# LSFG-NAVI (K153239)

_Softcare Co., Ltd. · HKI · May 12, 2016 · Ophthalmic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153239

## Device Facts

- **Applicant:** Softcare Co., Ltd.
- **Product Code:** [HKI](/productcode/HKI.md)
- **Decision Date:** May 12, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.1120
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The LSFG-NAVI system is intended to capture and display the blood flow distribution in the human retina in real time, and to monitor the blood flow in retinal vessels for their quantitative evaluation.

## Device Story

LSFG-NAVI is an ophthalmic camera system for non-invasive, real-time visualization and monitoring of retinal blood flow. The device uses a diode laser (830nm) to illuminate the retina; scattered light from moving blood cells creates a laser speckle interference pattern. A high-sensitivity CCD camera captures these patterns; the system calculates the rate of time variation of speckle intensity at each pixel, which is proportional to blood flow velocity. The system consists of a camera unit, 3D-stage with chin/head rest, and a PC running analysis software. It produces 2D blood flow maps and time-varying flow indices synchronized with the heartbeat. Used in clinical settings, the device is operated by trained personnel. Healthcare providers use the displayed perfusion maps and flow indices to monitor retinal vascular blood flow, aiding in the evaluation of ocular conditions. The device provides quantitative data without requiring mydriatic agents.

## Clinical Evidence

No clinical data. Effectiveness established via bench testing: validation of blood flow index measurement, validation of time variation of blood flow index measurement, and validation of retinal perfusion map. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60825-1, and ISO 15004-2 demonstrated.

## Technological Characteristics

Ophthalmic camera using laser speckle flowgraphy. Light source: 830nm diode laser; 940nm LED illumination. Sensing: CCD camera. Connectivity: PC-based system. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60825-1, ISO 15004-2, ISO 14971. Software: Measurement and analysis software for 2D flow mapping.

## Regulatory Identification

An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

## Predicate Devices

- Retinal Functional Imager ([K062416](/device/K062416.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles facing right, overlaid on top of each other. The profiles are connected by a flowing, ribbon-like element. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 12, 2016

Softcare Co., Ltd % Mr. Takahiro Haruyama President Globizz Corporation 1411 W. 190th St., Suite 200 Gardena, CA 90248

Re: K153239

Trade/Device Name: LSFG-NAVI Regulation Number: 21 CFR 886.1120 Regulation Name: Opthalmic Camera Regulatory Class: Class II Product Code: HKI, HLI Dated: April 8, 2016 Received: April 11, 2016

Dear Mr. Haruyama:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kesia Alexander

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |
|-----------------------------------------|
| Food and Drug Administration            |
| <b>Indications for Use</b>              |

| Form Approved: OMB No. 0910-0120  |
|-----------------------------------|
| Expiration Date: January 31, 2017 |
| See PRA Statement below.          |

| 510(k) Number (if known)       | K153239                                                                                                                                                                                                                            |
|--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name                    | Laser Speckle Flowgraphy LSFG-NAVI                                                                                                                                                                                                 |
| Indications for Use (Describe) | The Laser Speckle Flowgraphy LSFG-NAVI system is intended to capture and display the blood flow distribution in the human retina in real time, and to monitor the blood flow in retinal vessels for their quantitative evaluation. |

| Type of Use (Select one or both, as applicable) | <div> <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </div> <div> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </div> |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

 Food and Drug Administration  

 Office of Chief Information Officer  

 Paperwork Reduction Act (PRA) Staff  

 PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

| FORM FDA 3881 (8/14) | Page 1 of 1 |
|----------------------|-------------|
|----------------------|-------------|

{3}------------------------------------------------

# 510(k) Summary

### Submitter's Name:

Software Co., Ltd. 243-5 Yatsunami Fukutsu City, Fukuoka 811-3201 JAPAN Phone: +81-940-36-9693

### Contact Person:

Noriyoshi Takahashi Phone: +81-940-36-9693 Email: takahashi@softcare-Itd.co.jp

## Official Correspondence:

Takahiro Haruyama Globizz Corporation 1411 W. 190th St. Suite #200 Gardena, CA 90248

### Date of 510(k) summary preparation: November 3, 2015

Trade/Proprietary Name: Laser Speckle Flowgraphy LSFG-NAVI Common/Usual Name: Ophthalmic Camera Device /Ophthalmoscope Classification Name: Camera, Ophthalmic, Ac-powered (21CFR 886.1120, Product Code HKI) Ophthalmoscope, Ac-powered (21CFR 886.1570, Product Code HLI)

## Device Description:

The LSFG-NAVI can visualize and measure the distribution of blood flow in the human retina in a broad area from macular to the optic nerve head. The most useful feature of the instrument is the ability to non-invasively observe varying blood flow synchronized with the heartbeat as a series of 2D blood flow maps.

{4}------------------------------------------------

The LSFG-NAVI consists of following major units:

- (1) LSFG-NAVI Camera Unit: including an emitter of a diode laser, retinal camera imaging system, imager, CCD camera (video capturing)
- (2) Personal Computer (PC): including LSGF-NAVI System Software (Measurement Software and Analysis Software
- (3) 3D-Stage Unit: The LSFG-NAVI Camera Unit is attached to the 3D-Stage unit and a patient places his/her head and face at Chin Rest and Head Rest of the 3D-Stage Unit.
- (4) Electronic Unit: including the AC Adapter and Isolation Transformer.

#### Intended Use:

The LSFG-NAVI system is intended to capture and display the blood flow distribution in the human retina in real time, and to monitor the blood flow in retinal vessels for their quantitative evaluation.

### The LSFG-NAVI Technology:

The LSFG-NAVI is based on image analyses of the so called 'speckle' pattern produced by random laser interference. The system consists of an emitter of a diode laser, retinal camera imaging system, imager, hardware for video capturing and a personal computer (PC). The light from the diode laser is expanded and illuminates the retinal surface with a large spot. The laser is scattered from the moving blood cells flowing in the vessels and the capillary network, and comes back to the image plane of the camera, where the random interference pattern or laser speckles are produced. According to the motion of the blood cells, the structure of the image speckles varies over time, and the rate of variation at that point is proportional to the average flow velocity at the corresponding objective point. The image intensity is captured by a highly sensitive CCD camera and the rate of time variation is calculated at each pixel point. By displaying the results in 2D form, the time varying blood flow maps of the retina become observable as a motion picture.

#### Non clinical performance data:

LSFG-NAVI has been tested and met the standards for Electrical safety testing of IEC 60601-1, electromagnetic compatibility testing of IEC 60601-1-2, laser safety testing of IEC 60825-1, optical radiation levels of ISO 15004.

The LSFG-NAVI software was also verified and validated according the "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued by FDA in 2005. Risk mitigation of LSFG-NAVI software was also carried out based on ISO 14971.

Furthermore, the 3 non-clinical testing were performed to establish its effectiveness and performance,

{5}------------------------------------------------

namely, Validation of blood flow index measurement, validation of time variation of

blood flow index measurement, and validation of perfusion map of retina.

The LSFG-NAVI also complies with the radiation safety regulation and the radiation safety report (product report, 21CFR 1002.10) has been submitted and the accession number assigned (#1310447-000).

It is of importance to note that the LSFG-NAVI has not been tested clinically where this statement is included in the User Manual to inform the end user.

## Substantial Equivalent:

The Laser Speckle Flowgraphy LSFG-NAVI is considered to be substantially equivalent to the predicate device below:

- י Retinal Functional Imager (K062416) - Optical Imaging, Ltd.
The comparison of the technological characteristics between LSFG-NAVI and the Retinal Functional lmager (RFI) - K062416 is summarized below:

| Item                                | LSFG-NAVI                                                                                                                                                                                        | RFI                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|-------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) number                       | K153239                                                                                                                                                                                          | K062416                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Product Code /<br>Classification    | HKI/HLI                                                                                                                                                                                          | HKI/HLI                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Classification Name                 | Ophthalmic Camera Device; AC<br>powered.<br>Ophthalmoscope                                                                                                                                       | Ophthalmic Camera Device; AC<br>powered.<br>Ophthalmoscope                                                                                                                                                                                                                                                                                                                                                                                     |
| Class                               | II                                                                                                                                                                                               | II                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Method                              | Image using Diode Laser. A<br>mydriatic agent is not required in<br>a darkroom (displayed on PC<br>monitor).                                                                                     | Similar to conventional fundus<br>camera with eye drop (displayed<br>on PC monitor)                                                                                                                                                                                                                                                                                                                                                            |
| Item                                | LSFG-NAVI                                                                                                                                                                                        | RFI                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Intended Use                        | The LSFG-NAVI system is to capture and display the blood flow distribution in the human retina in real time, and to monitor the blood flow in retinal vessels for their quantitative evaluation. | The Retinal Functional Imager is a digital imaging and fundus camera system intended to observe, capture, display, and store images of patients' fundus (retina) under mydriatic conditions and provide information obtained from the images about blood flow (velocity) and path of flow in retinal vessels. The device is indicated for use as an aid in diagnosing or monitoring diseases of the eye that may be observed and photographed. |
| Modes of operation                  | Blood flow Index and Perfusion map of retina                                                                                                                                                     | Red-free manual multiflash –blood flow velocity and path by flow                                                                                                                                                                                                                                                                                                                                                                               |
| Target Population                   | General                                                                                                                                                                                          | General                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Anatomical region                   | Retina                                                                                                                                                                                           | Retina                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Light Source                        | Laser Diode (830nm)                                                                                                                                                                              | Halogen lamp 50W                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Field angle(degrees)                | 21                                                                                                                                                                                               | 50, 35, 20                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Viewing Magnification               | 1.2X                                                                                                                                                                                             | 12.1X, 17.1X, 29.9X                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Focusing                            | Matching two laser spots                                                                                                                                                                         | Matching two oscillating points                                                                                                                                                                                                                                                                                                                                                                                                                |
| Working distance                    | 44mm                                                                                                                                                                                             | 40mm                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Record Media                        | Digital data                                                                                                                                                                                     | Digital data / 35 mm film                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Principle of blood flow measurement | Laser Speckle Flowgraphy                                                                                                                                                                         | Frame Mapping of Xenon Flash lamp illumination                                                                                                                                                                                                                                                                                                                                                                                                 |
| Measurement duration                | 30Hz for data acquisition                                                                                                                                                                        | 125 ms for data acquisition                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Measurement results                 | Given as relative value                                                                                                                                                                          | Given in mm / sec                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Numerical data                      | Blood flow Index                                                                                                                                                                                 | Velocity flow rate                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Graphically data                    | Perfusion map of retina,<br>Time variation of blood flow index                                                                                                                                   | Path of flow                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Illumination Light                  | LED (940nm)                                                                                                                                                                                      | 12 V Halogen lamp                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Item                                | LSFG-NAVI                                                                                                                                                                                        | RFI                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Flash Exposure                      | Continuous exposure                                                                                                                                                                              | Automatic, 1 ms pulse, train of up<br>to 8 pulses, repetition rate 17.5ms.                                                                                                                                                                                                                                                                                                                                                                     |
| Electrical Safety                   | IEC 60601-1:2012<br>IEC 60601-1-2: 2007                                                                                                                                                          | IEC 60601-1: 1st Ed.<br>IEC 60601-1-2: 1st Ed.                                                                                                                                                                                                                                                                                                                                                                                                 |
| Performance Standard                | ISO 15004-2:2007<br>ISO 10940: 1998<br>ISO 60825-1: 2007<br>ISO 14971: 2007                                                                                                                      | ISO 15004: 1st Ed.<br>ISO 14940: 1998<br>ANSI RP-27-1-96: 1st Ed.                                                                                                                                                                                                                                                                                                                                                                              |

{6}------------------------------------------------

{7}------------------------------------------------

Although there is technological difference between two devices, the results of nonclinical testing, including electrical safety and electromagnetic compatibility along with the software validation and verification, demonstrated the safety and effectiveness of LSFG-NAVI with the predicate device and did not raise a new concern for safety and/or effectiveness.

Both devices are ophthalmic imaging management systems where intended use for both devices is to capture, display, and store images of the retina to monitor the blood flow in retinal vessels for their quantitative evaluation.

## The comparison of the technological characteristics

LSFG-NAVI selected The Retinal Functional Imager (RFI) - K062416 as the predicated device, and examined the substantial equivalence to its predicate device. The intended use and function are primarily equivalent and did not identify any new problem in safety and effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K153239](https://fda-staging.innolitics.com/device/K153239)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).