K153253 · GC America, Inc. · EBI · Mar 3, 2016 · Dental
Device Facts
Record ID
K153253
Device Name
GC Reline II
Applicant
GC America, Inc.
Product Code
EBI · Dental
Decision Date
Mar 3, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3760
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GC RELINE II is a resilient material for making soft relining of dentures.
Device Story
GC Reline II is a vinyl polysiloxane (VPS) silicone material used for chairside soft relining of dentures. The system consists of base and catalyst silicone pastes, which are extruded from a cartridge and automixed via a mixing tip. The device is available in three viscosities: Soft, Extra Soft, and Extra Extra Soft. A resin primer is used for adhesion, and a modifier is available to activate existing reline surfaces. The device is intended for use by dental professionals in a clinical setting. By providing a resilient lining, the device improves denture comfort and fit for the patient. The material cures via chemical polymerization of vinyl polysiloxane initiated by a platinum catalyst system.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance with ISO 10139-2:2009 (soft lining materials) and ISO 10993 (biocompatibility) demonstrated through testing of physical properties (hardness, bond strength, sorption, solubility) and biological safety (cytotoxicity, irritation, sensitization).
Technological Characteristics
VPS silicone-based resilient denture liner. Materials: high density polyethylene (cartridge/remover), polypropylene (cap/mixing tip/modifier bottle), glass (primer bottle). Curing mechanism: chemical polymerization of vinyl polysiloxane via platinum catalyst. Conforms to ISO 10139-2:2009 and ISO 10993-1, -2, -5, -10. Non-sterile, 2-year shelf life.
Indications for Use
Indicated for use as a resilient material for soft relining of dentures.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 3, 2016
GC America, Inc. Mr. Mark Heiss Director, Regulatory & Academic Affairs 3737 W. 127th St. Alsip. Illinois 60803
Re: K153253
Trade/Device Name: GC Reline II Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, or Rebasing Resin Regulatory Class: II Product Code: EBI Dated: February 1, 2016 Received: February 3, 2016
Dear Mr. Heiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Mark Heiss
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K153253
Device Name
GC Reline II
Indications for Use (Describe)
GC RELINE II is a resilient material for making soft relining of dentures.
Type of Use (Select one or both, as applicable)
| <span style="font-size:10pt;"><b> <label><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </b></span> |
|---------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size:10pt;"><b> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </b></span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the letters 'GC' in a stylized, teal-colored font. The letters are slightly italicized and have a bold appearance. There are quotation marks before and after the letters.
GC AMERICA INC.
3737 W 127 STREET
ALSIP, ILLINOIS 60803
TEL (708) 597-0900
FAX (708) 926-9100
www.gcamerica.com
## K153253 510(k) Summary
February 29, 2016 Date Prepared:
- Submitter Information: 1.
GC AMERICA INC. 3737 W. 127th Street Alsip, IL 60803
| Contact Person: | Mark Heiss, D.D.S. |
|--------------------|--------------------|
| Phone: | (708) 926-3090 |
| Alternate Contact: | Lori Rietman |
| Phone: | (708) 926-3092 |
| Fax: | (708) 926-9100 |
- 2. Device Name: Proprietary Name: Classification Name: CFR Regulation: Device Classification: Product Code:
GC RELINE II Denture relining, repairing or rebasing resin 21CFR 872.3760 Class II EBI
- 3. Predicate Device:
| Company | Device | 510(k) | Date Cleared |
|------------------|---------------------------|---------|---------------|
| GC America, Inc. | GC Reline Soft/Extra Soft | K990376 | March 5, 1999 |
{4}------------------------------------------------
- 4. Description of Device:
GC RELINE II is a VPS silicone for relining of dentures. The components consist of the base silicone paste and the catalyst silicone paste that are extruded from a cartridge and are automixed with a mixing tip. GC Reline II is available in the following viscosities: Soft, Extra Soft, and Extra Extra Soft.
The cartridge is made of high density polyethylene and the cap is made of polypropylene. The mixing tip is made of polypropylene.
GC Reline Primer for Resin is a resin adhesive that is packaged in a glass bottle.
GC Reline Modifier is available in Liquid A and Liquid B and is similar to the Base and Catalyst of the subject device. Liquid A and Liquid B are dispensed in equal amounts onto the mixing pad and mixed together. It is used to modify (activate) the existing reline surface to accept additional Reline II material.
న. Indications for Use: GC RELINE II is a resilient material for making soft relining of dentures.
- 6. Technological characteristics:
The components of GC RELINE II and the predicate device are similar. The subject device GC Reline II contains a new ingredient, Polydimethylsiloxane, which is used in each of the three viscosities. Bench testing demonstrates that the addition of the new ingredient does not impact technological characteristics when compared to the predicate K990376. The curing mechanism of the applicant and predicate devices is by polymerization of uncured vinyl polysiloxane. This reaction is caused by chemical polymerization initiator systems. Bench top testing indicates that even with different formula, applicant device and predicate devices, meet specifications listed in ISO 10139-2.
Therefore, the applicant and predicate device are substantially equivalent. There are no statistically significant differences between GC RELINE II and GC RELINE when comparing performance data.
{5}------------------------------------------------
| | Applicant device | Primary Predicate |
|---------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product category | CHAIRSIDE<br>VINYL<br>POLYSILOXANE<br>RESILIENT DENTURE LINER | CHAIRSIDE<br>VINYL<br>POLYSILOXANE<br>RESILIENT DENTURE LINER |
| Trade name | GC RELINE II | GC RELINE (Soft/ Extra Soft) |
| 510(k) Number | K153253 | K990736 |
| Indications for<br>Use | GC RELINE II is a resilient material for<br>making soft relining of dentures. | GC RELINE Soft Extra Soft is a resilient<br>material for making soft relining of dentures. |
| Device Description | GC RELINE II is a VPS silicone for relining<br>of dentures. The device consists of the base<br>paste and the catalyst paste Including VPS.<br>The material sets by mixing the catalyst paste<br>with the base paste. | GC RELINE is a VPS silicone for relining of<br>dentures. The device consists of the base paste<br>and the catalyst paste Including VPS. The material<br>sets by mixing the catalyst paste with the base<br>paste. |
| Device<br>Components | Base:<br>Silicon dioxide<br>Vinyldimethylpolysiloxane<br>Methylhydrogen dimethylpolysiloxane<br>Polydimethylsiloxane*<br>Colorant<br>Catalyst: Silicon dioxide<br>Vinyldimethylpolysiloxane<br>Polydimethylsiloxane*<br>Platinum Catalyst | Base:<br>Silicon dioxide<br>Vinyldimethylpolysiloxane<br>Methylhydrogen dimethylpolysiloxane<br>Methylphenylpolysiloxane<br>Colorant<br>Catalyst: Silicon dioxide<br>Vinyldimethylpolysiloxane<br>Platinum Catalyst |
| Similar Physical<br>Properties<br>(tested in<br>accordance with<br>ISO 10139-2) | Shore A hardness, 24h<br>Shore A hardness, 28d<br>Bond strength (MPa)<br>Sorption( $\[\mu$ /mm3)<br>Solubility ( $\[\mu$ /mm3) | Shore A hardness, 24h<br>Shore A hardness, 28d<br>Bond strength (MPa)<br>Sorption( $\[\mu$ /mm3)<br>Solubility ( $\[\mu$ /mm3) |
| Composition of<br>Materials | Cartridge - high density polyethylene<br>Cap - polypropylene<br>Mixing tip - polypropylene.<br>GC Reline II Primer for Resin-8g glass bottle.<br>GC Reline II Modifier-polypropylene bottle.<br>GC Reline II Remover-high density<br>polyethylene. | Cartridge - high density polyethylene<br>Cap-polypropylene<br>Mixing tip - polypropylene.<br>GC Reline II Primer for Resin - 8 g glass bottle.<br>GC Reline II Modifier-polypropylene bottle. |
| Standards of<br>Conformity | ISO 10139:2009, Sections 5.1, 5.2, 5.3<br>ISO10993-1:2009, Sections 6.2.2.2, 6.2.2.3,<br>6.2.2.4<br>ISO 10993-2:2006<br>ISO 10993-5:2009 Sections 8.5<br>ISO 10993-10:2010, Sections 6, 7 | Not available |
| Biocompatibility | 10993-5 Cytotoxicity - pass<br>10993-10 irritation - pass<br>10993-10 sensitization - pass | Not available |
{6}------------------------------------------------
### Biocompatibility Testing
| Test | Test Standard | Test Results<br>(pass or fail) |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------|
| Cytotoxicity Test | Under the conditions of ISO 10993-1:2009, Biological evaluation of medical<br>devices, sections 6.2.2.2, 6.2.2.3, 6.2.2.4, the test articles should meet the test<br>requirements | Pass |
| Cytotoxicity Test<br>In Vitro | Under the conditions of ISO 10993-5:2009, Biological evaluation of medical<br>devices, section 8.5 Determination of cytotoxicity, the test articles should meet<br>the test requirements | Pass |
| Irritation and<br>Sensitization | Under the conditions of ISO 10993-10:2010, biological evaluation of medical<br>devices, Sections 6, 7, the test articles should meet the<br>test requirements | Pass |
#### 7. Nonclinical Testing:
Benchtop testing demonstrates that the subject device GC Reline II, like its predicate device GC Reline Soft/Extra Soft, is in compliance with ISO 10139-2:2009 Dentistry-Soft lining materials for removable dentures-Part 2: Materials for long-term use, and as such, the subject device meets specific requirements for softness, adhesion, water sorption and water solubility as well as for packaging, marking and manufacturer's instructions for soft denture lining materials suitable for long-term use. The subject device GC Reline II was also tested and met the requirement for cytotoxicity, irritation and sensitivity as shown in the biocompatibility table above.
#### 8. Substantial equivalence:
The subject device GC RELINE II and its predicate device GC RELINE (Soft/ Extra Soft) share the same function and intended use. The curing mechanism of the new and predicate device is equivalent. Bench testing demonstrates that the addition of a new ingredient in the subject device does not impact technological characteristics when compared to the predicate K990376.
The subject device and predicate devices encompass the same range of physical dimensions, including hardness, bond strength, sorption, solubility, and also share equivalent biocompatibility. Both the subject and predicate devices are provided nonsterile and have a shelf life of 2 years. Any differences in the technological characteristics do not raise new questions.
GC AMERICA INC. has demonstrated that, for the purposes of FDA's regulation of medical devices, the subject device GC Reline II is substantially equivalent to the predicate device in intended use, material composition, fundamental scientific technology, principals of operation, and basic design.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.