# EasyPass Guidewire (K153264)

_Medi-Globe Corporation · OCY · May 19, 2016 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153264

## Device Facts

- **Applicant:** Medi-Globe Corporation
- **Product Code:** [OCY](/productcode/OCY.md)
- **Decision Date:** May 19, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The EasyPass Guidewire is used for selective cannulation of the biliary ducts including, but not limited to the common bile duct, cystic duct, and right and left hepatic ducts. The EasyPass guidewire is designed to be used during endoscopic biliary procedures for catheter introduction and exchanges.

## Device Story

EasyPass Guidewire; sterile, single-use, disposable instrument. Facilitates selective cannulation of biliary ducts; assists catheter introduction and exchange during endoscopic biliary procedures. Features Nitinol core wire; proximal 2-color spiral patterned PTFE jacket for visualization; distal stainless steel spring section with hydrophilic coating for flexibility. Used by physicians in clinical endoscopic settings. Device provides mechanical guidance for catheters; aids visualization via spiral pattern. Benefits patient by enabling precise access to biliary anatomy during diagnostic or therapeutic procedures.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via biocompatibility testing (cytotoxicity, irritation, sensitization, USP Class VI) and mechanical bench testing (fracture, bend, friction, scratch, tensile, and radiographic image tests).

## Technological Characteristics

Nitinol core wire; proximal 2-color spiral PTFE jacket; distal stainless steel spring with polymer jacket and hydrophilic coating. Sterile, single-use. Biocompatibility per ISO 10993; USP Class VI color additives. Mechanical testing: fracture, bend, friction, scratch, tensile, and radiographic imaging.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Taxi Endoscopic Guidewire ([K081708](/device/K081708.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an emblem featuring a stylized depiction of three human profiles facing right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 18, 2016

Medi-Globe Corporation Scott Karler Regulatory Affairs Coordinator 7850 South Hardy Drive, Suite. 112 Tempe, AZ 85284

Re: K153264

> Trade/Device Name: EasyPass Guidewire Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: OCY Dated: April 7, 2016 Received: April 13, 2016

Dear Scott Karler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

**Indications for Use**

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

| 510(k) Number (if known)       | K153264                                                                                                                                                                                                                                                                                                      |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                    | EasyPass Guidewire                                                                                                                                                                                                                                                                                           |
| Indications for Use (Describe) | The EasyPass Guidewire is used for selective cannulation of the biliary ducts including, but not limited to the common bile duct, cystic duct, and right and left hepatic ducts. The EasyPass guidewire is designed to be used during endoscopic biliary procedures for catheter introduction and exchanges. |

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

(AN/8) 1886 AQ3 MAO3

ခုေပြေ

ປີ ປະຈຸບັນ ( ໄປນີ້ ) ແລວ ເຄລາ ຄວາ ສາມາດ ສາມາດສານປະການ ລົງສຸ

{3}------------------------------------------------

ົວຂອງ ໄດ້ ເວີຍ ແລ່ງວັນນ້ອຍການປະກວດນາງງາມ ປີ ປຸກາ ຂອງໄປປຸ່ນ ແລະວ່າໄປປຸ່ນ ແວ່ງວ່າ ຂໍ້ມັ

iseqs redio your and exemites nebud sint pribremori in noilemorii in noileemore ent əri bulbuları ,əsrlədən vəd suvorların əsasən və bateminə sinə aid ini və əmir nəbud əri ətələr və bərən əsasən əsasən və qalını qarşının ətəb çərir. Bu və bir və illər və qalında və bir mill of the same and the mails are and of the common and and

:of ,nebrud sint prisuber 101 suoitzaggus pribuloni ,unitsellos noilsmonni siril 10 vog.zdy.spy@jtfillelsAAP 1912 (AR9) ISA noilloubes വാഗ്യാപക Office of Chief Information Officer noilerisinimbA pung bung boo รอว่างาอร์ ทรการ Home HillseH To เมือกการตุลอ

้า ติดใจอีกดว ธ , ot bnoqser of beiluper ใดต ร่านตราคาร เวทราคาราชอาณา เอก เยก บุคคล พ.ศ. 2019 บุคคลจากคำ

{4}------------------------------------------------

## 510(k) SUMMARY

Date of Preparation: 11/09/2015

#### A. Submitter Information:

Medi-Globe Corporation 7850 S. Hardy Drive, #112 Tempe, Arizona 85284

Contact Person: Scott Karler Telephone: (480) 897-2772 ext. 208 Fax: (480) 897-2878 Email: skarler@mediglobe.com

#### B. Proposed Device Information:

- Trade Name: EasyPass Guidewire
- 2 Common Name: Endoscopic Guidewire, Gastroenterology-Urology
- ಷ Classification Name: Endoscope and Accessories 21 CFR876.1500
- Σ Product Code: ΟCY

### C. Predicate Device:

- Trade Name: Taxi Endoscopic Guidewire
- ಷ Manufacturer: Lake Region Manufacturing, Inc.
- と Clearance Number: K081708
- Σ Common Name: Endoscopic Guidewire, Gastroenterology-Urology
- Classification Name: Endoscope and Accessories 21 CFR876.1500
- Product Code: OCY

#### D. Device Description:

The EasyPass Guidewire consists of a Nitinol core wire with a proximal 2-color spiral patterned PTFE jacket and a copolymer distal section. The distal stainless steel spring section of the guidewire provides flexibility and incorporates a hydrophilic coating. The EasyPass Guidewire is a sterile packed, disposable instrument intended for single-patient use only.

#### E. Indications for Use:

The EasyPass Guidewire is used for selective cannulation of the biliary ducts including, but not limited to the common bile duct, cystic duct, and right and left hepatic ducts. The EasyPass guidewire is designed to be used during endoscopic biliary procedures for catheter introduction and exchanges.

{5}------------------------------------------------

#### F. Technological Characteristics Summary:

The EasyPass Guidewire is a 2-color spiral patterned guidewire used in endoscopic procedures. It is comprised of a Nitinol core wire with a 2-color PTFE jacket as a visualization aid during use of the guidewire. The distal section of the EasyPass Guidewire consists of a stainless steel spring encased in a polymer jacket and is hydrophilically coated. The predicate device, Taxi Endoscopic Guidewire. (K081708) incorporates the same design. intended use. diagnostic/therapeutic effect, method of introduction/use, technical characteristics and general range of descriptive features as the proposed. EasyPass Guidewire. There are no technological differences when comparing the proposed EasyPass Guidewire to the predicate device.

#### G. Non-Clinical Performance Data:

Design verification data and biocompatibility testing has demonstrated that the proposed EasyPass Guidewire meets the same performance requirements and is as safe and effective as predicate device, Taxi Endoscopic Guidewire, (K081708).

The following non-clinical performance tests were conducted on the proposed EasyPass Guidewire:

Biocompatibility testing per ISO 10993: Cytotoxicity Irritation Sensitization

USP Class VI Testing performed on color additives

- Bench/performance testing consisted of: Fracture Test Bend Test Friction Test Scratch Test Tensile Test Radiographic Image Test
#### Conclusions of Non-Clinical Performance Data: H.

Medi-Globe Corporation has demonstrated that the proposed EasyPass Guidewire is as safe and effective as the predicate device, (Taxi Endoscopic Guidewire, K081708). The EasyPass Guidewire is considered to have the same intended use, diagnostic/therapeutic effect. method of introduction/use. technical characteristics and general range of descriptive features as the predicate device. (Taxi Endoscopic Guidewire. K081708). No clinical testing was relied on for evaluation of the proposed device.

---

**Source:** [https://fda-staging.innolitics.com/device/K153264](https://fda-staging.innolitics.com/device/K153264)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).