containers
Containers were processed 780 times to simulate their intended life cycle: filled with sharps, closed for transport, opened, decanted, rinsed and decontaminated, Results; all containers successfully completed 780 processing simulation cycles without damage.
### Life Span Transport Simulation per CAN/CSA Z316.6-14 BA.8.3.2 Sharps injury protection -Requirements and test methods – Sharps containers
Containers processed 780 times were filled with sharps and vibrated to simulate transportation. Results: all containers completed transportation without damage.
#### Leak Resistance Test per Health Devices, August - September 1993, ECRI, Vol 22 Nos. 8-9 pgs. 383-384
Containers processed 780 times were filled with water and checked after 24 hours for loss of fluid. Results: passed.
# Container Stability Test per ISO 23907 Sharps injury protection - Requirements and test methods – Sharps containers
Containers processed 780 times were filled with a material of density 0.20 ± 0.01kg/l and placed on an incline with a minimum angle of 15°. Testing was repeated for each side of the container. The container failed the criteria if it toppled over. Results: passed.
# Strength of Handles per ISO 23907 Sharps injury protection - Requirements and test methods -Sharps containers
Containers processed 780 times were filled with a mass equivalent to 150% of the manufacturer's maximum gross mass, closed, suspended by the handles for an hour at 23 ± 5℃ and examined for breakage or detachment. Results: passed.
# Resistance to Damage and Leakage after Dropping per ISO 23907 Sharps injury protection -Requirements and test methods – Sharps containers
Containers processed 780 times were conditioned for testing, filled with liquid and sharps, closed, and dropped a distance of 1 meter. The containers were oriented to land on the bottom, side wall and adjacent side wall. After landing they were inspected for leakage and /or loss of integrity. Results: passed.
# Resistance to Damage and Leakage after Dropping per ISO 23907 Sharps injury protection -Requirements and test methods – Sharps containers (modified)
Containers were conditioned for testing, filled with liquid, sharps, glass, and stainless steel. The fill weight of the container was 40.1 lbs. The containers were closed, and dropped a distance of 1 meter. The containers were oriented to land on the bottom, side wall and adjacent side wall. After landing they were inspected for leakage and /or loss of integrity. Results: passed.
{6}------------------------------------------------
# Puncture Resistance per ASTM F2132-01 Standard Specification of Materials Used in Containers for Discarded Medical Needles and Other Sharps
Each side of the container (bottom, and 4 sides) was overlaid with a 1" x 1" grid. The 12 thinnest areas were tested using a 21 gauge by 1 inch hypodermic needle to measure the force required to penetrate the sample. The criteria to pass the puncture resistance test are an average value not less than 3.4 lbf (15 N) with no value less than 2.8 lbf (12.5 N). Results: passed.
# Puncture Resistance per ASTM F2132-01 Standard Specification of Materials Used in Containers for Discarded Medical Needles and Other Sharps
Each side of the container processed 780 times (lid, bottom, and 4 sides) was overlaid with a 1" x 1" grid. The 12 thinnest areas were tested using a 21 gauge by 1 inch hypodermic needle to measure the force required to penetrate the sample. The criteria to pass the puncture resistance test are an average value not less than 3.4 lbf (15 N) with no value less than 2.8 lbf (12.5 N). Results: passed.
# Drop Test per 49 CRF 178.603 Specification for Packagings – Subpart M – Testing of Non-Bulk Packagings and Packages
Containers were filled, closed, conditioned, and oriented to land during multiple trials on its top, bottom, end, side and top corner. The containers were dropped 1.2 meters, and examined for loss of contents. Results: passed.
# Stack Test per 49 CFR 178.606 Specification for Packagings – Subpart M – Testing of Non-Bulk Packagings and Packages
Containers were subjected to a temperature of 104°F for 28 days with weights of 130 pounds stacked on top of them. The containers were examined for deformation or release of contents. Results: passed.
# Vibration Testing per 49 CFR 178.608 Specification for Packagings – Subpart M – Testing of Non-Bulk Packagings and Packages
Containers were placed on a vertical linear vibration table operated at an amplitude of 1 inch and a frequency of 4.25 Hz for 60 minutes, and inspected for deterioration or loss of contents. Results: passed.
# Cleaning Validation per FDA Guidance: Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling
A cleaning validation protocol was designed and performed. All containers successfully met or surpassed performance criteria.
# EQUIVALENCE TO MARKETED DEVICE
Snyder Industries, Inc. submits the following information in this Premarket Notification to demonstrate, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed predicate device:
K111085. Rehrig Pacific Company, Sharps Tank Reusable Sharps Container
{7}------------------------------------------------
| | Subject Device | Primary Predicate Device |
|---------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Comparison | Snyder Industries, Inc.<br>Med-Tainer Single Deposit<br>Container | Rehrig Pacific Company<br>Sharps Tank Reusable Sharps<br>Container<br>K111085 |
| Indications for Use | Med-Tainer Single Deposit<br>Containers are intended for the safe<br>collection and transport of medical<br>sharps. The reusable 10 gallon and<br>17 gallon containers are to be used<br>in supervised areas of healthcare<br>institutions such as hospitals, clinics,<br>veterinary offices and laboratories.<br>The Med-Tainer Single Deposit<br>Container is available in Kelly<br>green, Ascent blue, teal, red and<br>yellow. | Rehrig Healthcare Systems reusable<br>17 gallon sharps container is<br>intended to be used for the collection<br>and transportation of used medical<br>sharps. The container is intended to<br>be used in hospitals and clinics,<br>operating rooms, and laboratories by<br>technicians, doctors, dentists, and<br>veterinarians. The device is only<br>intended for use in areas with no<br>unsupervised patient access.<br>Empty Weight: 9.4 lbs<br>Color: Red with contrasting<br>Biohazard Symbol Dimensions:<br>20.00"L x 12.50"W x 25.8"H<br>Opening Dimensions:<br>Main lid opening (approximate) -<br>18.00" x 10.7"<br>Sub-lid opening (approximate) - 8.0"<br>x 6.0" Both the main lid and sub-lid<br>have temporary and final closures.<br>Temporary closures are intended to<br>allow opening and closing of the<br>container by the user and to protect<br>against spillage if toppled. The final<br>closure is designed to secure the<br>container during transport. |
| | Subject Device | Primary Predicate Device |
|------------------------------------------------|--------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| Manufacturing | Rotationally molded | Injection Molded |
| Material | Linear Low density polyethylene | High density polyethylene |
| Approximate Size<br>(inches) | 18 L x 13 W x 16 H (10 gallon)<br>18 L x 13 W x 25 H (17 gallon) | 20 L x 12.5 W x 16 H (10 gallon)<br>20 L x 12.5 W x 26 H (17 gallon) |
| Opening dimensions<br>(length x width, inches) | Large Lab Lid opening: 5.75 x 9.75<br>Restrictive Lid opening: 13.7 x 4.2 | Main lid opening: 18.00 x 10.7<br>Sub-lid opening: 8.0 x 6.0 |
| Thickness (inches) | 0.1 | unavailable |
| Tare Weight (pounds) | 10 gallon with lid: 5.5, 17 gallon with lid: 7 | 10 gallon: unavailable, 17 gallon: 9.4 |
| Color | Red, Teal, Yellow, Blue Green with embossed and inked fill line; contrasting label includes Biohazard symbol | Red with contrasting Biohazard Symbol |
# CONCLUSION
The subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.