← Product Code [LJT](/productcode/LJT) · K153238

# Dignity Dual Port (K153238)

_Medcomp ( Medical Components) · LJT · Mar 15, 2016 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K153238

## Device Facts

- **Applicant:** Medcomp ( Medical Components)
- **Product Code:** [LJT](/productcode/LJT.md)
- **Decision Date:** Mar 15, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5965
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Dignity® Dual Port is a power injectable implantable infusion port that is indicated for patient therapies requiring repeated access to the vascular system. The Dignity® Dual Port can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a power injectable needle, the Dignity® Dual Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s with a 19 or 20 gauge non-coring power injectable needle. The maximum recommended infusion rate is 2 ml/s with a 22 gauge non-coring power injectable needle.

## Device Story

Dignity Dual Port is an implantable, power-injectable infusion port providing repeated vascular access. Device consists of two reservoirs, self-sealing silicone septa, and a radiopaque polyurethane catheter. Used in clinical settings by healthcare professionals; accessed percutaneously via non-coring needle. Power injection of contrast media enabled at rates up to 5 ml/s (19/20G needle) or 2 ml/s (22G needle). Device identified subcutaneously by palpation and via 'CT' marking under radiography. Provides reliable access for long-term therapies; benefits include reduced need for repeated venipuncture and support for high-flow contrast imaging.

## Clinical Evidence

Bench testing only. No clinical data. Performance testing included MRI conditional testing (ASTM F2113, F2052, F2503), corrosion susceptibility (ASTM F2129), and mechanical integrity testing per ISO 10555-1, ISO 10555-3, and ISO 10555-6. Biocompatibility evaluated per ISO 10993-1 using data from reference devices.

## Technological Characteristics

Materials: Polysulfone, silicone, polyurethane, titanium, polycarbonate. Design: Double lumen, dual reservoir, in-line configuration. Sterilization: Ethylene Oxide (ETO) per ISO 11135. MRI conditional. Connectivity: None (mechanical implant).

## Regulatory Identification

A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.

## Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.

## Predicate Devices

- C.R. Bard, Inc. PowerPort duo M.R.I. Implanted Port ([K090512](/device/K090512.md))

## Reference Devices

- Medcomp Power Injectable Implantable Infusion Port; Pro-Fuse ([K070003](/device/K070003.md))
- Medcomp Dignity Power Injectable Titanium Port ([K120281](/device/K120281.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 15, 2016

Medcomp (Medical Components) Ms. Courtney Nix Regulatory Associate 1499 Delp Drive Harleysville, Pennsylvania 19438

Re: K153238

Trade/Device Name: Dignity Dual Port Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, implanted, intravascular infusion port and catheter Regulatory Class: II Product Code: LJT Dated: February 9, 2016 Received: February 12, 2016

Dear Ms. Nix:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/7 description: The image contains a signature and the name "Tina Kiang" in a simple, sans-serif font. The signature is complex and illegible, featuring looping lines and curves. The text is positioned to the right of the signature, and the letters "-s" are located below the name.

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K153238

Device Name Dignity® Dual Port

Indications for Use (Describe)

The Dignity® Dual Port is a power injectable implantable infusion port that is indicated for patient therapies requiring repeated access to the vascular system. The Dignity® Dual Port can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples,

When used with a power injectable needle, the Dignity® Dual Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s with a 19 or 20 gauge non-coring power injectable needle. The maximum recommended infusion rate is 2 ml/s with a 22 gauge non-coring power injectable needle.

Type of Use (*Select one or both, as applicable*)

| <label><input checked="true" type="checkbox"/> Preconstruction Use (Text CA SB 255, 2015, Statute)</label> | <label><input type="checkbox"/> Interim Use</label> |
|------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|
|------------------------------------------------------------------------------------------------------------|-----------------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

# CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a pink geometric shape that resembles a stylized letter "M" or a series of interconnected squares. Below the geometric shape, the word "med" is written in pink, followed by "COMP" in gray. The overall design is clean and modern.

1499 Delp Drive Harleysville, PA 19438 Tel: 215-256-4201 Fax. 215-256-1787 www.medcompnet.com

| Section 5 |                         | 510(k) SUMMARY<br>K153238                                                                                                            | Traditional 510K |
|-----------|-------------------------|--------------------------------------------------------------------------------------------------------------------------------------|------------------|
| A.        | Submitter Information:  |                                                                                                                                      |                  |
|           | Submitter:              | MEDCOMP®<br>1499 Delp Drive<br>Harleysville, PA 19438<br>Tel: (215) 256-4201, x2244<br>Fax: (215) 256-9191                           |                  |
|           | Contact:                | Courtney Nix<br>Regulatory Associate<br>Cnix@medcompnet.com                                                                          |                  |
|           | Date Prepared:          | March 11, 2016                                                                                                                       |                  |
| B.        | Trade Name:             | Dignity® Dual Port                                                                                                                   |                  |
|           | Common Name:            | Power Injectable, Implantable, Infusion Port                                                                                         |                  |
|           | Product Code :          | LJT - Subcutaneous, Implanted, Intravascular Infusion<br>Port and Catheter                                                           |                  |
|           | C.F.R. Section:         | 21 CFR 880.5965, Class II                                                                                                            |                  |
|           | Classification Panel:   | General Hospital                                                                                                                     |                  |
| C.        | Predicate Devices:      | K090512 C.R. Bard, Inc. PowerPort© duo M.R.I.TM<br>Implanted Port with 9.5 Fr, Dual Lumen ChronoFlex®<br>Catheter, class II 880.5956 |                  |
| D.        | Reference Devices:      | K070003 Medcomp®, Power Injectable Implantable<br>Infusion Port; Pro-Fuse®, class II 880.5965                                        |                  |
|           |                         | K120281 Medcomp®, Dignity® Power Injectable<br>Titanium Port, class II 880.5965                                                      |                  |
| E.        | Purpose for Submission: | Expansion of Medcomp®'s Dignity® product line to<br>include an in-line dual reservoir point with injection<br>capabilities.          |                  |

The Dignity® Dual power injectable implantable infusion port is an implantable access device designed to provide repeated access to the vascular system. Port access is performed by percutaneous needle insertion using a non-coring needle. Power injection is performed using a power injectable needle only. The Dignity® Dual Port device consists of two primary components: an injection port with a self-sealing septa and a radiopaque catheter. The Dignity® Dual Ports can be identified subcutaneously by feeling the top of the septum and the top rim of the port

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for medCOMP. The logo features a stylized, geometric design in pink, resembling a house or an abstract letter 'M'. Below the geometric shape, the word "medCOMP" is written in a stylized, italicized font, also in pink. 

Harleysville, PA 19438

Tel: 215-256-4201

Fax. 215-256-1787

www.medcompnet.com

housing. Power Injectable Implantable Infusion Ports can be identified by the letters "CT" under radiographic imaging.

The Dignity® Dual Port will be marketed in four kit configurations, which are listed below:

- 9.5F Standard Port Kit (Catalog number MRDP95ADN)
- 9.5F Port Kit with Micro-Stick (Catalog number MIDP95ADN)
- 9.5F Standard Port Kit with Silicone Filled Suture Holes (Catalog number MRDP95ADS)
- 9.5F Port kit with Silicone Filled Suture Holes and with Micro-Stick (Catalog number MIDP95ADS)

#### Indications for Use: G.

The Dignity® Dual Port is a power injectable implantable infusion port that is indicated for patient therapies requiring repeated access to the vascular system. The Dignity® Dual Port can be used for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples.

When used with a power injectable needle, the Dignity® Dual Port is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s with a 19 or 20 gauge non-coring power injectable needle. The maximum recommended infusion rate is 2 ml/s with a 22 gauge non-coring power injectable needle.

#### H. Comparison to Predicate Devices:

The Dignity® Dual Port is substantially equivalent to the predicate device, C.R. Bard PowerPort© (K090512), in terms of intended use, anatomical location, basic design, performance, labeling, manufacturing process and method of sterilization.

The Dignity® Dual power injectable implantable infusion port, is equivalent to the Medcomp® references devices, Medcomp® Pro-Fuse® (K070003) and Medcomp® Dignity® Power Injectable Titanium Port (K120281), in materials.

The difference between the reference device, Dignity® Power Injectable Titanium Port (K120281), and the Dignity® Dual Port assembly is that the Dignity® Dual Port has a polysulfone cap, base, two silicone septa and a titanium insert, while the titanium port is comprised of a one piece titanium housing and silicone septum. The Dignity® Dual Port utilizes a titanium insert to connect the distal reservoir to the catheter. Both ports use the same grade of titanium.

The difference between the Dignity® Dual Port and the predicate, Bard Access Systems, PowerPort™ duo M.R.I™ Implanted Port (K090512), is that the PowerPort™ duo M.R.I™ is a side-by-side reservoir design, whereas the Dignity® Dual Port has an in-line reservoir design.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with a geometric design in pink and the word "medCOMP" in a combination of pink and gray. The geometric design consists of a stylized shape resembling an incomplete square with rounded corners, positioned above two curved shapes that mirror each other. The word "medCOMP" is written in lowercase, with "med" in pink and "COMP" in gray.

Harleysville, PA 19438

Tel: 215-256-4201

Fax. 215-256-1787

www.medcompnet.com

## Table 1: 510(K) Summary: Design Comparison Matrix

| Device                        | Proposed Dignity® Dual Port<br>(K153238)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | C.R. Bard PowerPort®<br>(K090512)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Substantially<br>Equivalent<br>Comparison                                                                     |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Design                        | Double lumen open ended<br>Dual reservoir<br>Power injection                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Double Lumen<br>Dual reservoir<br>Power Injection                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Equivalent                                                                                                    |
| Dimensions                    | Bases Fits within:<br>41mm x 12mm<br>Internal Volume:<br>0.7ml and 0.6 ml<br>Depth Marking every 1cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Bases Fits within:<br>29mm x 39mm<br>Internal Volume:<br>0.6 ml each reservoir<br>Depth Marking every 1cm                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Equivalent                                                                                                    |
| Indications<br>for use        | The Dignity® Dual Port is a<br>power injectable implantable<br>infusion port that is indicated for<br>patient therapies requiring<br>repeated access to the vascular<br>system. The Dignity® Dual Port<br>can be used for infusion of<br>medications, I.V. fluids,<br>parenteral nutrition solutions,<br>blood products, and for the<br>withdrawal of blood samples.<br><br>When used with a power<br>injectable needle, the Dignity®<br>Dual Port is indicated for power<br>injection of contrast media. For<br>power injection of contrast<br>media, the maximum<br>recommended infusion rate is 5<br>ml/s with a 19 or 20 gauge non-<br>coring power injectable needle.<br>The maximum recommended<br>infusion rate is 2 ml/s with a 22<br>gauge non-coring power<br>injectable needle. | PowerPort™ Implanted Port is<br>indicated for patient therapies<br>requiring repeated access to the<br>vascular system. The port<br>system can be used for infusion<br>of medications, I.V. fluids,<br>parenteral nutrition soltuions,<br>blood products, and for the<br>withdrawal of blood samples.<br><br>When used with a PowerLoc™<br>Safety Infusion Set, The<br>PowerPort™ device is indicated<br>for power injection of contrast<br>media. For power injection of<br>contrast media, the maximum<br>recommended infusion rate is 5<br>ml/s.<br>(From 510(k) Summary<br>K090512) | Equivalent                                                                                                    |
| Catheter<br>Size              | 9.5F Double Lumen ChoronFlex<br>Polyurethane Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 9.5F Double Lumen ChoronFlex<br>Polyurethane Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | Equivalent                                                                                                    |
| Sterilization                 | ETO<br>ANSI/AAMI/ISO 11135-1:2007<br>SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | ETO<br>Reference Not Sited in<br>published 510(k) Summary                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Equivalent                                                                                                    |
| Power<br>Injectable           | Power Injectable<br>Rate: 5 ml/s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Power Injectable<br>Rate: 5 ml/s                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Equivalent                                                                                                    |
| Materials                     | Thermoplastic Polymer<br>Silicone<br>Polyurethane<br>Titanium<br>Polycarbonate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Plastic<br>Silicone<br>Polyurethane                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Equivalent to<br>References<br>Devices;<br>Medcomp®<br>Pro-Fuse®<br>(K070003)<br>and<br>Medcomp ®<br>Dignity® |
|                               |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                                                                                               |
| MRI<br>Performance<br>Testing | MRI Conditional<br>Power Injection:<br>19Ga 4.9ml/s<br>20Ga 4.9ml/s<br>22Ga 1.9ml/s<br>Infusion Testing: 1600ml/hr<br>Catheter Lock Disengagement:<br>6.9 lb<br>Needle Insertion:<br>19Ga 3.8lb<br>22Ga 3.6 lbs<br>Gravity Flow:<br>1760 ml/hr                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | MRI Conditional<br>Power Injection:<br>19Ga 4.9ml/s<br>20Ga 4.9ml/s<br>22Ga 1.9ml/s<br>Infusion Testing: 1700ml/hr<br>Catheter Lock<br>Disengagement:<br>10 lbs<br>Needle Insertion:<br>19Ga 2.88lb<br>22Ga 2.05 lbs<br>Gravity Flow:<br>1680 ml/hr                                                                                                                                                                                                                                                                                                                                     | Injectable<br>Titanium Port<br>(K120281)<br>Equivalent<br>Equivalent                                          |

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a red geometric shape that resembles a stylized house or building. Below the shape, the word "medCOMP" is written in a stylized font, with "med" in red and "COMP" in gray.

Harleysville, PA 19438

Tel: 215-256-4201

Fax. 215-256-1787

www.medcompnet.com

l,

Bench / Performance Data / Non-Clinical Testing:

The results of performance testing, in conjunction with the substantial equivalence claims, effectively demonstrate the proposed device is equivalent to the predicate devices. The performance testing was performed in accordance with the following standards:

- BS EN ISO 10555-1: 2013: Intravascular Catheters Sterile and Single-use Catheters Part 1: General Requirements
- ISO 10555-3: 2013; Intravascular Catheters Sterile and Single-Use -Catheters - Part 3: Central Venous Catheters (General Plastic Surgery/General Hospital)
- BS EN ISO 10993-1; 2009: Biological evaluation of medical devices --Part 1: Evaluation and testing within a risk management process (Biocompatibility)
- BS EN ISO 11135; Second Edition 2014; Sterilization of health-care products - Ethylene oxide-Requirements for the development, validation and routine control of a sterilization process for medical devices
- ISO 10555-6: 2015; Intravascular Catheters Sterile and Single-use -Catheters - Part 6: Subcutaneous implanted ports
- ASTM F2213-06: (Reapproved 2011), standard test method for , measurement of magnetically induced torque on medical devices in the magnetic resonance environment. (Materials)
- ASTM F2129-08: Standard test method for conducting cyclic = potentiodynamic polarization measurements to determine the corrosion susceptibility of small implant devices. (Materials)
- ASTM F2052-14: standard test method for measurement of magnetically induced displacement force on medical devices in the magnetic resonance environment. (Materials)
- ASTM F2503-13: standard practice for marking medical devices and other items for safety in the magnetic resonance environment. (Materials)

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for medCOMP. The logo consists of a pink geometric shape that resembles a stylized letter "M" or a house. Below the shape, the word "medCOMP" is written in a combination of pink and gray letters. 

Harleysville, PA 19438

Tel: 215-256-4201

Fax. 215-256-1787

www.medcompnet.com

### Biocompatibility:

J.

Materials of the Dignity® Dual Port are equivalent to the cleared, reference devices; Medcomp®, Power Injectable Implantable Infusion Port: Medcomp® Pro-Fuse® (K070003) and Medcomp ® Dignity® Power Iniectable Titanium Port (K120281). Biocompatibility on both reference devices we performed per ISO 10993-1, for a permanent implant device with tissue and both indirect and direct blood contact.

The Dignity® Dual Port is comprised of a polysulfone cap (with silicone filled suture holes, or open suture holes), two silicone septa, and a polysulfone base assembly with a titanium tube that provides a channel from the stem to the distal reservoir. The reservoir is plastic (polysulfone). The Dignity® Dual catheter locking assembly (makrolon/pellethane) locks the chronoflex lumen to the plastic stem of the Dignity® Dual Port. The materials that the Dignity® Dual Port is constructed of are the same as those cleared in the Power Injectable Implantable Infusion Port Pro-Fuse ™ (K070003), and the same formulation of titanium as cleared in the Dignity® Power Injectable Titanium Port (K120281). Therefore, as mentioned above, Medcomp® intends to use approved Medcomp® devices, Medcomp® Pro-Fuse® (K070003) and Medcomp® Dignity® Power Injectable Titanium Port (K120281), for biocompatibility testing results per ISO 10993-1, for a permanent implant device with tissue and both indirect and direct blood contact. Hence, all biocompatibility testing demonstrates the materials used meet the requirements of ISO 10993-1 and a summary of prior testing is contained in section 15.

#### Summary of Substantial Equivalence: K.

In conclusion, the proposed device, The Dignity® Dual power injectable implantable infusion port, is equivalent to the predicate device, C.R. Bard PowerPort© (K090512), in intended use, design, function, and performance. The proposed device. The Dignity® Dual power injectable implantable infusion port, is equivalent to the Medcomp® references devices, Medcomp® Pro-Fuse® (K070003) and Medcomp® Dignity® Power Injectable Titanium Port (K120281), in materials.

The proposed device, The Dignity® Dual power injectable implantable infusion port, meets the performance criteria of design verification as specified by ISO standards. quidance documents and internal test protocols. The proposed device is substantially equivalent to the indicated legally marketed predicate and reference devices as defined in paragraph above.

---

**Source:** [https://fda-staging.innolitics.com/device/K153238](https://fda-staging.innolitics.com/device/K153238)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
